The reconnect in ce441a0 opened a fresh connection and sent a bare 0x15 (the
identity check) before any SESSION_RESET + POLL. It exists for a unit that has
just started -- sensor check, then monitoring -- and a monitoring MiniMate
ignores a bare request until woken (measured 2026-04-08), so the identity read
would have timed out and the start reported unconfirmed anyway. Now: open,
wake (the same retried wake the read-back uses), read the serial, read 0x1C --
the order the status read and the legacy write routes use. Still one
reconnect, TCP only, nothing re-sent.
The fake gains bare_requests_ignored (off unless a test asks): a fresh
connection's reads time out, with the real client's TimeoutError, until woken.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
BE9441 over a modem, 2026-10-29: the start was acked, but the modem closed the
line during the ~30 s sensor check, so the read-back's first wake hit
'[Errno 32] Broken pipe' and the change was reported 503 'not checked' although
the unit was monitoring. On an OSError in the read-back, a TCP unit is now
reconnected once, its serial checked against the named unit, and its state read.
Nothing is re-sent: the start still goes out exactly once. A serial cable is
never reconnected (opening a MiniMate's port reboots it), a second failure or a
silence (a cleanly closed line) ends 'not checked' as before, and a different
unit after the reconnect is refused (409, change unconfirmed).
Store consequence: none.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Second start through SFM, BE9441 2026-10-26 23:54:41: acked 0x69, and Brian saw
'starting, performing sensor check, recording' -- but the read-back's single
SESSION_RESET + POLL went out 2.3 s after the ack, inside the ~30 s sensor
check, and was never answered or re-sent (504 after 120 s, audit ca7b5b42).
Brian: every unit is set to sensor-check before monitoring, ~30 s, then
'Recording...'.
- MiniMateProtocol.wake(attempt_s, deadline_s): SESSION_RESET + POLL re-sent
every attempt_s until the unit answers, then the POLL data step; a named
timeout at the deadline. Reads only. MiniMateClient.wake() wraps it.
- The Series-3 change: after a START, the first read waits S3_SENSOR_CHECK_S
(30 s) before asking; every read then wakes with retries (every 5 s, up to
60 s). A stop is unchanged -- its read-back was confirmed three times tonight.
Tests: a unit deaf for 4 probes is asked a 5th time and answers; a unit that
never answers ends in the named timeout after ~6 probes; a start's first read
waits the sensor check, a stop's does not; every read is preceded by a wake.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
c32a9e7 made the download check read the 5A body; the Series-3 fake handed
back placeholder frames (['frame']), so 4 plumbing tests in
test_device_access_s3_events.py and test_device_access_event_dispatch.py
failed with "'str' object has no attribute 'data'" -- and that commit was
pushed without the failure set being read. The fake's one-event download now
carries what a real one does: BE9441 event 1's recorded frames, their decoded
waveform and the unit's own peaks for it. Those tests now see verified=True
on a whole event. Suite back to the 16 known fixture failures.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Final fix wave for phase 3 (start/stop monitoring), per the whole-phase review.
- I1 + M2: every failure after the send has started is a ChangeUnconfirmed
(audit_id, sent = "acknowledged" | "unknown", the cause's HTTP status) whose
message says the frame was sent and names the audit record -- the ring in
place of the ack, the four lost-ack follow-ups, and the three read-back
paths. Series4Session releases the line on it. Every audit finish goes
through one helper; a finish that fails is AuditIncomplete (500) and still
says what happened to the unit. The line is kept.
- I2: device_commands gains `sent` (1 acknowledged, 0 nothing sent, NULL
unknown) and `states` (before / seen / after), with an idempotent ADD COLUMN
in _migrate for a table created earlier on this branch.
- M1: a RADIO RING already buffered is caught before 0x96/0x97 goes out
(pre_send = read_unsolicited, one 10 ms transport read). That is a
CallHomeInProgress with sent=0, not ChangeUnconfirmed.
- M3: Series 3 reads the raw 0x1C flag (new MonitorStatus.state_flag); a byte
other than 0x00/0x10, or a short section, is unknown(...) and refused.
- M8: even_if_already on a unit already in the state notes that the read-back
proves nothing.
- M10: a scan pins `.protocol.execute(` to sfm/device_access/s4.py; both
scans fail closed on a file that does not parse.
- M11: docs corrected (CLAUDE.md phase-2 sentence, phase-3 bullet, REST list;
micromate/protocol.py scope and the 0x49 read_state docstring; s4.py; route
docstrings). "No command has ever been originated" is NOT retracted.
- M12: GET /db/device_commands normalises the serial.
- Minors: unused imports, S3 dry-run frames and 120 s timeout pinned, test
rename, serial-cache and short-0x49 tests, _RUN_STATES comment.
Schema: two NULLable columns on the branch-new device_commands table, added
by _migrate on open; no backfill. No codec change, TOOL_VERSION unchanged.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
minimateplus's _extract_record_type returns "Waveform" for every 0C
record layout -- it cannot tell a histogram -- so a MiniMate histogram
was listed as "waveform". The fake's "Histogram" was a value the real
client never produces, which is why no test saw it.
- _s3_ref: record_type is always None, with
detail.record_type_note = "Series 3's event record does not say
waveform or histogram; not decoded yet".
- default_s3_events(): both events "Waveform", as the real client gives.
- The shared-shape contract test allows exactly {"record_type"} to be
None for Series 3 and nothing for Series 4.
Rows SFM stores for Series 3 still carry "Waveform" from the decode (the
same as the legacy blastware_file route); parked in the spec. No codec,
store or DB change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Series3Adapter.download_event compared the timestamp only after
download_event_by_key had run the whole 5A. A stale id cost a full
download of the wrong event (minutes on cellular), could surface the
64 KB-page 502 instead of 409, and a lost 0C reply was reported as "the
unit was erased".
- get_events() gains only_key_timestamp: with only_key, if the target's 0C
says another time (or none), the target takes the existing bypass --
no 1F(arm), no POLL, no 5A. A new branch only; it adds no command.
- download_event_by_key(key4, *, expected_timestamp=None) passes it on.
- The adapter converts the id's stamp to that ISO form and checks, in
order: no event -> 404; no readable 0C timestamp -> 502 MalformedReply
("could not read ... record (0C)"); a different time -> 409 StaleEvent;
no frames -> 502.
- The fake's download mirrors the real walk (no frames on a mismatch) and
records what it was asked for.
- Legacy get_events() (none of the new options) is pinned by a test of
its exact call sequence and 5A arguments, written before the change.
Read-only; nothing new on the wire. No codec, store or DB change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44