docs: phase 5 benched on both series (2026-11-04)

Both protocol references, the spec's bench section and CLAUDE.md record
the bench: every settings change acknowledged, read back and confirmed on
the unit's screen, both units restored byte for byte; the escaped-
checksum stray byte (fixed); a MiniMate's open setup menu writes its
whole-block copy back on Cancel or Save, losing a remote change; the
alarm write at channel +38; [26:28] re-set by the unit while monitoring;
the first Micromate setup push with the unit's CRC equal to SFM's; the
Micromate scheduler switch in none of the blocks SFM reads.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-11-04 16:16:40 -05:00
1 parent c790e20116
commit 3564c4a744
4 files changed
+79 -4

No files matched your search

@@ -1,7 +1,8 @@
# Phase 5: changing a unit's settings by serial (design)
**Date:** 2026-11-03. **Status:** approved by Brian 2026-11-03; built 2026-11-03
(plan `docs/superpowers/plans/2026-11-03-sfm-phase5-settings.md`); bench pending. Phase 5 of
(plan `docs/superpowers/plans/2026-11-03-sfm-phase5-settings.md`); **benched 2026-11-04** on both
series (results at the end). Phase 5 of
`docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md` (row 10, "change
recording settings"; build phase 5, "setup push: one field first, then several").
**Branch:** `feat/series4-sfm-parity`.
@@ -342,6 +343,37 @@ intervals 2/5/15/60/300/900 s, and each series' measured mode codes.
3. A several-field "backup unit" change.
4. A dry run while monitoring, expecting the refusal.
## Bench 2026-11-04 (results)
Brian at BE9441 and UM20147, each on its own modem; every send with his go.
Notes: `~/bench-logs/2026-11-04-sfm-settings/`.
- **Both series change correctly.** Each change was acknowledged, read back as
sent, and confirmed on the unit's own screen; both units ended byte for byte as
found.
- BE9441: trigger each way, a note, several fields at once, the alarm, a full
restore.
- UM20147: trigger, several fields (two notes, job number and its flag, record
time), a full restore. The unit's recomputed CRC equalled SFM's.
- **Fixed at the bench (`9992b25`):** the first live MiniMate read came back one
byte long. An escaped checksum byte leaves its `0x10` behind (about 1 frame in
85; the recordings never had it). The read now takes exactly the lengths it
asked for, and so does the waveform read (`c790e20`).
- **🔑 The keypad menu wins.** A MiniMate change sent while its setup menu is
open is acknowledged and read back, then lost when the operator presses Cancel
or Save: the menu writes its copy of the whole block back. SFM cannot see this
happen; the MiniMate preview now warns. Rule for the field-plus-office routine:
nobody in the unit's menus while the office sends.
- **The MiniMate alarm write is measured**, so it is now writable (`55c7b7b`).
- **Refused while recording, both series, nothing sent.** A recording MiniMate
sometimes leaves a preview at "no answer" rather than the refusal.
- **The Micromate scheduler switch is in no block SFM reads.** UM20147 carries a
leftover THOR schedule (Fay setup, 07:00 daily) with the switch Off; SFM blocks
it, a false alarm here. The bench used a scratch-only override.
- Next: a THOR bridge recording to find the switch (parked for work hours).
- Meanwhile, worth considering: let the operator confirm "scheduler off" in the
preview.
## Calls I made (Claude), for Brian to check
- **Series 3 `[26:28]` goes out as `00 00`, as Blastware writes it**, rather than