Release v0.32.0 — event times you can trust, and Series 4 mapped #54
@@ -1091,7 +1091,7 @@ zero-data response. Thor follows each with a `SUB 0x1C` status read to confirm.
|
|||||||
### `SUB 0x1C` — monitor status, 55-byte data
|
### `SUB 0x1C` — monitor status, 55-byte data
|
||||||
|
|
||||||
```
|
```
|
||||||
data[12] monitoring flag 0x0E monitoring / 0x00 idle
|
data[12] monitoring flag NON-ZERO = monitoring, 0x00 = idle
|
||||||
data[13] day
|
data[13] day
|
||||||
data[14] month
|
data[14] month
|
||||||
data[15:17] year, uint16 BE
|
data[15:17] year, uint16 BE
|
||||||
@@ -1103,8 +1103,11 @@ data[-8:-4] memory total, uint32 BE = 15,000,000 bytes exactly
|
|||||||
data[-4:] memory free, uint32 BE
|
data[-4:] memory free, uint32 BE
|
||||||
```
|
```
|
||||||
|
|
||||||
⚠ **The monitoring flag is `0x0E`, not Series III's `0x10`.** Do not reuse the
|
⚠ **Test `data[12]` for non-zero — do NOT compare against a constant.** It read
|
||||||
Series III constant. Only five bytes differ between the monitoring and idle
|
`0x0E` on 2026-09-24 and `0x0C` on 2026-09-25, both while monitoring, so it
|
||||||
|
carries sub-state in its low bits rather than being a flag. Series III's `0x10`
|
||||||
|
does not apply either. `SUB 0x49` `data[11]` read `0x02` in every monitoring
|
||||||
|
sample so far, but the same caution applies until more values are seen. Only five bytes differ between the monitoring and idle
|
||||||
responses: the flag, `data[17]`, the clock, and the memory-free field.
|
responses: the flag, `data[17]`, the clock, and the memory-free field.
|
||||||
|
|
||||||
Memory free dropped by exactly **4,096 bytes** across the ~70-second monitoring
|
Memory free dropped by exactly **4,096 bytes** across the ~70-second monitoring
|
||||||
@@ -1877,6 +1880,33 @@ It explains, too, why the field symptom resisted characterisation. The unit is
|
|||||||
perfectly reachable throughout. Poke it and it answers. Nothing is wrong with
|
perfectly reachable throughout. Poke it and it answers. Nothing is wrong with
|
||||||
the unit, the modem, or the link.
|
the unit, the modem, or the link.
|
||||||
|
|
||||||
|
### The stale display extends to monitoring state — a live unit shown as Idle
|
||||||
|
|
||||||
|
The operator's own test: start monitoring **from the unit's keypad**, and see
|
||||||
|
whether THOR ever notices.
|
||||||
|
|
||||||
|
At **13:21:15**, verified by an independent probe through the same relay:
|
||||||
|
|
||||||
|
```
|
||||||
|
unit, on the wire SUB 0x49 data[11] = 0x02 MONITORING
|
||||||
|
SUB 0x1C data[12] = 0x0c MONITORING
|
||||||
|
device clock 13:21:14
|
||||||
|
THOR's last contact 13:10:37 (a stop-monitoring command it issued itself)
|
||||||
|
THOR's display Connected · Monitoring Mode: Idle · Last Updated 1:04:22 PM
|
||||||
|
```
|
||||||
|
|
||||||
|
**A unit is actively recording and THOR shows it as Idle.** Had an event
|
||||||
|
triggered in that window, THOR would not have known and would not have collected
|
||||||
|
it. This is the operational consequence of the wedge: not merely a wrong
|
||||||
|
indicator, but a monitoring system that has silently stopped monitoring its
|
||||||
|
monitor.
|
||||||
|
|
||||||
|
One further detail. THOR **did** contact the unit at 13:10:37 and read
|
||||||
|
`MONITOR_STATUS` — yet `Last Updated` still reads 1:04:22. So a successful
|
||||||
|
exchange does not refresh that timestamp; only the full status check does. The
|
||||||
|
single honest field on the screen is honest about the wrong thing, which makes it
|
||||||
|
useless as a staleness indicator precisely when staleness is the problem.
|
||||||
|
|
||||||
### Design consequences for SFM
|
### Design consequences for SFM
|
||||||
|
|
||||||
0. **Connection status must EXPIRE.** The single highest-value fix here. If the
|
0. **Connection status must EXPIRE.** The single highest-value fix here. If the
|
||||||
|
|||||||