docs: phase 5 benched on both series (2026-11-04)

Both protocol references, the spec's bench section and CLAUDE.md record
the bench: every settings change acknowledged, read back and confirmed on
the unit's screen, both units restored byte for byte; the escaped-
checksum stray byte (fixed); a MiniMate's open setup menu writes its
whole-block copy back on Cancel or Save, losing a remote change; the
alarm write at channel +38; [26:28] re-set by the unit while monitoring;
the first Micromate setup push with the unit's CRC equal to SFM's; the
Micromate scheduler switch in none of the blocks SFM reads.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-11-04 16:16:40 -05:00
1 parent c790e20116
commit 3564c4a744
4 files changed
+79 -4

No files matched your search

+33
View File
@@ -2745,6 +2745,39 @@ were wrong.
established** — only one setting was varied. Series III's offsets are a starting
hypothesis, not a transfer.
#### Bench 2026-11-04: SFM's first setup push (UM20147)
SFM's audited `POST /device/settings/change`, one send per go (Brian at the
unit). Notes: `~/bench-logs/2026-11-04-sfm-settings/`.
**Measured:**
- **The setup push works as THOR's.** Trigger 0.6 → 0.5, then a five-field
change (Location, Client, job number + its report flag, record time), then a
full restore. Each send was THOR's seven frames, acknowledged in THOR's order
(`25 97 8C 7D 7C 8E 8D`), read back as sent, and confirmed in the unit's menu.
The setup block ended byte for byte as found, CRC included.
- **The unit's recomputed CRC equals SFM's** (`with_crc`): after the first push
the stored block equalled SFM's computed image byte for byte, CRC `ea8d`
included.
- **`schedule.dat` read on a real unit** (`0x94` → `0x6B`; `0x48` → descriptor
`04 00 … 01 … 01`, body, all-zero end). UM20147 holds a leftover THOR
schedule: repeat on, 07:00, start monitoring with
`Fay - Commercial Street - Irish Center.mmb`. The unit runs `test2.MMB`, so
SFM blocked the change, as designed.
- 🔑 **The scheduler On/Off switch is in none of the blocks SFM reads.** With
the keypad switch Off, then On (saved), these were byte for byte identical:
`0x08` prefs, `0x2E`, `0x1A`, `0x2C`, `0x41` and `schedule.dat`. The
firmware's `SysPref.bMonitorScheduler` lives elsewhere. THOR's end-of-push
`0x47` pair (`params[7]` 1 then 3; replies `04 00 … 0N … 00 01`) was only
ever captured with the switch On. **Next:** a THOR bridge recording with the
switch Off (parked for the work-hours project).
- **While monitoring**, a settings preview and a live attempt were both refused
on the run state before any block was read (409, nothing sent, 0 write frames).
- **With call-home off, a stop goes straight to idle** (`monitoring` →
`idle`): no calling-home state and no ring.
- Call-home `0x01` ↔ `0x00` was taken again, both ways. The plan's
"unobserved" warning is retired for those values.
## What the THOR manual settles about the ACH session (2026-09-24)
Source: `manuals/723U0201 THOR Operator Manual Rev 08.pdf` §6.2, §3.6.