From 1c6aa04687ddef929eadea103de6413c23a69361 Mon Sep 17 00:00:00 2001 From: serversdown Date: Tue, 20 Oct 2026 17:41:00 -0400 Subject: [PATCH] docs(spec): SFM Series-4 parity and a device safety layer -- design for review Parity of jobs with Series 3 for regular comms and settings (identify, status, settings, events, download, monitor log, start/stop, call-home enable/disable, setup push); Series-4 emergency tools deferred. A single device-access layer: series from the serial prefix, identity confirmed before any write, a per-unit lock, an 0xA2 opcode guard, read-back verification of every change, distinct error codes, and an audit table. Series-4 sessions ("stay on the line") now; Series-3 sessions after a bench test of MiniMate hangup/DCD behaviour. Includes the fix list found in the 2026-10-19 inventory, the to-be-added table, and the RX55 container idea as a parking-lot item. Co-Authored-By: Claude Opus 5.5 Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44 --- .../2026-10-20-sfm-series4-parity-design.md | 343 ++++++++++++++++++ 1 file changed, 343 insertions(+) create mode 100644 docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md diff --git a/docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md b/docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md new file mode 100644 index 0000000..0cbf114 --- /dev/null +++ b/docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md @@ -0,0 +1,343 @@ +# SFM: Series-4 parity and a device safety layer — design + +**Date:** 2026-10-20 · **Branch:** `feat/series4-sfm-parity` · **Status:** draft for review +**Decided with:** Brian, in conversation 2026-10-19/20. + +## In plain terms + +SFM can already *read* a Micromate almost as well as it reads a MiniMate, but +only through the Python library and bench scripts. SFM's web endpoints still +only speak Series 3. This design does three things: + +1. Gives Series 4 the same everyday jobs in SFM that Series 3 has: identify a + unit, check its status, read its settings, list and download events, + start/stop monitoring, and change its settings. +2. Adds a **safety layer** every device request goes through. It makes sure a + command only reaches the unit it was meant for, logs every change, and + checks every change by reading the unit back. +3. Lets SFM **stay on the line** with a Micromate. It calls up once, confirms + the serial, and keeps the connection open for later commands instead of + reconnecting each time. + +It does **not** cover the emergency tools (runaway-unit stops, rescue, erase), +the call-home receiver, or anything async. Those come later, as features +shared by both series. + +**Production reach, stated honestly:** while THOR is set to stay connected to +production Micromates, a modem's single connection is taken. SFM can't reach +those units until THOR lets go or SFM takes over call-home. In this phase, +Series-4 SFM is for the bench and for units THOR is not holding. + +--- + +## Decisions + +| decision | why | who | +|---|---|---| +| Match **jobs**, not endpoints | some Series-3 endpoints only exist to work around Series-3 quirks | agreed | +| **No Series-4 emergency tools** yet (runaway stop, rescue, erase) | Micromates rarely need them | Brian | +| The series comes from the **serial prefix**: `UM` = Micromate, `BE` = MiniMate Plus, `BA` = Blastmate | the prefixes are definitive, so there is no probing | Brian | +| Before any **write**, confirm the unit that answered is the one named | SFM addresses units by IP; a moved modem or stale IP means a different unit answers | agreed | +| **Synchronous** endpoints for now; async later, for both series together | parity first, then shared features | agreed | +| **No PC-side setup library** | a name is the only link between Thor's two copies, and nothing checks they match | Brian | +| **Sessions for Series 4 now; Series 3 only after a bench test** | THOR holds Micromate lines daily without harm; a MiniMate's hangup/DCD behaviour is untested | Brian | +| Audit log in **SFM's database** (new table) | Terra-View can show it; it fits the existing `CREATE TABLE IF NOT EXISTS` pattern | proposed | + +--- + +## Scope: the parity table + +Rows 8 and 11 of the original inventory (runaway stops, rescue/erase) are out +of scope by decision. + +| # | job | Series 3 in SFM today | Series 4: protocol status | Series 4 in this phase | +|---|---|---|---|---| +| 1 | Identify a unit | `/device/info` (walks the whole chain, slow) | implemented: `connect()` | dispatch to the Micromate client | +| 2 | Quick status (monitoring, battery, memory, clock) | `monitor/status`, `storage_range`, `events/index` | implemented: `get_state()`; the `0x06` event count is decoded only in the bench tool | same endpoints; move the count decode into the client | +| 3 | Read settings | compliance in `/device/info`; `GET /device/call_home` | `get_setup()` implemented and verified on hardware; call-home: enable + dial string observed, rest unmapped | setup in full; call-home returns enable + dial string + raw | +| 4 | List events | `/device/events` (downloads every event to list them) | implemented, and cheap: the chain walk returns sizes | dispatch | +| 5 | Download an event | `/event/{i}`, `/waveform`, `/blastware_file` | implemented: streams the `.IDFW`/`.IDFH` itself, with a decode self-check | save through the existing `/db/import/idf_file` path | +| 6 | Monitor log | only via `ach_server` (broken, see the fix list) | `0x0A` read exists; no wrapper or decoder | client wrapper + partial decode | +| 7 | Start / stop monitoring | `monitor/start`, `monitor/stop` | mapped, identical to Series 3 (`0x96`/`0x97`, acks `0x69`/`0x68`); never sent | **first write.** Verify with `0x49` state | +| 9 | Change call-home settings | `POST /device/call_home` | write mapped (`0x2C` → `0x7E`/`0x7F`, same as Series 3); field map mostly unknown | enable/disable only; the rest waits for the THOR round | +| 10 | Change recording settings | `POST /device/config` | sequence mapped (`0xDA` → `0x68/73` → `0x82/83` → `0x71/72`); no encoder | read-modify-write, stop → push → restart, verify with `get_setup()` | + +--- + +## Architecture + +### One door for every device request + +A new module, `sfm/device_access.py`, sits between the `/device/*` handlers and +the two device libraries. No handler opens a transport or picks a library +itself any more. + +``` +/device/* handler + │ (serial, address, action, tier) + ▼ +device_access ── series from serial prefix ── session / per-unit lock + │ │ │ + │ ▼ ▼ + │ minimateplus (S3) or micromate (S4) audit table + ▼ +read-back verification (changes only) +``` + +- **Series selection.** `serial` picks the library: `UM` → `micromate`; `BE` + or `BA` → `minimateplus`; any other prefix → 422. With no `serial`, the + request is Series 3, so existing Series-3 callers are unchanged. + **Series-4 requests must include `serial`.** +- **Endpoints.** The same `/device/*` paths, plus a `serial` query parameter. + Responses gain `"series": 3 | 4`. Where the two series genuinely differ (the + setup fields, the event identity), the body is series-specific under a + documented shape rather than forced into Series 3's. +- **Work separate from HTTP.** Each operation is a plain function taking the + session. The handler is a thin wrapper, so making an operation an async job + later means changing the wrapper, not the operation. + +### The safety layer + +**Tiers.** Every operation declares one: + +| tier | examples | what the layer does | +|---|---|---| +| read | info, status, setup, events, download | runs; no identity check | +| change | start/stop, call-home, setup push | identity check → audit → send → **read back** → audit the result | +| destructive | delete, erase (none in this phase) | everything above + the caller must send the unit's serial as `confirm` | + +**Identity check (changes only).** Before the first change in a connection, +compare the serial the unit reports with the one requested; on a mismatch, +refuse (409) and send nothing. Series 4 already reads the serial in +`connect()`. A Series-3 change adds one serial read (`SUB 0x15`), about one +round trip. + +**Opcode guard.** A send-side check that refuses known cross-series +collisions. It starts with **`0xA2`**, which is CONFIRM ERASE on Series 3 and +"end call-home" on Series 4. This phase never sends `0xA2`; the guard makes +that impossible by accident. The table of colliding opcodes lives in one +place. + +**One connection per unit.** A per-unit lock keyed by address +(`tcp:host:port` / `serial:port`). A request for a busy unit waits up to +**20 s** (configurable; under Terra-View's 60 s proxy timeout), then gets +**423 "unit busy"** instead of opening a second connection. This applies to Series 3 as well. The webapp's Diagnostics +refresh currently fires three device requests at once against a modem that +allows one; under the lock they queue. + +**Read-back verification.** A zero-data ack has never been proof that a write +applied. Every change is followed by a read: +- start/stop → `0x49` state: start expects `0x02`, stop expects `0x00`. Record + the value either way; `0x03` is the in-call-home state +- call-home → re-read `0x2C` and compare the edited bytes +- setup push → `get_setup()` and compare every field that was set + +A mismatch is reported as a failed change (and audited as one), even if the +ack was clean. + +**Errors that say what happened.** Distinct responses for: +- 503: couldn't connect +- 504: connected but the unit didn't answer +- 423: unit busy +- 409: wrong unit +- 422: bad input +- 502: the reply was malformed + +Today 502 covers both of the first two. + +**Audit table.** New table `device_commands`, created on startup like the +others. It is additive, but still a schema change, so the changelog says so. +One row per change attempt: + +| column | content | +|---|---| +| `ts`, `serial`, `series`, `address` | who and where | +| `action`, `tier` | e.g. `monitor_stop`, `change` | +| `request` | parameters, plus hex of what was sent | +| `ack` | reply summary | +| `verified` | read-back result: `ok` / `mismatch` / `not_checked` | +| `outcome`, `error` | final result | +| `caller` | best-effort origin (`webapp`, `terra-view`, `script`) | + +Reads are not audited. + +### Sessions (Series 4) + +1. **Call up.** Connect, `connect()` (POLL / SERIAL / STATE), identity check + against the requested serial. +2. **On the line.** Requests for that unit run in turn on the same connection, + behind the per-unit lock. +3. **Keep-alive.** A POLL every 60 s while idle. The ACEmanager TCP idle + timeout is 2 min. +4. **Hang up.** On `POST /device/hangup`, after 5 idle minutes, or on any + transport error. A reconnect is a new session with a new identity check. +5. **Ring safety.** If `RADIO RING` arrives on an SFM session, log it and hang + up **immediately**. Until SFM is the call-home destination, the unit's + retry should reach THOR. 🔶 *Inferred:* the unit retries; its `0x2C` + retries field reads 3. + +Series 3 keeps today's connect-per-request behaviour until the bench test +under *Testing*. + +### Series-4 write path + +**Encoder.** Extend `micromate/framing.py`: +- a request builder that carries a data section +- the same escape set `{02,03,04,10}`, applied to data, params and checksum +- plain SUM8 of the destuffed payload (*not* Series 3's write format) + +Rules: `micromate_protocol_reference.md` → *The escape set — exactly four bytes*, +*Checksum — plain SUM8 of the de-stuffed payload*, *Write-frame destuffing*. + +**Dry run.** Every Series-4 write accepts `dry_run=true`. It returns the exact +frames it would send and sends nothing. The first live send of each write is +reviewed this way first. + +**Per operation:** + +| operation | frames | verify | open questions it settles | +|---|---|---|---| +| start / stop | `0x96` / `0x97`, offset `0xFFFF`, zero params | `0x49` | the ack when already in that state | +| call-home enable/disable | read `0x2C` → `0x7E` (126 B content, offset `0x007E`) → `0x7F` | re-read `0x2C` | none new; bit 0 of `write[5]` | +| setup push | read `0x08`, `0x2E`, `0x1A` → `0xDA` (name) → `0x68`/`0x73` → `0x82`/`0x83` → `0x71`/`0x72` | `get_setup()` | the offset rule; whether `0x68`/`0x82` are needed | + +**Setup push, details:** +- Read-modify-write of **all three** blocks, as THOR appears to do. + 🔶 *Inferred from two captures:* `0x68` = the preferences content with + `[0:3]` zeroed; `0x82` = the `0x2E` content with `[0:3]` and the CRC slot + zeroed; `0x71` = the `0x1A` content `[0:2090]` with the length echo zeroed. +- Offsets come from THOR's observed constants. 🔶 The hypothesis "offset = + the read block's length + 2" fits all four, but each has been seen at one + size only. +- The edited fields come from `MicromateSetup` and are written back at the + offsets `setup_block.py` already uses. Only fields graded **measured** may + be edited. +- Sequence: stop monitoring (if running) → push → restart (if it was running) + → verify. THOR's "must be idle" is policy, not the device, but stopping + first removes a variable from the first live write. +- ⚠ **Edit-buffer hazard.** `0x1A` reads the keypad's unsaved edits, so a + read-modify-write taken while someone is in the menu would commit them. + This can't be detected from the wire; it is documented in the endpoint and + audited. +- **Target file:** the active setup, overwritten in place, by default. A + named target is allowed, but overwriting a *non-active* setup or + `factory.MMB` is untested, so it is refused in this phase. + +--- + +## Testing + +**Offline (CI-capable):** +- **Frame reproduction.** The new encoder must reproduce THOR's captured write + frames byte for byte: `0x96`/`0x97`, `0x7E`/`0x7F`, and the full setup push, + from the 9-24 / 9-25 / 10-14 captures. This is the same method that proved + the reads (218/218). The needed frames are copied into force-added fixtures, + since `bridges/captures/` is gitignored. +- **Safety layer on fakes:** + - series by prefix + - identity mismatch → 409, nothing sent + - `0xA2` refused + - lock → 423 + - read-back mismatch → failed change + - audit row written for every change attempt + - RADIO RING → session ends +- **Endpoints** through FastAPI's `TestClient` against the scripted transport + already used in `tests/test_micromate_client.py`. + +**Bench (UM20147, with Brian):** each write's first live send follows the same +protocol: +1. Dry run. +2. Review the frames. +3. Send once, with a capture running. +4. Read back. +5. Write the result into the reference. + +Order: stop/start → call-home off/on → setup push of one field → setup push of +several fields. + +**Series-3 session test (before Series-3 sessions are built):** hold a call-up +session open to a MiniMate that is monitoring, for several minutes. Record: +- does it keep monitoring (does an event still record)? +- after hangup, is it monitoring? +- does it behave differently on an RV55 with the DCD issue? + +--- + +## Build phases + +Each safety piece arrives with its first real use, so nothing sits untested. + +1. **Series-4 reads in SFM.** `device_access.py`, series selection, the + per-unit lock (Series 3 too), the error mapping, and Series-4 sessions with + ring safety. Rows 1–6. **No writes.** +2. **Encoder + opcode guard + dry run + frame-reproduction tests.** Still + nothing sent. +3. **Start/stop monitoring** (row 7): the first live write, at the bench. + Brings the identity check, the audit table and the read-back framework. +4. **Call-home enable/disable** (row 9). +5. **Setup push** (row 10): one field first, then several. + +Each phase is its own implementation plan and its own merge. + +--- + +## Out of scope (this phase) + +- Series-4 emergency tools: runaway stops, rescue, per-event delete, erase. + (Erase-all is also unmapped.) +- Async jobs; auth (Terra-View as jump host); the call-home receiver. These + are shared features for later. +- Scheduler read/write/enable; clock set; remote sensor check; restart; + firmware. +- Series-3 sessions, pending the bench test. +- Mapping the rest of `0x2C` (the THOR round). + +## Fix list (found 2026-10-19, separate small changes) + +- `bridges/ach_server.py:524`: `seen_keys` is undefined. Series-3 call-home + monitor-log collection has failed every session since `9afa348` + (2026-05-07). +- `bridges/mm_ach_server.py:262`: `_cstring`/`_content` aren't imported, so + the ring-answer path would raise `NameError`. From `8f67608` (2026-10-14). +- Series-3 `5A` 64 KB page-boundary bug (known; surfaces as HTTP 500). +- `POST /device/config` writes a fixed 29-byte `0x82` blob from one 3-11-26 + capture, whatever the unit's trigger settings. +- `/device/events` runs a full 5A download per event just to list headers. +- `0x08` is named `read_event_index` in `micromate/`; it is the preferences + block. + +## Table 2: to be added later (both series) + +| area | item | +|---|---| +| infrastructure | async jobs; auth via Terra-View | +| THOR has it, SFM doesn't | clock set; remote sensor check; scheduler; one setup to many units; restart | +| call-home receiver | hold lines open (S4); "already downloaded" tracking; rules by serial prefix (copy events, monitor log, delete, set clock); detect "Monitor While Calling Home" blocking deletes (⚠ one unexplained 21 → 5 count drop in THOR's log); hand off to `/db/import/*` instead of the watchers | +| settings, the no-PC-files way | intended settings per deployment; drift report (intended vs actual); push then verify | +| notifications | email/SMS by event type and PPV (THOR does it; may belong in Terra-View) | +| emergency tools for S4 | if a Micromate ever needs them | + +## Parking lot + +**A modem-side agent on the RX55.** AirLink +OS on the RX55 has "Container Applications (Early Adopter)", and Brian already +runs `slmm-proxy` there. A small program owning the unit's serial port could: +- let THOR and SFM share the line +- dial out to SFM, avoiding the IP whitelist +- handle `RADIO RING` locally +- **Brian's hunch:** make RX55s work with Series 3 by driving DTR/DCD itself + +Open questions: +- Can a container open the RS-232 port? +- Can it set the control lines? +- How is it deployed to a fleet, and does it Autostart? +- Will a firmware update keep the beta feature? + +## Open questions this design leaves for the bench + +- The ack for start/stop when the unit is already in that state. +- The offset rule, and whether `0x68`/`0x82` are required, for the setup push. +- Whether the device ever refuses a write while monitoring. Only THOR's policy + has been seen. +- How a failed write is signalled. It has never been observed. +- MiniMate behaviour with a held session (above).