From b023083c79ca34178744609dfabf25099bfe578b Mon Sep 17 00:00:00 2001 From: serversdown Date: Thu, 29 Oct 2026 18:30:27 -0400 Subject: [PATCH] docs(changelog): T438HGTZ.B30W is not a decoder truncation Its Tran, Vert and MicL end 3 samples before Long in Blastware's own XML sample table as well; the setup-field harness compared Tran's length with the table's row count. Co-Authored-By: Claude Opus 5.5 Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9 --- CHANGELOG.md | 6 ++++-- 1 file changed, 4 insertions(+), 2 deletions(-) diff --git a/CHANGELOG.md b/CHANGELOG.md index 9a36a88..4cd0fa1 100644 --- a/CHANGELOG.md +++ b/CHANGELOG.md @@ -25,8 +25,10 @@ All notable changes to seismo-relay are documented here. - Verified against Blastware's own `_XML.XML` reports for the 70,042 DL2 events that have one (2016–2026): sample rate 70,042/70,042, record time 6,524/6,524, pre-trigger 6,484/6,484, notes 70,042/70,042, sample - count 6,523/6,524 (one waveform the decoder ends 3 samples short — - separate). + count 6,523/6,524. The one miss is the harness's, not the decoder's: + `T438HGTZ.B30W`'s Tran, Vert and MicL end 3 samples before Long, in + Blastware's own sample table too, and the harness compared Tran's + length with the table's row count. - Events imported with an ASCII report are unchanged: the report still overlays. Not yet read: the job number (2,997 archive events carry one). The waveform-time estimate still finds the record time by the