diff --git a/docs/instantel_protocol_reference.md b/docs/instantel_protocol_reference.md index f4b118f..8a8115e 100644 --- a/docs/instantel_protocol_reference.md +++ b/docs/instantel_protocol_reference.md @@ -3536,26 +3536,65 @@ steps. A small common-mode diurnal wobble rides on top of it, but that wobble is present on the healthy channels too (it is the unit's normal thermal breathing) and is not the fault. -**Open: two candidate mechanisms.** +**The fault is on the geophone side.** Operator report (2026-08-25): +*attaching a different geophone to an affected unit makes the offset go +away.* That is the channel-swap test, already run in the field many +times, and it rules out the unit's analog front-end and any stored +per-channel zero constant — a constant lives in the unit and would +survive a sensor swap. -1. *A latched bad zero.* These units run Sensor Check "Before - monitoring", so a baseline is captured at session start. Disturb the - geophone at that instant and the unit bakes a non-zero reading in as - "zero". Fits the discrete steps, the rock-steadiness between them, - surviving power cycles, the sensor check still passing, and clearing - on a clean re-zero. Also explains why it is episodic rather than - constant — it needs a disturbance at the moment of zeroing. -2. *A degrading analog front-end.* Fits BE9558 better: it was clean for - nine months, then escalated 3 → 1 → 1 → 193 events over five weeks - **and moved from Tran to Long**, which a single bad channel would - not do. +The stored data agrees. MicL — a separate transducer on its own cable — +shows no offset during either episode (`|mean|/peak` = 0.17 and 0.02), +while the geo channels on the same unit at the same moment are pinned. -BE18438 (same channel twice, months apart) looks like (1); BE9558 -(escalating, channel-migrating) looks like (2). The decisive test is to -force a clean re-zero on a faulted unit: if the offset clears and stays -clear it is (1) and fixable in the field — potentially over the wire, -since we already speak SUB `0x0E` (channel sensor data) and `0x98` -(trigger test), see §7.10. +**Two distinct failure patterns**, both geophone-side: + +| unit | Tran | Vert | Long | MicL | reading | +|---|---|---|---|---|---| +| BE18438 | 0.16 | **0.97** | 0.18 | 0.02 | one element only | +| BE9558 | 0.90 | 0.81 | **0.99** | 0.17 | all three, Long dominant | + +(values are `\|mean\|/peak` on the dominant axis) + +BE18438 has a single faulted conductor pair with the other two clean. +BE9558 has all three geo channels offset by different amounts +(Long +0.340, Tran +0.037, Vert −0.020) — that points at a shared return +or ground path in the housing rather than one element. + +**Candidate mechanisms.** A geophone coil is passive and cannot +generate sustained DC on its own, so an offset originating on the sensor +side has to come from one of: + +1. *Galvanic corrosion at a connector or splice* — dissimilar metals + plus moisture form a literal cell. This generates DC directly, in + the tens-of-millivolts range that matches the measured offset + (0.29 in/s ÷ 6.206053 in/s per volt ≈ 46 mV referred to the ADC + input). Fits the discrete steps, surviving power cycles, and being + cured by swapping the sensor. +2. *A leakage path to shield or another conductor*, letting the unit's + own bias/reference divide into the signal pair. Best fit for the + BE9558 all-three-channels pattern. +3. *Changed coil DC resistance* (partial short or degraded winding) + interacting with the amplifier's input bias current — + `V_offset = I_bias × R_source`. Best fit for the BE18438 + single-element pattern. Note the on-device sensor check would not + catch this: it measures ring-down frequency and damping, not DC. + +**Caveat on the swap test.** Swapping a geophone requires stopping and +restarting monitoring, and these units run Sensor Check "Before +monitoring" — so the restart also re-zeros the channels. The swap +therefore does not cleanly separate *"new sensor"* from *"the restart +re-zeroed it"*. Two cheap controls settle it: restart monitoring +**without** swapping (if the offset clears, it was the re-zero), or swap +the **original** sensor back (if the offset returns, it is the sensor). +The operator's remedy being a swap rather than a restart is weak +evidence for the sensor, since a restart is the easier thing to try +first. + +**Best single measurement when it recurs:** DC voltage across the +suspect geophone's connector pins with the sensor **disconnected from +the unit**. A galvanic cell (mechanism 1) shows an open-circuit +voltage; mechanisms 2 and 3 do not. **Note:** the surveyed population is subject to survivorship bias — flooded events were routinely deleted before this snapshot, so units