docs(spec): what the phase 3 review left open, and why

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-10-23 19:36:36 -04:00
1 parent cdc6360ec0
commit b1c87ecccd
1 file changed
+49
@@ -540,6 +540,55 @@ matters to is named.
through `0x94`/`0x48` may detect unsaved keypad edits, which would undercut
"can't be detected from the wire". One bench check.
## Left open by the phase 3 review (2026-10-23)
The whole-phase review judged phase 3 ready for the bench, with nothing that
sends twice, sends without an audit row, or sends to an unverified unit. These
items were left for later on purpose:
- **Series-3 change path shape. Settle before the MiniMate bench check.**
- The change path makes one POLL at a 120 s receive timeout, with no retry. A
cold MiniMate that drops that POLL costs 120 s of a locked unit and then a
504. The legacy route retried `poll()` on the same connection at 10 s.
- The identity read is a full `connect()` (POLL, `0x15`, `0x01`, `0x1A`, `0x08`).
The spec asks for one serial read.
- Proposed: 10 s timeouts before the send, one identity-poll retry on the same
connection (a read, never the write), POLL plus `0x15` in place of
`connect()`, and the long timeout only on the read-back polls.
- That needs two measurements first: whether a *monitoring* MiniMate answers
the full `connect()` over TCP, and what `0x1C` reads during the ~40 s sensor
check.
- **For 1c (the webapp).**
- 409 means three things. "Wrong unit" and "unmeasured state" sent nothing;
"mismatch" sent a frame with a clean ack. Only the body tells them apart.
- `ChangeResult.detail` is a dict, so the webapp's `e.detail || r.statusText`
would print `[object Object]`.
- The 60 s verify window, plus the call-up, plus up to 20 s of lock wait,
exceeds the 60 s Terra-View proxy timeout. A mismatch could never be
delivered through that proxy.
- Nothing marks `pending` rows left by a restart mid-change. Reconcile them at
server start only, never on every `SeismoDb` open.
- **A late `0x69` after a lost ack.** When the ack is lost but the follow-up
read validates, the line is kept, and a late `0x69` then breaks the next
unrelated request with a 502. Safe, but confusing. Hanging up after any lost
ack needs a way to release the line on a path that returns a result.
- **Legacy routes take no per-address lock.** They can interleave with an
audited change on the same address. This belongs with the blind-stop-tools
decision, which is still open.
- **Test and scan gaps.**
- The `protocol.execute(` caller scan does not catch an aliased caller.
- Dropping the S4 ack's `data` field is not caught by any test.
- The fake's `fail` queue is also consumed by `read_unsolicited` on the change
path.
- **Docs nits.**
- `s4.py` says `ChangeUnconfirmed` follows "a link fault or a ring". It also
follows a `calling_home` read.
- The 2026-10-02 CLAUDE.md bullet still says "Read-only".
- The route docstrings' 409 line omits the held-line `WrongUnit`.
- **For the merge to dev.** The changelog states: a new `device_commands` table
(`CREATE TABLE IF NOT EXISTS`) plus a `_migrate` step that adds `sent` and
`states` to tables created from this branch. Prod: no migration, no backfill.
## Table 2: to be added later (both series)