docs(offset): mechanism campaign — five hypotheses dead, onset is a ramp

Records the mechanism investigation in journal 8c. The headline: still unknown,
but the shape is now constrained and a long list of dead ends is closed.

Onset is a ramp of minutes-to-hours, not a step — BE18438 Vert resolved to
one-minute cadence via the histogram corpus, 50% of the excursion in 7 minutes,
>=25 intermediates, validated 75/75 against Blastware's own ASCII. That kills
both poles of the original dichotomy: not a latched digital step, not slow
component wear. What survives is a reversible two-time-constant settling
process, which is a shape constraint and not a mechanism.

Thermal, ground-motion shock, handling/redeployment, accumulated duty, age,
firmware and a mechanical element fault are each refuted or explicitly bounded,
with the power behind every negative stated.

Retracts two claims this journal carried: polarity consistency was a tautology
of offset_scan3's spread gate, and the fleet is 8-9 units rather than 5 once
that gate is dropped. Also notes the gate is blind to onsets by construction --
it rejects a moving floor, and it rejected the one record where the ramp shows.

Co-Authored-By: Claude Opus 5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01HgTe8CamXAHcAmaQ6QNcog
This commit is contained in:
2026-09-06 09:33:12 +00:00
co-authored by Claude Opus 5
parent a27310c3c6
commit 58c1fe8a96
+189
View File
@@ -662,6 +662,191 @@ BE9557|Tran 2021-06-28, BE9557|Vert 2025-06-12.
---
## 8c. Mechanism — five hypotheses tested, all dead (2026-09-06)
**The mechanism is still unknown.** Five campaigns, ~105 effectively independent
tests, seven nominally significant results against **5.2 expected by chance**
under a global null. Every one died to its own confound analysis. What the
campaign bought is a set of *shape constraints* and a long list of dead ends.
### ⚠ Two things retracted from this journal
**1. "Polarity is perfectly consistent — 11 of 11, zero mixed cases."** That is
a **tautology of the spread gate**, not a property of the fault. `spread <= 0.02`
requires pre/mid/end to agree, which forces one sign. Amplitude-only at the same
0.025 threshold: **12 of 53 unit-channels are mixed**, including BE18438|Vert
(88+/1−) and BE9558|Vert (1+/35−). Withdrawn.
**2. "5 of 45 units, unchanged across a 2x threshold range."** The
threshold-insensitivity is also a property of the gate. Amplitude-only gives
**9 units at 0.020, 8 at 0.025** (adding BA10895, BE12844, BE18003), 5 at 0.040.
The fleet is **8–9 units, not 5**.
**3. "Persistent — it stays until the geophone is serviced."** Weakened, not
withdrawn. There are **23 recoveries after runs of >=3 flagged events, median
gap 6.03 days**, three inside ten minutes. BE18438|Vert reads `pre=mid=end=
+0.0000` on 2026-02-10, +0.185→+0.370 across 02-25/26, and `+0.0000` again on
2026-03-22 — identical Project, Seis Loc, calibration date, geo range and
trigger throughout. The one thing that cannot be excluded is a **field
autozero**: it is a button sequence at the unit and writes nothing into the
event header. So "persistent" may be "persistent unless somebody pressed the
buttons," and the archive cannot tell those apart.
### The one positive finding: onset is a RAMP, minutes to hours
Both onsets resolvable at minute cadence are ramps. **BE18438|Vert,
2026-02-20** — the histogram corpus collapses a 14 d 21 h waveform bracket to
**one minute**:
```
~14,200 consecutive quiet minutes at 0.000–0.005 (ten full daily files)
09:32 +0.005 09:39 +0.045 10:20 +0.125 16:00 +0.165
09:33 +0.010 09:42 +0.070 13:13 +0.150 20:17 +0.185 plateau
```
**50% of the excursion in 7 minutes**, the rest asymptotic over ~10 h, **>=25
distinct one-minute intermediates**. Validated **75/75** against Blastware's own
ASCII export. Its 2025-11-15 onset is the same shape over 2.7 h. BE13117 stage B
is a 91-minute monotone rise, +0.035 → +1.745 in/s over ~40 samples.
**This kills both poles of the original dichotomy** (journal Q1): not an
instantaneous latched step (a bad autozero, a stuck trim-DAC), and not slow
component degradation over days or weeks.
⚠ It rests on **2 of 45 instruments**. Clopper-Pearson on 4/4 resolved onsets
gives 95% CI [0.40, 1.00] — a mixed population with up to 60% true steps is not
excluded. BE13117 has zero paired ASCII, so its ramp rests on our decoder alone.
### The methodological corollary — more important than the finding
**A waveform-only bracket manufactures the appearance of a step, and the spread
gate is blind to onsets by construction.**
The offset is what fires the trigger, so no waveform event can exist until the
ramp has nearly reached the trigger level. BE18438's first event of each episode
sits at 0.280 against a 0.300 trigger, and 0.185 against 0.200. At daily cadence
against a 3 h ramp, P(catching an intermediate) = **0.125**.
And `spread <= 0.02` rejects any record in which the floor is *moving* — which
is exactly what an onset is. **The gate rejected the very BE18438 record where
the ramp is visible.** If the operational goal is catching a fault early, before
the unit floods the store with junk events, the current detector is the wrong
shape for the job.
### The surviving shape
An **electrical, reversible, two-time-constant settling process** (~10 min and
~hours), saturating at a ceiling, with occasional sub-3-minute discrete jumps
superposed (BE18438 2026-02-26: 13:24 pre +0.180 / mid +0.240 / end +0.255 →
13:27 +0.325, identical metadata). That is the signature of a **bias or leakage
path charging a high-impedance node** — the class of fault Instantel's autozero
recovers ~10% of the time, and what the X1/X8 gains measure.
**It is a shape constraint, not a mechanism. Do not write it up as one.**
### Dead — with the evidence, so none of this is re-derived
| Killed | Evidence |
|---|---|
| **Latched step at onset** | >=25 one-minute intermediates over ~10 h, ASCII-validated. Direct observation, not a test. |
| **Slow degradation over days/weeks** | Same observation — bulk of the excursion in 7 min to 2.7 h. |
| **Thermal driving of pedestal magnitude** | BE13117, 365-count pedestal, n=128: full-day modulation **−0.42% ± 0.42%**, 95% CI [−1.25%, +0.40%]. Healthy-fleet seasonal zero drift totals **~0.3 A/D counts** — 15x to 1200x too small. Best-powered result in the campaign. |
| **Ground-motion shock** | 30-day window-max percentile ranks 0.03/0.98/0.15/0.01/0.68/0.24, median **0.194** against a null of 0.5. **0 of 7 events >=9 in/s** was followed by an onset within 30 d. BE12599 hit 10.220 in/s (2023-11) and 10.005 (2025-04) and did not onset until 2026-08-14. |
| **Handling / redeployment** | **0 of 9** onsets had a Project/Client/Seis Loc change. Widened to 30 d: 2 observed vs 4.90 expected, P(X>=2)=0.995 — *depleted*, the wrong direction. The apparent gap effect (p=0.035) died on histogram coverage: BE18438's "59.7-day gap" contains 122 histogram files; true silence 0.52 d. |
| **Mechanical resonance / damping change** | BE18438|Vert at a 64-count pedestal (3x outside Instantel's ±21): ΔTest-Freq **CI [−0.090, +0.021]** against 0.127 Hz for a real calibration. Block permutation p=0.658. |
| **Accumulated-duty threshold** | ~4 clean units logged more monitoring than the largest positive onset dose; BE18193 logged **13.45M intervals, 6.2x**. A counterexample — no power argument weakens it. |
| **Firmware** | **14,338 of 14,340** exports read `V 10.72-8.17`. A constant cannot explain a variable. |
| **Unit age** | Serial rank-sum 118.0 vs null 115.0, p=0.549; unchanged on the 8-unit re-cut (p=0.586). Serial is a poor age proxy anyway (Spearman +0.113 against archive entry). |
| **Strong seasonal clustering** | 25 onsets, exposure-weighted permutation **p=0.59**. Excludes >=75%-in-one-season only; a 2x seasonal hazard is *not* excluded. |
Also retire two overstated bounds. H6's dose-response exclusion "|r| > 0.03" is
a **10x overstatement** once clustering is corrected — the honest bound is
|r| > 0.1–0.3, so a real r=0.2 is not excluded. And **any statistic quoted
per-event**: 512 flagged channel-events collapse to **4.9 effective independent
observations** (unequal-cluster design effect 104.6 at ICC=1), and **55% of the
flagged corpus is one instrument on two calendar days** (BE13117, 2023-05-03/04).
### Power — read every negative in this section as bounded
Fisher exact, 5 positives of 45, one-sided α=0.05, exposure a third of the fleet:
| relative risk | power |
|---|---|
| 1.5 | 0.059 |
| 2 | 0.112 |
| 3 | 0.231 |
| 6 | 0.497 |
| 15 | 0.753 |
80% power needs **RR ≈ 13–20**. Even a *perfect* split reaches p<0.05 only if
the exposed group is <=25 of 45 units. **This archive can detect only
near-deterministic unit-level causes.** Every negative above excludes a strong
effect, not a real one.
### What this archive can NEVER answer
- **The A/D zero and the X1/X8 gains.** The 2027–2069 numbers appear in no file,
header or decoded record. They exist only on a live device behind `SUB 0x0E`.
Q1 is structurally unanswerable from data.
- **Unit-level vs component-level cause.** **Zero of 14,340** exports carry a
geophone or sensor serial. Q4 is dead — there is no way to know whether the
same physical geophone came back after service.
- **Service history.** The only service-adjacent field is `Calibration: <date>`
— 30 distinct dates fleet-wide, none before 2023, ASCII corpus entirely
2025–26. BE9558's 2020 and BE13117's 2023 episodes have no calibration record.
- **Temperature.** Zero exports carry it. Battery Level is a verified coarse
thermometer (+0.204 V winter over summer, 20/20 unit-years, p=9.5e−7, matching
lead-acid tempco) but quantised at 0.1 V ≈ 10 °C — useless within a day. The
archive can *bound* thermal; it can never *test* it.
- **BE13117 specifically** — 55% of the flagged corpus, the largest pedestal at
1.92 in/s, **zero** ASCII exports, histogram record ending eight months before
its episode. The most informative case in the archive is permanently outside
every metadata test.
- **The mild-offset rate**, and therefore the base rate's denominator. Event
files only see offsets large enough to dominate the trace.
### The experiment to run — `SUB 0x0E`, one afternoon
Point Blastware at `bridges/ach_mitm.py` and run **Unit Channel Test** against
(1) a faulting unit, (2) a known-good control, (3) the same unit before and
after an autozero. BW's sequence is `0x0E x8 → 0x98 x2 → 0x0E x8`, the second
pass carrying live ADC. Eight 10-byte payloads with expected values near 2048 is
a very constrained puzzle.
- **Proves:** whether the X1/X8 gains are readable over the wire, and whether
the fault sits at or upstream of the ADC zero reference. Gains walk out of
2027–2069 with the pedestal → the fault *is* the zero reference, Q1 answered.
Gains hold while the trace moves → the fault is downstream, look at the front
end.
- **§8c hands it a falsifiable time course:** poll at ~1-minute cadence and the
numbers should **ramp over minutes-to-hours, not step**. If they step while
the trace ramps, the two are decoupled.
- **Payoff:** converts the 10%/90% ship-it-or-not gamble into a decision made
before packing a box, remotely, for the whole fleet.
- ⚠ In the MITM topology filenames are reversed — `raw_s3_*.bin` holds
Blastware's bytes.
**Second: swap the geophone** between a faulted base and a healthy one. Fault
follows the sensor → element or cable. Fault stays with the base → front-end
board. One afternoon, zero code, and it settles the one question the archive is
permanently blind to.
**Third: log a faulting unit for 72 h untouched.** Every recovery we have is
confounded by a possible field autozero. A shelf and a logger settles whether
the fault genuinely self-reverses.
**Fourth, free: re-cut the fleet label** — drop the spread gate, re-score
amplitude-only, screen the 11 unscreened thin-coverage units. Might reach 9–10
positives. Be honest about the gain: power against "older half carries 3x the
hazard" rises only 0.23 → 0.30.
**Highest-value item overall, and not an experiment: the RMA/repair records.**
Which unit went back, when, what was done (autozero vs geophone replaced vs
board), and the geophone serial fitted. "Same channel after a documented
geophone *replacement*" is component-level-negative in one observation.
---
### ⚠ Serial prefixes — four of these units are BlastMates, not MiniMates
Corrected 2026-09-06, after Brian queried "BA10895?" against a report that
@@ -707,3 +892,7 @@ decoder lost the geo threshold along with the serial. See commit `9ceff65`.
| 2026-09-04 | **BA10895 reclassified** from transient to a genuine Vert fault of a different subtype — 49.4% single-axis-dominant events, the highest in the fleet, 100% on Vert. The other six marginal units are clean. |
| 2026-09-04 | **Defect found in `offset_scan3.py`**: its `spread <= 0.02` gate discards 18.8% of rows with \|pre\| >= 0.025, concentrated on 41 negative unit-channels; 4 would be sustained positives without it. The fleet label is three-state, not two. |
| 2026-09-06 | **Four units relabelled BA, not BE** — BA9229, BA10060, BA10895, BA15957 are BlastMates. The BW filename carries only the serial number; the family prefix must be read from the file body. Fixed in the scanners and in two production sites. |
| 2026-09-06 | **Mechanism campaign — five hypotheses, all dead.** Thermal, ground-motion shock, handling/redeployment, accumulated duty, unit age, firmware and a mechanical element fault are each refuted or bounded. 7 nominally significant results against 5.2 expected by chance. |
| 2026-09-06 | **Onset is a RAMP of minutes-to-hours, not a step** — BE18438 Vert resolved to one-minute cadence, 50% of the excursion in 7 min, >=25 intermediates, ASCII-validated 75/75. Kills both a latched digital step AND slow component degradation. Surviving shape: a reversible two-time-constant settling process — a bias/leakage path charging a high-impedance node. |
| 2026-09-06 | **Polarity consistency RETRACTED** (a tautology of the spread gate; amplitude-only gives 12 of 53 unit-channels mixed) and the fleet **re-cut to 8–9 units, not 5**. "Persistent until serviced" weakened: 23 recoveries, median gap 6 days — though a field autozero cannot be excluded. |
| 2026-09-06 | The spread gate is **blind to onsets by construction** — it rejects a moving floor, which is what an onset is. It rejected the very record in which the ramp is visible. |