feat: add erase-all protocol and browse helpers to protocol/client layer

protocol.py:
- SUB_ERASE_ALL_BEGIN = 0xA3, SUB_ERASE_ALL_CONFIRM = 0xA2 (confirmed 4-11-26 MITM)
- SUB_CHANNEL_CONFIG (0x06) data length = 0x24 (36 bytes) in DATA_LENGTHS
- begin_erase_all()              — single frame, token=0xFE, response 0x5C
- confirm_erase_all()            — single frame, token=0xFE, response 0x5D
- read_event_storage_range()     — two-step read (probe+data), token=0xFE
  Response last 8 bytes = first/last stored event key; both 0x01110000 after erase

client.py:
- list_event_keys()              — browse-mode 1E→0A→1F walk, no waveform download;
  returns list of hex key strings; used as fast pre-check before get_events()
- get_events(skip_waveform_for_keys=set())
  — for already-seen keys: only 0A+1F(browse), skips 1E-arm/0C/POLL×3/5A entirely
- delete_all_events()            — orchestrates the confirmed erase sequence:
  0xA3 → 0x1C → 0x06 → 0xA2; logs first/last key from storage range response

Co-Authored-By: Claude Sonnet 4.6 <noreply@anthropic.com>
This commit is contained in:
2026-04-11 01:14:37 -04:00
committed by serversdown
parent 4921b0489a
commit a3b8d10fa8
2 changed files with 200 additions and 2 deletions
+80 -1
View File
@@ -57,7 +57,7 @@ SUB_POLL = 0x5B
SUB_SERIAL_NUMBER = 0x15
SUB_FULL_CONFIG = 0x01
SUB_EVENT_INDEX = 0x08
SUB_CHANNEL_CONFIG = 0x06
SUB_CHANNEL_CONFIG = 0x06 # Event storage range read (first/last key) ✅
SUB_MONITOR_STATUS = 0x1C # Monitoring status read (battery, memory, mode) ✅
SUB_EVENT_HEADER = 0x1E
SUB_EVENT_ADVANCE = 0x1F
@@ -82,6 +82,12 @@ SUB_TRIGGER_CONFIRM = 0x83 # Confirm trigger write ✅
SUB_START_MONITORING = 0x96 # Start monitoring → response 0x69 ✅
SUB_STOP_MONITORING = 0x97 # Stop monitoring → response 0x68 ✅
# Erase-all SUBs (confirmed from 4-11-26 MITM capture)
# Both use token=0xFE at params[7] and return minimal 11-byte acks.
# Standard response formula applies: 0xFF - SUB.
SUB_ERASE_ALL_BEGIN = 0xA3 # Begin erase all events → response 0x5C ✅
SUB_ERASE_ALL_CONFIRM = 0xA2 # Confirm erase all events → response 0x5D ✅
# Hardcoded data lengths for the two-step read protocol.
#
# The S3 probe response page_key is always 0x0000 — it does NOT carry the
@@ -96,6 +102,7 @@ DATA_LENGTHS: dict[int, int] = {
SUB_SERIAL_NUMBER: 0x0A, # 10-byte serial number block ✅
SUB_FULL_CONFIG: 0x98, # 152-byte full config block ✅
SUB_EVENT_INDEX: 0x58, # 88-byte event index ✅
SUB_CHANNEL_CONFIG: 0x24, # 36-byte event storage range (first/last key) ✅
SUB_MONITOR_STATUS: 0x2C, # 44-byte monitor status block (idle) ✅
SUB_EVENT_HEADER: 0x08, # 8-byte event header (waveform key + event data) ✅
SUB_EVENT_ADVANCE: 0x08, # 8-byte next-key response ✅
@@ -1137,6 +1144,78 @@ class MiniMateProtocol:
self._send(frame)
return self.recv_write_ack(expected_sub=rsp_sub)
def read_event_storage_range(self) -> S3Frame:
"""
Read event storage range (SUB 0x06 → response 0xF9).
Two-step read: probe (offset=0x00) then data (offset=0x24 = 36 bytes).
Uses token=0xFE at params[7] — same as the erase sequence.
The 36-byte response ends with two 4-byte event keys (first and last
stored event key). After a successful erase, both keys are 0x01110000
(device-empty sentinel). Confirmed from 4-11-26 MITM capture.
Returns:
S3Frame with 36 bytes of storage range data.
Raises:
ProtocolError: on timeout or wrong response SUB.
"""
rsp_sub = _expected_rsp_sub(SUB_CHANNEL_CONFIG) # 0xFF - 0x06 = 0xF9
params = token_params(0xFE)
log.debug("read_event_storage_range: probe step rsp_sub=0x%02X", rsp_sub)
self._send(build_bw_frame(SUB_CHANNEL_CONFIG, offset=0x00, params=params))
self._recv_one(expected_sub=rsp_sub)
log.debug(
"read_event_storage_range: data step offset=0x%02X",
DATA_LENGTHS[SUB_CHANNEL_CONFIG],
)
self._send(build_bw_frame(SUB_CHANNEL_CONFIG,
offset=DATA_LENGTHS[SUB_CHANNEL_CONFIG],
params=params))
return self._recv_one(expected_sub=rsp_sub)
def begin_erase_all(self) -> S3Frame:
"""
Send Begin-Erase-All command (SUB 0xA3 → response 0x5C).
Single frame with token=0xFE at params[7]. The device acknowledges with
a minimal ack and begins the erase process. Follow up with
read_monitor_status() + read_event_storage_range() + confirm_erase_all()
to complete the sequence. Confirmed from 4-11-26 MITM capture.
Returns:
S3Frame ack from device (SUB 0x5C).
Raises:
ProtocolError: on timeout or wrong response SUB.
"""
rsp_sub = _expected_rsp_sub(SUB_ERASE_ALL_BEGIN) # 0xFF - 0xA3 = 0x5C
log.debug("begin_erase_all: rsp_sub=0x%02X", rsp_sub)
self._send(build_bw_frame(SUB_ERASE_ALL_BEGIN, params=token_params(0xFE)))
return self._recv_one(expected_sub=rsp_sub)
def confirm_erase_all(self) -> S3Frame:
"""
Send Confirm-Erase-All command (SUB 0xA2 → response 0x5D).
Single frame with token=0xFE at params[7]. Must be preceded by
begin_erase_all() + read_monitor_status() + read_event_storage_range().
After this call the device memory is cleared. Confirmed from 4-11-26
MITM capture.
Returns:
S3Frame ack from device (SUB 0x5D).
Raises:
ProtocolError: on timeout or wrong response SUB.
"""
rsp_sub = _expected_rsp_sub(SUB_ERASE_ALL_CONFIRM) # 0xFF - 0xA2 = 0x5D
log.debug("confirm_erase_all: rsp_sub=0x%02X", rsp_sub)
self._send(build_bw_frame(SUB_ERASE_ALL_CONFIRM, params=token_params(0xFE)))
return self._recv_one(expected_sub=rsp_sub)
# ── Internal helpers ──────────────────────────────────────────────────────