@@ -2143,10 +2143,18 @@ the failure began when the pair was split across two cables and one of them was
the wrong chipset. **One working and one not, set up side by side, is the
signature of a mixed cable supply** — not of a unit fault.
⚠ Still worth confirming: whether that cable travelled home *with* the unit, or
was picked up separately at the bench. If it travelled with it, there is no
inference left. Either way the FTDI cable arriving 2026-09-26 settles it — same
unit, same modem, one variable.
✅ **The cable travelled home with the unit** (confirmed 2026-09-25). So the
Benfei on the bench is physically the cable from the failed office setup — not a
substitute picked up later. There is no inference left in the chain:
| | |
|---|---|
| the cable from the failed unit | **is** a Benfei / Prolific PL2303 |
| the Micromate's USB host | has **no** Prolific driver, in either firmware line |
| a laptop on that same cable + modem | round-trips perfectly |
| the Micromate on it | never answered |
The FTDI cable arriving 2026-09-26 is now a confirmation rather than a test.
**Corollary worth knowing.** The standing workaround — *"just deploy a Series III
instead"* — works partly because a **MiniMate Plus has a DB-9 port directly on the