serversdownandClaude Opus 5.5 5644dbd5ad fix(decode): one MicL count is 0.25 Pa -- the mic psi was ~95x low
SFM converted mic counts to psi with a '~0.5 psi full-scale' guess
(0.0125 / 32768 psi per count).  Measured against 1,729,253 DL2
histogram intervals' 'MicL Peak' columns: one count is exactly 0.25 Pa.
count x 0.25 Pa matches Blastware's psi(L) on every interval, and
20 log10(count x 0.25 Pa / 20 uPa) its dB(L) on all but 17 exact
half-way cases.

- event_file_io: _MIC_FS_PSI -> _MIC_PSI_PER_COUNT = 0.25 / 6894.757
  (3.626e-5 psi per count, was 3.81e-7).  The event's mic peak in psi,
  and with it the .h5 mic trace (event_hdf5 back-solves its scale from
  that peak), come right on events with no report.
- waveform_codec.mic_count_to_db: the offset is 20 log10(0.25 / 20e-6)
  = 81.938, not 81.94 -- the old value missed 49 of those intervals at a
  rounding edge (count 119 printed 123.4, 81.94 gave 123.451).

The decoded count was always right: count 16 on K441LTON.900 is
Blastware's 106.0 dB.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
2026-10-31 01:28:23 -04:00

150 lines
6.4 KiB
Python

"""
The decode checker's SFM side and comparison (decode checker, Tasks 2-4).
Expected values come from the fixtures' Blastware XML reports. SFM's side is
built from the event bytes alone -- no filename, no report.
"""
from __future__ import annotations
from pathlib import Path
import pytest
from minimateplus.decode_check import sfm_side
FIX = Path(__file__).parent / "fixtures/decode-checker-10-29-26"
def test_waveform_side_from_bytes():
s = sfm_side((FIX / "R352HIUI.F40W").read_bytes())
assert s.values["SampleRate"] == 1024 and s.values["RecordTime"] == pytest.approx(1.0)
assert s.values["Pre-triggerLength"] == pytest.approx(-253 / 1024)
assert (s.values["EventDate"], s.values["EventTime"]) == ("2018-07-30", "16:13:52")
assert s.notes == ["JV 241 - 1425 Spanlers Mill Rad", "Walsh Granite JV",
"Terra-Mechanics Inc- D. Harrison", "Loc #1 - 1425 Spanlers Mill Rd"]
assert len(s.counts["Tran"]) == 1277 and s.job_number is None
def test_histogram_interval_peak_in_counts():
s = sfm_side((FIX / "M529L4XQ.R30H").read_bytes())
# Blastware: Tran interval peak 2.430 in/s = 7776 counts at 10/32000 in/s per count
assert s.counts["Tran"] == [7776] and s.values["NumberofIntervals"] == 1
def test_event_type_comes_from_the_bytes():
# No filename reaches the decoder; the type is the STRT record-type byte (#41).
s = sfm_side((FIX / "M529L4XQ.R30H").read_bytes())
assert s.values["EventType"] == "Full Histogram"
def test_report_fields_stay_blank():
s = sfm_side((FIX / "T1905KDW.5F0W").read_bytes())
for tag in ("TranZCFreq", "TranPeakAcceleration", "BatteryLevel", "Calibration", "TranTestResults"):
assert s.values.get(tag) is None
# ── Task 3: the comparison ──────────────────────────────────────────────────
from minimateplus.bw_xml_report import parse_xml_report_file # noqa: E402
from minimateplus.decode_check import compare, matches_printed # noqa: E402
def _results(stem, carveouts=()):
raw = (FIX / stem).read_bytes()
ref = parse_xml_report_file(FIX / (stem.replace(".", "_") + "_XML.XML"))
return {r.field: r for r in compare(sfm_side(raw), ref, carveouts)}
def test_matches_printed_rounds_half_away():
assert matches_printed(-0.0625, "-0.063") and matches_printed(0.0015, "0.002")
assert not matches_printed(0.0026, "0.002")
def test_full_waveform_row_verdicts():
res = _results("R352HIUI.F40W")
for f in ("SampleRate", "RecordTime", "Pre-triggerLength", "EventDate", "EventTime",
"notes", "samples.Tran", "samples.Vert", "samples.Long"):
assert res[f].verdict == "match", (f, res[f])
assert res["TranZCFreq"].verdict in ("sfm_blank", "no_reference")
assert res["BatteryLevel"].verdict == "sfm_blank"
assert "FileName" not in res
def test_ragged_rows_match_by_running_channels():
# Tran/Vert/MicL end 3 samples before Long, in SFM and in Blastware's table alike
res = _results("T438HGTZ.B30W")
for ch in ("Tran", "Vert", "Long"):
assert res[f"samples.{ch}"].verdict == "match", res[f"samples.{ch}"]
def test_an_extra_reference_row_is_a_mismatch_with_detail():
raw = (FIX / "T1905KDW.5F0W").read_bytes()
ref = parse_xml_report_file(FIX / "T1905KDW_5F0W_XML.XML")
ref.sample_rows.append(["0.000", "0.000", "0.000", "0.000"])
res = {r.field: r for r in compare(sfm_side(raw), ref)}
assert res["samples.Tran"].verdict == "mismatch"
assert "row_length_disagreements=1" in res["samples.Tran"].detail
def test_job_number_na_and_above_range():
res = _results("T1905KDW.5F0W")
assert res["JobNumber"].verdict == "sfm_blank"
assert res["TranZCFreq"].verdict == "no_reference" # N/A
assert res["VertZCFreq"].verdict == "sfm_blank" # >100, SFM has no flag yet
def test_mic_units_pick_the_conversion():
res = _results("T004JTRF.ME0H") # psi(L)
assert res["intervals.MicL Peak dB(L)"].verdict in ("match", "mismatch")
assert res["intervals.Tran Freq Hz"].verdict == "sfm_blank"
fields = set(_results("R352HFZ7.530H")) # no mic
assert not any(f.startswith(("samples.MicL", "intervals.MicL")) for f in fields)
# ── Task 4: carve-outs ──────────────────────────────────────────────────────
from minimateplus.decode_check_carveouts import CARVEOUTS # noqa: E402
def test_clipped_normal_range_is_accepted():
# Blastware: Tran 5.445, Vert 6.310, Long OORANGE, PVS OORANGE; SFM Long 10.005, PVS 11.55
res = _results("T438HGTZ.B30W", CARVEOUTS)
assert res["TranPPV"].verdict == "match" and res["VertPPV"].verdict == "match"
for f in ("LongPPV", "PeakVectorSum"):
assert (res[f].verdict, res[f].carveout) == ("accepted", "saturation")
def test_sensitive_range_8x_stays_a_mismatch():
res = _results("M007HVWF.K30W", CARVEOUTS)
assert res["TranPPV"].verdict == "mismatch" and res["samples.Tran"].verdict == "mismatch"
assert res["PeakVectorSum"].verdict == "mismatch"
def test_every_carveout_has_evidence():
assert all(c.id and c.fields and len(c.evidence) > 40 for c in CARVEOUTS)
def test_pvs_carveout_needs_oorange_from_blastware():
raw = (FIX / "T438HGTZ.B30W").read_bytes()
ref = parse_xml_report_file(FIX / "T438HGTZ_B30W_XML.XML")
ref.scalars["PeakVectorSum"] = "10.392" # a vector sum may exceed the range unclipped
res = {r.field: r for r in compare(sfm_side(raw), ref, CARVEOUTS)}
assert res["PeakVectorSum"].verdict == "mismatch"
def test_pvs_carveout_needs_the_clipped_channel_at_the_ceiling():
# 1.250 range, Blastware PVS OORANGE; an UNclipped channel's 8x value happens to land
# in the 1.250 band -- that must not let SFM's 8x vector sum through.
raw = (FIX / "M007HVWF.K30W").read_bytes()
ref = parse_xml_report_file(FIX / "M007HVWF_K30W_XML.XML")
ref.scalars["TranPPV"] = "0.157"
side = sfm_side(raw)
side.values["TranPPV"] = 1.256
res = {r.field: r for r in compare(side, ref, CARVEOUTS)}
assert res["PeakVectorSum"].verdict == "mismatch"
def test_interval_mic_psi_matches():
# psi(L) report: every interval's "MicL Peak psi(L)" now converts right
assert _results("T004JTRF.ME0H")["intervals.MicL Peak psi(L)"].verdict == "match"