docs: where the 1995 default-clock events come from
Brian's account: units come back from calibration with the internal battery unplugged, so a comms-test trigger before the date is set carries the default. DL2 holds 17 such events on 6 units, all within 1995-09-01 12:02 .. 09-03 11:24. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
This commit is contained in:
1 parent
8a40eb638a
commit
e1e8a851c8
1 file changed
+5
@@ -2669,6 +2669,11 @@ Appears in event index blocks. Time-of-day fields (hour/min/sec) are absent.
|
||||
| `1E` | 30 | Day (0x1E = 30 decimal) | ✅ CONFIRMED |
|
||||
|
||||
> ✅ **2026-02-26 — CONFIRMED:** The year 1995 is the **MiniMate Plus factory default RTC date**, which the device reverts to whenever the internal battery is disconnected or the real-time clock loses power. Any event timestamped around 1995 means the clock was not set. This is known device behavior, not an encoding anomaly.
|
||||
>
|
||||
> **2026-11-03 — where they come from (operator account, Brian):** units come back from calibration with the internal battery unplugged, so an event triggered while testing communications, before anyone sets the date, carries the default.
|
||||
> - **DL2 (measured):** 17 such events on 6 units (BE9440, BE9557, BE11529, BE17354, BE18189, BE18190). All fall between 1995-09-01 12:02:02 and 1995-09-03 11:24:06, most of them between 12:02 and 13:01 on 1 September.
|
||||
> - 🔶 **Inferred:** the clock restarts at a fixed default near 1995-09-01 12:00 and runs from power-up.
|
||||
> - These files still end in a valid footer. SFM reads the 1995 times and reports them as `CLOCK_NOT_SET` (`models.Timestamp`).
|
||||
|
||||