Release v0.34.0 — Series-4 parity, and most of Blastware's report from the event file alone #59
@@ -418,6 +418,16 @@ Each safety piece arrives with its first real use, so nothing sits untested.
|
||||
wake repeats every 5 s up to 60 s). A third was cancelled at the
|
||||
keypad mid sensor check; SFM read it back honestly as a mismatch
|
||||
("unit reads idle", 30 reads), audit `b019ae6d`.
|
||||
- 🔶 **A stop starts a call-home when auto-call-home is on** (Brian,
|
||||
2026-10-27): stopping by command or keypad makes the unit call home and
|
||||
upload its events + MLG, which can take minutes. UM20147's `0x01` for
|
||||
~18 s after a stop (10-25) fits; BE9441 never called home (ACH
|
||||
presumably off). For SFM: after a stop the read-back may meet a
|
||||
call-home state, a ring, or silence, and on cellular the unit contends
|
||||
for the modem's one TCP session SFM may be holding. **Series 3's 0x1C
|
||||
value during a call-home is unmeasured** -- the change path would
|
||||
refuse it as unknown. To measure: a MiniMate stop with ACH on, through
|
||||
the bridge.
|
||||
- Still owed:
|
||||
- **before merging to dev, align `persist_download`'s already-stored
|
||||
check with dev's restamp fix** (fix/waveform-time-restamp,
|
||||
|
||||