serversdownandClaude Opus 5 0659543a15 docs(series4): the 2026-09-22 failure is accounted for -- the timeline was wrong
An earlier draft said the Prolific-cable finding did not fit the timeline, since
the unit was described as working and then degrading, which a cable that never
enumerates cannot do.  The timeline was what needed correcting.

From the operator: the initial setup was ONE modem and ONE cable, and it worked.
Two Micromates then had to be deployed, so a second cable was fetched to run both
at once -- and only one of the two ever worked.  The other never talked over its
modem at any point, was never deployed, and a MiniMate Plus went out in its place.
That Micromate came back to the bench, and the cable on it is a Benfei (Prolific
PL2303).

So nothing degraded.  The "worked at first" was the single-cable setup before the
split.  One unit working and one not, configured side by side, is the signature of
a mixed cable supply rather than a unit fault.

Still worth confirming whether that cable travelled home with the unit or was
picked up at the bench; if it travelled with it there is no inference left.  The
FTDI cable arriving 2026-09-26 settles it either way -- same unit, same modem, one
variable.

Records a corollary: "just deploy a Series III instead" works partly because a
MiniMate Plus has a DB-9 port directly on the unit, with no USB-to-serial adapter
in the path and therefore no chipset to get wrong.  That workaround has been
routing around this exact failure mode.