Release v0.32.0 — event times you can trust, and Series 4 mapped #54

Merged
serversdown merged 124 commits from dev into main 2026-10-27 11:01:01 -04:00
Showing only changes of commit 492b6683a4 - Show all commits
+42 -10
View File
@@ -50,18 +50,50 @@ machines, and units that will not cross over. It is not a misconfiguration.
**Instantel ships different firmware for different host software**, and the
wire protocol differs with it.
Versions observed so far:
### Fleet audit (physical, 2026-09-22/23)
| where | version | notes |
|---|---|---|
| bench unit UM12947, today | **`11.0CB`** | Blastware line — answers Series III |
| corpus, 932 event files | `11.0AK` | produced Thor-collected `.IDFW`/`.IDFH` |
| corpus, 83 event files | `10.90GC` | older; UM12947's own files from Sept 2025 |
| unit | firmware | line | location |
|---|---|---|---|
| UM11719 | `11.0CB` | Blastware | Ped Bridge Loc 2 |
| UM6047 | `11.0CB` | Blastware | Brookville Loc 9 |
| UM12947 | `11.0CB` | Blastware | **bench** |
| UM14133 | `11.0CB` | Blastware | Pitt-Music Bldg Loc 1 |
| UM11402 | `11.0BD` | Thor | Ped Bridge Loc 1 |
| UM20147 | `11.0BD` | Thor | **bench** |
| UM13981 | `11.0AK` | pre-split | RKM Loc 1 |
| UM20146 | `11.0AK` | pre-split | Karns Loc 2 |
| UM12420 | `10.90GC` | pre-split | RKM Loc 2 |
Note UM12947 produced `10.90GC` files in production last year and reports
`11.0CB` on the bench now — so **it has been reflashed at some point**, and
firmware is not stable per-unit over time. Any fleet-wide claim needs a
per-unit firmware audit first.
**4 Blastware / 2 Thor / 3 pre-split.** The Blastware line is already the
plurality, which makes "standardise on Blastware" less disruptive than it
first appeared.
A store-derived audit (firmware is recorded in every `.sfm.json` sidecar as
`extensions.idf_report.version`) agreed with the physical audit on **7 of 9**.
The two that differed — UM6047 and UM14133 — are the most recently deployed
units, reflashed after their last stored event. Useful technique: the fleet's
firmware history is reconstructable from the store without touching a unit,
but it lags reality by one deployment.
Firmware is **not stable per-unit over time** — five of nine have been
reflashed at least once. Any fleet-wide claim needs a fresh audit.
### ⚠ Retraction: firmware line does NOT determine Thor compatibility
An earlier draft of this document suggested that UM12947's trouble with Thor
was explained by its being on the Blastware build. **That is not supported.**
Ped Bridge runs UM11402 (`11.0BD`) and UM11719 (`11.0CB`) side by side, both
deployed 2026-04-20, and **both call Thor successfully** — UM11719 has 331
Thor-collected events in the store while on `11.0CB`, through 2026-08-23.
So a Blastware-line unit does feed Thor. Whatever the CB/BD split changes, it
is not "which host software can collect from it", and UM12947's specific
problem remains unexplained.
**What is actually established:** a `11.0CB` unit answers Series III command
frames. Whether a `11.0BD` unit does is **untested** — and UM20147 (`11.0BD`)
is on the bench, which makes that a direct A/B away.
The version is readable over the wire from `SUB 0x01` as two separate ASCII
runs — `"0CB"` then `"11"` — with no single concatenated string.