1 Commits
Author SHA1 Message Date
serversdownandClaude Opus 5.5 a8cc516601 fix(minimateplus): SUB 0x15's serial is a C string -- cut at the first NUL
BE9441 (MiniMate bench, 2026-10-25) answers b"BE9441\x00?": six characters,
the terminator, a leftover byte.  rstrip(b"\x00") kept the "?", so the serial
read "BE9441\x00?" and the serial-named path refused every request as the wrong
unit (409).  Seven-character serials fill the field to the terminator, which is
why no earlier unit showed it.  The legacy /device/info output is cleaned too.
No codec, store or DB change.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
2026-10-25 17:14:11 -04:00