- micromate reference: the watchdog can check every 5 s, hourly or never
(Brian, operator-reported). The 10-14 "reconnects continuously" describes
that THOR's setting, not fixed behaviour; whether a check holds the line,
and how a call-home reaches THOR with the watchdog off, are unmeasured.
- spec, ring-path test: the old setup ("a unit THOR is not otherwise
watching") would not test the production path, since THOR receives a
call-home on its own connection. Revised: UM20147 as a THOR test unit; Run 0
baseline with the watchdog off; Run A with SFM holding the line; Run A2 and
Run B at the fleet's watchdog setting; after the 1c bench pass.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
203 KiB
Micromate Protocol Reference — Thor / Micromate Series IV, live wire protocol
Sibling to instantel_protocol_reference.md
(Series III, "the Rosetta Stone") and
idf_protocol_reference.md (Series IV file
format). This document covers the Series IV live device protocol — what
the unit says over the wire, as opposed to what it writes into a .IDFW.
Status (2026-09-23): opening session. Everything below was established in a single bench session against one unit. Treat it as a strong start, not a settled spec — in particular, everything here comes from one unit, over USB, with no events stored.
The headline
A Micromate running the Blastware firmware answers Series III command frames.
An unmodified Series III POLL (SUB 0x5B), built by
minimateplus.framing.build_bw_frame with no changes at all, produced a
complete two-step probe/data cycle. Ten Series III read commands were then
tried and all ten answered, every one obeying the Series III response-SUB
rule.
⚠ That qualifier is load-bearing, and it was discovered after the fact. Instantel ships the Micromate in two firmware lines:
| firmware | Instantel's own description |
|---|---|
11.0CB |
"Utilize with Blastware" |
11.0BD |
"Utilize with THOR, Vision, Vision II" |
The bench unit reports 11.0CB — the Blastware build. So the clean
Series III behaviour above is very likely because the unit is in Blastware
mode, not because the Micromate natively speaks Series III. Nothing here
should be assumed to hold on a 11.0BD unit until tested.
This reframes the project. The question is no longer only "what is the Series IV protocol" but "which firmware line do we target, and does one of them let the existing Series III stack drive the whole fleet?"
Firmware — the variable nobody knew was a variable
This explains a production problem TMI has lived with: two ACH servers, two machines, and units that will not cross over. It is not a misconfiguration. Instantel ships different firmware for different host software, and the wire protocol differs with it.
Fleet audit (physical, 2026-09-22/23)
| unit | firmware | line | location |
|---|---|---|---|
| UM11719 | 11.0CB |
Blastware | Ped Bridge Loc 2 |
| UM6047 | 11.0CB |
Blastware | Brookville Loc 9 |
| UM12947 | 11.0CB |
Blastware | bench |
| UM14133 | 11.0CB |
Blastware | Pitt-Music Bldg Loc 1 |
| UM11402 | 11.0BD |
Thor | Ped Bridge Loc 1 |
| UM20147 | 11.0BD |
Thor | bench |
| UM13981 | 11.0AK |
pre-split | RKM Loc 1 |
| UM20146 | 11.0AK |
pre-split | Karns Loc 2 |
| UM12420 | 10.90GC |
pre-split | RKM Loc 2 |
4 Blastware / 2 Thor / 3 pre-split. The Blastware line is already the plurality, which makes "standardise on Blastware" less disruptive than it first appeared.
A store-derived audit (firmware is recorded in every .sfm.json sidecar as
extensions.idf_report.version) agreed with the physical audit on 7 of 9.
The two that differed — UM6047 and UM14133 — are the most recently deployed
units, reflashed after their last stored event. Useful technique: the fleet's
firmware history is reconstructable from the store without touching a unit,
but it lags reality by one deployment.
Firmware is not stable per-unit over time — five of nine have been reflashed at least once. Any fleet-wide claim needs a fresh audit.
⚠ Retraction: firmware line does NOT determine Thor compatibility
An earlier draft of this document suggested that UM12947's trouble with Thor was explained by its being on the Blastware build. That is not supported.
Ped Bridge runs UM11402 (11.0BD) and UM11719 (11.0CB) side by side, both
deployed 2026-04-20, and both call Thor successfully — UM11719 has 331
Thor-collected events in the store while on 11.0CB, through 2026-08-23.
So a Blastware-line unit does feed Thor. Whatever the CB/BD split changes, it is not "which host software can collect from it", and UM12947's specific problem remains unexplained.
What is actually established: a 11.0CB unit answers Series III command
frames. Whether a 11.0BD unit does is untested — and UM20147 (11.0BD)
is on the bench, which makes that a direct A/B away.
The version is readable over the wire from SUB 0x01 as two separate ASCII
runs — "0CB" then "11" — with no single concatenated string.
The strategic fork
- Option A — standardise the fleet on the Blastware line. Every unit,
MiniMate and Micromate alike, then speaks Series III, and the existing
minimateplus/stack drives all of it. One protocol, one call-home receiver. Dramatically cheaper if it holds up. - Option B — reverse-engineer the Thor line (
11.0BD) and support both.
Option A is the shortcut, but it is unproven and carries real unknowns, all of which are cheap to answer on the bench and expensive to discover later:
- What file format does a
11.0CBunit produce? If it emits Blastware binaries rather than.IDFW/.IDFH, the (now exact) Series III decoder applies and the IDF codec becomes a legacy path. Not a loss — we have both — but it changes what the ingest pipeline sees. - Does Thor still work with a
11.0CBunit? If not, flashing a production unit breaks data collection until the replacement path exists. - Are any Micromate-specific capabilities lost on the Blastware line?
- Is the flash reversible in the field, and what does it cost in downtime?
How this was obtained
No Thor, no modem, no Windows machine. The Micromate exposes its protocol on a USB CDC-ACM virtual serial port:
ID 2504:0300 Instantel Inc. MICROMATE COM PORT
driver: cdc_acm ATTRS{serial}=="V1.00"
→ /dev/ttyACM0
Plain CDC ACM, so no vendor driver and no proprietary USB layer — any host that can open a serial port can talk to the unit.
Baud is irrelevant over USB. Identical byte-for-byte responses at 38400 and 115200; CDC-ACM ignores the line rate. TMI provisions Micromate modem links at 115200 (Series III uses 38400) — that matters for the cellular path, not for USB.
The device never speaks first. 20 s of passive listening on an idle open port produced zero bytes. It is strictly request/response.
Physical + framing layer
Requests — Series III format, unmodified
Every frame in this session was produced by build_bw_frame(sub, offset) with
no Series IV changes, and the device accepted all of them:
[ACK 0x41] [STX 0x02] [10 10] [flags 00] [SUB] [00] [offset_hi] [offset_lo] [params×10] [chk] [ETX 0x03]
The payload layout is identical to Series III. The stuffing is not.
⚠ Correction, 2026-09-27 —
build_bw_frame()is NOT sufficientThis section used to say requests were Series III frames "unmodified", and the client spec accordingly planned to re-export
minimateplus.framing.build_bw_frame. Measured against Thor's own frames, that builder reproduces 161 of 218 captured read frames. It escapes only0x10; a Micromate escapes four bytes (see The escape set below).The 57 it gets wrong are not obscure:
frame why it breaks every SUB 0x5Adownloadoffset = 0x0400puts a literal0x04inoffset_hi, which must go out as10 04SUB 0x47scheduler enableparams[7] = 0x03→10 03An unescaped
0x03or0x04is a frame terminator, so the unit sees a short frame and does not answer — indistinguishable from a dead unit, and event download would have hit it on the first request ever sent.
micromate/framing.py:build_request()reproduces 218/218. Pinned per frame intests/test_micromate_framing.py.
✅ Settled 2026-09-27: a 0x10 inside params needs no special handling.
Previously flagged untested. Thor sends SUB 0x5A with
params = 00 00 10 00 … in five captured frames and the wire carries the
ordinary doubled 10 10. There is no Series-III-style partial-stuffing
carve-out here — one rule covers the whole payload including the checksum.
Responses — Series III minus the DLE prefix
Series III: [DLE 0x10] [STX 0x02] … [chk] [ETX 0x03]
Micromate: [STX 0x02] … [chk] [ETX 0x03] ← no leading DLE
This single byte matters operationally: Blastware's parser locates frames by
scanning for DLE+STX, so it will never find a frame boundary in Micromate
traffic no matter what else is correct. That is a structural reason a
Micromate cannot call into a Blastware ACH server, independent of any baud
mismatch.
Response payload header
[0] CMD 0x00 same as Series III
[1] flags 0xC5 ← Series III uses 0x10. Constant across all 10 SUBs.
[2] SUB 0xFF − request_SUB
[3] PAGE_HI
[4] PAGE_LO
[5+] data
Checksum — plain SUM8 of the de-stuffed payload
chk = sum(payload) & 0xFF # payload already de-stuffed
251/251 responses and 251/251 requests, across every capture in
bridges/captures/9-24-26 - micromate2/.
⚠ Correction, 2026-09-27 — "the DLE-aware variant"
This section previously specified
chk = sum(b for b in payload if b != 0x10) & 0xFF, and that is wrong as paired with the uniform de-stuffing rule below.The earlier claim was not a misreading; it was a correct rule attached to the wrong convention. The DLE-aware form belongs with Series III de-stuffing, which leaves an escaped byte in the payload as two bytes — there, skipping the
0x10is the necessary correction. De-stuffing10 XX → XXalready removes it, so excluding0x10as well subtracts the correction twice.Cost of the pairing, measured: it disagrees with the wire on 55 of 251 response frames — every frame whose payload holds a literal
0x10. A clean example, from the0x5Achunk at offset0x0070ofraw_s3_…_Download_events_then_delete_1_event.bin: the wire says0xC1, plain SUM8 says0xC1, the DLE-aware form says0x91.Why nobody noticed:
scratch/mm_frame_parse.pyaccepts a frame matching either rule and labels which one hit. It reported those 55 frames asSUM8and never flagged one bad, so "zero bad checksums across 24, 38 and 40-frame sessions" was true and told us nothing about which rule was right. A tool that tries every candidate cannot falsify any of them — if it is going to stay permissive, it has to report the split, not just the pass.Pinned by
tests/test_micromate_framing.py.
The escape set — exactly four bytes
A Micromate escapes 0x02, 0x03, 0x04 and 0x10, each prefixed with a
DLE, and nothing else — in both directions.
Established by re-stuffing every captured frame and comparing to the wire:
| candidate escape set | responses reproduced | requests reproduced |
|---|---|---|
{0x10} — the Series III rule |
130/251 | 177/251 |
{0x10, 0x03} |
135/251 | 184/251 |
{0x10, 0x02, 0x03} |
178/251 | 192/251 |
{0x10, 0x02, 0x03, 0x04} |
251/251 | 251/251 |
{0x10, 0x02, 0x03, 0x04, 0x41} |
196/251 | — |
Corroborated independently by the byte that follows a wire DLE: across all
251 responses it is only ever 0x02 (933×), 0x03 (510×), 0x04 (655×) or
0x10 (1179×). Nothing else ever appears there.
The last row matters because ACK looks like it ought to be escaped and is not —
a literal 0x41 in the data goes out bare.
The checksum byte is escaped too. Three captured responses have a checksum
of 0x02/0x03/0x04 and all three arrive as 10 XX immediately before the
terminating ETX. No captured request happened to land on one, so Thor's
behaviour there is unobserved — but a device and its host share one framing
routine, and the alternative is a frame the far end truncates, so escape it.
This supersedes nothing: the de-stuffing rule 10 XX → XX stays exactly as
documented. Knowing only four values are ever escaped is what makes that
uniform rule exact rather than merely convenient — and it is what an
encoder needs, which the write path will.
The probe response carries the data length
Series III hardcodes DATA_LENGTHS per SUB. On the Micromate the probe
response tells you, as a uint16 BE at payload[8:10]:
⚠ Corrected 2026-09-23. An earlier draft read this as a single byte at
payload[9]. That is right only while the high byte is zero, and it is
catastrophically wrong for SUB 0x1A, whose real length is 0x082C = 2092
— read as a byte it gives 44, a 47× under-read. Always read the pair.
| SUB | command | payload[9] |
Series III constant |
|---|---|---|---|
0x15 |
serial number | 0x0A |
0x0A ✓ |
0x01 |
device info | 0x98 |
0x98 ✓ |
0x1C |
monitor status | 0x2C |
0x2C ✓ |
0x06 |
storage range | 0x24 |
0x24 ✓ |
0x2C |
call-home config | 0x7E |
0x7C ✗ differs by 2 |
0x08 |
event index | 0x5A |
— |
0x1E |
event header | 0x08 |
— |
0x1A |
compliance config | 0x082C (2092) |
— |
0x0A |
waveform header | 0x00 |
— (no event context) |
0xFE |
full config | 0x00 |
— (see note) |
Four of four known Series III lengths match exactly. Read the length from the probe rather than hardcoding it — it is free, and it already caught the call-home divergence.
Confirmed commands (read-only)
All ten below answered with a correct 0xFF − SUB response. Nothing that
writes, erases, or changes monitoring state has been sent to a unit.
| SUB | RSP | Command | Data proven |
|---|---|---|---|
0x5B |
0xA4 |
POLL / handshake | yes — ID block |
0x15 |
0xEA |
Serial number | yes — UM12947 |
0x01 |
0xFE |
Device info | yes — 152 B |
0x2C |
0xD3 |
Call-home config | yes — 126 B |
0x1C |
0xE3 |
Monitor status | yes — 44 B |
0x06 |
0xF9 |
Event storage range | yes — 36 B |
0x08 |
0xF7 |
Event index | probe only |
0x1E |
0xE1 |
Event header / first key | probe only |
0x0A |
0xF5 |
Waveform header | probe only |
0x1A |
0xE5 |
Compliance config | probe only |
Decoded so far
SUB 0x15 — serial. ASCII, null-terminated: UM12947.
SUB 0x5B / 0x01 — identification strings.
Instantel\0 and MM/ISEE/S/IO (MicroMate / ISEE standard). 0x01 also
carries eight consecutive 3f 80 00 00 float32 values (= 1.0f) — almost
certainly per-channel calibration/scale factors, by analogy with Series III's
geo_hardware_constant. Unverified.
SUB 0x1C — monitor status. Series III field offsets apply unchanged:
| field | offset | read |
|---|---|---|
| battery × 100 | payload[-10:-8] uint16 BE |
0x017D → 3.81 V |
| memory total | payload[-8:-4] uint32 BE |
15,000,000 |
| memory free | payload[-4:] uint32 BE |
15,000,000 (empty) |
| date | payload[18:22] |
day 23, month 9, year 0x07EA = 2026 |
The battery reading independently corroborates: Thor's own event reports for
these units print BatteryLevel : 3.8 volts.
⚠ content[6] is a sub-second tick, not a status field (UM20147, 2026-10-15).
Long recorded only as "NOT the hour". Read back to back it advances
~4,020 counts/second in steps of exactly 4, wrapping at 256 — a period of
about 64 ms, consistent with the low 6 bits of a ~1 kHz millisecond counter
scaled by 4.
⚠ Practical consequence: a SUB 0x1C block is never byte-identical to itself.
Any check of the form "has the status block changed?" answers yes on every
read, roughly 15 times a second. Compare named fields, never whole blocks.
content[9] (seconds) is the only other byte that moves on an idle unit.
⚠ How this was nearly recorded wrong, because the trap is cheap to fall into. Sampled once per second it looks like a smooth, monotonic counter advancing ~16 per second with a 16-second wrap — a clean, plausible, entirely false reading, because once-per-second sampling of a 64 ms counter is aliased. Twelve consecutive samples supported it. Measure a counter back to back before believing any rate derived from spaced samples.
The config blocks are bit-stable — which is what makes a diff loop viable
Measured on UM20147 (11.0BD), four reads back to back with nothing touched:
| block | size | bytes that moved on their own |
|---|---|---|
0x1A compliance |
2,103 B | 0 |
0x2C call home |
137 B | 0 |
0x2E trigger |
39 B | 0 |
0x08 event index (= preferences, see 2026-10-19) |
101 B | 0 |
0x01 device info |
163 B | 0 |
0x1C monitor status |
59 B | 2 (content[6] tick, content[9] seconds) |
Every config block held every byte still. So the Series III method applies directly: change exactly one setting, re-read, and any byte that moved is that setting. No noise floor to subtract and no repeat-sampling needed to distinguish a field from a counter — which was the open question about whether this approach would work at all here.
Harness: scratch/mm_config_diff.py (volatile / snap / diff), read-only.
⚠ The settings do not have to be changed from THOR. They can be changed on the unit's own keypad, which avoids contending with anything for the modem's single TCP session and needs no Windows VM.
SUB 0x2C — call-home config. Contains the ASCII string RADIO RING.
Worth flagging: that is the exact string seen in the RV50 ALEOS_SERIAL
debug during the BE12599 incident —
'ATQ1^MATE0^MATS0=2^M^MRADIO RING^M'. So this block holds the modem dial /
answer strings, and it is the most directly relevant command to the
call-home-receiver goal. Field layout not yet mapped; Series III's map
(raw[5] enabled, raw[6:46] dial string) is a starting hypothesis only, and
the length already differs (0x7E vs 0x7C).
SUB 0x06 — storage range. All zeros on this unit, consistent with
memory free == memory total. Series III reads first/last event keys from
the final 8 bytes; untestable until the unit holds events.
The event chain — walked end to end (2026-09-23, 5 events)
With 5 events on the bench unit (4 waveform + 1 histogram), the Series III browse walk works unmodified:
1E (all-zero params) -> first key + size
0A (key) -> partial record, histogram only
0C (key) -> 210-byte waveform record
1F (all-zero params/browse) -> next key + size
... repeat ...
1F -> all-zero key = NULL SENTINEL, chain ends
The sentinel terminated correctly after exactly 5 events.
Event keys are sequential, not addresses
055d4a81 055d4a82 055d4a83 055d4a84 055d4a85
This is a real divergence. Series III keys are flash-buffer addresses
(01110000, 011121F2, …) that advance by the event's byte length, which is
why its 5A chunk walk is address-arithmetic. Micromate keys are a plain
incrementing counter. Any port of the Series III download walk must not
assume key arithmetic means anything.
The 4 bytes after the key are the event's size
1E/1F return [key 4B][size 4B]. Series III uses that slot as an offset
to the next key; here it is a byte count:
| key | size | kind |
|---|---|---|
055d4a81 |
4,076 | histogram |
055d4a82 |
11,032 | waveform |
055d4a83 |
11,502 | waveform |
055d4a84 |
13,424 | waveform |
055d4a85 |
8,746 | waveform |
Consistent with real file sizes (corpus .IDFH ≈ 3.7–25 KB, .IDFW ≈
8.6–15.8 KB), and the histogram is unmistakably the small one. ⚠ Inferred,
not proven: the sizes sum to 48,780 while monitor status reports 57,344 bytes
used, so ~8.5 KB of overhead is unaccounted for.
SUB 0x0C — waveform record, and it carries the job metadata
Length 0xD2 = 210 bytes — identical to Series III. Contents confirmed
across all 5 events:
- the event key, echoed
- date + time (
17 09 07 ea→ 23 Sep 2026, then10 21→ 16:33 — matching the actual bench recording time) - title note
"Location" - the project string —
"Univ of Pitt-1st Yr Housing-Loc1 Ruskin" - serial
"UM12947" - channel labels
Tran/Vert/Long/Mic— the same labels Series III uses, and the same label-relative float32 layout - per-event float32 peaks: 3.5152, 1.3720, 2.3542, 3.5152, 0.4227 in/s across the five events (varied deliberately during recording)
This closes the biggest open question for the call-home-receiver goal.
The job identity strings (project / client / operator / setup) that
today arrive only via Thor's .txt sidecar — and which no amount of sample
decoding can reconstruct — are available over the wire from 0x0C. A
direct-to-SFM event need not arrive with blank metadata.
SUB 0x0A — partial record, histogram only
0x0A returned len = 0x1E (30 B) for the histogram and len = 0x00 for all
four waveforms. The histogram payload carries two timestamps and the
ASCII string "\r Vert: 0.300 in/s" — structurally the Series III
monitor-log partial record (0x2C type), which likewise holds a start/stop
pair and a "Geo: <float> in/s" trigger string.
So on the Micromate the division of labour is: 0x0A describes interval-style
records, 0x0C describes triggered events. Series III uses 0x0A's
response length (0x46 vs 0x2C) to tell real events from boundaries;
that discriminator does not apply here.
DLE stuffing in responses — confirmed present
Earlier marked untested. The 0x0C timestamp field contains 10 10, which
destuffs to a single 0x10 and yields a sensible clock reading. Responses
are DLE-stuffed, so a parser must destuff before applying field offsets.
A/B: Blastware build vs Thor build (2026-09-23)
UM12947 (11.0CB) and UM20147 (11.0BD) were each put on the bench and given
the identical read-only sweep. Both answer Series III command frames.
UM12947 11.0CB |
UM20147 11.0BD |
|
|---|---|---|
| POLL answers | ✓ | ✓ |
| All 10 read SUBs answer | ✓ | ✓ |
response_SUB = 0xFF − req |
✓ | ✓ |
| DLE-aware checksum valid | ✓ | ✓ |
| Two-step probe/data read | ✓ | ✓ |
| ID string | MM/ISEE/S/IO |
MM/ISEE/S |
| flags byte | 0xC5 |
0x03 |
0x1C length |
0x2C |
0x30 |
The firmware line does not change the wire protocol. One protocol stack can drive the whole fleet regardless of which build a unit is on. This is the single most consequential finding so far: "standardise the fleet on one firmware" becomes an optional convenience rather than a prerequisite for building a call-home receiver.
The two differences that do exist
1. The flags byte identifies the build. Response payload[1] is 0xC5
on the Blastware line and 0x03 on the Thor line, constant across all ten
SUBs on both units. That makes firmware line detectable from any response,
without reading device info. ⚠ Two units, one each — treat as a strong
hypothesis, not a proven encoding.
Note 0x03 is ETX, so on Thor-line units it arrives DLE-escaped as 10 03.
A parser that fails to destuff will mis-locate every field by one byte on
exactly half your fleet.
2. SUB 0x1C (monitor status) is 4 bytes longer on the Thor line —
0x30 vs 0x2C — with four extra trailing bytes (0f a0 00 00, purpose
unknown).
⚠ This breaks relative-to-end parsing. Series III reads battery and
memory from the end of the 0x1C block ([-10:-8], [-8:-4], [-4:]).
Those offsets are correct on 11.0CB and wrong on 11.0BD — applying them
blindly to UM20147 yields a battery reading of 577.92 V. Parse forward
from the declared length instead of backward from the end.
With the offsets shifted by 4, UM20147 reads correctly: battery 3.81 V, memory 15,000,000 total and free (no events stored).
Divergences from Series III (running list)
- No
DLEprefix on responses — bareSTX. - Response flags byte is
0xC5, not0x10. - Call-home config is 126 bytes, not 124.
- Data lengths are discoverable from the probe response at
payload[9]. 4b. Event keys are a sequential counter, not flash addresses. 4c.1E/1Freturn the event size, where Series III returns an offset. 4d.0x0Avs0x0Csplit by record type, not by the0x46/0x2Clength discriminator Series III uses. 4e. Responsepayload[1](flags) encodes the firmware line —0xC5Blastware,0x03Thor — where Series III has a constant0x10. - Modem serial rate is 115200, not 38400 (per TMI provisioning practice; not independently verified here).
- Reads are single-step — Thor asks for
offset = 0xFFFFand gets the whole block; Series III always probes first. (2026-09-24) - Setup writes are preceded by
SUB 0xDA, which names the target.MMBfile. Series III has no equivalent — it has one config, not named files. - The compliance block is written in ONE
0x71frame, not Series III's three chunks. 0x69/0x74(waveform data write) are absent from a setup push.- Six channel blocks (
Tran/Vert/Long/Mic/LMic/SMic) on a 48-byte stride, against Series III's four. - The notes block is four fixed-width
[label:22][value:42]entries on a 64-byte stride, with different labels — not Series III'slabel: valuescan targets. - The geophone scale factor is
3.10308, exactly half Series III's6.206053, with an ADC of 10,000 counts per volt. - The monitoring flag is
0x0E, where Series III uses0x10— and it sits atSUB 0x1Cdata[12]. A second indicator lives atSUB 0x49data[11](0x02monitoring), which Series III has no equivalent of. - Setup files are enumerated with a
0x3F/0x40first/next walk. Series III has one config and nothing to enumerate. SUB 0x1Ccarries the device clock (day/month/year/h/m/s atdata[13:21]). Nothing else read so far reports the unit's own time.- Deletion is per-event (
0xA8+ a key, then0xAA), where Series III can only erase everything (0xA3/0xA2). SUB 0x93arms each event before1E/1F, replacing Series III's1E(token=0xFE)— no token, no params.- Event keys are a sequential counter (
055D4A81…86), not flash addresses. - The ACH enable is
0x05/0x04, not Series III's0x01/0x00— bit 0 is the flag, bit 2 is set in both states. - There is a generic file transfer addressed by full path —
0x94/0x48read,0x8D/0x8Ewrite. Series III has nothing comparable; its config is reachable only through dedicated commands. - The scheduler is a separate file,
\system\schedule\schedule.dat, and a schedule entry names a setup file to load.
The firmware images are unencrypted — and they document themselves
ref-stuff/micromate-firmware/MICROMATE(CB).BIN and MICROMATE(BD).BIN,
~2.77 MB each and within 192 bytes of one another.
- Entropy 6.08 bits/byte — neither encrypted nor compressed. Plain code and data.
- Header is a big-endian vector table, handlers at
0x4010_30xx. - ~16,700 extractable strings, including the developers' own debug
printfformat strings with function names intact.
This is a legitimate interoperability reference for hardware TMI owns, and it short-circuits work I had scoped as "only answerable from a live modem capture".
The call-home state machine, verbatim
ACH_NOT_STARTED → ACH_IDLE → ACH_INITIALIZING → ACH_CONNECTING
→ ACH_CONNECTED → ACH_TRANSFER_DATA
→ ACH_RETRY / ACH_QUITTING (also ACH_STARTED)
Supporting strings:
ACH: Entry StartCallHome()
ACH: CallHome_task ; CheckAliveTime CANCEL ; TimeBetweenRetries = %d
ACH: CallHome_task ; !ExpectedCommunicationsDetected() CANCEL ; TimeBetweenRetries = %d
ACH: CallHomeCommectionCompleteProcessing() ACH=%s EAMWC=%s
ACH: %s() three attempts and it's over
ACH: %s() Send CMD_START_MONITOR
ACH: %s() Send CMD_STOP_MONITOR
ACH: Start Ignore request, Call Home is in progress
What this tells us without a single captured packet:
- Retry limit is three — "three attempts and it's over".
ExpectedCommunicationsDetected()gates the session. If the host does not say something the unit recognises, the call is cancelled and rescheduled afterTimeBetweenRetries. A homebrew receiver must satisfy this check or units will retry forever — which is exactly the failure mode seen on BE12599.- The unit stops monitoring to call home and restarts afterwards
(
Send CMD_STOP_MONITOR/CMD_START_MONITOR). Relevant to any wedged-unit rescue: the monitoring state around a call is the device's own doing, not ours. - Calls are not re-entrant — "Call Home is in progress" is ignored.
Internal command table
CMD_CALLHOME, CMD_CALLHOME_CANCEL, CMD_CALLHOME_CONNECTION_CONFIRMED,
CMD_CALLHOME_CONNECTION_COMPLETE, CMD_CALLHOME_SET_SCHEDULE,
CMD_CALLHOME_CLEAR_SCHEDULE, CMD_STOP_CALLHOME_FILETRANSFER,
CMD_DUTYCYCLE_AUTOCALLHOME, CMD_PURGE_EVENT_FLASH.
CONNECTION_CONFIRMED as a distinct state from CONNECTION_COMPLETE implies
a handshake the host must complete before data flows — the concrete shape
of ExpectedCommunicationsDetected().
Event delivery — the mechanism, probably
All Events Uploaded
Mark/Unmark File Delete Marked Events Marked Events were Deleted
MONITOR::MESG PURGE_EVENT_FLASH BEGIN / END
A marking mechanism exists, alongside a distinct "all uploaded" terminal state. 🔶 Inferred: events are marked as transferred rather than deleted on send, and purging is a separate explicit act. If so, a receiver that fails to mark would see the same events re-offered every call — the question that gates a safe homebrew receiver. Not yet confirmed; needs either a live call-home capture or disassembly around these strings.
A full user manual is embedded
The firmware carries its own HTML help, which documents configuration we would otherwise have to infer:
- Modem mode:
Generic(through a modem) vsUSB to PC. - Modem baud: 9600 / 19200 / 38400 / 57600 / 115200 / 230400 — "must match the expected rate of the PC or modem". Confirms 115200 is a setting, not a fixed rate.
- Modem relay + warmup (0–300 s), auxiliary mode, warning/alarm hold.
- Record modes: Waveform, Waveform Manual, Histogram, Histogram-Combo; sample rates 1024 / 2048 / 4096.
- A scheduler downloaded from THOR that can start/stop monitoring, change
record mode, trigger a call home, or run a self check on a daily/weekly
schedule. Pairs with
CMD_CALLHOME_SET_SCHEDULE.
Still worth doing
Diffing the two images should isolate exactly what the CB/BD split changes —
we know the wire protocol is not it, and the flags byte (0xC5 vs 0x03)
gives a concrete anchor to search for.
SUB 0x5A — bulk download. It streams the .IDFW file verbatim.
The complete read path works with no Instantel software in the loop.
It needs no arming sequence
Series III ignores a 5A probe unless preceded by
1E → 0A → 1E(token 0xFE) → 0C → 1F(token 0xFE) → POLL × 3. The Micromate
answers a bare 5A request with nothing before it. That whole ritual is
gone.
⚠ Correction, 2026-09-27 — THOR uses a 1024-byte chunk loop
The section below ("One request returns the entire event; there is no chunk loop") describes our own probes, and it is not how THOR downloads an event. Read from the wire, THOR's sequence per event is:
0x93 (arm) → 0x1E / 0x1F → key + size
0x0C → the 210-byte record
0x5A × n → n = ceil(size / 1024)
and the 0x5A frames are a plain bounded chunk walk:
chunk i offset |
min(1024, size − 1024·i) — a byte count |
| chunk 0 params | [key4][6 × 0x00] — the key means "from the start" |
chunk i>0 params |
[00 00][uint16 BE of 1024·i][6 × 0x00] |
| response data | exactly offset + 11 bytes; the file bytes are data[11:] |
response page_key |
offset // 256 — a page count, not an address |
Verified on all six bench events, sizes 4,076 → 13,424 bytes:
sum(offsets) == size exactly in every case, with the predicted chunk
count and predicted final offset. No STRT parsing, no TERM frame, no
over-read — that part of the original claim holds, and it is still far simpler
than the Series III walk.
key size(1E) chunks sum(offsets) last offset
055d4a81 4076 4 4076 0x03ec
055d4a82 11032 11 11032 0x0318
055d4a83 11502 12 11502 0x00ee
055d4a84 13424 14 13424 0x0070
055d4a85 8746 9 8746 0x022a
055d4a86 6092 6 6092 0x03cc
⚠ RETRACTED 2026-10-02 — there is no streaming mode
This section previously hypothesised that
offset_hi = 0x10meant "stream until done" and returned several frames, reconciling THOR's chunk loop with our 2026-09-23 single-request observation. Measured directly on UM20147, it does not:
request one frame returns file bytes offset = 0x1014(4,116)4,127 B data 4,116 offset = 0x111c(4,380)4,391 B data 4,380
offsetis simply a byte count. The device returns exactlyoffset + 11bytes in one frame, andpage_keyisoffset // 256.0x1000is not a marker — it is part of the number. One model covers every0x5Arequest ever observed, THOR's and ours.So the 2026-09-23 note that
offset_word = 0x1000 + 2 × pagesreturned an entire 11 KB event is wrong:0x102Cis 4,140, and 4,140 bytes is what it would have returned. Most likely the 11,049 figure was the whole capture rather than one frame's payload. Those captures never landed in the repo, so the error cannot be traced further — but it does not need to be, because the live measurement is unambiguous.🔑 And the negative result turned up something better. See below.
🔑 1024 bytes per request is THOR's choice, not the device's limit
The retraction above has a payoff. In establishing that offset is a plain byte
count, UM20147 served 4,380 bytes in a single frame without being asked
twice. THOR uses 1024; nothing about the device requires it.
That matters because of the round-trip cost: ~0.65 s each over cellular, regardless of payload. A 72,560-byte event is 71 requests ≈ 46 seconds at THOR's chunk size. The offset field is a uint16, so the structural ceiling is 65,535 bytes per request — which would make that same event 2 requests, ~1.3 seconds.
✅ Confirmed over a cellular modem too (2026-10-02)
UM12947 on an RX55, every size checked byte-for-byte against a 1024 B control: 1,024 / 2,048 / 4,096 / 8,192 / 14,176 all served in one frame, intact. Its largest event is 14,176 B, so the 16,384 ceiling itself was not reached — but a 14 KB single response crossing the PAD unscathed is the thing that needed proving. The reader reads to frame completion rather than using idle-gap detection, which is why ~10 TCP segments reassemble without special handling.
✅ The ceiling is 16,384 bytes, and over it the device CLAMPS SILENTLY
Measured on UM20147, each size checked byte-for-byte against a known-good download:
| requested | served | |
|---|---|---|
| 1,024 | 1,024 | ✅ |
| 2,048 | 2,048 | ✅ |
| 4,096 | 4,096 | ✅ |
| 8,192 | 8,192 | ✅ |
| 16,384 | 16,384 | ✅ |
| 32,768 | 16,384 | ⚠ clamped |
| 65,535 | 16,384 | ⚠ clamped |
⚠ The clamp is silent and the bytes are correct. A 32,768-byte request returns 16,384 bytes of perfectly good data and no error. There is nothing in the response to say it was truncated — the length is the only signal.
That is what dictates how the download loop must be written. A loop striding
by a fixed chunk size would either fail on a clamp or, worse, skip the bytes it
never collected. read_event_file() therefore tracks its offset by bytes
received, which makes a clamp cost one extra request rather than corrupting the
file — and makes the loop self-correcting against any short response, which is
precisely the failure mode that has bitten the Series III side repeatedly.
CHUNK_SIZE is now 16,384. For UM20147's events:
| event | THOR's 1024 | 16,384 | cellular |
|---|---|---|---|
| 4,796 B | 5 requests | 1 | ~3.3 s → ~0.7 s |
| 30,230 B | 30 | 2 | ~20 s → ~1.3 s |
| 72,560 B | 71 | 5 | ~46 s → ~3.2 s |
⚠ Measured on one unit, over USB. THOR_CHUNK_SIZE = 1024 remains available
and the replay tests pin it, so reproducing THOR's exact traffic is still one
argument away. A unit that clamps lower than 16,384 costs extra requests, not a
failure.
Implement THOR's chunked form. It is verified byte-exact across six events and five distinct sizes, and it is what the firmware runs every day.
The offset word is a LENGTH, not a position
⚠ Superseded as the implementation path by the correction above; the reading of the field is right and is what makes THOR's chunk walk make sense.
This is the key divergence. Series III walks chunks by absolute flash
address, stepping 0x0200 per request. On the Micromate the offset word
requests how much to send:
offset_word = 0x1000 + 2 × pages pages = ceil(event_size / 512)
offset_word |
pages | destuffed file bytes |
|---|---|---|
0x1002 |
1 | 518 |
0x1004 |
2 | 1,030 |
0x1006 |
3 | 1,541 |
0x102C |
22 | 11,033 — the whole event |
event_size comes from the chain walk (the 4 bytes after the key in
1E/1F). One request returns the entire event ⚠ — true of our
offset_hi = 0x10 probes; see the 2026-09-27 correction above. THOR chunks —
there is no STRT end-offset parsing and no TERM frame in either form.
Over-requesting is safe — 0x1030 (24 pages) returned exactly the same bytes
as 0x102C, so the device caps at the real size.
Params are the Series III probe form: [0x00][key4][6 × 0x00].
The payload is the .IDFW file, byte for byte
[18-byte frame header] [ .IDFW file ] [chk] [ETX] ← raw wire
^ destuffed offset 16
The file begins 00 12 01 00 00 00 "Instantel\0" — _THOR_PREFIX +
_INSTANTEL_TAG from micromate/idf_file.py. The first 32 bytes are
identical to a production .IDFW pulled from the store.
⚠ Responses are DLE-stuffed. Destuff before locating the file, or the raw byte count overshoots (11,781 raw → 11,049 destuffed for an 11,032-byte event).
End-to-end proof
Event 055d4a82 downloaded over USB and fed straight to read_idf_file():
serial UM12947
timestamp 2026-09-23 16:33:19
samples Tran 3072 Vert 3072 Long 3072 MicL 3072
peaks Tran 0.2433 Vert 1.3706 Long 0.2672 in/s
All four channels equal length, and the timestamp matches the 0x0C record
for the same key. Independent cross-check: 0x0C reports a stored peak
of 1.3720 for this event; the decoded samples give 1.3706 — two
unrelated paths agreeing to 0.1%.
Consequence: no new codec work is needed. The bytes off the wire are the
same bytes thor-watcher forwards today, so /db/import/idf_file ingests a
directly-downloaded event unchanged. Everything the IDF decoder already does
per-sample-exact applies.
What a full read now looks like
0x93 → arm (THOR sends this before every 1E/1F)
1E → first key + size
0C(key) → project/client/operator, timestamp, peaks
5A × ceil(size/1024) → the .IDFW, 1024 bytes at a time
0x93 → 1F → next key + size (until null sentinel)
✅ Verified against real hardware, both transports (2026-09-29)
micromate/{framing,protocol,client}.py driven against UM12947 by
bridges/mm_client_check.py, read-only, over both physical paths:
| USB-B "PC" (CDC-ACM) | RX55 → TCP :9034 | |
|---|---|---|
| identity, state, setups | identical | identical |
| bytes in / out | 11,580 / 761 | 11,580 / 761 |
list_setups() (24 entries) |
0.46 s | 16.05 s |
| 4,076 B download | 394 KiB/s | 1.6 KiB/s |
| transport reads | 83 | 36 |
The wire bytes are identical on both paths — same counts to the byte. The protocol does not care which physical layer it runs over, which is what the modem path needed to prove.
⚠ The modem needs FEWER reads, not more — the prediction was backwards
mm_client_check.py originally printed "a modem path should show MORE reads
for the same bytes". It shows 36 against USB's 83. The modem coalesces:
it buffers for up to ~1 s and then forwards one large TCP segment, where
CDC-ACM delivers many small chunks as they arrive.
That is reassuring rather than alarming — the risk was a frame arriving split
across reads, and the modem splits less than USB does. The client reads to
frame completion rather than using read_until_idle's idle-gap detection, and
handled both without a retry.
🔑 The cellular cost model — ~0.21 s per request + ~2,350 B/s
⚠ CORRECTED 2026-10-02 — it is NOT independent of payload
This section previously read "~0.65 s per round trip, independent of payload size", and concluded "minimise round trips, not bytes". The first claim is wrong and the second is too strong.
Every measurement behind it had a small payload — 59 B status, 266 B setup records, 1024 B download chunks. With the byte term small and similar across all of them, per-command cost looked constant. It was a narrow-range fit extrapolated past its evidence.
Measured against a 14,176 B single response on UM12947 over an RX55:
request measured model 1,024 B 0.67 s 0.65 s 2,048 B 1.22 s 1.08 s 4,096 B 2.13 s 1.96 s 8,192 B 3.59 s 3.70 s 14,176 B 6.25 s 6.25 s
t ≈ 0.21 s + bytes / 2,350— fits within ±11% across a 14× size range.The old 0.65 s figure was right for a 1024-byte response and is simply that model evaluated at 1 KB. Round trips still cost real money (0.21 s each, and 24 of them for a setup walk is still 16 s), but bytes cost more than round trips on anything over ~500 B, and that reverses the design advice: for status work minimise commands; for downloads the floor is throughput and no amount of batching beats it.
This is the number that matters for SFM's design.
list_setups() 24 commands 16.05 s ≈ 0.67 s each
get_state() 1 command 0.64 s
connect() 4 commands 2.53 s ≈ 0.63 s each
download 4 chunks 2.40 s ≈ 0.60 s each (1024 B per chunk)
The cost is per command, not per byte: a 1,024-byte chunk and a 16-byte state read cost the same. Over cellular, minimise round trips, not bytes.
Concrete consequences:
- Enumerating setups costs 16 seconds on a unit with 24 of them. Cache it; do not refresh it on a timer.
- A 13 KB event is 14 chunks ≈ 8.4 s of latency against ~0.03 s of data. If
the single-request
offset_hi = 0x10streaming mode is real (seeSUB 0x5A), it would cut a download to one round trip — that is worth the two-minute bench test on its own. - THOR's measured polling cost should be re-read in this light: its per-unit status sweep is round trips, and round trips are what cellular charges for.
End to end: downloaded events decode with the existing codec
All six bench events, assembled by MicromateProtocol.read_event_file() from
captured 0x5A responses and fed to micromate.idf_file.read_idf_file():
| key | size | type | decode |
|---|---|---|---|
055d4a81 |
4,076 | histogram | ✅ |
055d4a82 |
11,032 | waveform | ✅ 12,288 samples |
055d4a83 |
11,502 | waveform | ✅ 12,288 samples |
055d4a84 |
13,424 | waveform | ✅ 12,288 samples |
055d4a85 |
8,746 | waveform | ✅ 8,192 samples |
055d4a86 |
6,092 | histogram | ✅ |
No new codec work is needed. The bytes off the wire are the bytes
thor-watcher forwards today, so /db/import/idf_file ingests a directly
downloaded event unchanged.
✅ 11.0BD confirmed on real hardware (2026-09-30)
UM20147 read over USB. Every inference about the Thor firmware line held, and two of the three were covered only by synthesised test frames until now:
| inference | predicted | UM20147 reported |
|---|---|---|
flags = 0x03 identifies the line |
thor |
✅ thor |
0x03 is ETX, so it arrives as 10 03 |
destuffs before indexing | ✅ implied — the line could not be identified otherwise |
| model string is shorter | MM/ISEE/S |
✅ MM/ISEE/S |
0x1C is 4 bytes longer, extras trailing |
forward offsets survive | ✅ battery 3.55 V, clock correct to the second |
That last row is the one that mattered. Series III's from-the-end offsets would
have produced 577.92 V on this unit — a number so obviously wrong it is a
cheap self-check, which is why mm_client_check.py watches for it. Reading a
plausible 3.55 V and a correct clock confirms the four extra bytes really are
trailing.
Also read cleanly: serial UM20147, 5 setups (factory.MMB → test2.mmb),
active setup test2.mmb, 15,000,000 bytes total and free. 21 reads, 2,165 B in.
One protocol stack drives the whole fleet regardless of firmware line — the headline finding of 2026-09-23 — is now demonstrated by a working client, not just by matching response SUBs.
✅ A MONITORING unit answers reads normally (2026-09-30)
UM20147 was read while actively monitoring — monitoring=True on both
indicators (0x49 content[0] and 0x1C content[1], which agree) — and
answered POLL, serial, state, monitor status and a 5-entry setup walk with no
special handling.
This confirms the SESSION_RESET finding with our own client. Series III
requires a bare 41 03 before POLL or a monitoring unit will not answer over
TCP. That was previously inferred from its absence in THOR's captures; it is now
demonstrated directly, on the other firmware line, by a client that never sends
one.
Note THOR refuses to send a setup to a monitoring unit — that restriction is about writes. Reads are unaffected.
Monitoring also moves two values, so neither is a stable thing to compare against: battery drifted 3.55 → 3.50 V, and memory free dropped 15,000,000 → 14,848,448 (monitoring allocates as it runs).
⚠ Setup-name case is NOT consistent between 0x41 and 0x40
Same unit, same session, same file:
0x41 active setup name -> test2.MMB
0x40 setup-list entry -> test2.mmb
An earlier run 45 minutes before reported test2.mmb from both, so the case
is not fixed per command either. What changed in between is that the unit
started monitoring — plausibly it rewrites the active-setup name in a canonical
form when loading a setup, but that is a guess from two samples.
Practical consequence, which does not depend on the cause: compare setup file names case-insensitively. An exact-string test of "is the active setup one I know about?" answers no on this unit. Worth fixing before it is a field bug — Thor's own UI couples the scheduler to a setup name, so name matching is on the path for anything that replaces it.
✅ Real BD bytes, and the exact numbers (2026-09-30)
Captured with mm_client_check.py --capture and read out with
scratch/mm_frame_parse.py: 11 frames each direction, 0 bad checksums. Both
BD-specific responses are now real fixtures in tests/, replacing synthesised
ones.
SUB 0x1C, BD, 59 B data — the block the from-the-end offsets break on:
content[ 1] 0x0e monitoring
content[ 2:6] 1e 09 07ea 30 Sep 2026
content[ 6] 0x60 = 96 ⚠ still unidentified (32/100/116 on CB)
content[ 7:10] 0e 00 31 14:00:49 ✓ matches the wall clock
content[34:36] 01 5e 3.50 V
content[36:40] 00 e4 e1 c0 15,000,000
content[40:44] 00 e2 91 c0 14,848,448
content[44:48] 0f a0 00 04 ⚠ the four extra bytes
data[0] = 0x30 = 48 = 59 − 11, against CB's 0x2C = 44. The low-byte length
rule and the +4 both hold.
⚠ The extra four bytes are 0f a0 00 04, not 0f a0 00 00 as recorded
earlier. Only the first two look fixed. Nothing reads them, but do not treat
the last as padding.
And the warning is now demonstrated on the unit it predicted:
data[-10:-8] on this block is e1 c0 → 577.92 V. Exactly the figure in
the A/B section. Forward offsets give 3.50 V.
SUB 0x5B POLL, BD:
content[ 3] 0x56 ⚠ 0x50 on UM12947 — THIS BYTE VARIES BETWEEN UNITS
content[ 4] "Instantel\0"
content[26] "MM/ISEE/S\0" (CB: "MM/ISEE/S/IO\0", same offset)
⚠ content[3] is not a constant. It is printable on both units — P and V
— which is the second reason a printable-run scan is the wrong way to read this
block: it would return PInstantel on one unit and VInstantel on the other.
Read the vendor at the fixed offset content[4]; the model is at content[26] on
both lines, but anchor on MM/ since only its tail varies.
✅ Download from a BD unit (2026-09-30)
UM20147, event 055d4a81, 4,796 B, type read as histogram from the 0x0C
byte and decoded as .IDFH on the first try. The chunk walk, the offset
arithmetic, the 11-byte chunk prefix and the record-type byte all hold on the
Thor firmware line unchanged.
The type byte is now 7 events across both firmware lines — 4 waveform, 3 histogram — and has predicted correctly every time. Still not a large sample, and the fallback stays.
🔑 EVENT KEYS COLLIDE ACROSS UNITS — dedup must be (serial, key)
UM20147's first event is 055d4a81. So is UM12947's. Different units,
different sizes (4,796 B vs 4,076 B), different contents, identical key.
This confirms keys are a sequential counter (already recorded) and adds the consequence: the counter starts from the same value on every unit, so a key is meaningless without its serial.
⚠ Anything that stores, deduplicates or addresses Series IV events must key on
(serial, event_key). A store keyed on the event key alone will silently
treat one unit's event as a duplicate of another's — and the failure is invisible,
because the second event is simply never ingested.
Series III has a related hazard the ACH server already handles: after an erase
its counter resets, so keys are reused within a unit, which is why
ach_state.json tracks max_downloaded_key per serial. Series IV inherits that
and adds cross-unit collision on top.
⚠ 🔑 The chunk offset is a uint32 — and a 64 KB cap was one event away
UM20147 holds a 72,560-byte event (055d4a83). The 0x5A chunk offset was
implemented as a uint16 at params[2:4], because every offset THOR was
observed to send fits in two bytes — the largest is 0x3400 (13,312). That caps
a download at 65,536 bytes, so that event could not have been fetched at all.
params[0:4] is demonstrably one 4-byte field: chunk 0 puts the 4-byte event
key there. Writing the offset as a uint32 BE in the same slot is byte-identical
for every offset below 65,536 — so it changes nothing that was verified against
THOR's 74 captured frames — and extends the range to 4 GB.
✅ Confirmed on hardware 2026-10-01. UM20147's 72,560-byte event
(055d4a83) downloaded in 71 chunks at exactly its promised size — chunks
64–70 carry a 0x01 in params[1], and the device served them without
complaint. The field is a uint32; it was inference for one commit.
🔑 And it prices a large event over cellular. 71 chunks × ~0.65 s ≈ 46
seconds for one 72 KB histogram, against 0.2 s over USB. That is the
round-trip cost model doing real work: the bytes are nothing, the commands are
everything. A unit holding several events this size is a multi-minute call, and
the untested single-request offset_hi = 0x10 streaming mode would collapse it
to one round trip — which moves that two-minute bench test up the list.
⚠ At offset 1 MiB params[1] is 0x10, which must be escaped on the wire as
10 10. Unreachable below 64 KB, so the uint32 change is what first makes that
case possible — and an unescaped 0x10 in 5A params is the exact bug that cost
the Series III walk a release.
This is the Series III 64 KB page-boundary bug wearing a different hat. There,
parse_strt_end_offset() discards the key's page byte and the walk crashes once a
unit's buffer crosses 64 KB; that is still open. The transferable lesson:
an address field whose high bytes are zero in every capture is not a narrow
field, it is an untested one. Both bugs are the same mistake, found twice, in
code written years apart.
The test is sitting on a desk: download 055d4a83 from UM20147 and see
whether 71 chunks come back.
Still not covered
- A unit that is monitoring, and a unit with a nearly-full event buffer.
- The inbound call-home session — still the one protocol unknown.
Setups are FILES, not a config block
Series III has one compliance config you overwrite. Series IV keeps named setup files on an on-device filesystem, with a pointer to the current one. From the firmware:
csetup.MMB the current setup
factory.MMB Factory Default Setup File
callhome.MMB call-home config is a file too
"Current Setup File: " "Can Not Delete Active Setup File"
GetSelectedSetupFilePathName() CSelectSetupFiles CSaveSetupFile
Names are up to 20 characters and may contain spaces, hyphens, underscores. The unit's help text describes selecting, renaming and deleting them, and the event list records which setup file produced each event.
Filesystem primitives exist internally (NS_ReadFile_internal,
NS_WriteFile_internal, NS_SeekFile_internal).
⚠ RETRACTED 2026-09-24. This section originally concluded "no generic file-transfer command is exposed on the wire", reasoning from the absence of firmware strings. That was wrong.
SUB 0x94/0x48/0x8D/0x8Eare exactly that — a read/write file transfer addressed by full filesystem path — and they were caught in the open on the first capture that touched the scheduler. See The scheduler, and a generic file transfer below.The lesson is the usual one: absence of a firmware string is not absence of a command. The dispatch is a 68K jump table and the commands carry no strings of their own.
The setups half of the original claim survives: setups are not pushed as raw
.MMBblobs. They go through0xDA+ the ordinary config block.
SUB 0x1A reads the whole active setup — 2,092 bytes
Structurally close to Series III's ~2,126-byte compliance block, and it carries everything a setup consists of:
- the setup file name —
Univ of Pitt-1st Yr. Housing-Loc1 Ruskin.MMB - all four title note/value pairs —
Location,Client,Company,General Notes, with their strings - the sensor location string (
Loc 1) - per-channel labels and units:
Tran in./s.,Vert in./s.,Long in./s.,Mic psi (L),LMic psi (L),SMic (A)
Note LMic / SMic — linear and sound-level microphone variants that
Series III does not have.
This is the read half of setup management, and it means a setup can be round-tripped: read the active config, modify, write it back. ✅ The write half was observed on 2026-09-24 — see The write path below, which confirms the round-trip: the written block is the read block, 91% byte-identical.
Inside 0x1A: six 48-byte channel blocks (UM20147, 2026-10-15)
⚠ Read the grades before the table. This came from one setup on one
unit, and the three grades below are not interchangeable. The repo's own
retraction trail — the 0x0C float from a single sample, the cost model fitted
to sub-1 KB responses, offset detector v1 — is all from treating the third grade
as the first.
ASCII landmarks place six channel labels at content[1708] with a stride of
exactly 48: Tran, Vert, Long, Mic, LMic, SMic. (Mic is 3
characters, so a 4+-printable scan misses it and the stride appears to be 96
across that gap.)
MEASURED — these bytes are in the dump:
| offset | Tran / Vert / Long | Mic / LMic | SMic |
|---|---|---|---|
+0 |
label, null-padded | label | label |
+24 f32 BE |
3.1030800342559814 | 0.022503 | 73.0 |
+34 |
"in." |
"psi" |
"dB" |
+38 f32 BE |
1.0 | 0.0205945 | 60.0 |
+42 |
"/s." |
"(L)" |
"(A)" |
+46:48 |
Tran/Vert 00 01, Long 00 00 |
00 00 |
00 00 |
Tran, Vert and Long blocks are byte-identical except +47.
🔑 MEASURED — +24 carries the geophone scale, and it independently confirms
_GEO_LSB_IPS:
channel+24 = 3.1030800342559814
/ 10000 = 0.00031030800342559815
documented _GEO_LSB_IPS = 0.000310308 → agree to 1.1e-8
10.0 / 3.10308 = 3.222605 V full scale
× 10,000 counts/V = 32226.05 counts → the documented figure exactly
This matters because _GEO_LSB_IPS was previously pinned only by statistical
inference — intersecting 991,415 rounding constraints from Thor's CSV exports
(docs/idf_protocol_reference.md). The unit reports the same number in its
own config block. Two independent derivations agreeing to eight digits.
It is also almost exactly half Series III's 6.206053 (ratio 1.99997; the
residual is S3's rounded display value, not a real difference).
INFERRED, one setup / one unit — plausible, not established:
- that
+24is the calibration constant, rather than another field that happens to hold the same number (the arithmetic proves the number, not the field's purpose) - that
+46:48is a per-channel enable or trigger-source flag. It is the only byte separating three otherwise identical blocks, which is suggestive and nothing more.⚠ 2026-10-17: not trigger source, and not enable. Trigger sources are a bitmask at
content[29]and active sensors one atcontent[25]; neither change moved+46:48, nor did anything else in a 47-snapshot pass. Its purpose is unknown.
GUESSED — do not build on these:
+4as a uint32 "channel type" (reads 1 / 1 / 1 / 2 / 2 / 9 across the six). Six observations from a single setup, correlating with channel kind. That is the entire basis.✅ 2026-10-17: promoted, with one asymmetry. Switching Mic Type to Sound Level rewrote block 4's code
2 → 9and nothing else in the config moved, so the mic type has no byte of its own: this field is it. Geophone blocks read0while the geophones are deactivated. ⚠ Deactivating the mic did not zero block 4's code. See The0x1Afield map from a keypad pass.- ⚠ 2026-10-17: "six channels" is itself wrong. Blocks 1–3 are the
geophones, block 4 is the active mic, and blocks 5 and 6 are the stored
settings of the linear and sound-level mic respectively.
LMicandSMicare not channels. +8:24as eight uint16 BE values. Geo reads 74, 10, 40, 21, 1275, 525, 3, 0; mic reads 200, 50, 10, 10, 1500, 500, 5, 0. Field boundaries unverified — the uint16 split is an assumption, not a reading.
To promote any of the above: the diff loop settles it. Change one setting, see which of these offsets moves. A field whose purpose is guessed and which never moves when the matching setting changes is simply wrong.
The 0x1A field map from a keypad pass (UM20147, 2026-10-17)
Method: scratch/mm_config_diff.py watch polling all six blocks every 2 s while
settings were changed on the unit's own keypad, one at a time. 47
snapshots. Every byte that moved is accounted for by a named setting, and
nothing outside 0x1A moved at all. Snapshots:
scratch/mm_config_snaps/UM20147/s1017-watch-* and watch-1017-0150-*.
⚠ Grades. "Measured" below means: the operator named the setting and its before/after value, and the bytes moved accordingly. Two attributions made before the operator confirmed what was pressed were wrong, and were retracted in the session — the mode enum and the sensors mask both briefly carried the wrong label. A diff tells you which bytes moved, never which setting moved them.
Offsets are into content (data[11:], 2,092 bytes).
| offset | field | encoding | grade |
|---|---|---|---|
[0:2] |
length echo, 0x082a = 2090 = content minus the CRC |
u16 BE | inferred (arithmetic); Thor zeroes it on write |
[2:5] |
64 00 00 |
— | unknown, never moved |
[5] |
record mode | 01 Waveform · 02 Waveform Manual · 03 Histogram · 04 Histogram Combo |
measured, all four, screen-confirmed |
[6:8] |
sample rate, sps | u16 BE | measured (1024 → 2048) |
[8:10] |
histogram interval, seconds | u16 BE | measured (2…11, 30, 300; 01 2c settles the width) |
[10:12] |
00 00 |
— | unknown, never moved |
[12:16] |
pre-trigger, negative seconds | f32 BE | measured, all four: be80=−0.25 bf00=−0.5 bf40=−0.75 bf80=−1.0 |
[17] |
record stop mode | 00 Fixed · 01 Auto Record |
measured; u8 vs u16 BE [16:18] unresolved |
[18:22] |
record time, seconds | f32 BE | measured (10 → 11) |
[23] |
auto stop after, seconds | u8 (or u16 BE [22:24]) |
measured (3 → 1) |
[25] |
active sensors | bitmask: 0x07 geophones, 0x08 mic |
measured, all three options |
[26:28] |
05 3f |
— | unknown, never moved |
[29] |
trigger sources | same bits as [25] |
measured, both directions; mic bit screen-confirmed |
[30:34] |
scaled-distance distance, metres | f32 BE; the keypad converts ft ÷ 3.2808 | measured 10-19 (123 ft → 4215f6a3, bit-exact, unique among 5 constants) |
[34:38] |
charge weight, kg | f32 BE; the keypad converts lb ÷ 2.21 ⚠ not 2.2046 | measured 10-19 (45 lb → 41a2e55c, bit-exact) |
[39] |
report flags | 0x02 Job Number on · 0x08 Scaled Distance on · 0x01 ? |
measured 10-19 for 0x02/0x08; 0x01 set throughout, unknown |
[40:42] |
job number | u16 BE, kept while switched off | measured 10-19 (0 → 1 → 1234) |
[42:86] |
setup file name | ASCII, null-padded | measured; spelling varies, see 10-19 |
[122 + 64n] / [144 + 64n] |
note label (22 B) / value (42 B), n = 0..3 | ASCII | measured 10-19 (Location) |
[2090:2092] |
CRC | see below | measured, 47/47 + a second unit |
Mode enum ⚠: the menu order is Histogram → Combo → Waveform → Manual
(03 04 01 02), so the screen order is not the encoding. The values match
Series III's (01 Continuous, 03 Histogram, 04 Histogram+Continuous) —
the byte sits one position later than the S3 analogy predicted, the
encoding does not differ. There is no Single Shot (S3 00).
A framing question, answered in passing: 0x03 is ETX, and it
appeared as a literal value twice (record mode 03, interval 3 s). The block
stayed 2,103 bytes both times — the Series-4 read path destuffs an escaped
ETX correctly, which is exactly the case that shifted every later field by
one on Series III.
The channel blocks are 3 geophones, 1 active mic, and 2 stored mic setups
content[1708] + 48n. The 10-15 section above called these six channels;
they are not:
| block | role |
|---|---|
| 1–3 | Tran, Vert, Long |
| 4 | the active microphone — a byte-for-byte copy of block 5 or 6 |
| 5 | the linear-mic settings, kept while not selected |
| 6 | the sound-level-mic settings, kept while not selected |
Measured: Mic Type → Sound Level copied block 6 into block 4 byte-for-byte;
→ Linear copied block 5 back, carrying a 140 dB trigger set while Linear was
active — so block 5 is a store, not a template. Nothing else in the config
moved: there is no separate mic-type byte; block 4's +4:8 code (2
linear, 9 sound level) is the type.
⚠ For a decoder: read the mic from block 4 only, and branch on its code. Blocks 5 and 6 are the settings of whichever mic is not fitted; decoding either as a channel reports a sensor that is not there.
| offset | field | grade |
|---|---|---|
+4:8 |
sensor code: 1 geophone, 2 linear mic, 9 sound-level mic, 0 geophone deactivated |
measured; ⚠ deactivating the mic left block 4 at 2 |
+30 |
trigger level, f32 BE — geophones in/s; mic psi, whatever the display unit | measured (geo 0.5 → 0.6; mic 131 → 140 dB) |
The keypad has one geophone trigger level and writes it to all three blocks; the per-channel slots exist (Thor may set them separately — untested). A mic trigger edit writes block 4 and its store (block 5).
Mic trigger: 0.0102915 psi → 70.957 Pa → 130.999 dB re 20 µPa (screen: 131);
0.0290064 psi → 139.9997 dB (screen: 140). Reading is solved. ⚠
Writing is not: the textbook conversion of 140 dB is 0.0290075, 617 ULPs
away, and the implied Pa/psi differs between the two values (6895.29 vs
6895.03), so the firmware's dB → psi rule is not a single constant. A writer
cannot byte-match it yet.
Labels and unit strings are cosmetic — keypad saves re-render them
- Block 5's label went
LMic→Micon a geophone trigger-level save. - All three geophone unit strings went
in.//s.→in//swhen Active Sensors changed. One-way: re-activating did not restore the dots.
The same setup reads differently after Thor wrote it and after the keypad saved
it. Key channels on block position and the +4:8 code, never on a string.
Why some saves re-render and others do not is unknown (inferred: the firmware
rebuilds whatever blocks the edited screen owns, in its own spelling).
The CRC
CRC-16/AUG-CCITT poly 0x1021, init 0x1D0F, refin/refout false, xorout 0
over content[0:2088] (content[2088:2090] = 00 00 is NOT covered)
stored big-endian at content[2090:2092]
Identified from a single sample-rate diff by testing all 65,536 polynomials
against crc(a) ⊕ crc(b) (linearity makes the range start irrelevant), then
confirmed on absolute values. 0x1021 was the only catalogued polynomial among
~30 chance hits. Equivalent forms — CCITT-FALSE (init 0xFFFF) over two zero
bytes plus the content — are the same algorithm; leading zeros only move the
effective init. Verified: 47/47 snapshots on UM20147, every state predicted
before comparing, and UM12947's 9-24 read (3cfd), a different unit and setup.
Revisiting a state reproduces its trailer exactly — no counter in the block.
⚠ A writer does not send it. Thor's 0x71 data is content[0:2090]
(aligns at shift 0, 1902/2090 equal) — it stops before the CRC and zeroes the
length echo. The unit computes both. Its use is read-side integrity: every
0x1A read can now check itself, which on a codec repeatedly bitten by silently
short data is the cheap guard worth wiring in.
0x1A reads the keypad's EDIT BUFFER, live
Measured, two independent tests: a scrolled-to, unsaved record mode appeared
in 0x1A immediately; backing out with discard restored it. A second discard
reverted two fields at once (mode and an interval changed 13 times) to a
block byte-identical to the last saved state. Pressing save moved no byte.
When an edit reaches the buffer depends on the widget — measured per widget:
| widget | reaches 0x1A |
seen on |
|---|---|---|
| list | on every scroll step | record mode, active sensors, mic type |
| spinner popup | on Enter | histogram interval |
| sub-window | only on back-out → save to setup file | Record Time window (pre-trigger, stop mode, auto stop) |
A fast list scroll can pass a value inside one poll; the watcher then never sees it (Histogram was skipped once). Pause on each value.
⚠ Consequence: a 0x1A read taken while someone is in the setup menu
returns values that may never be committed — and may differ from the .MMB file
of the same name. Rare in the field; worth one line wherever SFM reports a
unit's config.
Negative results
0x2Enever moved — not under trigger level, trigger enable, or anything else. Whatever "trigger" config it holds, these settings are not it.0x08,0x01,0x2Cnever moved. The bit-stability also holds across days: the 10-17 baseline was byte-identical to 10-15's.2026-10-19:
0x08and0x2Cdo move — under preferences, which 10-17 never touched. See the next section.
Not yet mapped
Location / Client / Company / Notes (Location confirmed 10-19), Job
Number, Scaled Distance (10-19), GPS, Aux I/O, the sound-level
mic's own options (response, weighting, LN1/LN2 — SMic +8:24 holds 10, 90,
the manual's example LN1/LN2, guessed), the four unknown head runs above,
[39] bit 0x01, and +46:48.
Setup loading, the preferences block, and a checksum rule (UM20147, 2026-10-19)
Same method, same unit, 23 more snapshots (watch-1019-1642-*).
Loading a setup is an exact replacement
Two keypad-made setups differing in one field (SWAP2 / SWAP3, record time
11 / 5 s), plus test2. Measured, three loads: each load moved exactly the
fields that differ between the two files, in one poll, and the loaded block was
byte-identical (all 2,103 B) to that file's state when saved — no
timestamp, counter or load marker. No other block moved.
So the operator's hypothesis holds: a .MMB load repopulates the active
config wholesale, and a load of a many-differences setup is a one-step check
of the whole map. 🔶 Inferred, untested: the .MMB on flash is this block;
reading the file directly (0x94) would settle it.
⚠ The name field's spelling depends on how the setup became active. After a
keypad save: test2.MMB. After loading from the list: test2.mmb. A
save-to-current-file flips it back to .MMB, moving no other byte. Same file.
Compare setup names ignoring case and extension — already noted 2026-10-13
from three commands spelling it three ways; this is a fourth.
Also measured in passing:
- Hidden settings persist. Record time is not shown in Histogram mode, and
[18:22]kept its value throughout. The UI gates; the block does not. - A literal
0x10(DLE) value reads cleanly. Sample rate 4096 puts0x10atcontent[6]; length unchanged, CRC ok. This is the Series III "DLE jitter" case that shifted every later field by one byte there. - Save-to-new-file switches the active setup at once: the name field moved the moment the file was saved.
- Sub-windows (Job Number, Record Time) commit to the edit buffer when the window's own Save is pressed; the setup-file save is a separate step.
Unit conversions use the firmware's own rounded constants
Distance and charge weight are stored metric whatever the display unit:
| entered | stored | constant that reproduces it bit-exactly | textbook |
|---|---|---|---|
| 123 ft | 4215f6a3 = 37.490856 m |
÷ 3.2808 | ÷ 3.28084 / × 0.3048 → 4215f62b |
| 45 lb | 41a2e55c = 20.361992 kg |
÷ 2.21 | ÷ 2.20462 → 41a34b1f |
⚠ For SFM: to reproduce what the unit displays, convert back with its
constants. 20.362 kg × 2.20462 shows 44.89 lb for what the operator
typed as 45. The 2.21 makes stored kilograms 0.24% light — the unit is
self-consistent, the metric figure is simply wrong by that much.
Scaled distance is not stored. It went 0 → 18.34 on screen (= 123 / √45) and no byte moved beyond the charge weight.
🔶 Inferred: the mic's psi values not matching the textbook dB conversion (see above) is probably the same story — a rounded constant of the firmware's own. Not yet identified.
0x08 is the preferences block, not an "event index"
0x08 content opens 00 58 = 88, the size of Series III's block of the same
role (written by 0x68; S3 keeps backlight and power-save there). Measured:
Measurement Units Imperial ↔ Metric moved 0x08 content[31] (00 / 01)
and its trailer, and nothing in 0x1A — so setup values are stored in fixed
units and a decoder never needs the display preference. The harness and this
document's older tables still call it "event index"; the name is wrong.
One checksum rule covers three blocks
| block | content[0:2] = N |
CRC-16/AUG-CCITT over [0:N-2], stored at [N:N+2] |
grade |
|---|---|---|---|
0x1A setup |
2090 | ✅ | 72 snapshots, 48 distinct states |
0x08 preferences |
88 | ✅ [0:86] → [88:90] |
2 states |
0x2E |
26 | ✅ [0:24] → [26:28] |
1 state; fits the same rule |
0x2C call-home |
124 | ❌ [124:126] reads 00 00 |
— |
In every checked block [N-2:N] is 00 00. 🔶 Inferred: that is the classic
augmented CRC — a shift register run over the message plus two zero bytes —
which is exactly what AUG-CCITT is. It changes nothing about how to verify.
0x2C call-home: Series III's layout, mostly Thor-only
Only three call-home settings are on the keypad (manual §11.4.4): session timeout, modem type, modem baud. "Monitor While Calling Home" is set from Blastware/Thor only, as are enable, call-after-event and the schedule.
Read from the block, against Series III's map — inferred by analogy, except where marked:
[5] 01 enabled (S3 raw[5])
[6:..] "RADIO RING" dial string (S3 raw[6:46])
[87] 01 call after event recorded (S3 raw[87])
[117] 03 number of retries (S3 raw[117..118])
[118:120] 8fab ← MEASURED: tracks session timeout, encoding unknown
[120:122] 00 1e = 30 wait for connection? (S3 raw[121..122])
- Session timeout 5 → 37 min moved only
[118:120],8fab → 096d; back to 5 returned exactly8fab. Deterministic, but neither value reads as 5/37 in any width, unit or endianness, and no CRC-16 of the value (6 polys × 4 inits × 9 encodings) produces it. More points (6, 7, 8) would likely crack it. Low priority: TMI does not depend on it. - Modem baud rate is in none of the six blocks. 115200 → 230400, saved, and
nothing moved; the link stayed up (no
FAILED) because the bench connection is the PC port.
⚠ Rule for any SFM write path: never write modem baud rate or modem type remotely. The remote link is the modem port. A baud mismatch with the RV55, or switching type to USB-to-PC, drops the session mid-command and nothing remote can undo it — the unit stays dark until a site visit. (We do not know where either is stored, so this is not a capability we have; it is recorded so it stays a decision.)
The write path — observed end to end (2026-09-24)
⚠ We have still never sent a write command to a unit. Thor did every write
below; seismo_lab's TCP bridge sat between Thor and the unit and recorded both
directions. Capture: bridges/captures/be12599-diag/9-24-26 - micromate2/
(raw_bw_* = Thor, raw_s3_* = unit), UM12947 on the USB PC port, relayed
to TCP by socat on mint-mac. Operation: push a setup named TEST1.mmb.
All 12 device responses checksum-validate and every write is acked.
The sequence
Thor: 5B │ 41 │ 08 │ 2E │ 1A │ DA │ 68 → 73 │ 82 → 83 │ 71 → 72
unit: A4 │ BE │ F7 │ D1 │ E5 │ 25 │ 97 8C │ 7D 7C │ 8E 8D
└──────── reads ────────┘ └──────────── writes ────────────┘
Series III, for comparison: 68→73 │ 71×3→72 │ 82→83 │ 69→74→72.
Same write SUBs, and every response SUB matches the Series III table
exactly (68→97, 73→8C, 82→7D, 83→7C, 71→8E, 72→8D).
Three differences:
SUB 0xDAis new and comes first — see below.0x71is a single write, not three chunks. Series III splits the compliance block into 1027 + 1055 + remainder; the Micromate takes all 2,090 bytes in one frame.0x69/0x74(waveform data write) do not appear at all.
Every write ack is a 16-byte zero-data frame, same shape as Series III's.
Three new read commands
| SUB | rsp | payload | what it carries |
|---|---|---|---|
0x41 |
0xBE |
271 B | the active setup's file name, e.g. TEST1.mmb |
0x2E |
0xD1 |
44 B | trigger-config block; mirrors what 0x82 writes |
0xDA |
0x25 |
272 B (write) | declares the target setup file — see below |
SUB 0xDA — name the setup file you are about to write
Data is exactly 256 bytes: the file name, null-padded, nothing else.
54 45 53 54 31 2e 6d 6d 62 00 00 … "TEST1.mmb" + 247 × 0x00
This is the missing link in setups are files: Thor names the target file, then
writes the ordinary config block into it. The unit acks with 0x25 before any
config bytes are sent.
Note 0xDA takes a bare filename null-padded to 256 bytes. That is a
different mechanism from the path-addressed file transfer (0x94 / 0x8D,
which carry \system\schedule\schedule.dat unpadded) — setups do not go through
the file transfer, and the file transfer is not how setups are written.
Reads are single-step — no probe
Series III sends every read twice (probe at offset=0x00 to learn the length,
then a data step). Thor skips the probe and asks for
offset = 0xFFFF, getting the whole block in one response:
41 02 10 10 00 1a 00 ff ff 00 … SUB 0x1A, offset 0xFFFF
→ 2,103-byte response
POLL is the exception — it uses offset = 0x0030 (48), its data length.
This does not contradict the probe response documented above; the probe still
works and still reports its length at payload[8:10]. Thor simply does not
need it.
Write-frame destuffing — 10 XX → XX, uniformly
Only the leading BW_CMD is doubled (10 10); after that every 10 XX pair
on the wire destuffs to XX, including 10 03.
This was settled by checksum, not by assumption. Four candidate rules were tested against all four data-carrying write frames; only this one makes all four checksums validate:
| rule | 0xDA |
0x68 |
0x82 |
0x71 |
|---|---|---|---|---|
10 XX → XX |
ok | ok | ok | ok |
only 10 03 → 03 |
ok | BAD | ok | BAD |
10 10→10, 10 03→03 |
ok | BAD | ok | BAD |
| nothing collapses | ok | BAD | ok | BAD |
0x71's data contains 4 literal 0x03 bytes, escaped as 10 03 on the
wire. A writer that does not escape 0x03 will emit a frame the device
terminates early — this is the same defensive ETX escaping Blastware does, and
it is mandatory, not optional.
Checksum is plain SUM8 of the destuffed payload; the DLE-aware form gives the same answer once destuffing is correct, so it does not discriminate.
The offset field is a per-command constant
| SUB | data bytes | offset |
note |
|---|---|---|---|
0xDA |
256 | 0x0100 = 256 |
= the name-field size |
0x68 |
88 | 0x005A = 90 |
identical to Series III's documented value |
0x82 |
28 | 0x001C = 28 |
identical to Series III's documented value |
0x71 |
2090 | 0x082C = 2092 |
= the length SUB 0x1A reports |
⚠ There is no single length formula — two are len, two are len + 2, and
Series III's data[1] + 2 rule does not reproduce either. Treat these as
observed constants the device accepted. Pinning the actual rule needs a second
capture whose payloads differ in size. (This document has already been wrong
once by inferring a length field; do not infer this one.)
The write body is the read body
The 0x71 write payload and the 0xE5 read response align at a fixed
11-byte shift with 1902/2090 bytes equal (91.0%) — the remaining 9% is
exactly what was edited.
So a setup is read-modify-write, the same shape as Series III, and a write client does not need to synthesise a config block from scratch.
Field map, from the read/write diff
Twelve differing regions, all accounted for. Offsets are into the 0x71
data section (= the read response's data + 11).
| offset | size | field | this capture |
|---|---|---|---|
0x0000 |
3 | block length echo | 08 2a (=2090) → zeroed on write |
0x0009 |
1 | unidentified | 3c → 02 |
0x0013 |
1 | unidentified | 00 → 40 |
0x002A |
44 | setup file name | → TEST1.mmb |
0x0090 |
42 | note 1 value — Location |
→ Test Location 1 - 1234 electric boogaloo |
0x00D0 |
42 | note 2 value — Client |
→ TMI |
0x0110 |
42 | note 3 value — Company |
→ ServersDownLabs |
0x0150 |
42 | note 4 value — General Notes |
→ Hopefully this works! |
0x0428 |
16 | sensor location | → Test Config Send |
0x06CA |
4 | Tran trigger level float32 BE | 0.3 → 0.5 in/s |
0x06FA |
4 | Vert trigger level | 0.3 → 0.5 in/s |
0x072A |
4 | Long trigger level | 0.3 → 0.5 in/s |
Notes block: four entries on a 64-byte stride, each [label: 22][value: 42],
labels at 0x008A + 64n. The labels are Location, Client, Company,
⚠ 2026-10-19: the label offset is
0x007A + 64n(122), not0x008A. Measured on UM20147:Locationatcontent[122], its value atcontent[144](0x90, as the table says), and a keypad edit of Location moved exactlycontent[144:151].0x8Acontradicts the[label: 22][value: 42]layout in the same sentence; it is a 16-byte slip, as is0x06BCbelow.General Notes— not Series III'sProject:/Client:/User Name:/Seis Loc:, and they are fixed-width fields, notlabel: valuepairs.
Channel blocks: six, 48 bytes each, from 0x06BC (⚠ 2026-10-19: 0x06AC
= 1708, where the labels actually sit; the trigger offsets below were right) — Tran, Vert, Long,
Mic, LMic, SMic. Trigger level sits at label + 30.
The geophone scale factor is in the config block — and it confirms our LSB
Each geo channel block carries a float32 BE at label + 24:
40 46 98 dd = 3.10308003…
Which closes a loop from the file-decode work:
3.10308003… / 10000 = 0.000310308 = _GEO_LSB_IPS, to 8 figures
10.0 / 3.10308003… × 10000 = 32226.046 = the 32226.05 full scale
_GEO_LSB_IPS was derived statistically — intersecting 991,415 rounding
constraints from Thor's own CSV exports (see idf_protocol_reference.md).
The unit reports the constant directly, and it agrees. That upgrades the
value from a fit to a reading, and explains the odd full-scale count: the
Micromate's ADC is 10,000 counts per volt, and 3.10308 in/s per volt is
exactly half Series III's documented 6.206053 (ratio 1.99997).
Do not retune _GEO_LSB_IPS — this is corroboration, not a correction.
Still unknown on the write path
- What
0x68and0x82actually contain. Both were written with near-zero payloads here and nothing in them changed, so no field is located. Series III maps backlight/power-save/LCD-cycle into0x68; unverified here. - The three header bytes at data
0x0000/0x0009/0x0013. - Scheduler / call-home writes.
callhome.MMBis a file too, so it may go through the same0xDA+ block-write shape with a different target name.
Write-plan dry run on a second unit and firmware line (2026-10-22)
Firmware, stated as line and version separately. UM20147: Thor firmware line
(flags 0x03, measured 2026-10-22), version 11.0BD per the 2026-09-30 check;
UM12947: Blastware firmware line (flags 0xC5 in all 152 responses of the
seven write sessions), version 11.0CB.
2026-10-22 — UM20147 block sizes equal UM12947's: 0x1A
content 2092 B, 0x08 90 B, 0x2E 28 B, 0x2C 126 B (all CRCs good where a
CRC applies). Its dry-run plans (bridges/mm_dry_run.py) have the same SUB
order, offsets and data lengths as THOR's frames from UM12947: 0x96 / 0x97
(offset 0xFFFF, 0 B); 0x7E @ 0x007E 126 B + 0x7F; 0xDA @ 0x0100 256 B,
0x68 @ 0x005A 88 B, 0x73, 0x82 @ 0x001C 28 B, 0x83, 0x71 @ 0x082C
2090 B, 0x72. So THOR's write offsets apply to both firmware lines as far
as size goes (measured, one unit each; nothing sent -- reads only; the
opcode gate was armed throughout and refused nothing, because no write was
attempted). Not shown: that 11.0BD accepts these
frames. The call-home on/off plans differ in exactly one data byte ([5]);
the setup-rewrite 0x71 data equals the unit's 0x1A content [3:2090] behind
three zero bytes.
🔶 Unobserved: UM20147 call-home content[5] reads 0x01 (THOR's UM12947:
0x04/0x05). Our "off" plan writes 0x00, which no capture shows being
written. Unobserved; settle at the phase-4 bench.
Breadth of the reproduction (measured): across the five pushes, the 0xDA,
0x68 and 0x82 inputs are byte-identical (one distinct name, prefs and
trigger block), and the three no-edit 0x71 reproductions are the same frame.
So each block was reproduced from one distinct input, not five.
✅ 0xDA CREATES setup files (confirmed on the device, 2026-09-24)
TEST1.mmb did not exist on the unit before the push. Afterwards it is
present in the unit's setup list and selected as the active config —
verified on the Micromate's own screen, not inferred from the ack.
So the four commands below are the complete setup-management path, and a homebrew client needs no file-transfer primitive and no pre-existing target:
0x41 read the active setup's name
0x1A read its config block ─┐ read
0xDA name the target .MMB ─┘ modify
0x71 → 0x72 write the block back write (creates the file if absent)
That closes the last open question about whether Series IV setup management is reachable from outside Thor. It is.
✅ Overwriting is protocol-identical to creating (2026-09-24)
The firmware carries strings that suggest an overwrite handshake:
Overwrite File MFS FILE EXISTS
Cannot be Overwritten Can Not Delete Active Setup File
There is no such handshake on the wire. A second push to TEST1.mmb — a
name that now existed, and which SUB 0x41 confirmed was the active setup —
produced a byte-for-byte identical command sequence:
| create | overwrite | |
|---|---|---|
| SUBs, and their order | 5B 41 08 2E 1A DA 68 73 82 83 71 72 |
identical |
every offset field |
— | identical |
0xDA / 0x68 / 0x82 data |
— | 0 bytes differ |
0x71 data |
— | 18 bytes differ = the one edited note |
| all seven write acks | 11 zero bytes | identical, still all-zero |
No extra command, no confirm step, no error status, and no dialog on Thor.
Those firmware strings belong to the on-device Save screen (the
CSaveSetupFile UI class), not to the protocol.
The overwrite was verified on the unit itself — the edited General Notes
string is present in the setup on the device, so the write applied, silently and
in place.
This is the case that mattered most, and it landed the right way round: the target was the active setup, which is what a real remote config push would hit. Being active bought it no protection — it was overwritten directly. A writer therefore needs no exists-check and no overwrite negotiation.
⚠ That cuts both ways. A remote push to the active setup of a monitoring unit changes the config it is recording with, with no prompt, no warning and no distinguishable ack. Whatever SFM eventually exposes should gate this on the operator, not on the protocol — the device will not stop anyone.
⚠ Two overwrite cases remain untested: a non-active setup file, and
factory.MMB, which Cannot be Overwritten probably guards. Neither blocks a
writer — the active setup is the one worth pushing to.
⚠ Note also that the write acks are all-zero in every case observed, across
a create and an overwrite. We have never seen this protocol report a failed
write, so do not treat a zero ack as proof a write was applied. Read the
config back and compare; SUB 0x41 plus SUB 0x1A make that cheap.
The scheduler, and a generic file transfer (2026-09-24)
Capture: bridges/captures/9-24-26 - micromate2/*_read_scheduler.bin.
25 request frames, 25 responses, every checksum valid.
Four operations in one capture, segmented by Thor's POLL preamble (marks are
not written into the raw .bin — in TCP mode seismo_lab logs them to the
on-screen log only):
| frames | operation |
|---|---|
| 0–4 | read the schedule off the unit |
| 5–18 | push a setup — the sequence already documented above, unchanged |
| 19–21 | write the schedule back |
| 22–24 | enable the scheduler |
🔑 SUB 0x94 / 0x48 / 0x8D / 0x8E — file transfer by path
The unit will read and write files addressed by full filesystem path.
0x94 <path> → 0x6B open for READ
0x48 → 0xB7 read next page; repeat until an all-zero response
0x8D <path> → 0x72 open for WRITE
0x8E <body> → 0x71 write the file body
The path is plain ASCII, unpadded, with offset = its exact length:
5c 73 79 73 74 65 6d 5c 73 63 68 65 64 75 6c 65 5c 73 63 68 65 64 75 6c 65 2e 64 61 74
\system\schedule\schedule.dat 29 bytes, offset = 0x001D
0x94 and 0x8D sent byte-identical 29-byte payloads — the same path, only
the command distinguishing read from write.
The 0x48 read is paged, and the page number comes back in the response header
at payload[3:5]:
| call | page | payload | content |
|---|---|---|---|
| 1st | 0x0000 |
15 B | descriptor — 04 00 … 01 00 00 00 00 00 01 |
| 2nd | 0x0002 |
535 B | the file body |
| 3rd | 0x0000 |
11 B | all zeros = end of file |
⚠ This is the single most consequential find of the session, because it is not
specific to the scheduler. callhome.MMB is a file too, and the call-home
config is the last unsolved piece of the project. Reading it should be a
matter of pointing 0x94 at the right path. That is a lead, not a result —
no path other than schedule.dat has been tried.
The schedule file itself
The write body is the read body minus an 11-byte response prefix (524 vs 535), so both describe the same structure:
03 00 00 00 0f 03 02 [namelen] [setup-file name] …zeros… 27 03 10
[namelen]is a length prefix:0x28= 40 for the 40-character name read off the unit,0x09= 9 forTEST1.mmbwritten back. Confirmed both ways.- The trailing
27 03 10sits at0x108in the write body and0x113in the read — the same offset once the 11-byte prefix is accounted for.
A schedule entry names a setup file. That is how the help text's "change the record mode" works: an entry says at this time, load this setup. It also means the scheduler and the setup list are coupled — deleting a setup that a schedule references is a foot-gun worth checking before we ever expose either.
✅ The entry internals ARE decoded — see The schedule record format below.
The file turned out to hold two 260-byte records, not one: 27 03 10 is the
second record's time/day fields, not a trailer. 1/2h is a half-hour slot and
both times check out against what the operator entered.
SUB 0x47 — the scheduler enable, probably
Step 4 is two bare 0x47 frames with no data, differing only in params[7]:
req params = 00 00 00 00 00 00 00 [01] 00 00 → rsp 04 00 … 00 [01] 00 00 00 00 00 01
req params = 00 00 00 00 00 00 00 [03] 00 00 → rsp 04 00 … 00 [03] 00 00 00 00 00 01
The response echoes the selector and ends 01. Its 15-byte shape is identical
to the descriptor 0x48 returns on page 0.
⚠ Whether 0x47 sets or merely reads is genuinely undetermined. Both calls
returned the same trailing 01, and there is no before/after to compare — the
scheduler was enabled in this same capture, so no "disabled" reading exists.
params[7] is also the token position Series III uses, which is suggestive but
not evidence. Do not implement an enable against this until a
disable-then-enable capture settles it.
The scheduler enable is NOT in the setup config block
A prediction made before this capture — that the Scheduler On/Off switch would
appear as a byte in the 0x71 block, because the unit's help text lists it in
the same edit screen as Record Mode and Sample Rate — did not hold.
The 0x71, 0x68 and 0x82 payloads are byte-identical to the previous
capture's, all three, zero differences. The scheduler was enabled without any
of them changing.
⚠ The test is weaker than it looks: the operator re-sent the same config, so
a byte-identical block is also what "nothing changed" looks like. What it does
establish is that enabling the scheduler did not require a config write —
whatever 0x47 does, it does alone.
Monitoring control and the setup list (2026-09-24)
Capture: bridges/captures/9-24-26 - micromate2/*_turn_on_monitormode_*.
Thor started monitoring, listed the unit's setups, and stopped monitoring.
40 request frames and 40 responses, every checksum valid.
Thor's per-operation preamble
Thor re-runs this before every operation — three times in this one capture:
POLL (0x5B) → SERIAL (0x15) → 0x49 → POLL (0x5B) → <the operation>
⚠ Whether the unit requires this is untested — see Thor's conventions vs
the protocol's requirements. Thor sends it before trivial reads too, so it may
be habit rather than handshake. Do not assume it is mandatory. Note POLL here carries offset = 0x0030 (its data
length), not 0xFFFF — POLL is the one read Thor still addresses by length.
⚠ Narrowed 2026-09-27 — there is no universal preamble
Across all 8 captured sessions, the only invariant is that the session opens with
POLL. What follows depends on the operation:
opening sequence sessions operation 5b 15 49 5b …3 status refresh / monitoring / ACH change 5b 41 08 2e 1a da …3 setup push 5b 94 48 48 48 …2 scheduler read So
POLL → SERIAL → 0x49 → POLLis Thor's connection check, not a handshake the protocol demands — it appears where Thor wants to refresh what it displays. TreatPOLLas the one thing to send first.
🔑 No SESSION_RESET — the Series III requirement does not carry over
Series III needs a bare 41 03 (ACK + ETX, no STX) to wake a unit that is
actively monitoring; without it the unit will not answer POLL over TCP, and
protocol.startup() sends it before and between the POLL frames.
Thor never sends it to a Micromate. Zero occurrences across all 8 sessions
— including raw_bw_20260924_191214_turn_on_monitormode_…, which exchanges 40
frames with a unit that was monitoring at the time.
Measured offsets and response lengths (all read off Thor's frames)
offset = 0xFFFF for everything except two commands. Data lengths are from
UM12947 (11.0CB) and are orientation, not assertions — 0x1C is 4 bytes
longer on the Thor line.
| SUB | rsp | offset | data | notes |
|---|---|---|---|---|
0x5B POLL |
0xA4 |
0x0030 |
59 | the one length-addressed read |
0x15 serial |
0xEA |
0x000A |
21 | |
0x49 state |
0xB6 |
0xFFFF |
16 | |
0x1C monitor status |
0xE3 |
0xFFFF |
55 | +4 on 11.0BD |
0x06 storage range |
0xF9 |
0xFFFF |
47 | |
0x08 event index |
0xF7 |
0xFFFF |
101 | ⚠ 2026-10-19: it is the preferences block (units at content[31], CRC) |
0x2E trigger config |
0xD1 |
0xFFFF |
39 | |
0x1A compliance |
0xE5 |
0xFFFF |
2103 | one frame, not Series III's four |
0x2C call-home |
0xD3 |
0xFFFF |
137 | |
0x3F/0x40/0x41 setups |
0xC0/0xBF/0xBE |
0xFFFF |
266 | |
0x93 arm |
0x6C |
0xFFFF |
11 | ack only |
0x1E/0x1F chain |
0xE1/0xE0 |
0xFFFF |
19 | ⚠ token 0xFE at params[7] |
0x0C event record |
0xF3 |
0xFFFF |
221 | full key at params[4:8] |
0x0A monitor log |
0xF5 |
0xFFFF |
297 | ⚠ a walk — see below |
0x5A download |
0xA5 |
computed | offset+11 | 1024-byte chunk loop |
An acknowledgement is an 11-byte data section, and that doubles as the end-of-list signal on the walks.
⚠ 0x1E/0x1F carry token 0xFE
params = 00 00 00 00 00 00 00 fe 00 00 on all 7 captured chain reads — the
same token_params(0xFE) form Series III uses to arm its bulk stream. The
event-chain section above documents all-zero params; that was our own browse
probing, which also worked. Both evidently do, but Thor's form is the one with
mileage on it, and it is sent on browse and download alike.
🔑 SUB 0x0A is the monitor-log walk, not a keyed read
The command table long described 0x0A as a keyed "waveform header / partial
record" read, by analogy with Series III. What the bytes show is a cursor
walk: the same request repeated, the device advancing its own position.
0x93 → 1E → 0x0A ×8 (297 B each: "UM12947", "Histo…", "Ver…", " 0.49", " 28.4")
0x0A (11 B ack = end of list)
All nine frames carry identical params (…00 00 4a 81 00 00), so nothing in the
request selects the record. Terminate on a response of ACK_DATA_LEN (11).
⚠ The 4a 81 is the low two bytes of the event key then in play
(055d4a81). One key cannot distinguish "the key's low half" from "a cursor
handle that happened to equal it" — both produce those bytes. It does not
matter operationally, since the walk works with the params held constant.
This is a genuine structural divergence: Series III reaches the same data
through a record-type discriminator (0x2C partial vs 0x46 full) on its
event walk, so partial records and events share one chain. Here the monitor
log has its own cursor and the event chain never sees it.
USB identification — the Micromate's own ports
| device | VID:PID | appears as |
|---|---|---|
Micromate PC port (USB-B) |
2504:0300 — Instantel Inc. MICROMATE COM PORT |
/dev/ttyACM*, CDC-ACM |
| Sabrent cable (works on the USB-A host port) | 0403:* — FTDI |
/dev/ttyUSB* |
| Benfei cable (does not work) | 067b:* — Prolific PL2303 |
/dev/ttyUSB* |
⚠ Address the unit by /dev/serial/by-id/, not /dev/ttyACM<n>. The number
shifts whenever any other CDC-ACM device is plugged in, and this has already cost
a debugging session — the tool was aimed at a TMI-Seismo MCU and timed out
looking like a dead unit. The stable name is
usb-Instantel_Inc._MICROMATE_COM_PORT_V1.00-if00.
The PC port is a USB device port, so it is safe to plug and unplug freely.
The cold-boot requirement applies only to the USB-A host port, where the
modem cable goes.
Passing one through to a Proxmox guest works and hot-plugs, no VM restart:
qm set <vmid> -usb0 host=2504:0300. Bind by VID:PID rather than bus-port so it
survives a replug.
🔑 Every response has an 11-byte prefix — and its length byte lies
Content starts at data[11]. One rule, every command.
⚠ data[0] is the content length & 0xFF, and there is no high byte
anywhere in the prefix. data[1] is zero on every frame examined. So it
reads as a perfectly good length field for any response under 256 bytes — which
is most of them — and then:
| command | true content | data[0] says |
|---|---|---|
0x1A compliance |
2,092 | 44 |
0x0A monitor log |
286 | 30 |
0x5A download chunk |
1,024 | 0 |
0x41 setup name |
255 | 255 ✓ (only just) |
182 of 251 captured responses agree with a naive uint16 LE read; the 69 that do not are exactly the ones ≥ 256 bytes.
This is the third time a length in this protocol has been read too narrow —
after payload[9] vs payload[8:10] in the probe response, and after
data[0] here. The pattern is worth naming rather than fixing case by case:
take the content as data[11:] and let the frame's own length bound it.
Nothing needs the declared length; the frame already knows how long it is.
SUB 0x5B POLL content — where the strings actually sit
content[3] 0x50 ⚠ printable as "P", immediately before the vendor string
content[4] "Instantel\0"
content[17:26] binary — 06 00 c3 f0 4a 00 e4 19 4f 00 74 02
content[26] "MM/ISEE/S/IO\0" ← "MM/ISEE/S" on the Thor line
⚠ Do not parse this by scanning for printable runs. That 0x50 at
content[3] is printable, so a run scan returns PInstantel. Nothing
distinguishes a length or tag byte from text by inspection. Read the vendor
from the fixed offset and find the model by searching for MM/ — structural
rather than positional, which is what survives the model string's
firmware-line difference.
SUB 0x3F/0x40 setup walk — the terminator is an empty name
23 responses on the bench unit: 22 names, then a record whose name field is all
zeros. The empty name is the end of list — not an error, not a setup.
factory.MMB first, TEST1.mmb last.
SUB 0x01 has no Thor frame behind it
Thor never reads device info in any captured session. 0xFFFF for 0x01 comes
from our own 2026-09-23 probes — it answered correctly on both firmware lines,
but it is the only read in the table with no Thor precedent.
SUB 0x96 / 0x97 — start and stop monitoring ✅
Identical to Series III, including the acks:
| request | ack | effect |
|---|---|---|
0x96 |
0x69 |
start monitoring |
0x97 |
0x68 |
stop monitoring |
Both are bare frames — no params, no data — and both ack with the usual 16-byte
zero-data response. Thor follows each with a SUB 0x1C status read to confirm.
SUB 0x1C — monitor status, 55-byte data
data[12] monitoring flag NON-ZERO = monitoring, 0x00 = idle
data[13] day
data[14] month
data[15:17] year, uint16 BE
data[17] ⚠ unidentified — 32 while monitoring, 100 when idle
data[18] hour ─┐ device clock, verified against the capture's
data[19] minute │ own wall time (19:12:25 → 19:13:34 EDT)
data[20] second ─┘
data[-8:-4] memory total, uint32 BE = 15,000,000 bytes exactly
data[-4:] memory free, uint32 BE
⚠ Test data[12] for non-zero — do NOT compare against a constant. It read
0x0E on 2026-09-24 and 0x0C on 2026-09-25, both while monitoring, so it
carries sub-state in its low bits rather than being a flag. Series III's 0x10
does not apply either. SUB 0x49 data[11] read 0x02 in every monitoring
sample so far, but the same caution applies until more values are seen. Only five bytes differ between the monitoring and idle
responses: the flag, data[17], the clock, and the memory-free field.
Memory free dropped by exactly 4,096 bytes across the ~70-second monitoring session — monitoring allocates as it runs, so free memory is not a stable value to compare against.
The device clock is free here, on a command Thor already sends. That is worth having: nothing else read so far reports the unit's own time.
SUB 0x49 → 0xB6 — a cheap state check
16-byte data, in Thor's preamble before every operation:
05 00 00 00 00 00 00 00 00 00 00 [ST] e8 00 0b 00
↑ data[11]: 0x02 monitoring, 0x00 idle
Amended 2026-10-25. There are four values now.
0x03is a call-home session; see Correction:SUB 0x49has a third state value.0x01reads for about 18 s after a stop; its meaning is inferred to be a call-home attempt that has not connected. See The first commands this project originated against a Micromate.0x02was also read 2.5 s after a start, while the screen showed the sensor check.
A second monitoring indicator, in a 21-byte response instead of 0x1C's 60.
Different encoding from 0x1C's flag (0x02 vs 0x0E), so they are separate
fields, not the same byte read twice. For a polling client this is the cheaper
of the two, and Thor evidently treats it as the routine one.
SUB 0x3F / 0x40 — walking the setup-file list ✅
The same first/next shape as Series III's 1E/1F event walk, applied to
setup files:
| request | ack | meaning |
|---|---|---|
0x3F |
0xC0 |
first setup record |
0x40 |
0xBF |
next setup record; repeat until the name is empty |
Every record is 266 bytes of data and carries nothing but the name:
ff 00 00 00 00 00 00 00 00 00 00 <null-terminated name> 00 …
└──────────── 11-byte header ─────┘
The walk ends on a record whose name is empty — 24 records for 23 setups.
factory.MMB comes first, from 0x3F.
There is no active-setup flag in the list. The header is byte-identical on
every record including the terminator, and the tail is all zeros. The active
setup is identified only by SUB 0x41, whose response uses this exact record
format. (On the unit's own screen the active setup is marked with a trailing
asterisk — that is a UI decoration, not a field.)
TEST1.mmb, created over the wire earlier the same day, appears in the list and
is what 0x41 reports as active — independent confirmation that a written setup
becomes a real, enumerable file.
Thor refuses to send a setup while a unit is monitoring
Send-to-unit is greyed out in Thor's UI when the unit is monitoring, and
nothing is transmitted — this capture contains no 0xDA or 0x71 at all.
That is a Thor-side policy, not a device refusal: nothing observed suggests the Micromate would reject the write. It squares with the earlier finding that a push to the active setup overwrites silently — Thor is preventing exactly the footgun the protocol leaves open.
The reason behind the rule is sound and SFM should honour it — but not
necessarily by copying the greyed-out button. Stop → push → restart as a single
operation is what an operator usually wants, and 0x49 data[11] (or 0x1C
data[12]) makes the state check cheap either way.
Static analysis of the firmware (2026-09-23, solo session)
Architecture
ColdFire / 68K, big-endian, Freescale MQX RTOS — not ARM as the vector
table first suggested. The giveaway is the function epilogue/prologue
4E 5E 4E 75 4E 56 = UNLK A6 / RTS / LINK A6, littered through both
images, plus an MQX_OK assertion string.
CB vs BD: the same source, ~10 lines apart
A byte diff is useless — 68% of bytes differ because the two are separately linked builds with everything relocated. A string-set diff is position-independent and tells the real story: 17,128 strings shared, and almost every "unique" string is the same message with a different source line number:
CB: MONITOR[3268]: STATUS_BATTERY_LOW
BD: MONITOR[3258]: STATUS_BATTERY_LOW ← consistently 10 lines apart
The offset is exactly 10 across STATUS_BATTERY_LOW, STATUS_BATTERY_CRITICAL,
Battery Critical Exit Monitor, histogram interval size of 0 and
Offsets by channel — so one ~10-line block differs in the monitor module and
essentially nothing else. The only functional string unique to either build is
CITIZEN (a receipt-printer brand) in BD.
This corroborates the bench A/B from the other direction: the CB/BD split is a tiny code delta, not two protocol stacks. Whatever drives Instantel to ship two downloads, it is not a different wire protocol.
⚠ The SUB dispatch is a 68K switch jump table (CMPI.L bounds check →
MOVE.W (table,PC,Dn) → JMP (d8,PC,Xn)). Byte-pattern hunting will not
isolate the write opcodes — that needs a real disassembler.
Call-home config field names, from the firmware's own debug dump
CallHome.Enable = %s
CallHome.DialString = "%s"
CallHome.Retries = %d
CallHome.SessionTimeout = %d
CallHome.WaitForConnection = %d
CallHome.WarmupTime = %d
CallHome.PowerSave = %s
Seven fields, which is what the 126-byte SUB 0x2C block has to encode. Note
SessionTimeout and PowerSave have no Series III equivalent, and Series III's
scheduled-time fields (time1/time2 hour/min) are absent here — consistent
with Series IV moving scheduling into the THOR-downloaded scheduler instead.
Also present: AT+CSQ (signal quality), so the firmware talks AT to the modem
directly.
All five bench events, downloaded and decoded
Read-only, over USB, no Instantel software:
| key | declared size | got | decoded |
|---|---|---|---|
055d4a81 |
4,076 | 4,076 | histogram, 1 interval, 16:33:16 |
055d4a82 |
11,032 | 11,032 | waveform, 3072 × 4 ch, 16:33:19 |
055d4a83 |
11,502 | 11,502 | waveform, 3072 × 4 ch, 16:33:27 |
055d4a84 |
13,424 | 13,424 | waveform, 3072 × 4 ch, 16:33:34 |
055d4a85 |
8,746 | 8,746 | waveform, 2048 × 4 ch, 16:33:36 |
Every event arrived at exactly its declared size, every channel came out equal length, and the timestamps are sequential across the recording session.
Record type + filename: generate it, don't detect it
🔑 Update 2026-09-29 — there IS a type field, in the
0x0Crecord
0x0Ccontent[11]:0x07= waveform,0x08= histogram.Found by diffing the six bench events'
0x0Crecords against their known types: exactly one byte separates the two groups and is constant within each. It then predicted the right suffix for 6 of 6 on a blind re-run.⚠ Six events, split 4/2. That is thin evidence for a byte that could be a counter, a channel count or a mode. Treat it as a strong candidate, not a settled field, and keep a fallback:
bridges/mm_client_check.pytries the other suffix on failure and reports when the guess was wrong, which is how this would be caught rather than silently mis-filing an event.This does not overturn the section below — the filename still has to be generated, and the type still comes out of the protocol rather than the payload. It just means the type is available from a command we already send for every event, instead of needing to be carried out of the chain walk separately.
⚠ The claim below that the type comes from "
SUB 0x0Alength0x1E= histogram,0x00= waveform" is not visible in the 2026-09-24 download capture, where0x0Aappears once as a standalone monitor-log walk after the last chain entry, not per event. Either it was observed in a session not in the repo, or the two are being conflated. The0x0Cbyte is reproducible from bytes on disk; prefer it.
read_idf_file() decides waveform vs histogram from the filename suffix —
and there is no filename when downloading over the wire.
⚠ Worth correcting a natural assumption: Series III does not detect this from
content either. event_file_io.derive_record_type_from_filename() reads the
last character of the extension (M529LKIQ.G10H → H → Histogram). Nothing
in the codebase infers record type from file content, for either family.
And there is no obvious type field to find. The first 64 bytes of a histogram
and a waveform are byte-identical; they diverge at ~0x0947 into wholly
different structures rather than differing by a flag.
The answer is the Series III pattern — generate the name. Series III has
blastware_filename(), which builds a name from serial + timestamp + type.
Series IV needs the same thing, and its convention is far simpler:
<serial>_<YYYYMMDDHHMMSS>.IDF{W,H} e.g. UM12947_20260923163319.IDFW
versus Series III's <letter><serial3><4-char base-36 stem><AB0T ext>, where
the stem is base-36 of seconds-since-1985 ÷ 1296.
All three inputs are already available on a direct download:
| input | source |
|---|---|
| serial | extract_binary_metadata() — decoded from the IDF header |
| timestamp | extract_binary_metadata() — same |
| type | the chain walk — SUB 0x0A length 0x1E = histogram, 0x00 = waveform |
Verified against all five bench events: the generated names match the
convention of real files in the production store byte for byte. A directly
downloaded event can therefore be filed under exactly the name Thor would have
given it, and /db/import/idf_file needs no change at all.
⚠ The type still comes from the protocol, not the payload — so a downloader must carry it out of the chain walk. Losing it means losing the ability to name the file correctly.
✅ RESOLVED 2026-09-30: the 0x0C float is the PER-SAMPLE peak vector sum
0x0C content[Tran label − 12], float32 BE, = the peak vector sum computed
sample by sample. Exact to 0.000% on all four bench waveforms, against a PVS
recomputed from the decoded samples:
| key | 0x0C float |
per-sample PVS | err | sqrt(Σpeak²) |
max(T,V,L) |
|---|---|---|---|---|---|
…82 |
1.37198 | 1.37198 | +0.000% | 1.41746 | 1.37063 |
…83 |
2.35414 | 2.35414 | +0.000% | 2.37824 | 2.22274 |
…84 |
3.51515 | 3.51515 | −0.000% | 3.78731 | 3.44194 |
…85 |
0.42267 | 0.42267 | −0.000% | 0.46678 | 0.41985 |
⚠ Retraction — why this looked unresolved
The previous entry (below, kept) observed the float running 2–5% above
max(T,V,L)and concluded it was not the peak vector sum, because "computed PVS runs higher than both".That computed PVS was
sqrt(T² + V² + L²)from the three reported channel peaks — and that is an upper bound on the real thing, not the real thing. The channel maxima do not occur at the same instant, so combining them overstates the true peak. Every measurement in the old table is consistent with the correct answer:max(T,V,L) ≤ stored ≤ sqrt(Σpeak²), in all five rows.A near-miss worth recording: on UM20147's event the stored float matched
sqrt(Σpeak²)to 0.048%, which briefly looked like proof it was that quantity. It is not — the peaks on that event simply happen to be nearly coincident in time. One sample agreeing with a formula is not evidence when a second formula fits it equally well; the six CB events separated them immediately (errors to −9.5%).
The offset is now established, not inferred — four independent confirmations at 0.000%, plus a fifth on the other firmware line.
Why this is worth having: it is a free self-check on the decoder. The device computed this number from the same samples we decode, independently of our codec. If a decoded event's per-sample PVS does not match the stored float, the decode is wrong. Given this codebase's history of silent channel truncation — walker bugs that shorten a channel and raise nothing — a per-event invariant that costs one float comparison is cheap insurance. Worth wiring into the download path.
⚠ It remains true that max(T,V,L) and sqrt(Σpeak²) are not
interchangeable with it, so a consumer wanting the vector sum must use this
field or decode the samples — it cannot be reconstructed from the three
per-channel peaks.
Superseded: the original "unresolved" entry
The float32 extracted from 0x0C runs 2–5% above max(Tran, Vert, Long) from
the decoded samples:
| key | 0x0C float |
max channel |
|---|---|---|
…81 |
3.5152 | 3.4419 |
…82 |
1.3720 | 1.3706 |
…83 |
2.3542 | 2.2227 |
…84 |
3.5152 | 3.4419 |
…85 |
0.4227 | 0.4198 |
It is not peak vector sum either (computed PVS runs higher than both). The
field may not be the peak at all — its offset was inferred from a byte marker,
not established. Do not rely on it until it is pinned properly.
Worth noting the histogram (…81) and the loudest waveform (…84) report
identical peaks to four decimals, in both measures. That is self-consistent:
the histogram's single 1-minute interval spans the whole thumping session, so
its maximum should equal the loudest event in it.
0x0C field map, from UM20147 (11.0BD, 2026-09-30)
Real bytes, 221 B response → 210 B content declared at data[0] = 0xd2:
content[ 0] day 0x1e = 30
content[ 1] month 0x09
content[ 2:4] year BE 0x07ea = 2026
content[ 4] ⚠ 0xb3 = 179 — unidentified, same slot as 0x1C's content[6]
content[ 5:8] h:m:s 13:27:33
content[11] RECORD TYPE 0x08 histogram / 0x07 waveform
content[12] "Location\0" sensor-location label
content[34] "test2\0" the setup file name, without its extension
content[76] "UM20147\0" serial
content[86] float32 BE the per-sample PEAK VECTOR SUM (= Tran − 12)
content[98] "Tran\0\0" + float32 BE at +6
content[112] "Vert" … content[126] "Long" … content[140] "Mic" …
⚠ The setup name here is test2 — no extension — where 0x41 reported
test2.MMB and 0x40 reported test2.mmb in the same session. Three commands,
three spellings of one file. See Setup-name case.
Event download, per-event delete, and ACH config (2026-09-25)
Three captures with operator-supplied ground truth, including Thor screenshots of the event list and the schedule. All Thor-originated; we still send nothing.
✅ Day is the day of week — and [0] is the schedule type
A weekly schedule settles both fields. Thor's screen: "multiday, Repeat Weekly: Disabled, Start Monitoring at 8:00 AM every day, alternating TEST1 / test2." The file is 7 × 260 + 4 = 1,824 bytes, and every record matches:
| rec | 1/2h |
time | Day |
[6] |
name | |
|---|---|---|---|---|---|---|
| @0 | 16 | 08:00 | 0 | Sun | 2 | TEST1.mmb |
| @260 | 16 | 08:00 | 1 | Mon | 2 | test2.mmb |
| @520 | 16 | 08:00 | 2 | Tue | 2 | TEST1.mmb |
| @780 | 16 | 08:00 | 3 | Wed | 2 | test2.mmb |
| @1040 | 16 | 08:00 | 4 | Thu | 2 | TEST1.mmb |
| @1300 | 16 | 08:00 | 5 | Fri | 2 | test2.mmb |
| @1560 | 16 | 08:00 | 6 | Sat | 2 | TEST1.mmb |
Day = 0 Sunday … 6 Saturday, and the alternating setup names line up with
the screen row for row.
[0] on record 0 read 4 here, against 2 and 3 in the daily schedules —
so it carries the schedule type as well as Repeat:
[0] |
meaning |
|---|---|
2 |
Daily, repeat off |
3 |
Daily, repeat on |
4 |
Weekly, repeat off |
5 |
Weekly, repeat on — ⚠ predicted, not observed |
Bit 0 is Repeat; 2 and 4 are the Day/Week bases. Same bitmask style as
[6].
⚠ In daily schedules Day reads 3 in every record of every capture. With
[0] already saying "daily", the field is presumably ignored there — but that is
inference, and 3 is unexplained.
Event download — SUB 0x93 arms each event
Thor's download of six events:
POLL 0x15 0x49 POLL 0x1C 0x15 0x49 0x06
0x93 0x1E 0x0C 0x5A×4 ← event 1
0x93 0x1F 0x0C 0x5A×11 ← event 2
0x93 0x1F 0x0C 0x5A×12 ← event 3
0x93 0x1F 0x0C 0x5A×14 ← event 4
0x93 0x1F 0x0C 0x5A×9 ← event 5
0x93 0x1F 0x0C 0x5A×6 ← event 6
0x93 0x06 0x93 0x1E 0x0A×9 ← re-walk the list
2026-10-21 -- direct-by-key download equals the interleaved download (measured, one unit over USB, 5 events listed, 1 waveform compared).
On UM20147 (/dev/ttyACM0), with list_events() having walked the cursor to the
end of the chain, download_event(ref, chunk_size=1024) of 055d4a82 returned
30,230 B, byte-identical to the same event fetched inside iter_events() at the
cursor, and to the file SFM stored through the server (30,230 B).
So 0x5A does not need the cursor parked at the event, at least on this unit
and event; not tried on a BD unit, over cellular, or on a histogram.
s4_download_event was not changed.
2026-10-21 -- in Histogram Combo, a histogram's 0x0C peaks equal the largest
waveform recorded inside it (measured, UM20147, 2 histogram/waveform pairs).
UM20147's list showed histogram 13:27:33 and waveform 13:27:43 with identical
peaks and PVS to the last float bit, and likewise histogram 13:28:09 and
waveform 13:59:33 (31 minutes apart, before the next histogram at 13:59:53).
Not a listing bug: the histogram file's own interval data peaks at vert 5,021 /
tran 735 / long 675 counts = 1.5581 / 0.2281 / 0.2095 in/s at 0.000310308 in/s
per count, the waveform's peaks exactly. The histogram's record reports the
maximum over its whole window. 🔶 Inferred: the record is finalised when the
window closes but stamped with its start, which is how it can carry the peak of
a waveform that triggered later inside it. Do not "fix" equal peaks across
such a pair.
SUB 0x93 → ack 0x6C is sent before every event, with empty params and an
all-zero ack. It is the Series IV analogue of Series III's 1E(token=0xFE) arm
step, and it is simpler: no token, no params.
Event keys are a plain sequential counter. The six events walked as
055D4A81 … 055D4A86, consecutive, at data[11:15] of the 0x1E/0x1F
response, with the event's byte size at data[17:19]. Nothing like Series III's
flash addresses.
SUB 0x0A walks the event list returning 30-byte records carrying start and
stop timestamps ([11] day, [12] month, [13:15] year BE …), terminating on
an all-zero record. The dates match Thor's event list exactly (five events on
09/23/2026, one on 09/24/2026).
🔑 Delete is PER-EVENT — SUB 0xA8 + 0xAA
0xA8 params[0:4] = <event key> → ack 0x57
0xAA params = zeros → ack 0x55
The operator deleted the top row of Thor's list — the newest event — and
0xA8 carried 05 5d 4a 86, the highest key from the walk. Thor lists
newest-first, so newest = highest key. Confirmed end to end.
This is the last piece a homebrew ACH receiver was missing. The manual established that collection state is server-side bookkeeping and that "delete events from unit" is a separate server-issued action; this is that action's opcodes.
Two things worth noting against Series III:
- Series III erases everything (
0xA3 → 0x1C → 0x06 → 0xA2). Series IV deletes one named event, which is strictly safer — a receiver can delete exactly what it has confirmed it stored. - The opcodes are different. Do not reach for
0xA3/0xA2here.
⚠ SUB 0x06 (storage range) read identically before and after the delete in
this capture, so it is not a quick way to confirm a deletion landed. Re-walk the
event list instead.
✅ ACH config — 0x2C / 0x7E / 0x7F, exactly Series III
Thor enabling then disabling Auto Call Home, twice through the same sequence:
POLL → 0x15 → 0x49 → POLL → 0x2C (read) → 0x7E (write, 126 B) → 0x7F (confirm)
Same SUBs and the same acks as Series III — 0x2C→0xD3, 0x7E→0x81,
0x7F→0x80. The write payload is 126 bytes with offset = 0x007E, and
the 0x2C read returns those same 126 bytes behind an 11-byte prefix, so
read[n + 11] is write[n].
The enable flag is write[5] (= read[16]):
| value | state |
|---|---|
0x05 |
Auto Call Home enabled |
0x04 |
disabled |
Bit 0 is the enable — the same bit position Series III uses at raw[5], but with
bit 2 set as well in both states. Do not test for 0x01/0x00.
write[6:] holds the dial string, null-padded — "RADIO RING" on this unit,
matching Series III's raw[6:46].
⚠ write[118:120] — a field the UNIT maintains
Three samples:
| when | enable | field |
|---|---|---|
| 00:50, idle (schedule session) | 0x04 |
43 23 |
| 01:16, ACH read #1 | 0x04 |
43 23 — unchanged after 26 min |
| 01:16, ACH read #2, after a config write | 0x05 |
2e 5e |
Thor echoes back whatever it last read, including a value that no longer matches the config it is sending — and the write is accepted anyway.
What that rules out:
- Not a clock or timer — identical across 26 minutes of idle. Only a config write moved it.
- Not a counter — it went down, 17187 → 11870.
- Not computed by Thor — Thor demonstrably sends a stale value.
- Not a standard CRC16. Brute-forced all 65,536 polynomials × init
{
0x0000,0xFFFF} × all four reflection combinations, over four candidate regions ([0:118],[0:118]+[120:],[5:118], and the payload with the field zeroed). No match. Recorded so nobody repeats the sweep.
It behaves like a hash or checksum the unit computes over the stored config: a
one-byte input change (0x04→0x05) scattered the output completely, XOR
0x6D7D, where a sum would have moved by 1. But two samples cannot separate
that from a nonce regenerated on each write.
✅ Operationally this does not matter. The unit does not validate the field on input — Thor proved that twice in one capture. Read the config, change your field, send everything else back byte-for-byte. Never synthesise or zero it. Same read-modify-write shape as the setup block.
Resolving it properly needs more samples: several config writes with the times noted, which is cheap to collect during any future ACH capture.
Neighbouring bytes: write[117] is 0x03, lining up with Series III's
num_retries of 3, and write[120:122] is 00 3c = 60.
⚠ Beyond the enable byte and the dial string, the field map is not established — only one setting was varied. Series III's offsets are a starting hypothesis, not a transfer.
What the THOR manual settles about the ACH session (2026-09-24)
Source: manuals/723U0201 THOR Operator Manual Rev 08.pdf §6.2, §3.6.
⚠ This is vendor documentation, not observed bytes. It tells us the shape of
the session and the vocabulary; it does not give opcodes. Treated as a strong
prior, not as confirmed protocol.
🔑 A unit does NOT learn that an event was accepted — the server decides
This document previously listed, as the thing gating any homebrew receiver:
how a unit announces itself, and how it learns an event was accepted so it stops re-sending it.
That was the wrong question. There is no acknowledgement mechanism to discover, because the unit is not tracking what has been collected. Per §6.2.2.2 the ACH session is a list of server-chosen actions, and two of them are independent:
| ACH action | what it does |
|---|---|
| Copy events | downloads events — "Only applies to events not previously downloaded" |
| Copy monitor log | downloads the monitor log |
| Delete events and Logs from Unit | explicitly deletes them from the unit |
| Set Date/Time | unit synchronises its clock from the computer |
| Send Events / Schedule to Vision | Instantel cloud, not relevant to us |
"Not previously downloaded" is computer-side bookkeeping. The unit keeps its events until a server tells it to erase them, and the manual's own warning proves the two are decoupled:
⚠ "If you enable 'Delete Events and Logs from Unit' but disable 'Copy Events'. The events and logs will be deleted without being uploaded."
A server that never issues the delete simply re-reads the same events forever.
This is exactly the model our Series III ACH server already implements —
ach_state.json with downloaded_keys / max_downloaded_key, and erase as a
separate deliberate step. No new mechanism is needed for Series IV.
⚠ But no Micromate has ever been observed calling home (noted 2026-10-03)
The 72
ach_inbound_*sessions inbridges/captures/are MiniMate Plus. The Series III setup is not known to transfer, and there is a concrete reason it might not:
Series III (observed, 72 sessions) Series IV (inferred) who dials the modem (destination set in ACEmanager) unknown the unit's role passive on serial until the socket exists sends AT strings to the modem session pull — server runs the ordinary command set pull, per manual §6.2 + firmware strings ⚠ The Micromate demonstrably talks AT to its modem —
ATQ1 ATE0 ATS0=2 … RADIO RINGfrom the RV50 debug, and"RADIO RING"is what0x2C's dial-string field actually contains on UM12947.ATS0=2is auto-answer-after-2-rings, so that exchange is the unit arranging to accept calls — the call-up path — which argues the ACH destination lives in the modem after all.But if the unit instead drives an AT dial for call-home, that exchange is on the serial link and a TCP listener sees none of it. So the first experiment has an ambiguous null: no connection could mean the trigger did not fire, or that the interesting part is invisible from the network. The escalation is a serial-level capture between unit and modem, not more work on the server.
So a homebrew receiver needs: accept the connection, identify the unit, walk the
events (already solved — 0x08/0x1E/0x0A/0x5A), keep our own high-water
mark, and optionally erase. The erase opcodes are the only genuinely missing
piece, and they remain on the unsafe list.
🔑🔑 CAPTURED 2026-10-14 — a complete, working ACH session
MITM between UM12947's modem and THOR 1.5.0.29. 101 request frames, 101
responses, 0 bad checksums, 34,872 B of event data in 65 seconds.
bridges/captures/mitm-october/ach_mitm_20261014_182612/
🔑 Every command in a real ACH session is one we already implement — except one.
| SUB | name | count |
|---|---|---|
0x5B |
POLL | 36 |
0x5A |
BULK_DOWNLOAD | 27 |
0x49 |
STATE | 10 |
0x0C |
EVENT_RECORD | 8 |
0x1F |
EVENT_NEXT | 7 |
0x06 |
STORAGE_RANGE | 4 |
0x1C |
MONITOR_STATUS | 3 |
0x93 |
ARM_EVENT | 3 |
0x15 |
SERIAL | 1 |
0x1E |
EVENT_FIRST | 1 |
0xA2 |
see below | 1 |
Session shape — note THOR polls between every command:
POLL POLL SERIAL POLL STORAGE_RANGE POLL MONITOR_STATUS POLL STATE POLL
ARM_EVENT EVENT_FIRST POLL EVENT_RECORD POLL STATE POLL
[ EVENT_NEXT POLL EVENT_RECORD POLL STATE POLL BULK_DOWNLOAD... ] x n
POLL STORAGE_RANGE POLL STATE POLL 0xA2
⚠⚠ SUB 0xA2 ends a call-home — and COLLIDES with Series III's ERASE
The session's final command is 0xA2 at offset = 0xFFFF, answered 0x5D.
THOR's log at that instant:
TX: SENT REQUEST: StartMonitoringAfterCallhome
WARN Switching unit back to monitoring; IP: 63.45.161.30
SESSION COMPLETED SECCESSFULLY
So on Series IV, 0xA2 = exit call-home and resume monitoring.
⚠⚠ On Series III, 0xA2 → 0x5D is CONFIRM ERASE — the last step of
0xA3 → 0x1C → 0x06 → 0xA2, after which the unit's memory is cleared.
Same opcode, same response SUB, opposite consequence.
That is the most dangerous collision found in this protocol. Any code path shared
between the two families — and minimateplus and micromate already share a
transport — must dispatch on series before 0xA2 goes near a wire. Sending it
to a MiniMate Plus to "end a session" erases it.
It also corroborates the firmware string found by static analysis in September:
CMD_EXIT_CALL_HOME_DELETE_EVENTS_START_MONITORING.
⚠ "Monitor While Calling Home" silently blocks erase — and that fills units
Manual §11.4.4: a Micromate can call home while it continues to monitor, or pause monitoring and send. UM12947 is on the former, which is why this session logged:
CANNOT DOWNLOAD MONITOR LOG. The unit is monitoring.
CANNOT DELETE ALL EVENTS. The unit is monitoring.
⚠ THOR reports the session as SESSION COMPLETED SECCESSFULLY anyway. The
refusals are INFO lines. From the outside, a unit that is quietly accumulating
looks identical to one being managed correctly.
TMI's normal fleet setting is delete-after-download, and it is not optional: without it a unit fills up and re-offers every event on every call. So a unit left on monitor while calling home is misconfigured relative to the fleet, and drifts toward a full buffer on its own.
🔑 Design consequence for SFM: a receiver cannot rely on erase to manage
storage. The erase can fail for a reason the unit considers normal, so the
server-side high-water mark is load-bearing, not an optimisation — it is the
only thing preventing a full re-download every session. Series III's
ach_state.json (downloaded_keys / max_downloaded_key) is the pattern; it
has to carry over deliberately.
🔑 And it is a detectable fleet-health signal. "Memory filling and deletes
refused" is visible on every call from 0x06's event count plus 0x1C's memory
fields — worth surfacing, precisely because THOR calls it success.
⚠ Correction: 0x06 content[4:8] is NOT a constant
Recorded on 2026-10-01 as "9 on both units regardless of count — a constant,
not a count". It reads 47 here, four times in one session, where THOR's log
twelve minutes earlier reported Monitor Log count: 45 on the same unit.
It is almost certainly the monitor-log record count. Both units reading 9 earlier simply meant both held 9 monitor-log records. ⚠ Two samples and a near-miss (47 vs 45, twelve minutes and an active monitoring session apart), so treat as a strong candidate rather than settled.
content[0:4] remains the event count — 8 here, consistent.
⚠ Correction: SUB 0x49 has a third state value
Recorded as 0x00 idle / 0x02 monitoring. Throughout this ACH session it reads
0x03, ten times out of ten — so there is a distinct in-call-home state.
This is exactly why the rule is "test for non-zero". A client checking
== 0x02 would report a unit mid-call-home as idle.
⚠ THOR rewrites a unit's stored IP from the call-home source
After this session THOR showed UM12947's connection details as
TCP: 127.0.0.1 - Port: 9034 — the address the MITM connected from. THOR
learns a unit's address from its inbound call, which makes sense for units on
dynamic IPs and means a relay silently poisons THOR's record of where that unit
lives.
🔑 Both ACH paths are real — and which one runs depends on timing
⚠ Softened 2026-10-14 — "mm_ach_server is built on the wrong model"
An earlier entry concluded the inbound-listener model was wrong because
ListenerAchTcphad never accepted a connection. It has now:
condition path taken THOR already holds a connection to the unit ListenerAchRadioRing— the ring rides that connectionnothing is connected the modem dials out, ListenerAchTcpacceptsBoth are real. The listener model is valid; it simply loses whenever something already holds the modem's single session slot. That also explains every failed probe and relay attempt of 2026-10-14: they were not bugs, they were contention with THOR's
UnitWatchDogTimer, which reconnects continuously.
🔑 Call-home can also be an IN-BAND WAKE SIGNAL, not a dial-out
From THOR's own ACH log (1.5.0.29, 4,414 lines, UM12947 over an RX55). This overturns the model this document has carried since the beginning, and it explains every confusing result of the preceding weeks.
A Micromate does not dial anybody. It emits the ASCII string RADIO RING
over its serial port, and that text travels up a TCP connection the server
already has open to the unit. The server recognises it and starts a session
on that same connection.
17:18:17,268 TcpClientSystemNet #1: RADIO RING IS RECEIVED FROM IP:63.45.161.30 and Port:9034
17:18:17,268 ListenerAchRadioRing #2 RADIO RING IS DETECTED BY Serial ListenerAchRadioRing
17:18:17,304 ConnectorWorkerThread Process Request: ReadSerialNumber <- who is this?
17:18:18,489 ListenerAchRadioRing #3 RADIO RING UNIT SERIAL NUMBER = UM12947
17:18:18,490 ListenerAchRadioRing #4 NOTIFICATION IS SENT TO ACH MANAGER
17:18:18,499 ACHManager #8 RADIO RING STARTED A NEW SESSION
⚠ TcpClientSystemNet is THOR's outbound TCP client. The ring arrives on
the connection THOR itself opened to unit:9034 — the ordinary call-up path.
There is no inbound connection, no dial, nothing to accept.
THOR keeps that connection alive deliberately. UnitWatchDogTimer reconnects
to every unit continuously — 1,039 lines of it in 37 minutes. The call-home
mechanism depends on that connection existing.
2026-10-29 (Brian, operator-reported; not measured by us): the watchdog's interval is a THOR setting. It can check every 5 s, every hour, or never. The 1,039 lines above were taken at whatever that THOR was set to, so "continuously" describes that setting, not THOR's fixed behaviour. Two things are still unmeasured: whether a check holds the line open afterwards or drops it, and how a unit's call-home reaches THOR when the watchdog is set to never (the modem dialling out to
ListenerAchTcp, presumably, which this capture never saw used). The ring-path bench test in the parity spec now varies this setting on purpose.
THOR has a second, unused ACH path
ListenerAchTcp logs Waiting for a ACH connection on :: Port 12346 — and in
4,414 lines it never accepted one. All twelve ACH events came through
ListenerAchRadioRing. So the inbound-TCP ACH listener exists (presumably for
units that genuinely dial, on some other modem arrangement) but is not what a
PAD-connected Micromate uses.
What this means for a homebrew receiver
⚠ bridges/mm_ach_server.py is built on the wrong model for this deployment.
It listens for an inbound connection, which is ListenerAchTcp's model — the one
THOR never exercises here. The right shape is:
- Hold an outbound connection to each unit (
unit:9034), reconnecting as needed — this is a poller, not a listener - Watch the byte stream for
RADIO RING - On seeing it, identify the unit (THOR sends
ReadSerialNumber) and run an ordinary session on the same socket
That also explains the single-session contention that dogged every experiment: the receiver must hold the modem's one session slot permanently, so anything else probing that unit will be met with silence. Those are not separate problems; holding the slot is the design.
The ACH session, from THOR's state machine
InitializeState ReadEventRecordCount, ReadOperatingMode
CheckEventsCount
DownloadEventState ReadUnitACHStatus, ReadEventAddressFirst/Next,
SaveEventByAddress
DownloadMonitorLogState ReadMonitorLogRecordCount
DownloadScheduler
SetDateTimeState
DeleteEventState / DeleteAllEventsState / DeleteMonitorLogState
CheckEventWhileMonitoringBeforeEndingState / ...AfterDownloadingState
LoopState
EndWithSuccessState / EndWithCommunicationErrorState / EndState
Request counts over the log: ReadUnitACHStatus 53, SaveEventByAddress 47,
ReadEventAddressNext 43, ReadOperatingMode 10, ReadEventRecordCount 10,
ReadEventAddressFirst 4, StartMonitoringAfterCallhome 3,
ReadMonitorLogRecordCount 3.
StartMonitoringAfterCallhome matches the firmware string
CMD_EXIT_CALL_HOME_DELETE_EVENTS_START_MONITORING found by static analysis in
September — independent corroboration that the unit resumes monitoring when the
server finishes.
✅ ReadEventRecordCount returned Event count: 19 — corroborating the
SUB 0x06 event-count finding from a completely different direction.
⚠ Events are addressed as 90000001 here (length: 4076), where the chain walk
we implemented returns 055d4a81 for an event of exactly that length. Same
event, different address space — unexplained, and worth care before assuming our
keys and THOR's addresses are interchangeable.
Superseded 2026-10-14 — "THOR's ACH side is POLL"
First capture of THOR's half of a Micromate call-home, via ach_mitm.py between
UM12947's modem and a THOR 1.5.0.29 ACH server on port 8070.
unit -> THOR 60 B " \nRADIO RING \nRADIO RING \nRADIO RING \nRADIO RING"
THOR -> unit 84 B 4 x 41 02 10 10 00 5b 00 00 30 00 00 00 00 00 00 00 00 00 00 9b 03
THOR sends POLL (SUB 0x5B, offset 0x0030) — and nothing else. Four
times over ~21 seconds, then it gives up. It does not send a greeting, does
not answer the RADIO RING, does not send CONNECT, and does not wait to be
spoken to beyond its initial silence.
🔑 That POLL is byte-identical to the one micromate.protocol builds.
Verified: build_request(SUB_POLL, 0x0030) reproduces all four frames exactly.
So the ACH session is pull, and our receiver already implements it correctly.
bridges/mm_ach_server.py does what THOR does. The "inbound call-home session"
was never a separate protocol — it is the ordinary command set on an inbound
socket, exactly as Series III's 72 captured sessions suggested.
⚠ But the unit answered neither — and that is the open problem
In the same capture the unit sent only RADIO RING, four times, and never a
protocol frame. THOR polled four times and got nothing. Our own server, in three
earlier attempts the same afternoon, saw the identical thing.
The vendor's own software fails the same way we do, which eliminates the receiver as a suspect and places the fault between the modem and the unit.
What is established about that fault:
| cellular link | ✅ healthy — TCP connect in 151 ms, outbound dial succeeds |
| modem → us | ✅ bytes arrive |
| unit serial TX | ✅ something reaches us — the RADIO RING text |
| unit protocol mode | ❌ never entered; no frame ever sent |
| cable | FTDI, physically checked, cold-booted with it attached |
| modem serial config | 115200 8N1, no flow control, PAD/TCP — matches spec |
⚠ "RADIO RING" is the literal string in this unit's own 0x2C call-home
config (the 40-byte dial-string field). Whether the unit emits it or the modem
does is not yet established — the cheap disambiguation is to disconnect the
unit and see whether the string still appears.
Next step is a config diff, not more protocol work. The deployed fleet runs call-home over PAD successfully, so a working unit's modem settings are the reference. Compare them against this one.
Which matches the firmware's state machine and its
CMD_EXIT_CALL_HOME_DELETE_EVENTS_START_MONITORING /
Call Home Deleting Events strings: the unit dials in, then waits to be told
what to do, and returns to monitoring when the server is finished.
Scheduled ACH and event-triggered ACH behave differently (§3.6.1):
- Event-triggered (ACH enabled on the unit, an event occurred) — "Records events + transfer data, no stopping to monitor." Governed by Monitoring While Calling Home; with MWCH enabled "the monitor log will not be copied, events will not be deleted, and time will not synchronize".
- Scheduled — "Stop monitoring + transfer data (or delete events, or synchronize)." Monitoring stops and in-progress events complete first.
⚠ Worth noting for SFM: with MWCH on, an event-triggered session cannot delete or sync. A receiver that relies on erase to avoid re-reading will silently never erase on those units. Our high-water mark must be the primary mechanism, with erase as an optimisation — which is how Series III already does it.
Session Time Out is unit-side only
§6.2 is explicit that THOR configures ACH on both ends "with one exception; the
Session Time Out. This must be configured directly on the unit." That matches
the firmware's CallHome.SessionTimeout field, and means it lives in
callhome.MMB rather than anywhere THOR reaches — one more reason to read that
file with 0x94.
Unit identification is by serial number, with wildcards
§6.2.2.2: filters match against the serial number the unit presents, *
wildcards allowed, best-match wins — UM* all Micromates, MP* all Minimate
Pros, BE/BC Minimate Plus variants.
So the unit announces its serial early enough for the server to route on it.
SUB 0x15 returns the serial and is in Thor's standard preamble, which is
consistent, though the inbound direction has still never been observed.
⚠ The manual's worked example contradicts its own table — it says "To filter Micromate units type MP"* and "To filter Minimate Pro units type UM"*, which is backwards. The table is right. Noted so nobody copies the error.
Thor requires Idle for configuration
§6.2.1 step 3: "The Monitoring Mode must be 'Idle'." Independent confirmation that the greyed-out send observed on the bench is deliberate Thor policy, applied to ACH setup as well as compliance setup — and still not evidence the device refuses.
The schedule's actions, from the UI side
§3.6 lists exactly four actions THOR exposes: Start monitoring, Stop
monitoring, Self-check, Auto Call Home. The firmware has five — it also has
a setup-less DUTYCYCLE_START_MONITOR.
§3.6.2 step 8 confirms the coupling is Thor's own requirement: "Select the
appropriate Action and the Unit Setup. (A Unit Setup must exist…)" — while §3.6.2
step 1 notes "The unit will execute any actions in a schedule using its current
settings." The two statements sit a paragraph apart and pull in opposite
directions, which is consistent with START_MONITOR existing and THOR never
emitting it.
Also from §3.6: schedules are Day or Week ("Select Day… Select Week"), have a
Repeat Daily / Repeat Weekly flag, carry a Name, Description and Unit Type,
and "Saving a schedule will only store it on the computer, it must still be sent
to the unit and enabled" — which is why the capture shows 0x8E (send) and
0x47 (enable) as separate steps.
Those are the fields to look for when the schedule entry is finally decoded:
action, setup name, time, day-or-week, day selection, repeat. The one
captured entry (03 00 00 00 0f 03 02 …) has seven bytes before the name, which
is the right order of magnitude for that field list.
What THOR's "status check" actually is (2026-09-25)
Captured with bridges/mm_link.py standing in for the modem, THOR configured
Communication: TCP, Check connection every 5 s, Check status every 5 s.
It is eleven commands, not one
POLL → DEVICE_INFO → 0x49 → 0x5C → MONITOR_STATUS → SETUP_NAME_READ
→ STORAGE_RANGE → 0x02 → OPERATOR → 0x47 → CALLHOME_CFG
Each check opens a new TCP connection, runs all eleven exchanges in ~300 ms, and closes it. Measured over 21 consecutive checks:
| request payload | 236 B |
| response payload | 936 B |
| TCP connect + teardown | ~1,000 B (fresh handshake every time) |
| per check | ~2.2 KB |
The cost
| cadence | checks/day | per unit |
|---|---|---|
| every 10 s (as observed) | 8,640 | 18.8 MB/day — 563 MB/month |
| every 60 s | 1,440 | 3.1 MB/day — 94 MB/month |
| every 15 min | 96 | 0.2 MB/day — 6 MB/month |
⚠ On a metered cellular plan this is real money, and most of it is waste.
A status check re-reads the call-home config, the operator name, the active setup
name and the full device info block every ten seconds — none of which changes
between checks. SETUP_NAME_READ alone returns 274 bytes each time.
A genuine liveness + state check is POLL + MONITOR_STATUS: two commands,
131 bytes of response. Same information about whether the unit is alive and
whether it is monitoring, for a ninth of the traffic.
There are TWO checks, and only one of the dials is honoured
The eleven-command sequence is the status check. There is also a much cheaper connection check — three commands:
connection check POLL → SERIAL → 0x49
status check POLL → DEVICE_INFO → 0x49 → 0x5C → MONITOR_STATUS
→ SETUP_NAME_READ → STORAGE_RANGE → 0x02 → OPERATOR
→ 0x47 → CALLHOME_CFG
⚠ With both dials equal, the connection check never runs separately at all — every connection observed was the full eleven. It only appears once the two intervals differ. That alone explains a lot of "changing the settings does nothing": at equal values you only ever get the expensive one.
With connection check = 10 s and status check = 30 s, the steady state over 14 consecutive cycles is:
FULL at T
short at T + 10.1
short at T + 25.3 (+15.2)
FULL at T + 30.4 (+5.1)
| dial | set to | observed | |
|---|---|---|---|
| status check | 30 s | 30.4 s | ✅ honoured |
| connection check | 10 s | 15.2 s | ❌ 52% slow |
27 consecutive connection-check gaps, all 15.1–15.3 s. That is systematic, not jitter.
⚠ A "half the status period" hypothesis was raised here and REFUTED. It predicted that with status at 60 s the connection check would run at 30 s. Measured: 60.5 s. Recorded rather than deleted — the prediction was sharp and it was wrong.
Modem settings live in SysParm.cfg, not in the setup — and that is deliberate
Settled by experiment (2026-09-25). The unit's call modem setting was
changed on the keypad from generic to USB to PC. Then the active setup was
switched from TEST1 to test2 on the device, and the setting was checked
again: unchanged.
So it does not travel with the setup. It is device-global, in the SysParm.cfg
the firmware strings already hinted at, alongside SysPref.bMonitorScheduler.
That closes a hypothesis this document was pursuing: the ~102 bytes a .MMB
file carries beyond the 2,090-byte config block are not where this lives.
Those bytes remain unexplained, but they are no longer a candidate for it.
⚠ It also means the earlier observation "the call-home block (0x2C) was
byte-identical across 218 samples today" does not bear on this — the last
0x2C sample was at 13:30:23 and the keypad change came around 13:35. No sample
exists on the far side of it.
Why the split is the right design
The operator's reading, and it is convincing: you do not want modem settings reachable remotely. Get them wrong over the air and you lose the connection you would need to put them back — the unit must then be visited physically.
That explains the shape of the protocol rather than treating it as an oddity:
| reachable over the wire? | |
|---|---|
Call Home enable, dial string, retries, timings (0x2C / 0x7E) |
yes — recoverable if wrong |
Modem type and the rest of SysParm.cfg |
no — a wrong value is unrecoverable remotely |
SUB 0x2C is not an incomplete view of the modem configuration. It is the
deliberately-chosen subset that is safe to change remotely, and the
unreachable remainder is unreachable on purpose.
Design principle for SFM
There is a class of setting whose misconfiguration destroys the channel you would use to fix it — modem type, baud, serial parameters, the call-home destination itself. For those:
- Do not expose them for remote write, or gate them behind an explicit confirmation that names the risk.
- If they must be writable, use a commit/confirm with automatic rollback — apply, require the unit to call back within N minutes, revert if it does not.
- Always allow reading them, so an operator can diagnose a unit they cannot reconfigure.
Instantel chose option 1. Given the failure mode, that is defensible, and worth copying rather than improving on.
The connection dial sets a COUNT, not a period
Three configurations, each measured over many cycles:
| status | connection | status observed | connection observed | separate conn checks per status cycle |
|---|---|---|---|---|
| 10 s | 10 s | 10.1 s ✅ | — | 0 (none ever ran) |
| 30 s | 10 s | 30.4 s ✅ | 15.2 s ❌ | 2 |
| 60 s | 30 s | 60.5 s ✅ | 60.5 s ❌ | 1 |
The status dial is honoured in all three — within about 1%.
The connection dial is not honoured in any of them. What actually holds is:
separate connection checks per status cycle = (status ÷ connection) − 1
1 → 0 checks, 3 → 2 checks, 2 → 1 check. So the field does not set a period; it sets how many checks fit inside a status cycle, and it is off by one.
Consequences worth naming:
- Setting the two dials equal yields zero connection checks. Every connection is then the expensive eleven-command status check. That is the default-looking configuration, and it is the worst one.
- "Every 30 s" with status at 60 s gives one check per minute — half the advertised rate.
- The observed cadences are not multiples of the connection setting at all (10 s → 15.2 s; 30 s → 60.5 s), so no simple scale factor describes it either.
⚠ The timing within a cycle is still not fully explained. At status 30 / connection 10 the two short checks landed at T+10.1 and T+25.3, where an evenly divided cycle would put them at T+10 and T+20. The count rule holds across all three configurations; the phase does not follow from it.
For SFM: this is the concrete argument against exposing a dial whose behaviour is not what it says. If a control sets an interval, it should set that interval — and if the effective rate is derived from something else, the UI should show the derived value, not the input.
Traffic at each configuration
| status / connection | per minute | per day | per month |
|---|---|---|---|
| 10 s / 10 s | ~13 KB | 18.8 MB | 563 MB |
| 60 s / 30 s (current) | ~3.4 KB | 4.9 MB | 147 MB |
| a minimal check at 60 s | ~0.2 KB | 0.3 MB | 9 MB |
The last row is POLL + MONITOR_STATUS only — enough to answer "alive?" and
"monitoring?".
⚠ What this means for the "won't stay connected" failure
Because idle polling opens a fresh connection each time, a silently-dead link is less dangerous here than expected — a dead socket fails at connect and the next cycle simply tries again. The earlier hypothesis (THOR wedged on a half-open socket during polling) is weakened by this.
The exposure is during operations, not polling: THOR was observed holding one connection from 00:30 to 00:47 — seventeen minutes — while downloading events and pushing setups. A link that dies silently mid-operation leaves THOR waiting on a socket the OS will not fail for roughly the default keepalive (~2 h). That remains the best candidate for a unit that will not come back and where refresh does nothing.
Testable with mm_link.py: set blackhole during a download or config
push rather than while idle, and watch whether THOR ever gives up, whether
refresh emits any bytes, and whether it recovers when pass is restored.
Design notes for SFM
Three things this argues for, all cheap:
- Separate liveness from inventory. Poll
POLL+MONITOR_STATUSfrequently; read config, setup name and call-home settings only when something says they changed, or on demand. - Show the connection attempt. Every check here has a visible outcome — connected, frames exchanged, closed. THOR surfaces none of it, which is why a failing unit is undiagnosable from the UI. The log this section is built from took one afternoon to produce and answers questions THOR cannot.
- Bound every operation with its own timeout, independent of TCP's. Do not rely on the socket to report a dead peer.
Reproduced: THOR stops polling after a connection drops mid-download
2026-09-25, one trial, on the bench link. This is the failure the operator described in the field on UM12947 — "just wouldn't stay connected… clicking the little refresh button did nothing… no way to view the status of a connection attempt."
What was done
THOR was downloading events. Partway through the bulk transfer the link was faulted, the connection died, and the link was then fully restored.
⚠ What THOR actually experienced was a TCP close mid-download, not the silent
link that was intended — mm_link.py had a bug (socket.timeout subclasses
OSError, so 200 ms of quiet was mistaken for a closed socket). Fixed, but the
run stands as a drop-mid-download test, which is arguably the more realistic
case: a modem losing its tower often does produce a real close.
What happened
13:02:49.0 THOR connects, bulk download running (165 BULK_DOWNLOAD frames in)
13:02:50.0 link faulted -> connection dies mid-transfer
13:03:10.4 THOR reconnects once, 20 s later, sends SUB_1F (advance event
pointer) — an attempt to resume
13:03:10.8 that attempt dies too
13:03:33.3 link fully restored and healthy
... 48 s of complete silence ...
13:04:21.4 operator clicks Refresh -> full 11-command check, every response
correct, unit perfectly healthy
... 2 min 6 s of complete silence ...
13:06:27 still nothing. That refresh is the ONLY connection since 13:03:10.
Before the fault THOR was connecting every 30 s without a miss for over an hour.
What this establishes
- One retry, then give up. THOR retried once after 20 s and stopped. No backoff, no further attempts.
- The automatic poll loop dies with it. Not just the download — the 30 s connection check and 60 s status check both stopped entirely.
- It does not recover when the link returns. Sixteen minutes of healthy link produced no automatic connection attempts at all — only ones an operator initiated by hand.
- Refresh works, but only once. The manual check succeeded completely — all eleven commands, correct responses. It did not restart the automatic loop; two minutes later there was still nothing.
⚠ Point 4 is the dangerous one. Refresh makes the UI report a healthy unit while nothing is actually watching it. The operator sees a successful check and moves on; the background loop stays dead. That is worse than a refresh button that plainly fails, because the failure is silent and looks like success.
🔑 "Not polling" and "unreachable" are different states — and THOR conflates them
Demonstrated directly. At 13:10:44 an independent POLL through the same
relay succeeded with a correct response while THOR's automatic polling was dead.
⚠ A correction, recorded because the first version of this claim was sloppy.
It originally read "THOR connections since 13:04:22 : 0". That count was an
artifact of a time filter (13:0[4-9]) that silently dropped everything from
13:10 onward. THOR had connected four times, at 13:10:03–13:10:37.
Inspecting those four shows the conclusion survives — but for a better reason than the one first given. They were user-initiated commands, not polling:
13:10:03 POLL → SUB_96 (start monitoring) 13:10:05 MONITOR_STATUS
13:10:35 POLL → SUB_97 (stop monitoring) 13:10:37 MONITOR_STATUS
Single commands a human clicked, each followed by one status read. Neither the eleven-command status check nor the three-command connection check appears.
Automatic polling did not resume, and by 13:19:53 had been absent for 16 minutes. Every THOR connection in that window was something an operator clicked: the refresh at 13:04:21, then start and stop monitoring at 13:10.
The unit is reachable. The link is fine. THOR is simply not asking.
THOR reports Connection Status: Connected. Captured at 13:15, eleven
minutes after its last contact with the unit:
Connection Status : Connected <- false; nothing checked in 11 min
Last Updated : 09/25/2026 01:04:22 PM <- true, and that is the REFRESH
Notification : "Unable to download event(s) from UM12947 due to:
Did not receive response from unit." (01:03:31 PM)
Three things worth separating:
- The green tile is false. It reports a live connection that has not been exercised for eleven minutes.
Last Updatedis true — and it is the only honest field on the screen. THOR knows when it last succeeded. It renders that as small grey text under the unit name, unhighlighted, with nothing marking it stale, beneath a large green Connected tile. The operator must read a timestamp and do mental arithmetic to discover the headline is wrong.- The failure THOR did report was the download, not the poller stopping. Those are treated as unrelated events; nothing says "automatic checking has ceased."
So the state "I have given up checking this unit" is not merely undisplayed — THOR holds the data that would reveal it and presents a contradicting summary instead. So an operator cannot separate the unit is down from the poller is asleep — and those demand completely different responses: a site visit versus a mouse click.
This also settles a question the previous section left open. It does not much matter whether stopping after one retry is an intentional design choice or a defect: the reporting is wrong either way, because both plausible displays misrepresent reality.
It explains, too, why the field symptom resisted characterisation. The unit is perfectly reachable throughout. Poke it and it answers. Nothing is wrong with the unit, the modem, or the link.
The stale display extends to monitoring state — a live unit shown as Idle
The operator's own test: start monitoring from the unit's keypad, and see whether THOR ever notices.
At 13:21:15, verified by an independent probe through the same relay:
unit, on the wire SUB 0x49 data[11] = 0x02 MONITORING
SUB 0x1C data[12] = 0x0c MONITORING
device clock 13:21:14
THOR's last contact 13:10:37 (a stop-monitoring command it issued itself)
THOR's display Connected · Monitoring Mode: Idle · Last Updated 1:04:22 PM
A unit is actively recording and THOR shows it as Idle. Had an event triggered in that window, THOR would not have known and would not have collected it. This is the operational consequence of the wedge: not merely a wrong indicator, but a monitoring system that has silently stopped monitoring its monitor.
One further detail. THOR did contact the unit at 13:10:37 and read
MONITOR_STATUS — yet Last Updated still reads 1:04:22. So a successful
exchange does not refresh that timestamp; only the full status check does. The
single honest field on the screen is honest about the wrong thing, which makes it
useless as a staleness indicator precisely when staleness is the problem.
Design consequences for SFM
- Connection status must EXPIRE. The single highest-value fix here. If the last successful check is older than a small multiple of the check interval, the state is stale / unknown — never Connected. THOR already has the timestamp; it simply does not let it invalidate the summary.
- Retry must be unbounded with backoff, never one-and-done. A transient link fault should not take a unit out of service until a human notices.
- A manual check must restart the automatic loop, or the UI must state clearly that automatic checking is stopped. A green tick that means "it worked when you asked" is a trap.
- Surface the poll loop's own state — last successful check, last attempt, next scheduled attempt, consecutive failures. All four were invisible here, and every one of them would have made this diagnosable in seconds.
- Distinguish "unit unreachable" from "we stopped checking." Proven above to be different states that THOR cannot tell apart. A unit's displayed state should be one of: checks passing, checks failing (with the attempt count and last error), or not being checked (with why, and a way to resume). Collapsing the last two into one indicator is the root of this whole failure being undiagnosable.
Caveats
- One trial. Not yet repeated, and not yet tried with a true silent link
(
blackholewith the fixed tool) or with a cleandrop. - THOR's UI state during the dead window was not recorded — worth capturing on a repeat, since what the operator sees is half the problem.
- Whether a THOR restart is required, or whether it eventually recovers on a longer timescale, was not tested.
THOR leaks event subscriptions — evidence from its own log
Source: thor.log.2026-09-22 from the office THOR PC, supplied 2026-09-25.
209 lines covering roughly ten hours.
What the log contains
178x UnitOperatingModeViewModel - Start/stop monitoring request timed out: False
6x UnitOperatingModeViewModel - Callback of start/stop monitor request received
5x UnitOperatingModeViewModel - Start/stop monitoring request timed out: True
3x SchedulerViewModel - Retrieving unit setup ...
2x BWLicenseManager - License State: IsKeyValid 0, IsActivated 0, ...
85% of a day's logging is one message. Only 12 lines describe an actual operation.
The message is emitted N times for ONE event, and N grows
Grouping the repeated message by exact timestamp to the millisecond — so each group is one logical event, not repeated events:
| time | identical lines |
|---|---|
| 11:48 | 1–4 |
| 13:05 – 13:31 | 1–2 |
| 13:36 – 13:40 | 2–4 |
| 15:16 | 3 |
| 16:03 – 16:07 | 6 |
| 16:14 – 16:16 | 4–6 |
| 22:00 – 22:06 | 12 |
1 → 12 over ten hours of uptime. Twelve identical lines sharing one millisecond timestamp is not twelve events; it is one event dispatched to twelve handlers.
What it is
A subscription leak, and the class name identifies it:
UnitOperatingModeViewModel. THOR is a .NET/WPF application ("App thread",
ViewModel naming), where this is the archetypal leak — a view model subscribes to
an event when its view opens and never unsubscribes when the view closes. Each
open adds a live handler; every later notification runs them all.
Consequences that follow directly:
- N grows with uptime and usage, without bound. Ten hours of moderate use reached 12×; a THOR left running for days across a fleet will be far worse.
- Every notification performs N× the work — N timers, N callbacks, N state updates racing one another.
- A restart resets N to 1, which is consistent with the operator's report that only restarting THOR recovers a degraded session.
- The only
timed out: Trueentries in the file — five of them — sit at the very top, an episode captured just before log rotation.
⚠ Claim discipline. Established from this log: the handler count grows. Strong inference: it is a subscription leak. Not established: that it caused the specific field outage on 2026-09-22. This log does not cover that window, and the connection between leaked handlers and a dead TCP path is unproven.
How to confirm it in ten minutes
Restart THOR, note the burst size for one event, open and close a unit's detail view ten times, then trigger another event and re-count. If the burst grows, the leak is reproducible and attributable to a specific UI action.
For SFM
- Unsubscribe on teardown, or use weak event patterns. This is the single most common long-running-UI defect and it is entirely preventable.
- Log once per event, not once per handler. The duplication here is what made the leak visible, which is lucky — but a log that is 85% one repeated line is also a log nobody reads, which is why this went unnoticed for a year.
- Log what changed, not what did not.
timed out: False178 times is noise; the fiveTrueentries are the signal, and they are buried.
The Micromate's USB host supports FTDI and CDC-ACM only — not Prolific
Established 2026-09-25, by isolating a bench modem link layer by layer and then confirming against both firmware images.
The USB-A port is a HOST port with a fixed driver set
usbHostDelete_CDCACM CDC-ACM — the standard USB-serial class
usbHostDelete_FTDI FTDI chips (FTDISER handlers, set_line_coding, etc.)
usbHostDelete_MFS mass storage (thumb drives)
usbHostDelete_PRINTER printers
usbHostDelete_HUB / HC / USBH
Identical in 11.0CB and 11.0BD — 11 FTDISER strings and 22 CDCACM
strings in each, and zero matches for prolific, pl2303, cp210, ch34
or silabs in either. The firmware line makes no difference here.
⚠ So a Prolific PL2303 cable (VID 067b) very probably cannot work with a
Micromate — no driver, so the device never enumerates and the serial path does
not exist. Use an FTDI cable (VID 0403) or a CDC-ACM one.
Calibration. This rests on absence of strings, plus the observation that a
laptop answers on the same cable where the unit does not. usbHostDelete_* reads
like a complete per-class list, which is good evidence — but driver code can exist
without a matching string, and the USB enumeration path has not been
disassembled.
The test that would remove the inference: try an FTDI cable on the same modem. Either it works or it does not, and nothing is left to argue about.
✅ Corroborated by purchasing history (2026-09-25). TMI has bought two kinds of USB-to-serial cable:
| cable | chipset | works with a Micromate? |
|---|---|---|
| Sabrent USB 2.0 to serial | FTDI | ✅ driver present |
| Benfei USB to RS-232 | Prolific PL2303 | ❌ no driver |
So both types are in circulation and indistinguishable by eye. A unit handed the wrong one has no working modem port, and nothing about the cable's appearance says so.
The cable on the bench during this investigation was confirmed to be a Benfei — the Prolific one.
⚠ Identify cables by chipset, not by looks. lsusb on any Linux box:
FTDI is VID 0403, Prolific is 067b. Note also that counterfeit FTDI chips
are common in cheap cables — they carry FTDI's VID but may not behave like one,
and an embedded host with a single driver is far less forgiving than Linux.
✅ This accounts for the 2026-09-22 failure
An earlier draft of this section said the timeline did not fit, because the unit was described as working at first and then degrading — which a Prolific cable cannot do, since it never enumerates at all.
The timeline was the thing that needed correcting, not the finding. From the operator, 2026-09-25:
- Initial setup — one modem, one cable. Worked: configs sent, status read, no trouble.
- Two Micromates then had to be deployed, so a second cable was fetched to run both at once.
- Only one of the two ever worked. The other would not talk over its modem at any point.
- That unit was never deployed — a MiniMate Plus was put out in its place, and the Micromate came back to the bench.
- The cable on that bench unit is a Benfei (Prolific PL2303).
So nothing degraded. The "worked at first" was the original single-cable setup; the failure began when the pair was split across two cables and one of them was the wrong chipset. One working and one not, set up side by side, is the signature of a mixed cable supply — not of a unit fault.
✅ The cable travelled home with the unit (confirmed 2026-09-25). So the Benfei on the bench is physically the cable from the failed office setup — not a substitute picked up later. There is no inference left in the chain:
| the cable from the failed unit | is a Benfei / Prolific PL2303 |
| the Micromate's USB host | has no Prolific driver, in either firmware line |
| a laptop on that same cable + modem | round-trips perfectly |
| the Micromate on it | never answered |
The FTDI cable arriving 2026-09-26 is now a confirmation rather than a test.
Corollary worth knowing. The standing workaround — "just deploy a Series III instead" — works partly because a MiniMate Plus has a DB-9 port directly on the unit. No USB-to-serial adapter anywhere in the path, so there is no chipset to get wrong. That workaround has been quietly routing around this exact failure mode.
✅ RESOLVED (2026-09-26) — and it took a cold boot as well as the cable
Swapping to a genuine FTDI cable (lsusb: 0403:6001, FT232) did not work on
its own. A probe immediately after the swap returned the same connected, no
reply.
A cold boot of the unit was also required. Holding the power button for five
seconds — through the two-stage prompt that disconnects the internal battery —
and powering back up with the cable already attached, brought it straight up.
THOR at the office connected the moment it finished booting, and an independent
probe returned UM12947, idle over cellular.
⚠ The USB host does not rescan on hot-swap. The Micromate enumerates its USB-A host port at boot; once it has failed to identify a device there it does not appear to try again. Changing that cable is a power-cycle operation, and anyone swapping one in the field who probes immediately will conclude the new cable is faulty too.
So the full chain for a Micromate on a modem is:
- an FTDI or CDC-ACM cable — a Prolific PL2303 gives the unit no serial port
- a cold boot with that cable attached
✅ These modems bridge ONE session at a time — confirmed
Long suspected, never demonstrated. Demonstrated now, by accident:
| probe result | |
|---|---|
| THOR connected to the unit | TCP connect ok … NO REPLY |
| THOR disconnected, nothing else changed | reply 68 B, UM12947, idle |
The PAD accepts a second TCP connection and then does not forward it. It does not refuse, and it does not close — it simply never bridges.
⚠ This means "connected but no reply" has two causes that are indistinguishable
from the client side: a genuinely broken serial path, and somebody else
already has the unit. bridges/mm_probe.py originally reported only the first,
which would send a diagnosis in exactly the wrong direction; its verdict now
names both and tells you to eliminate contention first.
For SFM this is a design constraint, not a footnote. Our receiver and THOR cannot both hold a unit, and during any migration both will exist. "Another client holds this unit" needs to be a distinct, visible state — not folded into a failure, and certainly not into a green tick.
How this was isolated — the method is reusable
A Micromate on an RX55 was unreachable from THOR. Rather than guess, each layer
was eliminated in turn with bridges/mm_probe.py:
| layer | test | result |
|---|---|---|
| unit's public IP | static APN mw01.VZWSTATIC |
ruled out |
| firewall / trusted IPs | probe from a whitelisted source | TCP connect ok, 268 ms |
| network path | same | ruled out |
| baud | 115200 both ends | ruled out |
| unit modem-type setting | set back to generic |
ruled out |
| THOR | probe bypasses it entirely and still failed | ruled out |
| modem TCP→serial | swapped the unit for a Linux box on the same cable | POLL arrived byte-perfect |
| modem serial→TCP | that box answered with a canned POLL reply | round trip in 700 ms |
The last two are the decisive pair. With a laptop standing in for the unit on the same cable and modem, the full round trip works — so the modem, the PAD configuration, the firewall and the network are all clean, and the only element that changed is what sits at the end of the cable.
scratch/fake_unit.py-style substitution is worth remembering: replace the
device with something known-good and re-run the same test. It converts "the
unit is not answering" into "the unit is not receiving", which are very
different problems.
⚠ This is the best explanation for the bench setup. Whether it explains the 2026-09-22 field outage is not established — that unit reportedly worked at first and degraded, which a wrong cable would not do. Treat them as separate until the field unit's cable is identified.
Known-good RV55 config for a Micromate (2026-09-25)
Read off a unit deployed and working for months. This is the reference to diff a misbehaving modem against — the fastest diagnostic available, since TMI has seven working units and only one that misbehaves.
Serial interface — Port 1
| field | value | note |
|---|---|---|
| Mode | PAD | |
| Baud rate | 115200 | Micromate rate. Series III units use 38400 — a modem moved between series needs this changed |
| Data bits / Parity / Stop bits | 8 / None / 1 | |
| Flow control | None | hardware FC blocks TX when the pins are not wired |
PAD settings — Port 1
| field | value | note |
|---|---|---|
| PAD mode | TCP | |
| Listen for connections (auto-answer) | On | this is what lets THOR dial in. Off, and only call-home works — which would present exactly as "unreachable but it still calls home" |
| Listening port | 9034 | the port THOR dials |
| Destination address | 50.197.32.91 | the office public IP — where the unit calls home to |
| Destination port | 12345 | the office ACH listener |
| MTU | 1304 | |
| Data forwarding timeout | 500 ms | |
| Idle timeout | 2 min | matches the value this project settled on for Series III |
| Connect timeout | ≈1.08 min | |
| Start only when WAN available | Off | |
| TCP reconnect on disconnect | Off (locked) | |
| TCP keepalive | Off | ⚠ see below |
| Keepalive idle / interval / count | 30 s / 10 s / 3 | configured but inactive while keepalive is Off |
Cellular
APN mw01.VZWSTATIC — Verizon's static-IP APN, which confirms the fleet's
addresses are fixed and rules out any "the IP moved" explanation. IPv4 only,
MTU 1428, radio firmware VERIZON.
⚠ Two things worth questioning in this config
TCP keepalive is Off. Nothing detects a half-open TCP session from the modem's side; the 2-minute idle timeout is the only reaper.
And an idle timer can be held open indefinitely by a client that keeps writing. THOR polls every 30 s (measured — see the polling section). If it holds a session the unit has stopped answering on, and keeps writing into it, each write plausibly resets the modem's idle timer. The session then never ages out, the PAD's single slot stays occupied, and every new inbound connection fails — while ACEmanager keeps answering on its own separate service.
⚠ Hypothesis. Whether traffic in one direction resets that timer is not
confirmed, and the single-slot behaviour has not been tested on a real modem.
bridges/mm_probe.py --slots N tests the second part directly.
Turning keepalive On would give the modem an independent way to notice a dead peer, rather than depending on an idle timer a chatty client can hold open. That is a change to make deliberately and one modem at a time.
What to diff on a misbehaving modem
In rough order of likelihood for "worked at first, then stopped":
- Idle timeout — 0/disabled lets a stale session hold the slot forever
- Listen for connections — Off means inbound never works, but call-home still does
- Baud rate — 115200 for Micromate, 38400 for Series III
- Listening port — must match what THOR dials
- Destination address/port — wrong here breaks call-home but not inbound
- Data forwarding timeout and MTU — affect framing, not reachability
Note 2 and 5 fail in opposite directions, which is diagnostic in itself: inbound broken with call-home working points at the listener; call-home broken with inbound working points at the destination.
Thor's conventions vs the protocol's requirements
SFM is not meant to reimplement Thor. Thor is the only available teacher of the wire protocol, but almost nothing about how it sequences its work has been shown to be required by the device. Those are two different things and this document should not blur them.
The distinction matters because several observations above were written as "a client should do X" when the honest statement is "Thor does X, and we have not checked whether the unit cares."
| Thor does this | required? | what it means for SFM |
|---|---|---|
POLL → 0x15 → 0x49 → POLL before every operation |
unknown | Series III needed POLL×3 before 5A specifically, so a preamble requirement is plausible — but Thor sends this before trivial reads too. Testable: issue one operation cold and see if it answers. |
Reads with offset = 0xFFFF, skipping the probe |
no — the probe works | We have a genuine choice, and the probe is arguably better: it reports the length instead of making us trust a fixed one. This is the one place Thor's shortcut is probably worse. |
Sends 0x68 + 0x82 in every setup push |
unknown | In both setup captures these carried near-zero payloads and changed nothing. If they are optional, a setup write is 3 frames instead of 7, with less to get wrong. Worth testing before we build the writer. |
| Rewrites the whole 2,090-byte config for a one-field change | unknown | No narrower write has been observed. Read-modify-write is safe and known; a targeted write would be nicer but is unevidenced. |
Reads 0x41 twice in a row (scheduler capture, frames 6 and 8) |
no | Plainly redundant. A reminder that Thor's sequence is not a minimal one. |
| Greys out send-to-unit while monitoring | Thor policy | The reason is real — a push to the active setup overwrites silently. But "the button is grey and you figure it out" is a UX choice, not the only answer. SFM could offer stop → push → restart as one operation, which is what an operator actually wants. |
Names the target with 0xDA before a config write |
almost certainly required | The unit has to know which file to write. Closest thing here to a genuine protocol requirement. |
The schedule↔config coupling is Thor's, not the protocol's
In Thor, putting "start monitoring" in a schedule forces you to attach a setup, and sending the schedule pushes that setup at the same time — overwriting whatever already has that name. The protocol does not require any of this.
The evidence, from the scheduler capture:
-
The two writes are separate operations, not one transaction. The config write ends at frame 18 (
0x72), then Thor sends a freshPOLLpreamble, and only then opens the schedule for writing at frame 20. Different commands, different paths, no shared state:config 0xDA → 0x68/0x73 → 0x82/0x83 → 0x71/0x72 schedule 0x8D → 0x8E -
The schedule stores a NAME, not a config. A length-prefixed filename is all an entry carries. It is a reference, and references do not require the referent to be rewritten.
-
The config Thor pushed was already on the unit, unchanged. Its 2,090-byte
0x71payload is byte-identical to the previous capture's — zero differences. Thor spent a full block write re-sending a setup the device already had, purely to satisfy its own coupling.
So the sequencing is a Thor UI decision, and a costly one.
What SFM can do instead, with today's protocol and nothing new:
- Enumerate the unit's setups with
0x3F/0x40— cheap, read-only, and it yields the exact names a schedule may legally reference. - Write only the schedule (
0x8D/0x8E) when the referenced setup is already present. - Push a config only when it is actually missing, or when the operator deliberately edited it — and say so explicitly rather than silently.
That removes the write entirely from the common case: scheduling against a setup that already exists becomes a 524-byte schedule write instead of 524 bytes plus a 2,090-byte config overwrite.
⚠ And it removes a real hazard. Because the reference is by name, and because a same-name write overwrites silently with an indistinguishable ack, Thor's pattern means scheduling something can quietly rewrite a setup that other schedules — or the operator's own saved work — depend on. Nothing in the protocol or the ack reports that this happened. Validating the reference instead of rewriting the referent avoids the whole class of problem.
✅ The schedule record format — DECODED (2026-09-24)
A debug printf in the scheduler names the fields outright:
SCHEDULER : _ReadRecord(%d) -> %s (Action=%u, 1/2h=%u, Day=%u, Setup="%s") [WDAY=%d]
The captured file is two 260-byte records plus four zero bytes — 524 bytes exactly. Every non-zero byte in the file falls inside those two records.
offset field size notes
[0] schedule flags 1 record 0 ONLY — 3 = repeat on, 2 = repeat off
[1:4] padding 3 zero in every record observed
[4] 1/2h 1 half-hour slot, 0–47
[5] Day 1 3 in every record observed — undetermined
[6] Action 1 bitmask: 2 start, 4 stop, 8 self-check, 16 ACH
[7] name length 1
[8:260] Setup name 252 null-padded; empty for actions that take no setup
⚠ [0] and [6] were mislabelled twice before a five-entry schedule
separated them — see CORRECTED below.
Decoded against the operator's own description of what they entered — "start monitoring TEST1 at 7:30 AM, Auto Call Home at 7:30 PM":
| record @0 | record @260 | |
|---|---|---|
Action |
3 | 0 |
1/2h |
15 → 07:30 ✅ | 39 → 19:30 ✅ |
Day |
3 | 3 |
[6] |
2 | 16 |
| name length | 9 | 0 |
| Setup | TEST1.mmb |
(none) ✅ |
Five independent confirmations, no fitting required:
- Slot 15 = 07:30 and slot 39 = 19:30, both matching the operator's stated times, on a 30-minute grid — and 39 − 15 = 24 slots = exactly 12 hours.
- The name-length byte reads 9 for
TEST1.mmb, 40 for the 40-character name in the read capture, and 0 for the Auto Call Home entry. - Auto Call Home carries no setup name — which is what it should do, and direct proof that a schedule entry can exist with no setup attached.
- The 260-byte stride lands the second record's
1/2hexactly at[264]. 2 × 260 + 4 = 524, the whole body, with nothing left over.
⚠ Correction: the 27 03 10 bytes at [264] were previously recorded here as
a possible trailer. They are record 2's 1/2h, Day and [6] fields. Not
a trailer — a second record hiding behind an assumption that the file held one.
Action codes, so far:
| value | action |
|---|---|
0 |
Auto Call Home |
3 |
Start monitoring, with setup |
⚠ CORRECTED — [6] is the Action, and it is a bitmask (2026-09-25)
A five-entry schedule settled this, and it overturns the two previous readings of this record. The operator's Thor screen, captured alongside:
actioncheck — Micromate — Repeat Daily: Disabled
7:30 AM Start Monitoring TEST1
8:00 AM Stop Monitoring
8:30 AM Self Check
9:00 AM Start Monitoring test2
7:30 PM Auto Call Home
Five entries, and the file holds exactly five 260-byte records:
| rec | [0] |
1/2h |
Day |
[6] |
len | name |
|---|---|---|---|---|---|---|
| @0 | 2 | 15 = 07:30 | 3 | 2 | 9 | TEST1.mmb |
| @260 | 0 | 16 = 08:00 | 3 | 4 | 0 | |
| @520 | 0 | 17 = 08:30 | 3 | 8 | 0 | |
| @780 | 0 | 18 = 09:00 | 3 | 2 | 9 | test2.mmb |
| @1040 | 0 | 39 = 19:30 | 3 | 16 | 0 |
[6] tracks the action exactly, and the values are powers of two:
[6] |
action |
|---|---|
2 |
Start Monitoring |
4 |
Stop Monitoring |
8 |
Self Check |
16 |
Auto Call Home |
Bits 1–4 of a bitmask. Bit 0 (value 1) is unobserved — a natural home for
the setup-less DUTYCYCLE_START_MONITOR, but that is a guess, not a finding.
Two retractions:
[6]was recorded as "the one unidentified field" and predicted to be the Repeat flag. It is the Action.[0]was labelled Action, then "Action with repeat folded in". Both wrong.[0]is non-zero only on record 0 — including here, where record 0 and record 3 are the same action with different[0]values. It is a schedule-level field carried in the first record, and it holds the Repeat flag:3repeat on,2repeat off, matching "Repeat Daily: Disabled" on the screen.
The earlier repeat capture was consistent with both readings because it had one start entry and changed one byte. A single-variable test is not always enough — it took a schedule with four distinct actions to separate the two fields.
Day is 3 on all five records across every capture. Still undetermined.
✅ SUB 0x47 is the scheduler enable — confirmed by Thor's own notifications
Previously recorded as "genuinely undetermined" whether 0x47 sets or reads.
Thor's notification pane timestamps it:
00:51:29 schedule write (0x8E) completes
00:51:31 Thor: "actioncheck has been successfully SENT to ... UM12947"
00:51:31 → 0x47 params[7] = 0x01
00:51:33 → 0x47 params[7] = 0x03 (on the wire as 10 03 — DLE-escaped)
00:51:35 Thor: "actioncheck was successfully ENABLED on ... UM12947"
The only frames between sent and enabled are the 0x47 pair. 0x47
performs the enable.
⚠ The meaning of params[7] ∈ {1, 3} is still open — a single 0x47 with
params[7] = 3 also appears at session start, where nothing is being enabled,
so the byte is more likely a selector than a value. Note it must be
DLE-escaped: a bare 0x03 in params truncates the frame at what the device
reads as ETX.
🔑 Setups ARE written as raw .MMB files — second retraction
This document twice stated that setups are not pushed as raw .MMB blobs and
go only through 0xDA + the config block. Wrong. In this session Thor did
both, choosing by whether the setup is the active one:
| setup | how it was written |
|---|---|
TEST1.mmb (active) |
0xDA → 0x68/0x73 → 0x82/0x83 → 0x71/0x72 |
test2.mmb (not active) |
0x8D \system\setups\test2.mmb → 0x8E (2,192 B) |
So \system\setups\<name>.mmb is the setup directory, and the generic file
transfer writes there directly.
The .MMB file is nearly the compliance block: aligning the two gives
1,968 / 2,086 bytes equal (94.3%) at a 4-byte shift. The file is 102 bytes
longer (2,192 vs 2,090) and puts the setup name at offset 38 where the 0x71
block has it at 42. So the on-disk setup and the wire compliance block are the
same structure with different framing — not two formats.
This is the cleaner path for SFM. Writing a setup as a file needs two frames
and no 0xDA/0x68/0x82 ritual, and it does not disturb the active setup.
File writes are chunked
The schedule's 1,304 bytes went as two 0x8E frames, 1,024 + 280, each with
offset = that chunk's own length. The 2,192-byte setup file went in one
frame, so 1,024 is not a hard ceiling — the chunking rule is unexplained and
recorded as observed.
Six duty-cycle actions, not five
The full _PSA() dispatch — the scheduler's own action processor, reading records
via _ReadRecord:
DUTYCYCLE_START_MONITOR ← no setup
DUTYCYCLE_START_MONITOR_WITH_SETUP
DUTYCYCLE_START_MONITOR_WITH_SETUP_STOP_COMPLETE ← a third start variant
DUTYCYCLE_STOP_MONITOR
DUTYCYCLE_CALLHOME
DUTYCYCLE_SELF_CHECK
THOR exposes four. Two start variants it never offers, one of which needs no setup file at all.
_PSA() send ->> CMD_DUTYCYCLE_START_MONITOR shows the setup-less action is live
code that the scheduler genuinely dispatches, not a dead case — it sends a real
message. And since every one of these cases sits in the function that consumes
_ReadRecord's Action field, a schedule record can carry it.
An alternative explanation was checked and ruled out: the Micromate does have a
separate Timer Mode (MODE_TIMER, MODE_MONITOR_TIMER, Monitor Once Only,
under Special Setup), so START_MONITOR could have belonged to that path
instead. It does not — it is in _PSA(), the scheduler's dispatcher.
The scheduler enable lives in SysPref, not in the setup
SysPref.bMonitorScheduler = %s
_Task() CMD_SET_SCHEDULE - MonitorScheduler ENABLED
_Task() CMD_SET_SCHEDULE - ! MonitorScheduler NOT Enabled
This confirms the failed prediction above from the other side: the
Scheduler On/Off switch is a system preference, which is why the 0x71
compliance block was byte-identical when the scheduler was turned on. It was
never going to be in there.
It also suggests 0x47 — the two bare frames at the end of the capture — is a
SysPref get/set rather than anything scheduler-specific, which would explain its
params[7] selector and its 15-byte response shape being shared with 0x48's
page-0 descriptor. ⚠ Still a hypothesis; the disable/enable capture settles it.
The schedule has a setup-less start action — Thor just never uses it
The coupling above looked like it might be a workaround for the unit crashing on a missing setup. The firmware says otherwise: there are two distinct start-monitoring actions, and only one of them involves a setup file.
_PSA() case DUTYCYCLE_START_MONITOR
_PSA() case DUTYCYCLE_START_MONITOR_WITH_SETUP
The full duty-cycle action set, from the scheduler task's own dispatch strings:
| action | notes |
|---|---|
DUTYCYCLE_START_MONITOR |
no setup involved — uses whatever config is loaded |
DUTYCYCLE_START_MONITOR_WITH_SETUP |
the variant Thor always emits |
DUTYCYCLE_STOP_MONITOR |
|
DUTYCYCLE_CALLHOME |
send ==> CMD_SCHEDULE_CALL_HOME |
DUTYCYCLE_SELF_CHECK |
call DailySelfCheck() |
DUTYCYCLE_ON / OFF / NEXT |
scheduler state, not per-entry actions |
So the schedule↔config coupling is not a crash workaround — Thor picks the more
demanding of two available actions, every time. SFM can emit
START_MONITOR and skip the config entirely.
On whether a missing setup would break the unit: the firmware suggests
graceful degradation rather than a crash — Setup File Not Found exists, and so
does Invalid parameters reset to factory default - please review setup, which
is a deliberate fallback. ⚠ Untested, and not worth relying on until it is.
⚠ Hypothesis, not a finding: the schedule entry's leading byte may be the
action code — the one captured entry reads 03 00 00 00 0f 03 02 [namelen] [name], and 03 would fit START_MONITOR_WITH_SETUP in an enum of the above.
Unverified; there is one entry and nothing to diff against.
The capture that would settle it (more valuable than the 0x47
disable/enable test): build a schedule whose entry is stop monitoring or
call home — an action with no setup attached. That would (a) confirm or kill
the action-code hypothesis, since a setup-less action should change the leading
byte and drop the name, and (b) prove from the other direction that a schedule
can be sent with no config push at all.
Note also DUTYCYCLE_CALLHOME → CMD_SCHEDULE_CALL_HOME: a scheduled
call-home is a way to make a unit dial out on demand, which is the one
remaining lever on the unsolved call-home direction.
What this implies for the build order
The read path is fully known and needs no Thor-shaped decisions, so a read-only
client can be written now with confidence. The write path should not be built
by transcribing Thor's sequence — the 0x68/0x82 question above decides
whether our setup write is 3 frames or 7, and it is answerable with one capture
plus one careful experiment.
Two reliability problems worth designing against, both observed rather than assumed:
- A zero ack does not mean a write applied. Every ack seen is 11 zero
bytes, across creates and overwrites alike, and no failing write has ever been
observed. Whatever SFM does, it should read back and compare rather than
trust the ack.
0x41+0x1Amakes that cheap. - Nothing warns before clobbering a monitoring unit's active setup. The device will not stop it and the ack will not distinguish it. That guard has to live in SFM.
✅ The first commands this project originated against a Micromate: start and stop on UM20147 (2026-10-25)
These were the first frames this project ever sent to a Micromate that
change its state. Every Micromate write before this one was THOR's, recorded
while we watched. (Series III is different: the minimateplus stack has sent
writes — start/stop, config push, erase, call-home config — since April 2026.)
Brian gave a go at the bench for each send.
Setup.
- Unit: UM20147, firmware
11.0BD(Thor line, reply flags0x03), on USB CDC-ACM/dev/ttyACM0. - Sensors: none attached.
- Active setup:
BENCH01. - Sender: SFM's
POST /device/monitor/start|stop?serial=UM20147(branchfeat/series4-sfm-parityatb1c87ec), with every TX frame and every validated RX frame logged at millisecond resolution. - Path: each send went through the phase-3 safety core in this order:
- a fresh serial read;
- a
0x49state read; - the audit row written;
- one
execute(); - read back every 2 s.
What was sent and what came back (measured)
| time (EDT) | sent | ack | ack latency | 0x49 content[0] afterwards |
|---|---|---|---|---|
| 15:43:47.430 | 0x96 41021010009600ffff00000000000000000000a403 |
0x69, 11 zero bytes |
464 ms | 0x02 at the first read, 2.0 s after the ack |
| 15:45:51.922 | 0x97 41021010009700ffff00000000000000000000a503 |
0x68, 11 zero bytes |
116 ms | 0x01 on all 9 reads from +2.1 s to +18.2 s, then 0x00 from +20.3 s |
| 15:52:42.645 | 0x97 (unit already idle; even_if_already) |
0x68, 11 zero bytes |
10 ms | 0x00, unchanged |
| 15:53:51.871 | 0x97 (idle again, screen watched) |
0x68, 11 zero bytes |
10 ms | 0x00, unchanged |
- The frames. Both are THOR's own bytes, offset
0xFFFFincluded. The phase-2 bench question "is offset0x0000honoured" was not tested and does not need to be. 0x1Cdata[12]agrees. It read0x00idle before the start,0x0Eat the start's read-back and at the stop's preamble, and0x00again at +20 s.0x0Eis the same monitoring value UM12947 showed on 2026-09-24.0x1Cwas not read during the0x01window.- The start's screen showed a sensor check, then monitoring. With no sensors attached, the check passed without complaint (Brian, watching).
- The start's state.
0x49already read0x02at +2.5 s, so the state byte very probably reads monitoring during the sensor check. That is inferred, not measured: the check's length on screen was not timed. Consequence: a second start sent during the check would see "monitoring" and be skipped, not sent. - One histogram was recorded and closed. The event is timestamped
15:44:01, 14 s after the start, which is roughly the length of the sensor
check. The event count went 5 → 6 at the stop (
UM20147_20261025154401.IDFH, 7,976 B).- Its peaks are 0.00186 in/s on all three channels, identical, because no sensors were attached.
- Free memory fell 4,096 B at the start and another 4,096 B by the stop. The same 4 KiB step was seen in the 2026-09-24 session.
- The stop's screen. The unit tried to call home right after the stop. Brian
says it always does this when taken out of monitor mode. It never connected,
because it has no modem, and it ran out of attempts. A watch of
0x49once a second from 15:47:19 to 15:49:50 read0x00throughout. - A stop sent to an idle unit is harmless. It was acked in 10 ms, twice. Nothing changed: not the state, the event count, free memory or the screen, and there was no call-home attempt. The ~10x shorter ack suggests the firmware skips its close-down work. This settles the spec's open question "what does a stop-while-stopped do".
🔶 0x49 content[0] = 0x01: a fourth state value (value measured, meaning inferred)
The values are now 0x00 idle, 0x02 monitoring and 0x03 in a call-home
session (10 of 10 reads, 2026-10-14). The new one is 0x01, for about 18 s
right after a stop.
Inferred meaning: a call-home attempt in progress, dialling and not
connected. It is distinct from 0x03, a call-home session that has
connected. The timing fits: the 0x01 window opened 2 s after the stop, on
screen the unit was trying to call home, and it closed when the attempts ran
out.
Not excluded: 0x01 could instead mean "stopping, closing the record", with
the call-home attempts overlapping it. Two tests would separate them:
- a stop with call-home-after-event switched off;
- a stop on a unit whose modem connects, where the inference predicts
0x01→0x03. This fits into the ring-path Run A/B.
What SFM does with it. read_run_state() names it unknown(0x01), so the
change path refuses any start or stop in that window ("never been measured --
nothing sent", HTTP 409). That is deliberate: name it only once its meaning is
measured. The stop's read-back simply kept polling through it, because 0x01
is not idle, and the stop verified at +20 s, well inside the 60 s window.
Audit rows (SFM device_commands, scratch bench DB)
All four rows read outcome=sent, verified=ok, sent=1, caller
bench-brian. Their states columns:
- start:
idle → [idle, monitoring] → monitoring - stop:
monitoring → [monitoring, 9 × unknown(0x01), idle] → idle - the two forced stops:
idle → [idle, idle] → idle, each carrying the note that the read-back cannot show the send changed anything.
⚠ Untested and unsafe-until-agreed
✅ Retracted in part (2026-10-25): the opening line below was true until 2026-10-25. Start and stop (
0x96/0x97) have now been originated by SFM against UM20147. See The first commands this project originated against a Micromate. Every other item in this section remains never-sent.
Nothing below has been sent to a unit by us, and nothing should be without an explicit decision.
Note the distinction introduced on 2026-09-24: the setup-write sequence
(0xDA, 0x68/0x73, 0x82/0x83, 0x71/0x72) has now been observed,
because Thor performed it while we recorded. Observed is not the same as
exercised — we have still never originated a write frame. The wire format
is known; our encoder is unwritten and unproven.
- Writes (
0x68–0x83) — format known, never sent by us - Call-home write (
0x7E/0x7F) — ✅ observed 2026-09-25, acks0x81/0x80; enable flag isolated - Per-event delete (
0xA8/0xAA) — ✅ observed 2026-09-25, acks0x57/0x55. Destructive. Series III's0xA3/0xA2erase-all has not been seen on Series IV and may not exist - Start / stop monitoring (
0x96/0x97) — observed via Thor 2026-09-24, acks0x69/0x68. ✅ Originated by SFM 2026-10-25 on UM20147 (4 sends, all acked and verified) — no longer untested - Scheduler enable (
0x47) — observed;params[7]semantics still unclear
Every one of these is now observed. None has ever been originated by us, and that is the line that still matters: the wire format is known, our encoder is unwritten and unproven. (2026-10-25: true still of every item here except start/stop, which SFM originated at the bench — see the retraction note at the top of this section.)
0x1F(advance event pointer) — non-destructive on Series III but it does move device state, so it is parked with the rest
Also unknown:
-
Whether
0x10bytes inside request params need stuffing. (Write-frame data stuffing is now settled —10 XX→XX; see The write path.) -
The call-home session's bytes — the device-initiated direction has still not been observed at all, and cannot be over USB.
✅ The second half of this item is resolved (2026-09-24, from the THOR manual): "how it learns an event was accepted so it stops re-sending it" was the wrong question — it does not learn. Collection state is server-side bookkeeping and erase is a separate, optional, server-issued action. See What the THOR manual settles about the ACH session. What remains unknown is the framing of the inbound session and the erase opcodes, not any acknowledgement mechanism.
Session provenance
Unit UM12947, firmware 11.0CB (Blastware line), on the bench via
USB. Every response in this document was checksum-validated.
Three sittings, all on the same unit:
| date | state | what was exercised |
|---|---|---|
| 2026-09-23 | zero events stored | read commands, direct from Python |
| 2026-09-23 | 5 events (4 waveform, 1 histogram) | the event chain + 0x5A |
| 2026-09-24 | 5 events | Thor pushing a setup, via a recording relay |
The 2026-09-24/25 sittings used a topology worth keeping, because it is
reusable: socat on mint-mac shares /dev/ttyACM0 on TCP, seismo_lab's TCP
bridge relays Thor to it and records both directions. Thor is configured with
the unit at 127.0.0.1:<bridge port> exactly as if it were a field modem. No
modem, no SIM, and nothing on the production Thor box is touched.
Run the relay with -x and keep its log. It hex-dumps every forwarded byte
in both directions, which makes it a complete second copy of every capture,
independent of whether seismo_lab was recording:
socat -d -d -x TCP-LISTEN:12345,reuseaddr,fork /dev/ttyACM0,raw,echo=0,b115200 \
> ~/mm-captures/socat_<ts>.log 2>&1
That paid off on 2026-09-25, when a capture's .bin files never left the Windows
machine — the session was rebuilt from the relay log with
scratch/socat_log_split.py. When the real bins arrived later the
reconstruction was byte-for-byte identical in both directions (3,595 and
4,004 bytes), so the log is a validated fallback rather than an approximation.
Sessions concatenate in one log; split them on the accepting connection
markers.
⚠ Firmware is the single biggest caveat on this document. Every finding here
is from one unit on the Blastware build. A 11.0BD unit has not been
touched.
✅ The original "empty unit" limitation is gone — 0x08, 0x1E, 0x0A and
0x06 were all exercised against 5 stored events on 2026-09-23.