Brian: use the actual unit limits, not what has been used in the past. Trigger 0.005-10 in/s in 0.001 steps (capped by the range), record time 1-90 s (Micromate) or the MiniMate manual's 1 MB table by sample rate (13 s in Histogram Combo), job number 1-9999. The read-back reports any value a unit does not take. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
21 KiB
Phase 5: changing a unit's settings by serial (design)
Date: 2026-11-03. Status: draft for Brian. Phase 5 of
docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md (row 10, "change
recording settings"; build phase 5, "setup push: one field first, then several").
Branch: feat/series4-sfm-parity.
Intent (Brian, 2026-11-03)
Change any of a unit's settings from SFM, and make it as easy and reliable as possible. The goal is to fix what makes Blastware finicky. Three real cases:
- One wrong number on a deployed unit. UM6047 was out monitoring with its geophone trigger at 0.03 instead of 0.300. Brian remoted into the THOR PC, stopped it (it called in), changed the setup, made sure it was sent, then restarted monitoring.
- A typo in the notes.
- A backup unit. The unit set up at the office fails in the field, so a backup from the truck goes in with blank or stale notes and settings. There is no computer in the field, so the office re-does the whole setup remotely, by hand, while the person in the field stands by to press Start or power-cycle if the link breaks.
What is wrong with THOR, as Brian and his dad experience it:
- The same screen edits a setup on disk and views a unit's setup.
- Setups get saved under random number strings.
- Editing a setup does not change the unit; you verify separately that it was sent.
Saved setups as templates would be acceptable. Not knowing whether you are changing the unit or a file is not.
Goals, non-goals, anti-goals
Goals
- Edit the unit's live settings, both series.
- Show exactly what will change, old → new, before anything is sent.
- After sending, re-read the unit and show what it now holds.
- Make every refusal say why, in plain words.
Non-goals (this phase)
- Templates, saved setups, copying from another unit.
- Restarting monitoring. Brian presses Start, as today; SFM never restarts on its own.
- Setting the unit's clock. That is a later call-home receiver rule.
- Writing the Micromate scheduler. Next project, agreed 2026-11-03: "record during work hours" as a plain, configurable setting on the unit's page, both series, no schedule library and no setup files. On the Micromate, it relies on the firmware's setup-less scheduler start (found in the firmware, never sent; one bench session). On the MiniMate, its own timer setting (undecoded). Brian's dad despises THOR's scheduler; Brian is willing to retire it in SFM.
- Fields not yet decoded. "Any and all" grows as decoding grows.
Anti-goals: things that would make it a failure even if it "works"
- Any doubt about whether the unit changed.
- A screen where a file and a unit look alike, or a name SFM made up.
- Writing a byte SFM did not just read. Every byte SFM does not deliberately change
goes back exactly as read. The one exception is a byte that Blastware's (or
THOR's) own writes always set to a fixed value. Each such byte is named in the
codec with its evidence (so far: Series 3
[26:28]→00 00; my call, below). - Sending to a monitoring unit.
- A retry. A write is sent once; what happened is read back.
Measured basis
1. Series 4 (Micromate): the 0x1A setup content map. Measured from 47
keypad snapshots on UM20147 and checksum-verified. micromate/setup_block.py
(decode_setup) reads it, with 69 tests on 15 fixtures.
| content | field |
|---|---|
[5] |
record mode |
[6:8] |
sample rate |
[8:10] |
histogram interval |
[12:16] |
pre-trigger, f32, negative |
[17] |
stop mode |
[18:22] |
record time, f32 |
[23] |
auto-stop |
[25] |
active sensors |
[29] |
trigger sources |
[30:34] |
distance |
[34:38] |
charge |
[39] |
report flags |
[40:42] |
job number |
[42:86] |
setup name |
[122+64n] / [144+64n] |
notes (label / value) |
from [1708] |
48-byte channel blocks: sensor code +4:8, trigger +30 (geo in/s, mic psi) |
The CRC-16/AUG-CCITT over [0:2088] is stored at [2090:2092]; THOR's write
omits it and the unit recomputes it.
- The push sequence (from THOR's captures) is
0xDA(name) →0x68/0x73(0x08preferences) →0x82/0x83(0x2Etrigger) →0x71/0x72(0x1Asetup).micromate.writes.plan_setup_pushreproduces THOR's frames (phase 2). - Not yet mapped on Series 4: alarm level, geophone range, GPS, aux I/O, SLM options.
2. Series 3 (MiniMate) uses the same layout. Measured 2026-11-03. The
compliance image as Blastware writes it is logical (unescaped), 2,092 bytes, with
N = 2090 at [0:2] and the channel block "Tran" at [1708].
micromate.setup_block.decode_setup reads it correctly:
-
BE9441's 10-26 write (
tests/fixtures/minimate-bench-10-26-26-histogram, two whole sends): name, mode, 4096 sps, interval, record time 20 s, trigger 0.04 in/s, mic linear, job number and the notes all decode. Only the CRC fails (Series 3's tail goes out stale). -
BE11529's nine 2026-04-20 Blastware writes (
bridges/captures/4-20-26/, every one with a whole image) each changed exactly the expected bytes:Blastware changed bytes that moved histogram interval (×4) [8:10]recording mode (×3) [5]geophone range 1737/1785/1833, the range byte at each geophone channel label+29 -
The live
SUB 1Aread starts the same block. Measured on the first data frame of each of the four recorded reads (_settings_repliesintests/test_s3_compliance_anchor_fields.py): unescaped, each carriesN = 209012 bytes before the old "anchor", and content[5]reads 1 / 4 / 3 / 3. That anchor (be 80 00 00 00 00) is content[12:18], pre-trigger −0.25 plus stop mode Fixed. So Series 3 can be read by content offset rather than by anchor search, which survives an Auto-stop unit (the open question of 2026-10-17). -
A whole
1Aread assembles to exactly the 2,090 content bytes (measured 2026-11-03 on BE9441's 10-26 recording, 12 reads). Unescape each data frame and takedata[11:]. Every frame but the last then starts with one00byte that is not content; drop it. This holds for Blastware's read (four 512-byte frames plus 42) and SFM's (two 1024-byte frames plus 42).- The round trip is exact. Blastware re-read the block after each of its two writes, and each re-read equals the image it had just written, all 2,090 bytes.
- While monitoring, the read returns
0e 'BE9441' …(the device-info block), as the 10-29 finding says. The unit was monitoring for 5 of the 12 reads.
-
Bytes
[26:28]: Blastware always writes00 00(all 11 writes), though the unit reads01 7fafter a monitoring session (BE9441 10-26, BE11529 4-20 15:19). The unit sets them again by itself. Meaning unknown. -
The two bytes after the block are Blastware's own. Blastware sends a per-unit constant:
7c eaon all nine of BE11529's writes,0c e9on both of BE9441's. It is not the block's CRC-16/AUG, and no reply in BE9441's three-hour session contains it. SFM'swrite_imagesends00 00there. Since a read never returns those bytes, the read-back cannot show what the unit did with them. -
The mode byte is shared, but its codes are not. Series 3:
00Single Shot,01Continuous,03Histogram,04Histogram + Continuous. Series 4:01Waveform,02Manual,03Histogram,04Combo. -
Rules from the checker's file-side work (seismo-relay-b6, 2026-11-03; it measured the same content offsets on 70,042 event files and found nothing that contradicts one shared map):
- A bitmask byte is patched bit by bit, never rebuilt. The trigger-sources
byte
[29]showed an unidentified0x80bit on one unit. - On Series 3, stop mode (
0Fixed) and mic type (2linear) are proven for one value each. They are not offered for writing. - An event file's
[12:16]is that event's captured pre-trigger, not the setting. SFM writes only from a live read, so this matters only to anyone tempted to build a write from a file.
- A bitmask byte is patched bit by bit, never rebuilt. The trigger-sources
byte
-
Series-3-only fields (measured on Series 3, not Series 4): the geophone range byte at channel
+29; the alarm level at channel+38(checker-measured on event files; live raw+42).
3. Series 3 write frames are measured (#46). Writes use uniform escaping with
a length field. The compliance image goes in 512-byte addressed chunks. 0x68 is
written as read plus 02 00, 0x82 plus 00 00. The sequence is
68→73 | 71×n→72 | 82→83 | 69→74→72. push_config_raw refuses any block that
is not whole. tests/test_minimate_write_frames.py rebuilds Blastware's frames
byte for byte.
4. Monitoring.
- Series 3: while monitoring, the unit answers
1A(and0x2C) with its device-info block (measured 10-29 and 11-03). - Series 4: it refuses call-home setup while monitoring ("Feature disabled while monitoring", measured 11-03).
- A stop with call-home on makes the unit call in (measured 10-25, 10-29, 11-03).
5. Series 4 specifics.
0x1Areads the keypad's edit buffer live: unsaved keypad edits show in the read, and a push would commit them.- Loading a
.MMBreplaces the active config byte for byte. The push writes under the active setup's name (0xDA).
The flow
- Read.
GET /device/settings?serial=…(it exists) returns the unit's settings. The page shows them on the Config tab under "On the unit BE9441 — read 14:03", never anything that looks like a file. - Edit and preview. The operator changes fields, then Preview: a dry run. SFM re-reads the unit's blocks on the same line, builds the new blocks, and returns one row per field: old → new, plus any warnings. Nothing is sent.
- Send. The same request, live, is an audited change:
- identity check, audit row, one send, never retried, read back;
- refused with "stop monitoring first" if the unit is monitoring, noting that the stop makes it call in when call-home is on;
- refused if the unit's settings changed since the preview. The preview carries a revision (a hash of the blocks read), and the live call must match it.
- Verify. Re-read the blocks. Every changed field must read its new value.
Every other decoded field, and every byte SFM did not deliberately change, must
read as before. The page shows "the unit now holds": field, old, new, now.
Known unit-maintained bytes (the Series-4 CRC) are compared as the unit's own
checksum (
crc_ok), not byte for byte. - Start stays the operator's button.
Fields in the first version (Brian approved the list)
| field | Series 3 | Series 4 |
|---|---|---|
| geophone trigger level | yes (channel +30) |
yes (channel +30) |
| alarm level | yes (channel +38; measure live) |
not mapped, so not offered |
| notes: the four note slots, by the unit's own labels | yes (Project:, Client:, User Name:, Seis. Loc: on BE9441) |
yes (Location, Client, Company, General Notes) |
| recording mode | yes ([5]; values as measured on Series 3) |
yes ([5]; Series-4 codes) |
| sample rate | yes | yes |
| record time | yes | yes |
| histogram interval | yes | yes |
| geophone range | yes (channel +29) |
not mapped, so not offered |
| job number (+ shown on report) | yes | yes |
A field is offered only where it is measured for that series. Values are refused unless they are among the measured choices: sample rates 1024/2048/4096, intervals 2/5/15/60/300/900 s, and each series' measured mode codes.
- Numbers: the units' own limits (Brian, 2026-11-03: "I want to use the actual
unit limits, not just what we have done in the past"). From Instantel's
operator manuals in
~/seismo-relay/manuals/:- Geophone trigger: 0.005–10 in/s in steps of 0.001 (MiniMate Plus manual 716U0101 Rev 15, Appendix Series III Specifications; Micromate manual 721U0201 Rev 06, Geophone Trigger Level). It is also capped at the range's full scale (10 / 1.25 in/s). When range and trigger change in one request, check the trigger against the new range.
- Record time:
- Micromate: 1–90 s (Record Time).
- MiniMate: from 1 s to the manual's §4.4 Maximum Record Times for a 1 MB unit, four channels, at the chosen sample rate: 104 s at 1024 sps, 52 s at 2048, 26 s at 4096. In Histogram Combo, 13 s (§3.8.9). The units measured report about 960 KB in total memory (BE11529: 983,026 bytes), the 1 MB model.
- Job number: 1–9999 (both manuals).
- The read-back is the safety net. A unit that does not take a value (for instance, cuts a record time to what its memory allows) shows as not taken, with what it actually holds.
- Values seen across 75,958 Blastware reports (2016–2026) sit well inside these limits: trigger 0.005–0.500 in/s, record time 1–10 s, job number 1.
- Units: refuse a level write unless the channel's unit string reads the
measured one (
in.for a geophone). A unit set to Metric is unmeasured.
- Notes: edited by slot, under the label the unit itself shows (labels can be
renamed and are not identifiers). The value is at most its field width less
one byte, printable ASCII.
- Not in v1: Series 3's fifth slot,
Extended Notes(value\non BE9441). Blastware's extended notes are unmeasured and may live elsewhere. - Not in v1: Series 4's 16-byte field at content
[1064](Loc 1on UM12947), undecoded.
- Not in v1: Series 3's fifth slot,
Architecture
- One setup-block codec for both series. Keep
micromate/setup_block.pyas the field map and addencode_setup(content, changes) -> content: field patches only, every other byte as given. It stays a pure codec library: no I/O, nosfmimports, no import cycle withminimateplus. Series-specific fields (the Series-3 range and alarm) are marked by series. - Series 3 read. Assemble the
1Aread into the logical 2,090-byte block: unescape, strip the 12-byte reply prefix and the inter-frame headers (read_compliance_config). Decode it by content offset. - Series 3 write. The patched logical block plus the tail becomes the
compliance image for
push_config_raw. Event index (0x08), trigger (0x2E) and waveform data (0x09) are echoed as read, in Blastware's sequence.- On
0x09, Blastware echoed the unit's bytes on all nine April writes ([200:202]=8c 76read and written). Only BE9441's 10-26 writes changed them (17 17read,78 76written, and read back so). The meaning is unknown, so SFM echoes. - The raw-offset
_encode_compliance_configis retired.
- On
- The legacy write (
/device/config,/device/config/project) is retired (my call). It retries a whole write on a TCP error, is not audited, does no read-back, does not refuse a monitoring unit, and needs no serial: every anti-goal at once. Its only user is the webapp's by-host Config form, and Terra-View does not call it. Its routes answer with a pointer to the by-serial change, and the by-host form becomes read-only. - Series 4 write.
plan_setup_push(prefs, trigger, patched_setup, name).- No setup files in the operator's view (Brian, 2026-11-03). "We treat the Micromate setups just like Minimate setups … the unit just has its settings, and we change it based on the job. We don't send setups to the unit." A backup unit is usually on whatever setup its previous job left. So SFM rewrites the setup the unit is running, in place, under its own name. It never asks for a file or a name, and never creates one. The name may show as a read-only detail, since the unit's screen shows it.
- The scheduler can undo a change (Brian, 2026-11-03, with THOR's Manage
Schedules screen). Work-hours Micromates run a THOR schedule, e.g. Start
Monitoring 07:00 with setup
Mele-PWSA-Carroll -Loc 4, stop 18:30, repeat daily. THOR sends the schedule with its setup, and each scheduled start loads the setup it names. So the preview and the send readschedule.dat(format decoded, reference The schedule record format), read-only. A start action naming a setup other than the one being changed blocks the send: "this unit's schedule loads at 07:00, which would replace this change." When the schedule names the active setup (the usual case after a scheduled start), the change survives, and the preview says when the next scheduled start is. SFM never writes the schedule in this phase. - The name comes from the
0x41read, unstripped. - Refuse
factory.MMBand any target that is not the active setup (spec). The refusal says it plainly: "this unit is running its factory setup; load any other setup on its keypad, then try again." Brian says the team does not use the factory setup, so this should be rare. ENABLEABLE_WRITESgains0xDA 0x68 0x73 0x82 0x83 0x71 0x72.- The keypad edit buffer: warn in the preview that a unit with someone in its
menu would have those edits committed. The check the spec names,
0x1Aagainst the active.MMBvia0x94/0x48, is a bench question first.
- The change path. It keeps run_change's guarantees: identity, audit, one
send, read back,
ChangeUnconfirmedwithsent/audit_id. The "state" is a set of fields, so eitherrun_changetakes a verify function or a sibling core shares its parts. The builder decides; the guarantees do not change. The audit row keeps the per-field changes, both blocks (hex) and the verdict. - Routes.
POST /device/settings/change?serial=…(serial required) with body{changes: {field: value}, revision}anddry_run,caller. The response is per-fieldold/new/now, through_change_response. A dry run needs no revision and returns one. A live call without the revision of a preview is refused (my call): nothing reaches a unit unless someone saw old → new. - Page. The Config tab by serial becomes the unit's edit form: Preview → Send
(confirm) → "the unit now holds". New logic goes in
sfm/static/device_routing.jswith Node tests, and glue insfm/static/sfm_webapp.js(no inline JS).
Verification
- Series 3, byte for byte. For each of the 11 Blastware compliance writes,
apply the new encoder to the read before it (or the previous image), with the
fields Blastware changed. It must reproduce Blastware's content
[0:2090]. The two trailing bytes are compared separately, since they are Blastware's constant. Ten writes change one field. BE9441's 16:26 write changes five things: mode, sample rate 2048 → 4096, interval, the mic in the sensors byte[25], and[26:28]. That one is tested with all of them, through the codec, though v1's page does not offer the sensors byte. The frames rebuild throughpush_config_raw's path (the #46 tests). - Series 3, read side. The assembly rule above, pinned as a test on the 12 reads of BE9441's 10-26 recording. Each read-back equals the write before it.
- Series 4. THOR's setup-push captures still reproduce, with no field changed. Any capture where THOR changed a field becomes a byte-for-byte test.
- Bench, with Brian's go per send, per series:
- A dry run of a trigger change, then send, read back, check the unit's menu, and change it back.
- A notes change.
- A several-field "backup unit" change.
- A dry run while monitoring, expecting the refusal.
Calls I made (Claude), for Brian to check
- Series 3
[26:28]goes out as00 00, as Blastware writes it, rather than the01 7fthe unit may read. Measured beats inferred: Blastware's value is on 11 accepted writes, and writing back01 7fhas never been tried. - The two bytes after the Series 3 block stay
00 00, aswrite_imagesends them (a unit accepted that in the April 2026 test). Blastware's per-unit constant cannot be derived from anything SFM reads. The first bench write checks that the unit takes the block, and its menu shows the change. - The
0x09block is echoed as read, as Blastware did on all nine April writes. - Level writes are refused unless the channel's unit string is the measured
in./in(a unit set to Metric is unmeasured). - Notes are edited by slot under the unit's own labels, four slots on each series. Series 3 extended notes and Series 4's separate location field wait.
- The legacy by-host write is retired; the by-host form becomes read-only.
- A live send needs a preview's revision.
- Preview re-reads the unit rather than trusting the page's copy, and Send refuses if the unit changed since the preview.
- The edit form lives on the Config tab, by serial, replacing the legacy read-modify-write form for units reached by serial.
Left to the builder
- The module layout for the shared codec, within the no-cycle, pure-codec rule.
run_changegeneralised versus a sibling core.- The exact revision hash.
- Page layout details.