fix(decode): the mic's own range -- a range-1 mic reads an eighth
The mic's channel block carries a range byte at the geophones' offset (label + 29, reference D.6.4). At 1 each count is an eighth, 0.25 / 8 Pa: set on 7 DL2 events, all of them Blastware's values divided by 8, and on none of the other 70,026 -- independently of the geophone range (28 events have Sensitive geophones and a range-0 mic). SFM read those 7 mics 8x high: K557J1XQ.GF0W is 79.4 dB(L), not 97.5. Event.mic_range; the decoded mic peak (and so the .h5 mic samples, whose scale is the peak's) and the checker's mic conversions follow it. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
This commit is contained in:
1 parent
af8dde5114
commit
836c9e864b
7 files changed
+3266
-12
No files matched your search
@@ -608,6 +608,7 @@ class Event:
|
||||
trigger_level_mic_psi: Optional[float] = None
|
||||
record_stop_mode: Optional[int] = None # 0 = Fixed
|
||||
mic_type: Optional[int] = None # 2 = Linear Weighting (as Series IV)
|
||||
mic_range: Optional[int] = None # 0; 1 = each mic count is an eighth (2026-11-09)
|
||||
stop_time: Optional[datetime.datetime] = None # the footer's stop
|
||||
# A histogram's per-interval detail (2026-11-03); None for a waveform.
|
||||
histogram_interval_sec: Optional[int] = None # as ComplianceConfig.histogram_interval_sec
|
||||
|
||||
Reference in new issue