Release v0.32.0 — event times you can trust, and Series 4 mapped #54
@@ -362,6 +362,53 @@ are DLE-stuffed**, so a parser must destuff before applying field offsets.
|
|||||||
|
|
||||||
---
|
---
|
||||||
|
|
||||||
|
## A/B: Blastware build vs Thor build (2026-09-23)
|
||||||
|
|
||||||
|
UM12947 (`11.0CB`) and UM20147 (`11.0BD`) were each put on the bench and given
|
||||||
|
the identical read-only sweep. **Both answer Series III command frames.**
|
||||||
|
|
||||||
|
| | UM12947 `11.0CB` | UM20147 `11.0BD` |
|
||||||
|
|---|---|---|
|
||||||
|
| POLL answers | ✓ | ✓ |
|
||||||
|
| All 10 read SUBs answer | ✓ | ✓ |
|
||||||
|
| `response_SUB = 0xFF − req` | ✓ | ✓ |
|
||||||
|
| DLE-aware checksum valid | ✓ | ✓ |
|
||||||
|
| Two-step probe/data read | ✓ | ✓ |
|
||||||
|
| ID string | `MM/ISEE/S/IO` | `MM/ISEE/S` |
|
||||||
|
| **flags byte** | **`0xC5`** | **`0x03`** |
|
||||||
|
| **`0x1C` length** | **`0x2C`** | **`0x30`** |
|
||||||
|
|
||||||
|
**The firmware line does not change the wire protocol.** One protocol stack
|
||||||
|
can drive the whole fleet regardless of which build a unit is on. This is the
|
||||||
|
single most consequential finding so far: "standardise the fleet on one
|
||||||
|
firmware" becomes an *optional* convenience rather than a prerequisite for
|
||||||
|
building a call-home receiver.
|
||||||
|
|
||||||
|
### The two differences that do exist
|
||||||
|
|
||||||
|
**1. The flags byte identifies the build.** Response `payload[1]` is `0xC5`
|
||||||
|
on the Blastware line and `0x03` on the Thor line, constant across all ten
|
||||||
|
SUBs on both units. That makes firmware line detectable from *any* response,
|
||||||
|
without reading device info. ⚠ Two units, one each — treat as a strong
|
||||||
|
hypothesis, not a proven encoding.
|
||||||
|
|
||||||
|
Note `0x03` is ETX, so on Thor-line units it arrives DLE-escaped as `10 03`.
|
||||||
|
A parser that fails to destuff will mis-locate every field by one byte on
|
||||||
|
exactly half your fleet.
|
||||||
|
|
||||||
|
**2. `SUB 0x1C` (monitor status) is 4 bytes longer on the Thor line** —
|
||||||
|
`0x30` vs `0x2C` — with four extra trailing bytes (`0f a0 00 00`, purpose
|
||||||
|
unknown).
|
||||||
|
|
||||||
|
⚠ **This breaks relative-to-end parsing.** Series III reads battery and
|
||||||
|
memory from the *end* of the `0x1C` block (`[-10:-8]`, `[-8:-4]`, `[-4:]`).
|
||||||
|
Those offsets are correct on `11.0CB` and wrong on `11.0BD` — applying them
|
||||||
|
blindly to UM20147 yields a battery reading of **577.92 V**. Parse forward
|
||||||
|
from the declared length instead of backward from the end.
|
||||||
|
|
||||||
|
With the offsets shifted by 4, UM20147 reads correctly: battery **3.81 V**,
|
||||||
|
memory 15,000,000 total and free (no events stored).
|
||||||
|
|
||||||
## Divergences from Series III (running list)
|
## Divergences from Series III (running list)
|
||||||
|
|
||||||
1. **No `DLE` prefix on responses** — bare `STX`.
|
1. **No `DLE` prefix on responses** — bare `STX`.
|
||||||
@@ -372,6 +419,8 @@ are DLE-stuffed**, so a parser must destuff before applying field offsets.
|
|||||||
4c. **`1E`/`1F` return the event size**, where Series III returns an offset.
|
4c. **`1E`/`1F` return the event size**, where Series III returns an offset.
|
||||||
4d. **`0x0A` vs `0x0C` split by record type**, not by the `0x46`/`0x2C`
|
4d. **`0x0A` vs `0x0C` split by record type**, not by the `0x46`/`0x2C`
|
||||||
length discriminator Series III uses.
|
length discriminator Series III uses.
|
||||||
|
4e. **Response `payload[1]` (flags) encodes the firmware line** — `0xC5`
|
||||||
|
Blastware, `0x03` Thor — where Series III has a constant `0x10`.
|
||||||
5. **Modem serial rate is 115200**, not 38400 (per TMI provisioning practice;
|
5. **Modem serial rate is 115200**, not 38400 (per TMI provisioning practice;
|
||||||
not independently verified here).
|
not independently verified here).
|
||||||
|
|
||||||
|
|||||||
Reference in new issue
Block a user