feat(decode): a Blastware file's identity, battery and trigger settings

Group A of the file-only decode, every field matched against every DL2
report (2026-11-03):

- Identity block before the setup name, what the live read gets from
  SUB FE: serial k-202, firmware k-147, DSP version k-138, calibration
  date k-132 + 'by' k-128, manufacturer k-94, model k-72, firmware
  minor k-193.  Event.device_info (DeviceInfo gains calibration_date /
  calibrated_by).  Serial, Version and Calibration 69,983 / 69,983.
- Battery: uint16 hundredths of a volt at k-190 (Event.battery_volts).
  69,729 exact; the other 254 are Blastware rounding half up.
- Trigger source: the two bytes between the embedded event record's
  lead and fe 80, a bitmask (0x01 Tran 0x02 Vert 0x04 Long 0x08 MicL
  0x200 Manual): 6,493 / 6,493.  Event.triggered_by.
- Trigger settings in Series IV's 0x1A layout: sources bitmask k-19,
  stop mode k-31 (0 = Fixed), each channel's float32 level at
  label+30 (geo in/s, mic psi), mic type code at MicL+4 (2 = Linear
  Weighting).  Geo level 6,447 / 6,447, mic 4,403 / 4,403; Blastware
  prints a level when its channel is a trigger source and the event
  was not Manual.  The mic level's dB uses 6895 Pa/psi (4,272 / 4,272
  against 4,271 at 6894.757).
- .h5 attrs for all of them (additive).
- Checker: SFM's side reports the ten fields; the mic trigger level
  follows the report's dB/psi.  Carve-out
  units-is-a-display-preference: dB or psi is Blastware's PC setting
  when it wrote the report -- DL2's reports switch fleet-wide in 61
  clean runs by write time, and no byte of the file separates them.
  Two tests that pinned BatteryLevel / Calibration as blank now pin the
  match.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-11-03 13:38:39 -05:00
1 parent be5bab4787
commit 6c2eda5f63
14 files changed
+8384 -15

No files matched your search

+49 -3
View File
@@ -13,6 +13,7 @@ Pure: no I/O, and no import from `sfm/`.
"""
from __future__ import annotations
import math
import re
from dataclasses import dataclass
from typing import Callable, Optional, Sequence
@@ -30,6 +31,50 @@ GEO_IPS_PER_COUNT: float = _GEO_NORMAL_FS_INS / _GEO_INT16_FS
_CHANNELS = ("Tran", "Vert", "Long", "MicL")
_EVENT_TYPE = {"Waveform": "Full Waveform", "Histogram": "Full Histogram"}
# How Blastware prints the setup fields SFM reads (measured 2026-11-03).
_MONTHS = ("January", "February", "March", "April", "May", "June", "July",
"August", "September", "October", "November", "December")
_TRIGGERED_BY = {0x0001: "Tran", 0x0002: "Vert", 0x0004: "Long", 0x0008: "MicL", 0x0200: "Manual"}
_STOP_MODES = {0: "Fixed"}
_MIC_TYPES = {2: "Linear Weighting"}
_GEO_TRIGGERS, _MIC_TRIGGER = 0x07, 0x08
# Blastware turns the mic trigger level from psi to dB with 6895 Pa per psi, not
# 6894.757: 4,272 / 4,272 DL2 levels against 4,271 (the odd one 0.01 psi,
# printed 130.8 dB(L), is 130.7498 at the exact constant).
_TRIGGER_PA_PER_PSI = 6895.0
def _setup_values(ev: Event) -> dict[str, object]:
out: dict[str, object] = {}
di = ev.device_info
if di is not None:
out["SerialNumber"] = di.serial
fw = di.firmware_version or ""
if di.dsp_version and fw.startswith("S33") and di.model:
# "V 10.72-8.17 MiniMate Plus" from "10.72", "S338.17" and the model --
# the only version in DL2, so the rule is proven for it alone.
out["Version"] = f"V {di.dsp_version}-{fw[3:]} {di.model}"
if di.calibration_date and di.calibrated_by:
c = di.calibration_date
out["Calibration"] = f"{_MONTHS[c.month - 1]} {c.day}, {c.year} by {di.calibrated_by}"
out["BatteryLevel"] = ev.battery_volts
out["Trigger"] = _TRIGGERED_BY.get(ev.triggered_by)
out["RecordStopMode"] = _STOP_MODES.get(ev.record_stop_mode)
out["Microphone"] = _MIC_TYPES.get(ev.mic_type)
# dB or psi is Blastware's preference, not the unit's (carve-out
# units-is-a-display-preference); SFM reports dB, the default.
out["Units"] = "in/s and dB(L)" if ev.mic_type is not None else "in/s"
# Blastware prints a trigger level only for a channel that is a trigger
# source, and none for a Manual trigger.
src, manual = ev.trigger_sources or 0, ev.triggered_by == 0x0200
if src & _GEO_TRIGGERS and not manual:
out["GeoTriggerLevel"] = ev.trigger_level_geo
if src & _MIC_TRIGGER and not manual and ev.trigger_level_mic_psi:
psi = ev.trigger_level_mic_psi
out["MicTriggerLevel psi(L)"] = psi
out["MicTriggerLevel dB(L)"] = 20 * math.log10(psi * _TRIGGER_PA_PER_PSI / 2e-5)
return out
@dataclass
class SfmSide:
@@ -72,6 +117,7 @@ def sfm_side(raw: bytes) -> SfmSide:
if counts["MicL"]:
values["MicLPSPL dB(L)"] = mic_count_to_db(max(abs(c) for c in counts["MicL"]))
values["MicLPSPL psi(L)"] = pv.micl if pv is not None else None
values.update(_setup_values(ev))
return SfmSide(
event=ev,
@@ -144,10 +190,10 @@ def _mic_converter(ref: XmlReport) -> Optional[Callable[[int], float]]:
def _compare_scalar(tag: str, printed: str, sfm: SfmSide, ref: XmlReport) -> FieldResult:
if not printed.strip() or printed.strip() == "N/A":
return FieldResult(tag, "no_reference", printed, None)
if tag == "MicLPSPL":
if tag in ("MicLPSPL", "MicTriggerLevel"):
units = ref.event_units or ""
key = ("MicLPSPL dB(L)" if "dB(L)" in units else
"MicLPSPL psi(L)" if "psi(L)" in units else None)
key = (f"{tag} dB(L)" if "dB(L)" in units else
f"{tag} psi(L)" if "psi(L)" in units else None)
value = sfm.values.get(key) if key else None
else:
value = sfm.values.get(tag)