Group A of the file-only decode, every field matched against every DL2
report (2026-11-03):
- Identity block before the setup name, what the live read gets from
SUB FE: serial k-202, firmware k-147, DSP version k-138, calibration
date k-132 + 'by' k-128, manufacturer k-94, model k-72, firmware
minor k-193. Event.device_info (DeviceInfo gains calibration_date /
calibrated_by). Serial, Version and Calibration 69,983 / 69,983.
- Battery: uint16 hundredths of a volt at k-190 (Event.battery_volts).
69,729 exact; the other 254 are Blastware rounding half up.
- Trigger source: the two bytes between the embedded event record's
lead and fe 80, a bitmask (0x01 Tran 0x02 Vert 0x04 Long 0x08 MicL
0x200 Manual): 6,493 / 6,493. Event.triggered_by.
- Trigger settings in Series IV's 0x1A layout: sources bitmask k-19,
stop mode k-31 (0 = Fixed), each channel's float32 level at
label+30 (geo in/s, mic psi), mic type code at MicL+4 (2 = Linear
Weighting). Geo level 6,447 / 6,447, mic 4,403 / 4,403; Blastware
prints a level when its channel is a trigger source and the event
was not Manual. The mic level's dB uses 6895 Pa/psi (4,272 / 4,272
against 4,271 at 6894.757).
- .h5 attrs for all of them (additive).
- Checker: SFM's side reports the ten fields; the mic trigger level
follows the report's dB/psi. Carve-out
units-is-a-display-preference: dB or psi is Blastware's PC setting
when it wrote the report -- DL2's reports switch fleet-wide in 61
clean runs by write time, and no byte of the file separates them.
Two tests that pinned BatteryLevel / Calibration as blank now pin the
match.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
The setup block has Series IV's 0x1A layout here too: report flags at
k - 9 (bit 0x02 = job number shown) and the number, uint16 BE, at k - 8,
kept while switched off. The flag matches Blastware's JobNumber on
2,997 / 2,997 DL2 events that print one and 66,986 / 66,986 that don't.
Every printed value is 1 and matches; hidden ones hold real numbers
(802, 1012, 1013, 622), so the value is only proven for 1.
- ProjectInfo.job_number_enabled / job_number, named as in
micromate.setup_block.
- write_event_hdf5 stores both as .h5 attrs when read (additive).
- The checker reports the number only when shown, and scores a number
only SFM shows as a mismatch. The old test that pinned JobNumber as
sfm_blank now checks only the ZC cases it also covered.
Not carried into the DB, the API or the PDF report yet (needs a column).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
In the setup's channel blocks, the byte 29 after each geophone label is
the range: 0x00 Normal (10 in/s), 0x01 Sensitive (1.25 in/s). It matches
Blastware's GeoRange on 70,042 / 70,042 DL2 events, all three channels
agreeing on every one (35 Sensitive). SFM's live 1A has the same byte
at label+33. SFM assumed Normal, so every Sensitive event's peaks,
vector sum and .h5 samples read 8x high.
- Event.geo_range (0 / 1, None when not read -- channels disagreeing,
an unknown value or a missing block -- with a decode_problems entry).
- decode_blastware_bytes scales peaks and the histogram vector sum by it.
- event_hdf5.write_event_hdf5: geo_range defaults to None = the event's
own, else Normal; an explicit argument still wins.
- waveform_store.save_imported_bw and scripts/backfill_sidecars.py stop
forcing 'normal', so imports and the backfill store the file's range.
The waveform_store line was agreed with the parity session, which owns
that file; its own branch keeps a geo_range parameter, defaulting to
None at its next dev merge.
- The checker's SFM side converts at the event's range and reports
GeoRange; tests that pinned the old 8x gap now pin the fix, and the
saturation carve-out is still tested against an 8x value.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Brian's requirements: the instrument's data is the only time source; a filename
is a cross-check, never a fallback (an event SFM receives itself has none).
The record (found by reading a seismo_lab Inspector dump; Series-III reference
§7.8.9): 00 28 05 00 ?? ?? | fe 80 | trigger | stop, times dd mm yyyy(BE) 00 hh
mm ss. Measured on 7,935 waveforms (DL2 + prod store): exactly one candidate
validates in every file; trigger = Blastware's report 1,320/1,320, = filename
7,935/7,935, = the unit's 0C record 10/10.
Provenance (requirement 6, measured): in Blastware's download of 8 BE9441
events the record -- with 24 B before and 22 after -- occurs exactly once in the
unit's own A5 reply payloads; SFM-only downloads (no Blastware) carry it too.
It is instrument data, not Blastware's.
- _embedded_event_record(): every fe 80 before the footer is a candidate; it
counts only with the lead bytes, two well-formed times (zero byte zero),
footer ts1 <= trigger <= stop, stop = footer ts2 or a second before, and a
length matching the record time. Exactly one must count; none or several is
a decode problem, never a guess.
- Without one: the footer estimate, timestamp_source 'estimate', a
decode_problems entry, a WARNING. No filename fallback.
- A filename that disagrees is logged; the decoded time is kept. A paired
report fills in only without a validated record (apply_report_to_event, the
sidecar dict overlay, event_to_sidecar_dict) and is logged when it disagrees.
- decode_blastware_bytes(raw, name=None): no file or name needed;
save_imported_bw uses it (no temp file). Event gains timestamp_source and
decode_problems; the sidecar records both.
- Histograms unchanged (#51): their record (fe f5) is noted, not used.
Tests: raw bytes with no name (ZI0, 7 BE12844 oracles), a disagreeing name, each
validation rule alone, two valid candidates, no filename fallback, report
precedence, sidecar. 7 mutants, each caught. Real data, decoded with no name:
7,961/7,961 waveforms on the record, 0 disagreements with filename or report;
10,433/10,433 prod histograms unchanged. Suite: only the 16 known failures.
Open (reported, not changed): 3 DL2 files named as waveforms whose body the
waveform codec does not recognise keep the footer start, 4 s off, as on dev.
None in prod.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01MUHR251NLznsFvPSrfegqL
Make the sensor self-check a first-class part of the standardized decoded event
so SFM stops decoding it at report time — device-agnostic, per the store's
decoder→standardized-.h5→SFM model.
* Event gains a `sensor_check` field; both decoders attach the traces where
they set raw_samples — series-3 in event_file_io.read_blastware_file
(minimateplus.sensor_check), series-4 in waveform_store's IDF path
(micromate.sensor_check). Covers ingest and backfill (both re-decode).
* event_hdf5 bumps schema_version 1→2 and writes an optional /sensor_check
group (raw counts, int32, per channel present). read_event_hdf5 returns
it; plot_json_from_hdf5 carries it as a top-level key. Old v1 files still
read cleanly (no group → None), so nothing breaks before the backfill.
* gather_report_data reads sensor_check_waveforms from the .h5 and drops the
report-time series-3 decode — the report no longer reaches into a decoder,
and a series-4 event now lights up the same strip automatically.
Stored as raw counts (a shape diagnostic, rendered fit-to-box): the per-series
count scale differs and a physical mic unit is ill-defined, so conversion would
add complexity for no display benefit — easy to add later if a numeric use
appears.
Tests: .h5 roundtrip + backward-compat + plot_json + real series-3 decode
attaches to the Event.
Co-Authored-By: Claude Opus 4.8 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01YDXjZCr4RqT2U3QvMDhgzf
- Added `CallHomeConfig` model to represent the Auto Call Home settings.
- Introduced methods in `MiniMateClient` for reading (`get_call_home_config`) and writing (`set_call_home_config`) the call home configuration.
- Updated `MiniMateProtocol` with new commands for call home operations (SUB 0x2C for read, SUB 0x7E for write, and SUB 0x7F for confirm).
- Created API endpoints for retrieving and updating call home settings in the server.
- Enhanced the web interface with a new "Call Home" tab for user interaction with call home settings.
- Implemented JavaScript functions for reading and writing call home configurations from the web app.
Add full decode pipeline for 0x2C partial records from the device's event
list, representing continuous monitoring intervals where no threshold was
crossed. These records appear interleaved with full triggered events in the
browse walk and were previously ignored.
minimateplus/models.py
- Add MonitorLogEntry dataclass: key, start_time, stop_time, serial,
geo_threshold_ips, raw_header, duration_seconds property
minimateplus/protocol.py
- read_waveform_header() now returns (data_rsp.data, length) — full payload
including the record-type byte at position 0 — instead of the sliced header.
Callers that need the old slice use raw_data[11:11+length] as before.
minimateplus/client.py
- Add _decode_0a_partial_header(): auto-detects 9-byte (sub_code=0x10) vs
10-byte (sub_code=0x03) timestamp format, handles 1-byte inter-timestamp
gap, extracts serial via BE anchor and geo threshold via Geo: anchor.
- Add get_monitor_log_entries(skip_keys=None): browse walk (1E → 0A → 1F),
decodes partial records, skips full records and already-seen keys.
minimateplus/__init__.py
- Export MonitorLogEntry
bridges/ach_server.py
- After get_events(), call get_monitor_log_entries(skip_keys=seen_keys) and
save new entries to monitor_log.json in the session directory.
- Add _monitor_log_entry_to_dict() helper.
- Include monitor log keys in downloaded_keys for state persistence.
CLAUDE.md / CHANGELOG.md
- Document 0x2C partial record layout (timestamp format, ASCII metadata
region, 1-byte gap edge case) confirmed from 4-11-26 MITM capture.
- Version bump to v0.10.0; update What's next.
Co-Authored-By: Claude Sonnet 4.6 <noreply@anthropic.com>
- Introduced new SUBs for monitoring status, start, and stop commands in protocol.py.
- Implemented read_monitor_status, start_monitoring, and stop_monitoring methods in MiniMateProtocol class.
- Added new API endpoints for monitoring status retrieval and control in server.py.
- Enhanced the web application with a monitoring panel, including battery and memory status display.
- Created a new Python script to parse SUB 0x1C response frames for monitoring status.
- Documented the monitoring status response format and field locations in markdown and text files.
- Added `_decode_a5_waveform()` to parse SUB 5A frames into per-channel time-series data.
- Introduced `download_waveform(event)` method in `MiniMateClient` to fetch full waveform data.
- Updated `Event` model to include new fields: `total_samples`, `pretrig_samples`, `rectime_seconds`, and `_waveform_key`.
- Enhanced documentation in `CHANGELOG.md` and `instantel_protocol_reference.md` to reflect new features and confirmed protocol details.
- DeviceInfo.event_count: Optional[int] = None (new field in models.py)
- connect() now calls proto.read_event_index() after compliance config and
stores the decoded count in device_info.event_count
- _serialise_device_info() exposes event_count in /device/info and /device/events
JSON responses
event_count is decoded from uint32 BE at offset +3 of the 88-byte F7 payload
(🔶 inferred — needs live device confirmation against a multi-event device).
Any ProtocolError from the index read is caught and logged; event_count stays
None rather than failing the whole connect().
Co-Authored-By: Claude Sonnet 4.6 <noreply@anthropic.com>
Adds full support for reading device compliance configuration (2090-byte E5
response) containing record time, trigger/alarm levels, and project strings.
protocol.py:
- Implement read_compliance_config() two-step read (SUB 1A → E5)
- Fixed length 0x082A (2090 bytes)
models.py:
- Add ComplianceConfig dataclass with fields: record_time, sample_rate,
trigger_level_geo, alarm_level_geo, max_range_geo, project strings
- Add compliance_config field to DeviceInfo
client.py:
- Implement _decode_compliance_config_into() to extract:
* Record time float at offset +0x28 ✅
* Trigger/alarm levels per-channel (heuristic parsing) 🔶
* Project/setup strings from E5 payload
* Placeholder for sample_rate (location TBD ❓)
- Update connect() to read SUB 1A after SUB 01, cache in device_info
- Add ComplianceConfig to imports
sfm/server.py:
- Add _serialise_compliance_config() JSON encoder
- Include compliance_config in /device/info response
- Updated _serialise_device_info() to output compliance config
Both record_time (at fixed offset 0x28) and project strings are ✅ CONFIRMED
from protocol reference §7.6. Trigger/alarm extraction uses heuristics
pending more detailed field mapping from captured data.
Sample rate remains undiscovered in the E5 payload — likely in the
mystery flags at offset +0x12 or requires a "fast mode" capture.
Co-Authored-By: Claude Sonnet 4.6 <noreply@anthropic.com>
Confirmed 2026-04-01 against Blastware event report for BE11529 thump
event ("00:28:12 April 1, 2026", PVS 3.906 in/s).
models.py:
- Timestamp.from_waveform_record(): decode 9-byte format from 0C record
bytes[0-8]: [day][sub_code][month][year:2BE][?][hour][min][sec]
- Timestamp: add hour/minute/second optional fields; __str__ includes
time when available
- PeakValues: add peak_vector_sum field (confirmed fixed offset 87)
client.py:
- _decode_waveform_record_into: add timestamp decode from bytes[0:9]
- _extract_record_type: decode byte[1] (sub_code), not ASCII string
search; 0x10 → "Waveform", histogram TBD
- _extract_peak_floats: add PVS from offset 87 (IEEE 754 BE float32)
= √(T²+V²+L²) at max instantaneous vector moment
sfm/server.py:
- _serialise_timestamp: add hour/minute/second/day fields to JSON
- _serialise_peak_values: add peak_vector_sum to JSON
docs: update §7.7.5 and §8 with confirmed 9-byte timestamp layout,
PVS field, and byte[1] record type encoding; update command table;
close resolved open questions.
Co-Authored-By: Claude Sonnet 4.6 <noreply@anthropic.com>
- _decode_serial_number: read from data[11:] not data[:8] — was returning
the LENGTH_ECHO byte (0x0A = '\n') instead of the serial string
- _extract_peak_floats: search for channel label strings ("Tran" etc) and
read float at label+6; old step-4 aligned scan was reading trigger levels
instead of PPV values
- get_events: add debug=False param; stashes raw 210-byte record on
Event._raw_record when True for field-layout inspection
- server /device/events: add ?debug=true query param; includes
raw_record_hex + raw_record_len in response when set
- models: add Event._raw_record optional bytes field