fix(decode): one MicL count is 0.25 Pa -- the mic psi was ~95x low

SFM converted mic counts to psi with a '~0.5 psi full-scale' guess
(0.0125 / 32768 psi per count).  Measured against 1,729,253 DL2
histogram intervals' 'MicL Peak' columns: one count is exactly 0.25 Pa.
count x 0.25 Pa matches Blastware's psi(L) on every interval, and
20 log10(count x 0.25 Pa / 20 uPa) its dB(L) on all but 17 exact
half-way cases.

- event_file_io: _MIC_FS_PSI -> _MIC_PSI_PER_COUNT = 0.25 / 6894.757
  (3.626e-5 psi per count, was 3.81e-7).  The event's mic peak in psi,
  and with it the .h5 mic trace (event_hdf5 back-solves its scale from
  that peak), come right on events with no report.
- waveform_codec.mic_count_to_db: the offset is 20 log10(0.25 / 20e-6)
  = 81.938, not 81.94 -- the old value missed 49 of those intervals at a
  rounding edge (count 119 printed 123.4, 81.94 gave 123.451).

The decoded count was always right: count 16 on K441LTON.900 is
Blastware's 106.0 dB.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-10-31 01:28:23 -04:00
1 parent dc42ae041a
commit 5644dbd5ad
5 files changed
+56 -18

No files matched your search

+14 -9
View File
@@ -628,23 +628,28 @@ def decoded_to_adc_counts(decoded: dict) -> dict:
}
# One MicL count is 0.25 Pa (measured 2026-10-31): count x 0.25 Pa matches
# Blastware's "MicL Peak psi(L)" on all 1,729,253 DL2 histogram intervals
# checked, and 20 log10(count x 0.25 Pa / 20 uPa) its dB(L) on all but 17
# exact half-way cases. The dB offset below is that, 81.938 -- not the 81.94
# first fitted to a handful of values, which missed 49 of those intervals.
MIC_PA_PER_COUNT = 0.25
_MIC_DB_REF = 20.0 * math.log10(MIC_PA_PER_COUNT / 20e-6)
def mic_count_to_db(count: int) -> float:
"""Convert a MicL ADC count to dB(L) for BW-display-compatible output.
Empirical formula (confirmed 2026-05-11 against V70 fixture: count=813
→ 140.1 dB; count=±1 → ±81.94 dB; count=±24 → ±109.5 dB):
dB = sign(count) × (20 log10(0.25 Pa / 20 µPa) + 20 × log10(|count|))
= sign(count) × (81.938 + 20 × log10(|count|)) for |count| ≥ 1
dB = 0.0 for count == 0
dB = sign(count) × (81.94 + 20 × log10(|count|)) for |count| ≥ 1
dB = 0.0 for count == 0
The constant 81.94 corresponds to 10^(81.94/20) ≈ 12490 mic ADC counts
being the dB(L) reference level — almost certainly a calibration
constant from the device's mic.
count=±1 → ±81.94 dB; count=813 → 140.14 dB (the V70 fixture).
"""
if count == 0:
return 0.0
sign = 1.0 if count > 0 else -1.0
return sign * (81.94 + 20.0 * math.log10(abs(count)))
return sign * (_MIC_DB_REF + 20.0 * math.log10(abs(count)))
# ── A5-frame entry point ────────────────────────────────────────────────────