# Phase 3: start / stop monitoring, the first live write — Implementation Plan > **For agentic workers:** REQUIRED SUB-SKILL: Use superpowers:subagent-driven-development (recommended) or superpowers:executing-plans to implement this plan task-by-task. Steps use checkbox (`- [ ]`) syntax for tracking. **Task 7 is not for a subagent**: the controller runs it at the bench, with Brian approving the send. **Goal:** Start and stop monitoring by serial, for both series, through the safety layer. The layer covers: - an identity check right before the write; - the unit's state, read first; - a dry run that returns the exact frames; - an audit row for every attempt; - a read-back that decides success; - never a retried write. Then make the project's first live send to a unit (UM20147), with Brian approving it at the bench. **Architecture:** - **Series 4.** `MicromateProtocol.execute(plan)` becomes the only write path. - It transmits exactly a `WritePlan`'s frames, granting permission for that call alone. - It refuses any plan beyond a code-reviewed ceiling, `ENABLEABLE_WRITES = {0x96, 0x97}`. - It refuses on a connection whose replies did not carry Micromate flags. - **The change operation**, one per series: 1. re-reads the serial; 2. reads the run state, `0x49` on Series 4 and `0x1C` on Series 3; 3. skips or refuses as appropriate; 4. audits the attempt; 5. sends; 6. polls the state until it matches or a window expires; 7. audits the result. It returns a `ChangeResult`. - **No retry.** `DeviceAccess` runs change operations with no retry of any kind. - **Audit.** The table lives in SDM (`device_commands` in SeismoDb). The device side writes to it through an injected sink, so `sfm/device_access/` still imports no DB. - **Endpoints.** The existing `POST /device/monitor/start` and `/stop` gain `serial`, `dry_run`, `even_if_already` and `caller`. Without `serial` they behave exactly as today. **Tech Stack:** Python 3, FastAPI, pytest, SQLite. **Spec:** `docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md`, in particular: - *The safety layer* (tiers, identity, read-back, audit table); - *Series-4 write path* (dry run, start/stop row); - *Testing → Bench* (the first-live-send protocol); - **Build phases item 3, including its "Preconditions from the phase-2 final review" block**, which is binding. ## Global Constraints - **Writes.** The only writes are `0x96`/`0x97`. - On Series 4 they go through `MicromateProtocol.execute(plan)` only. - On Series 3 they go through `MiniMateClient.start_monitoring()`/`stop_monitoring()`, which are unchanged. - Nothing else may be transmitted that changes a unit. - Tests use fakes and scripted transports only. **No task but Task 7 touches hardware**, and Task 7 sends only with Brian's explicit go-ahead in the conversation. - **A write is never retried.** - This covers the retry-once in `DeviceAccess._dispatch`, the TCP retry in `Series3Adapter._run`, and any loop. - The structure must make retry impossible for a change operation. A comment saying so is not enough. - **Identity before any write.** - The serial is read from the unit immediately before the write: `0x15` on Series 4, `connect()` on Series 3. - A mismatch means 409 and nothing is sent. - **0xA2 is never sent.** - The ceiling is `{0x96, 0x97}`. - Raising it is a spec change. - **The audit is mandatory.** - A live change without an audit sink is refused (501), with nothing sent. - Every live attempt, including a skip, gets one row. - Dry runs are not audited: they send nothing. - **Legacy routes are unchanged without `serial`.** - The only permitted change: `baud` becomes `Optional[int] = Query(None)` with 38400 applied at the top of the legacy path. - **Schema.** The `device_commands` table is additive (`CREATE TABLE IF NOT EXISTS`). At merge, the changelog must say "schema: new table, no migration, no backfill". - **Tiers.** `sfm/device_access/` must not import `sfm.database`, `sfm.waveform_store` or `sfm.server`. - **The suite.** Its 16 pre-existing failures stay exactly that set. - **Branch.** Work on `feat/series4-sfm-parity`. Never commit to `main` or `dev`, do not edit `CHANGELOG.md`, and run `git push` after each commit. - **Never move or modify `bridges/captures/`.** Port 8200 is another SFM. - **Trailers.** Every commit message ends with: ``` Co-Authored-By: Claude Opus 5.5 Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44 ``` ## Review Focus These are the five things most likely to bite, each with the task that tests it: 1. **A lost ack.** It must never cause a second send, whether through the held-line retry, the Series-3 cold-boot retry, or anything else (Tasks 4, 5). 2. **A different unit at the address**, for example after a modem moved. It must be refused before any write, on a fresh serial read rather than the call-up's cached serial (Tasks 4, 5). 3. **A unit that is calling home** (`0x49` = `0x03`). It is refused, with nothing sent (Task 4). 4. **A clean ack where the state never changes.** This also covers a sensor check that delays the change. It is reported as a failed change after the verify window, audited as `mismatch`, with every state value seen recorded (Tasks 4, 5). 5. **Write permission leaking.** - After `execute()`, the same connection cannot send `0x96` again without another `execute()`. - A plan holding `0x7E` (call-home) is refused by the ceiling. - A connection that has seen no Micromate-flagged reply cannot execute anything (Task 1). --- ## File map | file | change | |---|---| | `micromate/opcodes.py` | `ENABLEABLE_WRITES` | | `micromate/protocol.py` | `execute(plan)`; `_send(frame, *, permit=)`; `_validate` records `last_flags` | | `micromate/models.py` | `MicromateRunState` | | `micromate/client.py` | `serial(refresh=)`; `read_run_state()` | | `sfm/database.py` | `device_commands` table; `begin_device_command`, `finish_device_command`, `query_device_commands` | | `sfm/audit.py` | **new**: `DbAudit`, the SDM sink | | `sfm/device_access/models.py` | `ChangeResult` | | `sfm/device_access/s4.py` | `s4_change_monitoring` | | `sfm/device_access/s3.py` | `Series3Adapter.change_monitoring`; `_run(retry=)` | | `sfm/device_access/access.py` | `start_monitoring`/`stop_monitoring`; `_dispatch(change=)`; `audit=`, `sleep=`, `verify_*` | | `sfm/server.py` | serial branch on `/device/monitor/start` and `/stop`; guard paths; `GET /db/device_commands`; audit wiring | | `tests/device_access_fakes.py` | run states, `execute`, Series-3 start/stop | | tests | `test_micromate_execute.py`, `test_device_commands.py`, `test_device_access_change.py`, plus additions to `test_micromate_client.py` and `test_server_device_access.py` | --- ### Task 1: `execute(plan)`, the ceiling and the flags check **Files:** - Modify: `micromate/opcodes.py` - Modify: `micromate/protocol.py` (`__init__`, `_validate`, `_send`, new `execute`, module docstring) - Test: `tests/test_micromate_execute.py` (new) **Interfaces:** - Produces: - `ENABLEABLE_WRITES: frozenset` - `MicromateProtocol.execute(plan: WritePlan) -> list[MicromateFrame]`, returning the ack frames - `MicromateProtocol.last_flags: Optional[int]` - [ ] **Step 1: Write the failing tests.** Create `tests/test_micromate_execute.py`: ```python """execute(plan): the only write path -- exact frames, one call's permission, a ceiling.""" from __future__ import annotations import os import sys import pytest sys.path.insert(0, os.path.dirname(os.path.dirname(os.path.abspath(__file__)))) from micromate.framing import FLAGS_THOR, build_request # noqa: E402 from micromate.opcodes import ENABLEABLE_WRITES, OpcodeRefused # noqa: E402 from micromate.protocol import ACK_DATA_LEN, UnexpectedResponse # noqa: E402 from micromate.writes import ( # noqa: E402 plan_call_home_enabled, plan_start_monitoring, plan_stop_monitoring, ) from test_micromate_protocol import frame, proto # noqa: E402 ACK_START = frame(0x69, bytes(ACK_DATA_LEN)) POLL_RSP = frame(0xA4, bytes(48)) def _ready(responses): """A protocol that has already seen one Micromate-flagged reply (a POLL).""" p, t = proto([POLL_RSP] + responses) p.poll() t.written.clear() return p, t def test_the_ceiling_is_start_and_stop_only(): assert ENABLEABLE_WRITES == frozenset({0x96, 0x97}) def test_execute_sends_exactly_the_plans_frame_and_returns_the_ack(): p, t = _ready([ACK_START]) acks = p.execute(plan_start_monitoring()) assert t.written == [build_request(0x96, 0xFFFF)] assert [a.sub for a in acks] == [0x69] def test_permission_ends_with_the_call(): p, t = _ready([ACK_START]) p.execute(plan_start_monitoring()) with pytest.raises(OpcodeRefused): p._send(build_request(0x96, 0xFFFF)) assert len(t.written) == 1 def test_a_plan_beyond_the_ceiling_is_refused_and_nothing_is_sent(): content = bytes(5) + b"\x05" + bytes(120) p, t = _ready([]) with pytest.raises(OpcodeRefused, match="ceiling"): p.execute(plan_call_home_enabled(content, False)) assert t.written == [] def test_execute_refuses_before_any_micromate_reply_has_been_seen(): p, t = proto([ACK_START]) with pytest.raises(OpcodeRefused, match="Micromate"): p.execute(plan_start_monitoring()) assert t.written == [] def test_execute_refuses_after_a_non_micromate_flags_reply(): p, t = proto([frame(0xA4, bytes(48), flags=0x10), ACK_START], strict_checksums=False) p.poll() t.written.clear() with pytest.raises(OpcodeRefused): p.execute(plan_stop_monitoring()) assert t.written == [] def test_a_wrong_ack_sub_raises_after_one_send(): p, t = _ready([frame(0x68, bytes(ACK_DATA_LEN))]) # stop's ack, for a start with pytest.raises(UnexpectedResponse): p.execute(plan_start_monitoring()) assert len(t.written) == 1 def test_thor_flags_are_accepted_too(): p, t = proto([frame(0xA4, bytes(48), flags=FLAGS_THOR), frame(0x68, bytes(ACK_DATA_LEN))]) p.poll() p.execute(plan_stop_monitoring()) assert t.written[-1] == build_request(0x97, 0xFFFF) ``` Before you write any code, make two checks: - In `tests/test_micromate_protocol.py`, check that `ScriptedTransport` exposes `.written` as a list you can `.clear()`, and that `frame(..., flags=)` exists. - Check that `micromate.framing` exports `FLAGS_THOR` and `FLAGS_BLASTWARE`. If either differs, adapt the helper calls only, and say so. - [ ] **Step 2: Run the tests and watch them fail.** Run: `.venv/bin/python -m pytest tests/test_micromate_execute.py -q` Expected: an ImportError on `ENABLEABLE_WRITES`. - [ ] **Step 3: Implement.** In `micromate/opcodes.py`, add after the tables: ```python # The write SUBs any code may transmit at all -- through MicromateProtocol.execute, # one call at a time. Raising this is a spec change, reviewed like one: delete # (0xA8/0xAA), raw file writes (0x8D/0x8E -- they could reach the modem # settings) and 0xA2 never belong here without one. ENABLEABLE_WRITES = frozenset({0x96, 0x97}) ``` In `micromate/protocol.py`: - Imports: add `FLAGS_BLASTWARE` and `FLAGS_THOR` to the `from .framing import ...` line. Add `ENABLEABLE_WRITES` and `OpcodeRefused` to the `.opcodes` import. - In `__init__`, add `self.last_flags: Optional[int] = None`. - In `_validate`, before `return frame`, add `self.last_flags = frame.flags`. - Replace `_send` with: ```python def _send(self, frame: bytes, *, permit: frozenset = frozenset()) -> None: # The gate sees the bytes about to go out, not the caller's intent. # `permit` is passed only by execute(), for exactly one step. frame = bytes(frame) check_transmit(request_sub(frame), self._allowed_writes | permit) log.debug("TX %d bytes: %s", len(frame), frame.hex()) self._transport.write(frame) ``` - Add after `_exchange`: ```python def execute(self, plan) -> list: """Transmit a `micromate.writes.WritePlan` -- the only write path. Permission is granted for this call alone, one step at a time, and only for SUBs under `ENABLEABLE_WRITES`. Refuses unless the last reply on this connection carried Micromate flags (0xC5 / 0x03). Each step's ack is awaited before the next is sent. Never retries: a timeout raises, and what happened is the caller's to find out by reading the unit back. """ beyond = plan.subs - ENABLEABLE_WRITES if beyond: raise OpcodeRefused( "plan holds SUB(s) above the write ceiling: " + ", ".join(f"0x{s:02X}" for s in sorted(beyond))) if self.last_flags not in (FLAGS_BLASTWARE, FLAGS_THOR): raise OpcodeRefused( "no Micromate-flagged reply seen on this connection -- " "refusing to write") acks = [] for step in plan.steps: self._parser.reset() self._pending.clear() self._send(step.frame, permit=frozenset({step.sub})) acks.append(self._recv_one(expected_sub=step.ack_sub, reset_parser=False)) return acks ``` - Module docstring: in the scope paragraph, add one sentence. `execute(plan)` is the only write path: one call's permission, at most `ENABLEABLE_WRITES`, and only on a connection that has had Micromate-flagged replies. - [ ] **Step 4: Run the tests and watch them pass.** Run: `.venv/bin/python -m pytest tests/test_micromate_execute.py tests/test_micromate_opcodes.py tests/test_micromate_protocol.py -q` Expected: all pass. - [ ] **Step 5: Full suite, then commit and push.** Message: `feat(micromate): execute(plan) -- the only write path, one call's permission, a start/stop ceiling`. --- ### Task 2: The unit's run state and a fresh serial **Files:** - Modify: `micromate/models.py`, `micromate/client.py` - Test: `tests/test_micromate_client.py` (append) **Interfaces:** - Produces: - `MicromateRunState(code: int, name: str)`, where `name` is one of `"idle"` / `"monitoring"` / `"calling_home"` / `"unknown(0xNN)"`. - `MicromateClient.read_run_state() -> MicromateRunState` - `MicromateClient.serial(*, refresh: bool = False) -> str` - [ ] **Step 1: Write the failing tests.** Append to `tests/test_micromate_client.py`: ```python # ── run state (0x49) and a fresh serial ─────────────────────────────────────── import json as _json # noqa: E402 _WRITES = Path(__file__).parent / "fixtures" / "micromate-writes" def _captured_0x49(): ex = _json.loads((_WRITES / "monitor-start-stop.json").read_text())["exchanges"] return [bytes.fromhex(e["response_data"]) for e in ex if e["sub"] == 0x49] def test_run_state_names_thors_captured_states(): datas = _captured_0x49() names = [] for d in datas: mm, _t = client([frame(0xB6, d)]) names.append(mm.read_run_state().name) assert "idle" in names and "monitoring" in names @pytest.mark.parametrize("code,name", [(0x00, "idle"), (0x02, "monitoring"), (0x03, "calling_home"), (0x07, "unknown(0x07)")]) def test_run_state_names_every_value(code, name): data = bytes(11) + bytes([code]) + bytes(4) mm, _t = client([frame(0xB6, data)]) st = mm.read_run_state() assert (st.code, st.name) == (code, name) def test_serial_refresh_reads_the_unit_again(): mm, t = client([frame(0xEA, bytes(11) + b"UM20147\x00\x00\x00")]) mm._serial = "UM99999" # a stale cached serial assert mm.serial(refresh=True) == "UM20147" assert len(t.written) == 1 ``` Before running, check two things: - `client()` and `frame()` exist in this file with those shapes. If they differ, adapt the calls. - The response SUBs: `0x49` gives `0xFF-0x49 = 0xB6`, and `0x15` gives `0xEA`. - [ ] **Step 2: Run the tests and watch them fail.** Run: `.venv/bin/python -m pytest tests/test_micromate_client.py -q -k "run_state or refresh"` Expected: `AttributeError: ... 'read_run_state'`. - [ ] **Step 3: Implement.** In `micromate/models.py`, add: ```python @dataclass(frozen=True) class MicromateRunState: """`SUB 0x49` content[0]. 0x00 idle, 0x02 monitoring, 0x03 calling home (ten of ten in the 10-14 ACH session); anything else is named, not guessed.""" code: int name: str ``` In `micromate/client.py`: - Import `MicromateRunState`. - Change `serial` to `def serial(self, *, refresh: bool = False) -> str:`, with the condition `if refresh or self._serial is None:`. Document that a change operation refreshes it right before writing. - Add after `serial`: ```python _RUN_STATES = {0x00: "idle", 0x02: "monitoring", 0x03: "calling_home"} def read_run_state(self) -> MicromateRunState: """`SUB 0x49` -> the unit's run state, with the in-call-home value kept (connect()'s `monitoring` bool flattens 0x03 into True).""" c = _content(self._proto.read_state()) if not c: raise ProtocolError("0x49 state reply has no content") code = c[_STATE_FLAG] return MicromateRunState(code, self._RUN_STATES.get(code, f"unknown(0x{code:02X})")) ``` - [ ] **Step 4: Run the tests and watch them pass.** Then run the full suite, commit and push. Message: `feat(micromate): read the unit's run state and refresh its serial on demand`. --- ### Task 3: The audit table and its sink **Files:** - Modify: `sfm/database.py`. Add to `_SCHEMA`, plus three methods. - Create: `sfm/audit.py` - Modify: `sfm/server.py`, for `GET /db/device_commands` - Test: `tests/test_device_commands.py` (new) **Interfaces:** - Produces: - `SeismoDb.begin_device_command(*, serial, series, address, action, tier, request: dict, caller: str) -> str` - `SeismoDb.finish_device_command(command_id, *, outcome, verified, ack: Optional[dict], error: Optional[str]) -> None` - `SeismoDb.query_device_commands(serial=None, limit=100) -> list[dict]`, newest first, with `request` and `ack` parsed back from JSON. - `DbAudit(get_db: Callable[[], SeismoDb])` with `begin(**fields) -> str` and `finish(command_id, **fields)`. - [ ] **Step 1: Write the failing tests** in `tests/test_device_commands.py`: ```python """The device_commands audit table: one row per change attempt, pending -> result.""" from __future__ import annotations import os import sys sys.path.insert(0, os.path.dirname(os.path.dirname(os.path.abspath(__file__)))) from sfm.audit import DbAudit # noqa: E402 from sfm.database import SeismoDb # noqa: E402 ROW = dict(serial="UM20147", series=4, address="serial:/dev/ttyACM0", action="start monitoring", tier="change", request={"plan": {"steps": [{"sub": "0x96"}]}}, caller="test") def test_a_command_is_pending_until_finished(tmp_path): db = SeismoDb(tmp_path / "s.db") cid = db.begin_device_command(**ROW) row = db.query_device_commands(serial="UM20147")[0] assert (row["id"], row["outcome"], row["verified"]) == (cid, "pending", None) assert row["request"]["plan"]["steps"][0]["sub"] == "0x96" db.finish_device_command(cid, outcome="sent", verified="ok", ack={"subs": ["0x69"]}, error=None) row = db.query_device_commands(serial="UM20147")[0] assert (row["outcome"], row["verified"], row["ack"]) == ("sent", "ok", {"subs": ["0x69"]}) assert row["finished_ts"] def test_the_table_survives_reopening(tmp_path): SeismoDb(tmp_path / "s.db").begin_device_command(**ROW) assert len(SeismoDb(tmp_path / "s.db").query_device_commands()) == 1 def test_db_audit_is_a_thin_sink(tmp_path): db = SeismoDb(tmp_path / "s.db") audit = DbAudit(lambda: db) cid = audit.begin(**ROW) audit.finish(cid, outcome="failed", verified="mismatch", ack=None, error="unit reads idle") row = db.query_device_commands()[0] assert (row["outcome"], row["verified"], row["error"]) == ("failed", "mismatch", "unit reads idle") ``` - [ ] **Step 2: Run them, and watch them fail** (ImportError). - [ ] **Step 3: Implement.** - Append to `_SCHEMA` in `sfm/database.py`: ```sql CREATE TABLE IF NOT EXISTS device_commands ( id TEXT PRIMARY KEY, ts TEXT NOT NULL, -- when the attempt began (UTC ISO) finished_ts TEXT, serial TEXT NOT NULL, series INTEGER NOT NULL, address TEXT NOT NULL, action TEXT NOT NULL, -- e.g. "start monitoring" tier TEXT NOT NULL, -- "change" | "destructive" request TEXT, -- JSON: parameters + the planned frames (hex) ack TEXT, -- JSON: ack summary verified TEXT, -- "ok" | "mismatch" | "not_checked" outcome TEXT NOT NULL, -- "pending" | "sent" | "skipped" | "failed" error TEXT, caller TEXT -- "webapp" | "terra-view" | "script" | ... ); CREATE INDEX IF NOT EXISTS idx_device_commands_serial ON device_commands(serial, ts); ``` - Add the methods to `SeismoDb`. Use the existing `_new_id()` and `_connect()`, `json.dumps`/`json.loads` for `request`/`ack`, and a UTC ISO string for `ts`/`finished_ts`. - `query_device_commands` orders by `ts DESC`. - Create `sfm/audit.py`: ```python """The SDM side of the device audit: DeviceAccess calls this sink, which writes SeismoDb.device_commands. Kept here so sfm.device_access never imports the DB.""" from __future__ import annotations from typing import Any, Callable class DbAudit: def __init__(self, get_db: Callable[[], Any]) -> None: self._get_db = get_db def begin(self, **fields: Any) -> str: return self._get_db().begin_device_command(**fields) def finish(self, command_id: str, **fields: Any) -> None: self._get_db().finish_device_command(command_id, **fields) ``` - Add `GET /db/device_commands`, with params `serial: Optional[str] = Query(None)` and `limit: int = Query(100)`. It returns `{"commands": _get_db().query_device_commands(serial, limit)}`. Put it near the other `/db/*` reads. - [ ] **Step 4: Run the tests and watch them pass.** Then run the full suite, commit and push. Message: `feat(sfm): device_commands audit table -- one row per change attempt`. --- ### Task 4: The Series-4 change operation and `DeviceAccess.start_monitoring` / `stop_monitoring` **Files:** - Modify: `sfm/device_access/models.py` (add `ChangeResult`), `sfm/device_access/s4.py` (add `s4_change_monitoring`), `sfm/device_access/access.py`, `sfm/device_access/__init__.py` - Modify: `tests/device_access_fakes.py` - Test: `tests/test_device_access_change.py` (new) **Interfaces:** - Consumes: Task 1 (`execute`), Task 2 (`read_run_state`, `serial(refresh=)`), and the phase-2 `plan_start_monitoring`/`plan_stop_monitoring`. - Produces: - `ChangeResult`, with the fields shown below. - `DeviceAccess(..., audit=None, sleep=time.sleep, verify_timeout_s=60.0, verify_interval_s=2.0)` - `DeviceAccess.start_monitoring(address, serial, *, dry_run=False, even_if_already=False, caller="api") -> ChangeResult`, and the same for `stop_monitoring`. - [ ] **Step 1: Extend the fakes.** In `tests/device_access_fakes.py`, give `FakeMicromateClient` these keyword arguments: - `run_states=None`, a list of codes consumed one per `read_run_state()` call, with the last one repeating; it defaults to `[0x00]`; - `serial_reported=None`, defaulting to `serial`; - `execute_error=None`. Then add these methods: ```python def serial(self, *, refresh=False): self._step("serial") return self.serial_reported def read_run_state(self): from micromate.models import MicromateRunState self._step("read_run_state") code = self.run_states.pop(0) if len(self.run_states) > 1 else self.run_states[0] names = {0x00: "idle", 0x02: "monitoring", 0x03: "calling_home"} return MicromateRunState(code, names.get(code, f"unknown(0x{code:02X})")) ``` On `_FakeProtocol`, add: ```python def execute(self, plan): self._owner._step("execute") self._owner.executed.append(plan) if self._owner.execute_error is not None: raise self._owner.execute_error return [type("Ack", (), {"sub": s.ack_sub, "data": bytes(11)})() for s in plan.steps] ``` Also add `self.executed = []` to `FakeMicromateClient.__init__`. - [ ] **Step 2: Write the failing tests** in `tests/test_device_access_change.py`: ```python """Start/stop through the safety layer: identity, state, dry run, audit, read-back, no retry.""" from __future__ import annotations import os import sys import pytest sys.path.insert(0, os.path.dirname(os.path.dirname(os.path.abspath(__file__)))) import micromate.protocol as mm_proto # noqa: E402 from device_access_fakes import FakeMicromateClient # noqa: E402 from sfm.device_access.access import DeviceAccess # noqa: E402 from sfm.device_access.errors import ( # noqa: E402 CallHomeInProgress, NotSupported, WrongUnit, ) from sfm.device_access.models import UnitAddress # noqa: E402 ADDR = UnitAddress(host="10.0.0.5", tcp_port=9034, port=None, baud=None) class Audit: def __init__(self): self.rows = {} def begin(self, **f): cid = f"c{len(self.rows) + 1}" self.rows[cid] = dict(f, outcome="pending") return cid def finish(self, cid, **f): self.rows[cid].update(f) def access(fake, audit="default"): return DeviceAccess(s4_client_factory=lambda a, **kw: fake, start_reaper=False, audit=Audit() if audit == "default" else audit, sleep=lambda s: None, verify_timeout_s=6.0, verify_interval_s=2.0) def test_a_dry_run_returns_the_exact_frames_and_sends_nothing(): fake = FakeMicromateClient(run_states=[0x00]) da = access(fake) r = da.start_monitoring(ADDR, "UM20147", dry_run=True) assert (r.dry_run, r.sent, r.outcome, r.state_before) == (True, False, "dry_run", "idle") assert r.plan["steps"][0]["frame"] == "41021010009600ffff00000000000000000000a403" assert fake.executed == [] and da._audit.rows == {} def test_start_sends_once_and_verifies_by_reading_back(): fake = FakeMicromateClient(run_states=[0x00, 0x00, 0x02]) da = access(fake) r = da.start_monitoring(ADDR, "UM20147") assert (r.sent, r.outcome, r.verified, r.state_after) == (True, "sent", "ok", "monitoring") assert len(fake.executed) == 1 row = da._audit.rows[r.audit_id] assert (row["outcome"], row["verified"], row["action"]) == ("sent", "ok", "start monitoring") assert r.detail["states_seen"][-1] == "monitoring" def test_already_in_state_is_skipped_and_audited_as_skipped(): fake = FakeMicromateClient(run_states=[0x02]) da = access(fake) r = da.start_monitoring(ADDR, "UM20147") assert (r.sent, r.outcome) == (False, "skipped") and fake.executed == [] assert da._audit.rows[r.audit_id]["outcome"] == "skipped" def test_even_if_already_sends_anyway(): fake = FakeMicromateClient(run_states=[0x02]) r = access(fake).start_monitoring(ADDR, "UM20147", even_if_already=True) assert r.sent and len(fake.executed) == 1 def test_a_unit_calling_home_is_refused_and_nothing_is_sent(): fake = FakeMicromateClient(run_states=[0x03]) with pytest.raises(CallHomeInProgress): access(fake).stop_monitoring(ADDR, "UM20147") assert fake.executed == [] def test_a_different_unit_on_the_fresh_serial_read_is_refused_before_writing(): fake = FakeMicromateClient(serial_reported="UM12947") with pytest.raises(WrongUnit): access(fake).start_monitoring(ADDR, "UM20147") assert fake.executed == [] def test_a_live_change_without_an_audit_sink_is_refused(): fake = FakeMicromateClient() with pytest.raises(NotSupported, match="audit"): access(fake, audit=None).start_monitoring(ADDR, "UM20147") assert fake.executed == [] def test_a_clean_ack_but_no_state_change_is_a_failed_change(): fake = FakeMicromateClient(run_states=[0x00]) # never leaves idle da = access(fake) r = da.start_monitoring(ADDR, "UM20147") assert (r.outcome, r.verified, r.state_after) == ("failed", "mismatch", "idle") assert len(r.detail["states_seen"]) == 4 # 6 s window / 2 s + the first read assert da._audit.rows[r.audit_id]["verified"] == "mismatch" def test_a_lost_ack_is_never_retried(): fake = FakeMicromateClient(run_states=[0x00, 0x02], execute_error=mm_proto.TimeoutError("no ack")) da = access(fake) r = da.start_monitoring(ADDR, "UM20147") assert len(fake.executed) == 1 assert (r.outcome, r.state_after) == ("failed", "monitoring") assert "no ack" in r.detail["error"] assert da._audit.rows[r.audit_id]["outcome"] == "failed" def test_a_change_on_a_dropped_held_line_is_not_retried(): fake = FakeMicromateClient(run_states=[0x00]) da = access(fake) da.status(ADDR, "UM20147") # hold the line fake.fail = [mm_proto.TimeoutError("peer gone")] # next call fails with pytest.raises(Exception): da.start_monitoring(ADDR, "UM20147") assert fake.executed == [] # never reached the write, never re-tried ``` Before running, check `FakeMicromateClient.status`'s call sequence, so the `fail` queue in the last test hits the first call of the change. Adapt the queue position if needed, and say so in the report. - [ ] **Step 3: Run the tests and watch them fail** (AttributeError: `start_monitoring`). - [ ] **Step 4: Implement.** `sfm/device_access/models.py`, append: ```python @dataclass class ChangeResult: """The outcome of a change request. `outcome`: "dry_run" | "skipped" | "sent" | "failed". `verified`: "ok" | "mismatch" | "not_checked". `plan` is the exact frames planned (dry run) or sent.""" serial: str series: int action: str dry_run: bool sent: bool outcome: str verified: str state_before: Optional[str] state_after: Optional[str] audit_id: Optional[str] plan: Dict[str, Any] = field(default_factory=dict) detail: Dict[str, Any] = field(default_factory=dict) ``` `sfm/device_access/s4.py`, add: - imports: `from micromate.writes import plan_start_monitoring, plan_stop_monitoring`, plus `ChangeResult`, `CallHomeInProgress`, `NotSupported` and `WrongUnit` where not already imported; - the function below. It is a plain function; `DeviceAccess` calls it inside `session.run`. ```python def s4_change_monitoring(client: Any, serial: str, *, start: bool, dry_run: bool, even_if_already: bool, audit: Any, address: UnitAddress, caller: str, sleep: Callable[[float], None], verify_timeout_s: float, verify_interval_s: float) -> ChangeResult: """Identity -> state -> (dry run | skip | audit -> send -> read back -> audit).""" action = "start monitoring" if start else "stop monitoring" want = "monitoring" if start else "idle" reported = normalise_serial(client.serial(refresh=True)) if reported != serial: raise WrongUnit(f"asked for {serial} but {reported} answered at {address.key} -- " "nothing sent") before = client.read_run_state() if before.name == "calling_home": raise CallHomeInProgress(_RING_MESSAGE) plan = plan_start_monitoring() if start else plan_stop_monitoring() base = dict(serial=serial, series=4, action=action, plan=plan.describe(), state_before=before.name) if dry_run: return ChangeResult(dry_run=True, sent=False, outcome="dry_run", verified="not_checked", state_after=None, audit_id=None, **base) if audit is None: raise NotSupported("no audit log is configured -- SFM refuses to change a unit " "without one (nothing sent)") request = {"plan": plan.describe(), "even_if_already": even_if_already} cid = audit.begin(serial=serial, series=4, address=address.key, action=action, tier="change", request=request, caller=caller) if before.name == want and not even_if_already: audit.finish(cid, outcome="skipped", verified="not_checked", ack=None, error=f"already {want}") return ChangeResult(dry_run=False, sent=False, outcome="skipped", verified="not_checked", state_after=before.name, audit_id=cid, detail={"reason": f"the unit is already {want}"}, **base) try: acks = client.protocol.execute(plan) except Exception as exc: # noqa: BLE001 -- recorded, never retried after = _try_run_state(client) audit.finish(cid, outcome="failed", verified="not_checked", ack=None, error=f"no clean ack: {exc}") if after is None: raise # the line is suspect: let _exchange hang up return ChangeResult(dry_run=False, sent=True, outcome="failed", verified="not_checked", state_after=after, audit_id=cid, detail={"error": f"no clean ack: {exc}"}, **base) ack = {"subs": [f"0x{a.sub:02X}" for a in acks], "data": [bytes(a.data).hex() for a in acks]} seen = [before.name] polls = int(verify_timeout_s // verify_interval_s) for _ in range(polls): sleep(verify_interval_s) seen.append(client.read_run_state().name) if seen[-1] == want: break ok = seen[-1] == want audit.finish(cid, outcome="sent" if ok else "failed", verified="ok" if ok else "mismatch", ack=ack, error=None if ok else f"unit reads {seen[-1]}, expected {want}") return ChangeResult(dry_run=False, sent=True, outcome="sent" if ok else "failed", verified="ok" if ok else "mismatch", state_after=seen[-1], audit_id=cid, detail={"states_seen": seen, "ack": ack}, **base) def _try_run_state(client: Any) -> Optional[str]: try: return client.read_run_state().name except Exception: # noqa: BLE001 return None ``` Check the poll count against the test's expectation. With a 6 s window and a 2 s interval, the loop gives 3 reads after the first, so 4 in total. Keep the test and the code consistent. If you change one, say which in the report. In `sfm/device_access/access.py`: - Constructor: add the keyword arguments `audit: Any = None`, `sleep: Callable[[float], None] = time.sleep`, `verify_timeout_s: float = 60.0` and `verify_interval_s: float = 2.0`, and store them, with `self._audit = audit`. - Add `change: bool = False` to `_dispatch`. Make the Series-4 retry condition `if was_held and not change and isinstance(exc, (NoAnswer, ConnectFailed)):`, and update the comment: a change is never retried, by construction. - Add the public methods: ```python def start_monitoring(self, address: UnitAddress, serial: str, *, dry_run: bool = False, even_if_already: bool = False, caller: str = "api") -> ChangeResult: return self._change_monitoring(address, serial, start=True, dry_run=dry_run, even_if_already=even_if_already, caller=caller) def stop_monitoring(self, address: UnitAddress, serial: str, *, dry_run: bool = False, even_if_already: bool = False, caller: str = "api") -> ChangeResult: return self._change_monitoring(address, serial, start=False, dry_run=dry_run, even_if_already=even_if_already, caller=caller) def _change_monitoring(self, address, serial, *, start, dry_run, even_if_already, caller): kw = dict(start=start, dry_run=dry_run, even_if_already=even_if_already, audit=self._audit, address=address, caller=caller, sleep=self._sleep, verify_timeout_s=self._verify_timeout_s, verify_interval_s=self._verify_interval_s) return self._dispatch( address, serial, lambda a, s: self._s3.change_monitoring(a, s, **kw), lambda c, s: s4_change_monitoring(c, s, **kw), change=True) ``` `self._s3.change_monitoring` arrives in Task 5. Until then, add a stub on `Series3Adapter` that raises `NotSupported("Series 3 start/stop through DeviceAccess arrives in Task 5")`, so this task's tests run. - Export `ChangeResult` from `__init__.py`. - [ ] **Step 5: Run the tests and watch them pass.** Run: `.venv/bin/python -m pytest tests/test_device_access_change.py tests/test_device_access*.py -q` Expected: all pass. - [ ] **Step 6: Full suite, then commit and push.** Message: `feat(sfm): start/stop monitoring for Series 4 -- identity, state, dry run, audit, read-back, never retried`. --- ### Task 5: The Series-3 change operation **Files:** - Modify: `sfm/device_access/s3.py` - Modify: `tests/device_access_fakes.py` - Test: `tests/test_device_access_change.py` (append) **Interfaces:** - Produces: - `Series3Adapter.change_monitoring(address, serial, **kw) -> ChangeResult`, with the same keyword arguments as `s4_change_monitoring`. - `Series3Adapter._run(address, fn, timeout=None, retry=True)` - [ ] **Step 1: Extend the fakes.** On `FakeMiniMateClient`: - add the keyword arguments `monitor_states=None` (a list of bools consumed by `get_monitor_status()`, with the last one repeating; it defaults to `[monitoring]`) and `write_error=None`; - add `self.writes = []`; - add `start_monitoring()` and `stop_monitoring()`. Each `_step`s, appends its name to `writes`, and raises `write_error` if one is set; - make `get_monitor_status()` consume `monitor_states`. - [ ] **Step 2: Write the failing tests** (append to `tests/test_device_access_change.py`): ```python # ── Series 3 ────────────────────────────────────────────────────────────────── import minimateplus.protocol as mmp_proto # noqa: E402 from device_access_fakes import FakeMiniMateClient # noqa: E402 TCP3 = UnitAddress(host="10.0.0.6", tcp_port=12345, port=None, baud=None) def access3(fake, audit="default"): seen = [] def factory(a, **kw): seen.append(kw) return fake da = DeviceAccess(s3_client_factory=factory, start_reaper=False, audit=Audit() if audit == "default" else audit, sleep=lambda s: None, verify_timeout_s=6.0, verify_interval_s=2.0) da.factory_calls = seen return da def test_series3_dry_run_returns_its_own_frame_and_sends_nothing(): fake = FakeMiniMateClient(monitor_states=[False]) r = access3(fake).start_monitoring(TCP3, "BE18193", dry_run=True) assert (r.series, r.outcome, r.state_before) == (3, "dry_run", "idle") assert r.plan["steps"][0]["sub"] == "0x96" and fake.writes == [] def test_series3_start_is_verified_by_monitor_status(): fake = FakeMiniMateClient(monitor_states=[False, False, True]) da = access3(fake) r = da.start_monitoring(TCP3, "BE18193") assert (r.outcome, r.verified, r.state_after) == ("sent", "ok", "monitoring") assert fake.writes == ["start_monitoring"] def test_series3_wrong_unit_is_refused_before_writing(): fake = FakeMiniMateClient(serial="BE11529") with pytest.raises(WrongUnit): access3(fake).stop_monitoring(TCP3, "BE18193") assert fake.writes == [] def test_series3_a_lost_ack_is_never_retried_even_over_tcp(): fake = FakeMiniMateClient(monitor_states=[False, True], write_error=mmp_proto.TimeoutError("no ack")) da = access3(fake) r = da.start_monitoring(TCP3, "BE18193") assert fake.writes == ["start_monitoring"] # one send, no cold-boot retry assert r.outcome == "failed" and r.state_after == "monitoring" ``` - [ ] **Step 3: Run them and watch them fail** (the Task 4 stub raises NotSupported). - [ ] **Step 4: Implement.** In `sfm/device_access/s3.py`: - `_run` gains `retry: bool = True`. In its `except NoAnswer:`, add `if not retry or not address.host: raise`. - Replace the stub with `change_monitoring`. It follows `s4_change_monitoring` step for step: 1. `info = client.connect()` and `self._check(info.serial, serial, address)`. This is the fresh identity read: `connect()` reads `0x15`. 2. `before = "monitoring" if client.get_monitor_status().is_monitoring else "idle"`. 3. The dry-run plan, built from `minimateplus.framing.build_bw_write_frame(SUB_START_MONITORING or SUB_STOP_MONITORING, b"")`. Take the SUB constants from `minimateplus.protocol`. Shape it as `{"action", "sends_nothing": True, "steps": [{"sub": "0x96", "what": ..., "offset": "0x0000", "data_len": 0, "expect_ack": "0x69", "frame": }]}`, the same shape as Series 4. 4. The audit, skip, send (`client.start_monitoring()`/`stop_monitoring()`), read-back poll and audit-finish logic, identical in shape to Series 4. There is no `calling_home` state on Series 3. 5. Call `self._run(address, op, timeout=S3_EVENT_TIMEOUT, retry=False)`. - Factor the shared poll/audit tail out of `s4_change_monitoring` into a helper in a new module, `sfm/device_access/change.py`, if that keeps both short. Do not duplicate 40 lines. - [ ] **Step 5: Run the tests and watch them pass**, then the full suite, then commit and push. Message: `feat(sfm): start/stop monitoring for Series 3 through the same safety layer`. --- ### Task 6: The endpoints **Files:** - Modify: `sfm/server.py`. Change `device_monitor_start` and `device_monitor_stop`, add to `_SERIAL_AWARE_PATHS`, and build `_device_access_instance` with `audit=DbAudit(_get_db)`. - Test: `tests/test_server_device_access.py` (append) **Interfaces:** - Produces: `POST /device/monitor/start` and `POST /device/monitor/stop`. Each gains `serial`, `dry_run: bool = False`, `even_if_already: bool = False` and `caller: Optional[str] = None`, and returns the `ChangeResult` as JSON. The HTTP status: | case | status | |---|---| | dry_run, skipped, sent and verified ok | 200 | | failed with `verified` "mismatch" | **409** | | any other failed | **502** | | a `DeviceError` | its own status | - [ ] **Step 1: Write the failing tests.** Append direct-call handler tests. Build each `DeviceAccess` with the fakes and an `Audit` recorder, and monkeypatch `srv._device_access_instance`. Cover: - a dry run returns 200 with the frame; - a live start that verifies returns 200; - a mismatch returns 409 with the body still holding `states_seen`; - a lost ack where the state read back fine returns 502; - `/device/monitor/start` without `serial` runs the legacy code unchanged. Use the same "legacy untouchable" pattern as the existing events test: monkeypatch `_device_access` to raise, and monkeypatch `_build_client` to raise a sentinel. - Add both paths to the guard test's "allowed" list. - [ ] **Step 2: Run them and watch them fail.** - [ ] **Step 3: Implement.** - **`baud` parameter:** `baud` becomes `Optional[int] = Query(None)` on both routes. - **New parameters:** append `serial: Optional[str] = Query(None, description=_SERIAL_DOC)`, `dry_run: bool = Query(False)`, `even_if_already: bool = Query(False)` and `caller: Optional[str] = Query(None)`. - **Docstring:** add a paragraph. With `serial`: identity check, state first, dry run, audit, read-back, never retried. The response is the ChangeResult. Without `serial`: the legacy behaviour below. - **Serial branch:** put it first in each route, then `baud = 38400 if baud is None else baud` before the legacy code. ```python if serial: return _change_response(lambda: _device_access().start_monitoring( _unit_address(port, baud, host, tcp_port), serial, dry_run=dry_run, even_if_already=even_if_already, caller=caller or "api")) ``` - **The helper:** ```python def _change_response(fn) -> JSONResponse: try: result = fn() except DeviceError as exc: raise HTTPException(status_code=exc.http_status, detail=str(exc)) from exc status = 200 if result.outcome == "failed": status = 409 if result.verified == "mismatch" else 502 return JSONResponse(status_code=status, content=dataclasses.asdict(result)) ``` - **Wiring:** `_device_access_instance = DeviceAccess(audit=DbAudit(_get_db))`. Import `DbAudit` from `sfm.audit`. - **Guard:** add `"/device/monitor/start"` and `"/device/monitor/stop"` to `_SERIAL_AWARE_PATHS`. - [ ] **Step 4: Run the tests, watch them pass, then the full suite, commit and push.** Message: `feat(sfm): start/stop monitoring by serial -- dry run, audit, read-back`. --- ### Task 7: The first live send, at the bench (controller and Brian, not a subagent) **Run only with Brian present and saying go.** UM20147 is on `/dev/ttyACM0` (`sg dialout`). SFM runs on port 8299 with a scratch database, through the phase-1b launcher pattern (patch `_DEFAULT_DB_PATH`, write a PID file, start and stop in one command). Log `micromate.protocol` at DEBUG, so every TX/RX frame is recorded. 1. **Dry run, both actions:** - `POST /device/monitor/start?serial=UM20147&port=/dev/ttyACM0&dry_run=true` - the same for `/stop` Show Brian the state and the exact frame. It must be `41021010009600ffff00000000000000000000a403` for start, byte-identical to THOR's. 2. **Brian says go:** a live `POST /device/monitor/start`. Record: - the HTTP status and body; - the debug TX/RX lines; - `states_seen`, including any sensor-check interval; - the `device_commands` row. Confirm on the unit's screen that it is monitoring. 3. **Brian says go:** a live `POST /device/monitor/stop`. Record the same, and confirm on the screen. 4. **Optional, if Brian wants to settle the open question:** - Stop again with `even_if_already=true` while idle. This records the ack of a "stop while stopped" (spec: open bench question). - The plain stop-while-idle path is a skip and sends nothing. 5. **Record the results:** - **`docs/micromate_protocol_reference.md`:** a dated, graded entry with the frames, acks, states and timings. It answers "does `0x49` reflect the change immediately", and records the sensor-check delay if any. - **`CLAUDE.md`:** retract, in place, "No command has ever been originated against a unit by this project". Add: the first commands were `0x96`/`0x97` to UM20147 on , sent by SFM at the bench with Brian's approval. - **Trilium:** a dev-log entry. ## Self-review notes - **Spec coverage.** Build phases item 3 and its six preconditions: - per-call permission: Task 1; - ceiling: Task 1; - never retry: Tasks 4 and 5; - plan built inside the lock right before sending, with no stored dry-run plan ever sent: Task 4, where the plan is built inside the dispatched operation; - read state first: Tasks 4 and 5; - Micromate flags: Task 1. The safety layer's identity check, audit table and read-back are Tasks 3–5. The dry run is Tasks 4–6. The bench protocol is Task 7. - **Not in phase 3:** call-home writes (phase 4), setup push (phase 5), the dry-run digest echo (optional in the spec, deferred), Series-3 held sessions, and the blind-stop-tool decision (Brian). - **The verify window.** It is 60 s by default. Series 3's sensor check took about 40 s (4-8-26). Series 4's is unmeasured, and Task 7 measures it.