docs: the 1c bench pass and the ring-path results (2026-10-29)

- Spec: the 1c bench pass on BE9441 and UM20147, each on its own modem, and
  its 16 findings (a new 'Left open by the 1c bench' section).  Ring-path
  results, measured: with THOR's watchdog off a call-home reaches THOR when
  nobody holds the line (Run 0, via ListenerAchTcp) and is lost when SFM holds
  it (Run A: the unit times out and does not retry); with THOR polling, SFM gets
  no reply and THOR is undisturbed (A2, step 1).  Decision owed before merge:
  the Micromate line hold.
- Micromate reference: retracts 'ListenerAchTcp never used' in place (it
  accepted tonight's call-home; port 12345); a call-home that rings on another
  receiver's connection is lost; a second TCP connection while THOR holds the
  line is accepted but silent.
- CLAUDE.md: 1c status, the open decision, THOR's log location.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-10-29 20:49:28 -04:00
1 parent 3528dde970
commit 7491cd4b6f
3 files changed
+123 -12

No files matched your search

+31 -1
View File
@@ -2960,7 +2960,26 @@ RX55 at `63.45.161.30`):
bytes** back, and THOR's own reconnect at 19:57:10 logged `Mismatch connection
found for 63.45.161.30:9034` four times, then "lost connection" at 19:57:38.
Cause not yet known (a unit reset and a modem power cycle did not restore it
by 20:01). See the parity spec's ring-path bench notes.
by 20:01; it answered again at 20:16:51 after another reset and time). See the
parity spec's ring-path bench notes.
#### A call-home that rings on someone else's connection is lost (2026-10-29)
Measured with THOR's watchdog off and auto call-home on. SFM held UM20147's line
and sent a stop (`0x97`, 20:21:13.001); the unit answered with `RADIO RING`
instead of the ack (20:21:13.7). SFM hung up, then reconnected 0.1 s later for a
status read, and the unit answered it. The unit's screen: "connecting..." →
"call home data transfer..." for under a minute → idle. **It did not retry**,
and THOR never saw the call. The same stop from the keypad with nobody on the
line reached THOR through `ListenerAchTcp` (above).
🔶 Inferred: the unit counts the connection its ring rode up as the call and
waits for the server to run the session; a receiver that is not THOR (or not
running an ACH session) swallows the call-home. One observation.
While THOR **holds** the line (polling every ~5–15 s), a second TCP connection to
the modem is accepted but receives **nothing** (SFM: 0 bytes in 10 s), and THOR's
own reads continue to be answered -- measured 20:44:35.9.
#### THOR has a second, unused ACH path
@@ -2970,6 +2989,17 @@ RX55 at `63.45.161.30`):
units that genuinely dial, on some other modem arrangement) but is not what a
PAD-connected Micromate uses.
> ⚠ **RETRACTED 2026-10-29 (measured):** the listener *is* used by a
> PAD-connected Micromate, whenever nothing holds the modem's session. With
> THOR's watchdog off and nobody connected, UM20147's call-home (keypad stop)
> arrived as `ListenerAchTcp - Connection accepted from 63.45.161.30:36648`
> (20:37:28.5), THOR read the serial, and `RADIO RING` then came **up that
> connection** to start the session. So both paths are real and timing decides:
> the ring rides an existing connection (10-14, THOR's watchdog), or the modem
> dials its destination (here `50.197.32.91:12345`, the office) and the ring
> follows. THOR's listener port is **12345** (Brian, 2026-10-29); the 12346 in
> the 10-14 log was that day's setup.
#### What this means for a homebrew receiver
⚠ **`bridges/mm_ach_server.py` is built on the wrong model for this deployment.**
@@ -505,11 +505,21 @@ Each safety piece arrives with its first real use, so nothing sits untested.
`POST /device/call_home`) refuse 409 before any write unless the unit
answers as Series 3 (and as the named unit, given `serial=`). The stop
tools stay exempt: 0x96/0x97 mean the same on both series.
- ⚠ **The ring-path Runs A/B have not been run.** The page does not poll
(every request is a click), SFM hangs up on RADIO RING, and Hang up
releases a held line, but whether a call-home reaches THOR while SFM
holds the line is unmeasured. Run them before serial mode is used on a
unit that calls home to THOR, i.e. before this branch merges.
- **Bench pass 2026-10-29 (BE9441 and UM20147, each on its own modem):**
identify, status, settings, list, one download, start and stop through the
page on both series. Sixteen findings, listed under *Left open by the 1c
bench* below.
- **Ring path, 2026-10-29 (measured; details under *Open questions ... for
the bench*):** with THOR's watchdog off, a call-home **reaches THOR** when
nobody holds the line (Run 0), and **is lost** when SFM holds it (Run A):
the unit takes SFM's connection for the call, sits in "call home data
transfer", times out, goes idle and does not retry. With THOR polling, SFM
cannot reach the unit at all and does not disturb THOR (A2, step 1).
⚠ **Decision owed before merge:** the 5-minute Micromate line hold makes
SFM a call-home sink whenever THOR is not holding the line. Options: a
much shorter hold or none (hang up after each request), or SFM runs the
call-home session itself (the ADR 0003 call-in server). A cool-down alone
does not help: the unit does not retry.
- Of the items parked for 1c below: **CORS ordering** does not bite the
page, which SFM serves itself (same origin); the **Series-4 mic** is not
shown (the event list shows geophone peaks only). The rest are server-side
@@ -644,6 +654,46 @@ matters to is named.
- **Series-3 stored `record_type` (MiniMate bench check).** Rows SFM stores for Series 3 still say "Waveform" for histograms — pre-existing, the same as the legacy `blastware_file` route; fix it with the `0C` decode.
- **A fresh Series-4 download's gaps (1c).** No `sample_rate` is stored, and the sidecar has no vector sum; the DB row has the unit's PVS (overlaid from `0x0C`).
- **Orphan files (SDM).** An event file in the store with no DB row is still overwritten by a download.
## Left open by the 1c bench (2026-10-29)
From the page bench on BE9441 (MiniMate, modem `63.43.212.233`) and UM20147
(Micromate, RX55 `63.45.161.30`). Numbers as given at the bench. Fixes wait for
a short plan; #14 waits for the line-hold decision above.
1. **S3 settings decode reads before escapes are removed** -- BE9441 showed 0 sps
and a 68.3-min interval (true: Histogram, 4096 sps, 2 s, per the 10-26 fixture
README); while monitoring the `1A` anchor is not found at all. Old bug, the
same family as #46's C1. Fix: decode after unescaping; this reply is the
fixture.
2. **A 1.7 KB mic-off histogram is refused** -- this branch lacks #47/#49 (`dev`
decodes `K441LTOK.QZ0`, this branch does not). Bring them over before merge.
3. ✅ The 71 KB histogram crossing two 64 KB pages downloaded over the modem
byte-identical to Blastware's (`f10f39a8…`).
4. List vector sum 0 / −0 for three events whose peaks reach 1.08 in/s --
suspected escapes in the `0C` record (not checked).
5. Stored S3 event: client "Terra-Mechanic" (last letter missing) -- unexplained.
6. Page: the Config tab shows the empty write form until Read in serial mode.
7. `/db/events` 500 "database is locked" on the first page load against a new DB.
8. **S3 downloads are stored from the live waveform decoder** (`store.save`), not
the file reader that Series 4 uses (`save_imported_idf`): a histogram gets no
samples and is labelled Waveform. Fix: store S3 downloads through the
file-import path (also aligns event times with v0.32.0).
9. **S3 start over a modem:** the line dropped during the sensor check, the
read-back hit `Broken pipe` and did not reconnect → 503 though the start
worked. Fix: reconnect for the read-back (reads only; the send stays single).
10. Page: "Start monitoring failed:" on a sent-but-unconfirmed change -- wording.
11. S3 stop ~19 s, 11.6 s of it the identity check (a full connect) vs ~1 s for a
serial-only read. **Identified only** (speed is secondary to reliability).
12. Page, serial mode: the Device tab's empty legacy compliance table; S4
Firmware "—" (should say Thor line).
13. S4 mic in the event's peaks "0.0 dBL" vs 96.54 dB(L) in the file -- the
`0x0C` mic denormal parked above, now visible.
14. **After hanging up for a RADIO RING, SFM called the unit straight back up**
(the page's post-change status read). Part of the line-hold decision.
15. **Audit:** the UM20147 stop's record says `sent: null` although `0x97` went out.
16. Page: a 504 on a unit THOR holds should say "connected, but no reply --
another program may be holding this unit's line".
- **Same-second collisions (SDM).** Two events of one unit in the same second collide on (serial, timestamp) — a DB-wide dedup property, not 1b's.
- **423 without `Retry-After` (1c).** Add `Retry-After` to the 423 a request gets while a download holds the unit.
- **A lost `0C` during a Series-3 list (1c).** The strict walk raises on a lost `0A`
@@ -881,6 +931,31 @@ Open questions:
- *Also settles:* whether the modem routes a call-home to its destination
on its own (Run 0), in which case SFM never sees a ring and THOR gets the call
while SFM is still connected.
- **Results, 2026-10-29** (UM20147 on a fleet-config RX55 at `63.45.161.30`,
added to THOR as a test unit; THOR log `C:\THORDATA\Logs\`; all measured
unless marked):
- **Run A (watchdog off, SFM holding the line).** SFM's stop (`0x97`,
20:21:13.001) was answered by `RADIO RING` instead of its ack (20:21:13.7);
SFM hung up. The page's post-change status read called the unit straight
back up (20:21:13.8 → held to 20:21:47; the reaper listens every 5 s, no
second ring). The unit showed "connecting..." → "call home data
transfer..." for under a minute → idle, **no retry**; THOR logged nothing
for the unit through 20:23:57.
- **Run 0 (watchdog off, nobody on the line, keypad start/stop).** The modem
dialled `50.197.32.91:12345`: `ListenerAchTcp - Connection accepted from
63.45.161.30:36648` (20:37:28.5) → serial → `RADIO RING IS RECEIVED` on that
connection → a full THOR session (count 8, download, **DeleteEventState**
per event, reports, email, Vision). SFM sent nothing 20:22–20:37.
- **A2, step 1 (THOR polling, ~5–15 s).** An SFM status read at 20:44:35.9:
the modem accepted the connection, 0 bytes back in 10 s (504). THOR's reads
of the unit at 20:44:37.4 and 20:44:44.3 were answered; no disconnect.
- **Conclusion:** the risk is confined to when THOR is not holding the line
(watchdog off, THOR down, reconnect gaps), and a cool-down would not fix
it. Run B (a poller) is deprioritised: while THOR polls it only collects
504s; the open point is a reconnect gap.
- Also seen: after a manual THOR Disconnect (19:48:07) the unit answered
nobody until a reset (back 20:16:51); THOR's reconnect logged `Mismatch
connection found`. Cause unknown.
- The ack for start/stop when the unit is already in that state.
✅ **Answered for stop (2026-10-25, UM20147):** a stop sent to an idle unit
is acked `0x68` in 10 ms (twice) and changes nothing: not the state, the