# Runbook — Recovering a wedged unit stuck in a call-home loop
**Original incident:** BE9558H at `166.246.130.1:9034`, recovered 2026-05-17.
**Incidents:** BE9558H at `166.246.130.1:9034`, 2026-05-17 (Method B) ·
BE12599 at `166.246.64.226:9034`, 2026-09-16 (Method A).
A field unit with a stuck-triggered geophone (or any hardware fault causing
constant event triggering) will record events back-to-back, and if Auto Call
@@ -14,24 +15,30 @@ This runbook describes how to break the loop and recover control.
---
## ⚠ There are TWO failure modes under "wedged unit"
## ⚠ Two cures for one disease — intercept first
Same root cause — an offset/connector fault drives the geophone above trigger,
the unit records back-to-back, ACH set to "after event recorded" dials
constantly — but the *recovery* differs, because the thing blocking you is
different.
Both incidents below are the **same failure**: a geophone offset crosses the
trigger level, the unit records back-to-back, ACH set to "after event
recorded" dials continuously, and the unit becomes unreachable because its
modem is in client mode almost all of the time.
| | **BE9558H (2026-05)** | **BE12599 (2026-09)** |
There are two ways to get a Stop Monitoring command into it.
| | **A — intercept the call** (preferred) | **B — catch it between calls** (original) |
|---|---|---|
| What blocks you | Modem mode-flipping kills inbound TCP | Device never enters S3 mode at all |
| Device state | Alive, in S3 mode, responsive once reached | Stuck repeating an **AT modem-init** string, deaf to S3 |
| Fix direction | **Inbound** — clear Destination, slow-drip a Stop | **Inverted** — point Destination at our own ACH server and let the unit call *us* |
| Winning tool | `scripts/slow_drip.sh` | `bridges/ach_server.py --stop-monitoring` |
| Idea | Be the server it dials. Point the modem's Destination at our own ACH server and answer it. | Clear the Destination so it stops dialing, then race a Stop into the gap. |
| Needs inbound? | **No — the unit calls us** | Yes: working inbound TCP to the modem |
| Determinism | Deterministic — it dials every ~75 s, we only have to be listening | A race. BE9558H took ~7 h of attempts before one landed. |
| Tool | `bridges/ach_server.py --stop-monitoring` | `scripts/slow_drip.sh` |
| Proven on | BE12599, 2026-09-16 | BE9558H, 2026-05-17 |
**Tell them apart with the ALEOS serial log** (see "Turn on ALEOS_SERIAL debug"
below). If the device is emitting `ATQ1/ATE0/ATS0=2` every ~75 s, it is in the
BE12599 mode and **no amount of inbound work will reach it** — skip to
"Second incident" below.
**Method A is the standard procedure now.** The unit won't answer us because
it is on the phone — so stop dialing it and be the one it calls. It rings,
we pick up, take its data, and tell it to stop calling here.
Method B is kept because it is proven, and because A needs a listener the
modem can actually reach (public IP + forwarded port). When you have that,
don't race it — intercept it.
---
@@ -52,9 +59,85 @@ If you see *all* of these, the unit is in this exact failure mode.
---
## Quick reference — how to recover
## Method A (preferred) — intercept the call
You need **ACEmanager access** to the unit's modem.
You need **ACEmanager access** and a host the modem can dial: public IP with
the listener's port forwarded to it.
### A1 — start the listener BEFORE touching the modem
```bash
cd /home/serversdown/seismo-relay
tmux new -s rescue
.venv/bin/python -u bridges/ach_server.py --port 12345 \
-o bridges/captures/<unit>-diag --stop-monitoring -v
```
⚠ **Listener first, always.** A Destination pointed at a dead port is the
worst state available — the device still dials, the modem still flips to
client mode, inbound stays blocked, and nothing gets delivered.
Do **not** add `--events-only` (it silently breaks dedup — see gotchas), and
do **not** add `--disable-ach` yet (see A4).
### A2 — point the modem at it
ACEmanager → **Serial → Port Configuration**:
| Field | Set to |
|---|---|
| **Destination Address** | the listener's public IP |
| **Destination Port** | the listener's port (e.g. `12345`) |
Apply. The modem auto-dials its Destination whenever serial data arrives
while the serial port is closed — so the unit's own retry cycle now lands on
you instead of nowhere.
### A3 — answer, and stop the bleeding
Within ~75 s you should see a call-in. `--stop-monitoring` fires SUB 0x97 at
step 1.5 — after the handshake, **before** the event walk — so the recording
halts at the earliest possible moment in the session. Confirm via
`rescue.json` in the session directory:
```json
{"peer": "166.246.64.226:60921", "stop_monitoring": "ok"}
```
That is the bleeding stopped. Everything after this is cleanup.
### A4 — drain the backlog, THEN disable ACH
⚠ **Order matters, and it is counter-intuitive.** Stopping monitoring also
removes your call-in trigger: ACH fires on "after event recorded", so with
recording stopped the unit has no reason to dial again. The backlog sitting
in its memory does **not** re-arm it.
So if the stored events are worth keeping — and on a fault unit they usually
are, they're the evidence — drain them across however many call-ins it takes
*before* you silence it. Only then add `--disable-ach` (or use
`scripts/rescue_device.sh <host> <port> --no-erase`).
If the unit has gone quiet and you still need it, cycling the modem produces
a call-in, and a unit with a scheduled daily call will dial at its configured
time regardless.
### A5 — restore the Destination, and confirm you did
Put `Destination Address` back to `0.0.0.0` (or the office Instantel ACH
server) once you are finished, and only stop the listener after that is done.
### A6 — do NOT re-enable ACH until the hardware fault is repaired
Otherwise the loop restarts the moment monitoring resumes and you run this
runbook again.
---
## Method B (fallback) — catch it between calls
The original 2026-05 procedure. Use when you cannot stand up a listener the
modem can reach. You need **ACEmanager access** to the unit's modem.
### Step 1: stop the modem's mode-flipping
@@ -287,9 +370,14 @@ recovery: ~7 hours of intermittent debugging across one evening.
level. Constant triggering → constant recording → ACH "after event recorded"
→ continuous dialing. Same disease as BE9558H.
**But the recovery was the opposite direction**, and none of the Step 1–4
procedure above worked. Total time ≈ 5 h, of which ~90 min was spent on two
red herrings documented below.
**Same disease, inverted cure.** Method B's Step 1 *did* work — clearing the
Destination stopped the dial-outs, confirmed in the ALEOS log. It was Step 2
that didn't land, and rather than keep racing we turned the rescue around:
gave the unit a different server to call, and answered it.
Total time ≈ 5 h, of which ~90 min went to two red herrings documented below.
Much of the rest was rediscovering the May procedure, which is why the
"two cures" table now sits at the top of this file.
---
@@ -302,7 +390,7 @@ DEBUG**, then view the serial log.
It is the only thing that tells you what the *device* is actually saying.
Everything before we did this was guesswork.
## What the log showed — the device was never in S3 mode
## What the log showed — the unit is on the phone
Every ~75 seconds, verbatim:
@@ -326,9 +414,15 @@ Read that carefully:
- The device therefore never receives `OK`, never progresses, and **retries
the identical 29 bytes forever**.
**Consequence: the device is not running the S3 protocol parser.** You can
land a byte-perfect S3 frame on it and it will be ignored. This is why every
inbound approach failed, and it is the structural difference from BE9558H.
**While it is in this state it is busy placing a call, not listening for
us.** This is almost certainly what BE9558H was doing too — we simply never
turned on ALEOS_SERIAL debug in May to look. It is not a different disease;
it is the same one, seen properly for the first time.
It is also the argument for Method A in one picture: the unit is mid-dial
every ~75 s, and our inbound Stop has to thread the gaps between those
attempts. Give it somewhere to dial and the problem inverts into a
deterministic one.
### Why `slow_drip` lied
@@ -351,6 +445,12 @@ nobody was reading.
⚠ **`send_error: null` + full duration is NOT success. Only
`bytes_received > 0` is success.**
⚠ **In fairness to slow_drip: it got exactly one attempt here**, run ~90 s
after a modem reboot, with a dead session visible in the log at 20:19:17 in
that same window. BE9558H took hours of attempts before one landed. Method B
was not ruled out on BE12599 so much as abandoned in favour of something that
doesn't need luck.
---
## ⚠ Two red herrings that cost ~90 minutes