Release v0.32.0 — event times you can trust, and Series 4 mapped #54
@@ -2097,8 +2097,39 @@ without a matching string, and the USB enumeration path has not been
|
||||
disassembled.
|
||||
|
||||
**The test that would remove the inference: try an FTDI cable on the same modem.**
|
||||
Either it works or it does not, and nothing is left to argue about. Until then
|
||||
this is well-supported, not proven.
|
||||
Either it works or it does not, and nothing is left to argue about.
|
||||
|
||||
✅ **Corroborated by purchasing history (2026-09-25).** TMI has bought **two**
|
||||
kinds of USB-to-serial cable:
|
||||
|
||||
| cable | chipset | works with a Micromate? |
|
||||
|---|---|---|
|
||||
| **Sabrent** USB 2.0 to serial | **FTDI** | ✅ driver present |
|
||||
| **Benfei** USB to RS-232 | **Prolific PL2303** | ❌ no driver |
|
||||
|
||||
So both types are in circulation and indistinguishable by eye. A unit handed the
|
||||
wrong one has no working modem port, and nothing about the cable's appearance
|
||||
says so.
|
||||
|
||||
⚠ **Identify cables by chipset, not by looks.** `lsusb` on any Linux box:
|
||||
FTDI is VID `0403`, Prolific is `067b`. Note also that counterfeit FTDI chips
|
||||
are common in cheap cables — they carry FTDI's VID but may not behave like one,
|
||||
and an embedded host with a single driver is far less forgiving than Linux.
|
||||
|
||||
### Does this explain the 2026-09-22 field outage?
|
||||
|
||||
⚠ **Not on the timeline as reported.** A Prolific cable does not fail
|
||||
*gradually* — it never enumerates, so the modem port never works at all. The
|
||||
unit was described as working at first and degrading.
|
||||
|
||||
There is one story where it fits: if the initial success was over a **different
|
||||
path** — the USB **PC** port, or a bench test before deployment — that would work
|
||||
regardless of which serial cable was attached. The modem path would then have
|
||||
been broken from the moment it was deployed, and "it worked and then stopped"
|
||||
would be a recollection conflating two connection types days after the fact.
|
||||
|
||||
Plausible, unverified, and recorded as such. Identifying that unit's cable would
|
||||
settle it.
|
||||
|
||||
### How this was isolated — the method is reusable
|
||||
|
||||
|
||||