serversdownandClaude Opus 5 711277131a feat(micromate): a config diff harness -- and the blocks turn out to be bit-stable
The Series III config map was built by changing one setting and seeing which
bytes moved.  Whether that method transfers to Series IV depended on an
unmeasured question: how much of a config block moves ON ITS OWN.  If a block
carries a clock or a counter, every diff arrives polluted and there is no way to
tell the real answer from the noise in a single pair.

MEASURED ON UM20147 (11.0BD), FOUR READS BACK TO BACK, NOTHING TOUCHED:

  0x1A compliance    2103 B    0 bytes moved
  0x2C call home      137 B    0
  0x2E trigger         39 B    0
  0x08 event index    101 B    0
  0x01 device info    163 B    0
  0x1C monitor status  59 B    2  (and those two are explained below)

Every config block held every byte still, so the method transfers directly and
needs no noise floor subtracted.  That was the open question.

scratch/mm_config_diff.py drives it: `volatile` establishes the floor, `snap`
saves a labelled read of all five blocks, `diff` reports changed runs with each
plausible field interpretation (u8/u16be/u32be/f32be) and reports any volatile
bytes separately rather than silently hiding them.

CORRECTION TO content[6] OF 0x1C, long recorded only as "NOT the hour": it is a
sub-second tick, ~4020 counts/second in steps of exactly 4, wrapping at 256 --
about a 64 ms period, consistent with the low 6 bits of a millisecond counter.
Consequence worth knowing: a 0x1C block is never byte-identical to itself, so
any "has the status block changed?" check answers yes ~15 times a second.
Compare named fields, never whole blocks.

I nearly recorded that byte wrong, and the trap is cheap enough to document:
sampled once per second it reads as a smooth monotonic counter at ~16/s with a
16-second wrap.  Twelve consecutive samples supported it.  It is aliasing -- a
64 ms counter read once a second is noise that happens to look tidy.  Measure a
counter back to back before trusting any rate from spaced samples.

Also: the settings do not have to be changed from THOR.  The unit's own keypad
works, which needs no Windows VM and does not contend for the modem's single
TCP session.

Co-Authored-By: Claude Opus 5 <noreply@anthropic.com>