Brian: use the actual unit limits, not what has been used in the past. Trigger 0.005-10 in/s in 0.001 steps (capped by the range), record time 1-90 s (Micromate) or the MiniMate manual's 1 MB table by sample rate (13 s in Histogram Combo), job number 1-9999. The read-back reports any value a unit does not take. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
40 KiB
Phase 5: Changing a Unit's Settings by Serial — Implementation Plan
For agentic workers: REQUIRED SUB-SKILL: Use superpowers:subagent-driven-development (recommended) or superpowers:executing-plans to implement this plan task-by-task. Steps use checkbox (
- [ ]) syntax for tracking.
Goal: An operator edits a unit's live settings in SFM, by serial, on either series: preview old → new, send once (audited), and see what the unit now holds.
Architecture:
- One pure settings codec for both series (
micromate/setup_fields.py), on the measured shared 2,090-byte layout. - Each series' adapter supplies reads, the plan and the send to the existing
run_changecore, unchanged.- Its "state" is a read-back verdict:
as asked/unchanged/differs. - Series 3 sends through
push_config_raw; Series 4 throughplan_setup_push. - Series 4 also reads the unit's schedule, read-only, to refuse a change that a scheduled start would undo.
- Its "state" is a read-back verdict:
- One audited route,
POST /device/settings/change. The legacy by-host write is retired.
Tech Stack: Python 3 (FastAPI, pytest), vanilla JS tested in Node (tests/test_webapp_device_routing.py).
Spec: docs/superpowers/specs/2026-11-03-sfm-phase5-settings-design.md (approved by Brian 2026-11-03). Parent spec: docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md.
Global Constraints
- Worktree and branch: work in
~/seismo-relay-parityonfeat/series4-sfm-parity. Checkgit branch --show-currentbefore every commit; never commit tomain; push every commit. - Python:
/home/serversdown/seismo-relay/.venv/bin/python -m pytest …from the worktree root. - The suite has 16 known failures.
- Run the full suite once before Task 1 and save the failure list in the plan workspace as
baseline-failures.txt. - Before every commit, run the suite, read the failure list, and confirm it equals the baseline.
- Commit in a separate call, never chained to the test run.
- Run the full suite once before Task 1 and save the failure list in the plan workspace as
- No live device traffic in any task. The bench is after the final review, with Brian's explicit go per send.
- Never port 8200.
- Never write modem baud or modem type.
- Never send
0xA2.
CHANGELOG.mdis not touched on this branch; it is written ondevat the merge.- No inline JavaScript. New page logic goes in
sfm/static/device_routing.js(DOM-free, Node-tested); glue goes insfm/static/sfm_webapp.js. docs/instantel_protocol_reference.mdis CRLF. Edit it only with the Edit tool. Keep its rows in date order.- Off-limits (the checker session's files):
minimateplus/event_file_io.py,histogram_codec.py,waveform_codec.py,bw_xml_report.py,decode_check*.py. TOOL_VERSIONis unchanged: nothing here changes event decoding. Operational consequence: none; no migration, no backfill.- Grade every protocol claim in comments and docs as measured or inferred.
- Mutation checks: after a mutate/restore, delete the module's
__pycache__/*.pycbefore the confirming run.
Review Focus
- A Micromate with no schedule file, or a schedule read that fails.
- A body with no records means no conflict, and the change proceeds.
- A read that raises refuses the change in plain words.
- Owner: Task 4; tests
test_no_schedule_records_is_no_conflictandtest_an_unreadable_schedule_refuses.
- A float that is stored as f32. 0.3 reads back as 0.30000001192.
- The verdict compares bytes with the image written, never floats, so it reads
as asked. - Owner: Task 2; test
test_f32_values_read_back_as_asked.
- The verdict compares bytes with the image written, never floats, so it reads
- A stale or foreign revision after a keypad edit, or a preview of another unit.
- The send is refused with 409, nothing is sent, and no audit row is written.
- Owner: Tasks 4 and 5; tests
test_s4_stale_revision_sends_nothingandtest_s3_stale_revision_sends_nothing.
- Range and trigger changed in one request. The trigger is checked against the new range.
- Owner: Task 2; test
test_trigger_checked_against_the_new_range.
- Owner: Task 2; test
- Notes at the edges.
""clears a note.- 41 bytes is accepted; 42 is refused; non-ASCII is refused.
- Labels are never written.
- Owner: Task 2; tests
test_note_edgesandtest_note_labels_are_never_written.
File Structure
| file | responsibility |
|---|---|
minimateplus/protocol.py |
Series 3 assembly: assemble_compliance_content, is_compliance_content, MiniMateProtocol.read_compliance_content. Writes: push_config_raw's pre-check raises ImageRefused. |
minimateplus/client.py |
Added: read_compliance_content() and read_settings_blocks(). Deleted: apply_config, set_project_info, _encode_compliance_config. |
micromate/setup_block.py |
Added: with_crc(content). |
micromate/setup_fields.py (new) |
The shared, pure settings codec: field registry, editable_fields, check_changes, encode_setup, field_values, settings_state, revision. |
micromate/schedule.py (new) |
Pure: parse_schedule, ScheduleEntry, schedule_conflicts. |
micromate/protocol.py |
Added: read_file(path), using 0x94 then 0x48 pages. |
micromate/client.py |
Added: read_schedule(), get_active_setup_raw(), get_prefs_content(), get_trigger_content(), get_setup_content(). |
micromate/opcodes.py |
ENABLEABLE_WRITES gains the setup-push SUBs. |
sfm/device_access/errors.py |
Added: SettingsChanged and ChangeRefused, both 409. |
sfm/device_access/change.py |
Added: settings_change(). |
sfm/device_access/s4.py |
s4_settings moves onto the codec. Added: s4_change_settings. |
sfm/device_access/s3.py |
Series3Adapter.settings moves onto the codec. Added: Series3Adapter.change_settings. |
sfm/device_access/access.py |
Added: DeviceAccess.change_settings. |
sfm/server.py |
Added: POST /device/settings/change. /device/config and /device/config/project return 410. |
sfm/static/device_routing.js, sfm/static/sfm_webapp.js, sfm/sfm_webapp.html |
The edit form. |
tests/device_access_fakes.py |
The fakes gain settings blocks, a schedule, and writes. |
Task 1: Series 3 — assemble a settings read into the logical block
Measured basis (spec, 2026-11-03): unescape each 1A data frame and take data[11:]. Every frame but the last starts with one non-content 00 byte; drop it. The result is 2,090 bytes with N = 2090 at [0:2].
Files:
- Modify:
minimateplus/protocol.py, nearread_compliance_config(≈ line 913) - Modify:
minimateplus/client.py - Modify:
docs/instantel_protocol_reference.md(one dated row, 2026-11-03) - Test:
tests/test_s3_settings_read.py(new)
Interfaces:
-
Produces:
assemble_compliance_content(frames: list[bytes]) -> bytes.framesis each data reply'sframe.data, escapes kept, in order.is_compliance_content(block: bytes) -> bool:len == 2090andint.from_bytes(block[0:2], "big") == 2090.COMPLIANCE_CONTENT_LEN = 2090.MiniMateProtocol.read_compliance_content() -> bytes:- sends the same probe and B/C/D requests as
read_compliance_config; - a missing data frame raises
ProtocolError(nocontinue); - returns the assembled bytes without judging them, so a monitoring unit's device-info answer comes back for the caller to recognise.
- sends the same probe and B/C/D requests as
MiniMateClient.read_compliance_content() -> bytes.MiniMateClient.read_settings_blocks() -> dictwith keysevent_index,compliance,triggerandwaveform. Each value is the bytes its existing read returns:event_index:proto.read_event_index()compliance:read_compliance_content()trigger:proto.read_trigger_config()waveform:proto.read_waveform_data_raw()
-
Step 1: Write the failing tests in
tests/test_s3_settings_read.py. Usetests/fixtures/minimate-bench-10-26-26-histogram(unit_replies.binparsed withS3FrameParser). Group the0xE5frames into reads: a frame withlen(data) == 11starts each read. Take the write images fromtests/test_minimate_write_frames.py(WRITES[0:13],WRITES[13:26]; the0x71payloads[16:]joined).
def test_reads_assemble_to_2090_bytes(): # all 12 reads, both frame patterns (5 and 3 frames)
assert all(len(assemble_compliance_content(r)) == 2090 for r in READS)
def test_a_read_after_a_blastware_write_equals_the_write():
assert assemble_compliance_content(READS[1]) == IMAGES[0][:2090]
assert assemble_compliance_content(READS[2]) == IMAGES[1][:2090]
def test_a_monitoring_units_answer_is_not_settings(): # READS[3:8] start 0e 'BE9441'
assert [is_compliance_content(assemble_compliance_content(r)) for r in READS[3:8]] == [False] * 5
assert all(is_compliance_content(assemble_compliance_content(r)) for r in READS[:3] + READS[8:])
def test_read_compliance_content_refuses_a_missing_frame():
... # a fake transport that answers the probe and frame B, then times out: ProtocolError, nothing assembled
- Step 2: Run them.
…pytest tests/test_s3_settings_read.py -q. Expected: FAIL (ImportError:assemble_compliance_content). - Step 3: Implement the Produces block in
protocol.pyandclient.py. Useunescape_replyfromminimateplus.framing. The docstring cites the measurement and its fixture. - Step 4: Run them. Expected: PASS (4 passed).
- Step 5: Add a reference row to
docs/instantel_protocol_reference.md(Edit tool; CRLF), dated 2026-11-03, measured. Cover:- the assembly rule;
- the round trip (2 of 2);
[26:28]: Blastware writes00 00(11 of 11), and the unit sets01 7fafter monitoring;- the trailing two bytes: a per-unit Blastware constant (
7c eaBE11529 ×9,0c e9BE9441 ×2), not the CRC, and never in a reply; 0x09 [200:202]: echoed 9 of 9 in April; changed on 10-26 (17 17→78 76).
- Step 6: Run the full suite and compare the failure list with the baseline. Then commit (separate call):
feat(s3): assemble a settings read into the 2,090-byte block.
Task 2: The shared settings codec
Files:
- Create:
micromate/setup_fields.py - Modify:
micromate/setup_block.py: addwith_crc(content: bytes) -> bytes, which returns the content withCRC-16/AUG-CCITTover[0:N-2]written BE at[N:N+2]. - Test:
tests/test_setup_fields.py(new)
Interfaces:
- Consumes: Task 1's
assemble_compliance_content, used in tests only. The codec itself imports nothing fromminimateplusand nothing fromsfm. - Produces (all pure). Each
contentis the 2,090-byte (Series 3) or 2,092-byte (Series 4) block.editable_fields(content: bytes, series: int) -> list[dict]. Each dict hasfield,label,kind("number","choice","text"or"bool") andvalue, plus as relevantunit,min,max,choices(a list of{value, label}) andmaxlength.field_values(content: bytes, series: int) -> dict[str, Any]: the same fields, values only.check_changes(content: bytes, series: int, changes: dict) -> dict: returns the normalised changes, or raisesSettingsValueError(ValueError)with a plain sentence. It has two kinds of check: static (names, types, choices, limits) and content-dependent (unit strings, the trigger against the range).encode_setup(content: bytes, series: int, changes: dict) -> bytes:- same length as
content; - only the changed fields' bytes move;
- bit fields are patched bit by bit;
- Series 3 also sets
[26:28] = 00 00(S3_BLASTWARE_FIXED); - Series 4 returns
with_crc(...).
- same length as
settings_state(before: bytes, written: bytes, now: bytes, *, series: int) -> str. It compares[0:2090]:"as asked"ifnowequalswritten(Series 4 also needscrc_ok(now));- else
"unchanged"ifnowequalsbefore; - else
"differs".
revision(blocks: list[bytes]) -> str=hashlib.sha256(b"".join(blocks)).hexdigest()[:16].S3_ALARM_WRITE_MEASURED = False. While False,geo_alarm_ipsis shown (read-only,kindplus"readonly": True) and refused incheck_changes.
The field registry (content offsets; channel k = 1708 + 48·k, k = 0, 1, 2 for the geophones):
| field | series | bytes | values (exact) |
|---|---|---|---|
record_mode |
3 | [5] |
single_shot=0, continuous=1, histogram=3, histogram_continuous=4 |
record_mode |
4 | [5] |
waveform=1, manual=2, histogram=3, combo=4 |
sample_rate |
3, 4 | [6:8] u16 BE |
{1024, 2048, 4096} |
histogram_interval_s |
3, 4 | [8:10] u16 BE |
{2, 5, 15, 60, 300, 900} |
record_time_s |
3 | [18:22] f32 BE |
1.0 ≤ v ≤ S3_MAX_RECORD_S[sample_rate] = {1024: 104, 2048: 52, 4096: 26}, and ≤ 13 when record_mode is histogram_continuous (the new mode and rate if they change too) |
record_time_s |
4 | [18:22] f32 BE |
1.0 ≤ v ≤ 90.0 |
geo_trigger_ips |
3, 4 | ch k +30 f32 BE, all three |
0.005 ≤ v ≤ 10.0, a multiple of 0.001 (within 1e-9), and ≤ the (new) range's full scale (10.0 / 1.25) |
geo_range |
3 | ch k +29 u8, all three |
normal=0, sensitive=1 |
geo_alarm_ips |
3 (gated) | ch k +38 f32 BE, all three |
0.005 ≤ v ≤ full scale |
job_number |
3, 4 | [40:42] u16 BE |
1 ≤ v ≤ 9999 |
job_number_shown |
3, 4 | [39] bit 0x02 |
bool |
note_1 … note_4 |
3, 4 | value [144+64n : 186+64n]; label [122+64n : 144+64n] read only |
printable ASCII 0x20–0x7E, ≤ 41 bytes, zero-filled to 42 |
The limits are the units' own, from Instantel's manuals (Brian, 2026-11-03), cited in the module as comments:
- Trigger 0.005–10 in/s, steps of 0.001: MiniMate Plus manual 716U0101 Rev 15, Appendix Series III Specifications; Micromate manual 721U0201 Rev 06, Geophone Trigger Level.
- Record time:
- Micromate 1–90 s (Record Time).
- MiniMate: §4.4's Maximum Record Times, 1 MB unit, channels 1–4; §3.8.9, 13 s in Histogram Combo.
- Job number 1–9999: both manuals.
A level write is refused unless all three geophone channels' unit string, ch +34:+38 with NULs and . stripped, equals in. Measured: in. and in both occur on UM20147. A unit set to Metric is unmeasured.
- Step 1: Write the failing tests in
tests/test_setup_fields.py:
def test_s3_reproduces_every_blastware_write():
# BE11529's 9 April images (bridges/captures/4-20-26/**/raw_bw_*.bin), in time order (the HHMMSS in
# the name): before = the previous image; the first (151921) uses the assembled 1A read in its own
# raw_s3 capture if that read is whole, else it is skipped (say so in the test). BE9441's 2 images:
# before = the assembled READS[0], then image 1. changes = field_values(after) for every field whose
# value differs. Assert encode_setup(before, 3, changes)[:2090] == after[:2090].
# The 16:26 write also moves [25] (mic on): include `_sensors_mic` as a test-only patch helper,
# not a registry field.
def test_s4_fixture_pairs_reproduce():
# for every ordered pair (a, b) of tests/fixtures/micromate-setup/*.bin whose differing bytes outside
# [2090:2092] all lie inside registry fields: encode_setup(a, 4, {f: field_values(b)[f] for differing f}) == b
def test_trigger_checked_against_the_new_range(): ... # Series 3: {"geo_range": "sensitive", "geo_trigger_ips": 2.0} refuses (above 1.25); 1.25 passes; {"geo_trigger_ips": 2.0} alone on Normal passes
def test_limits_refuse_outside_values(): ... # 0.004 and 10.001 in/s; 0.3005 (not a 0.001 step); job 0 and 10000; S4 0.9 and 90.1 s;
# S3 at 4096 sps: 26 ok, 26.1 refused; S3 histogram_continuous: 13 ok, 13.1 refused;
# S3 {"sample_rate": 1024, "record_time_s": 100} passes from a 4096 sps block (the new rate counts);
# sample rate 8192; interval 30
def test_field_not_offered_for_the_series(): ... # geo_range on series 4 -> SettingsValueError naming the series
def test_alarm_is_read_only_until_measured(): ... # Series 3 geo_alarm_ips -> SettingsValueError while S3_ALARM_WRITE_MEASURED is False
def test_metric_unit_string_refuses_level_writes(): ... # ch +34 = b"mm\x00\x00" -> SettingsValueError; b"in.\x00" and b"in\x00\x00" pass
def test_note_edges(): ... # "" clears; 41 bytes ok; 42 refused; "café" refused
def test_note_labels_are_never_written(): ... # changing note_2 moves only [208:250]
def test_bit_fields_keep_unknown_bits(): ... # [39] = 0x81: job_number_shown True -> 0x83, False -> 0x81
def test_s3_fixed_bytes(): ... # any S3 encode sets [26:28] = 00 00
def test_s4_encode_recomputes_the_crc(): ... # crc_ok(encode_setup(fixture, 4, {"job_number": 7}))
def test_f32_values_read_back_as_asked(): ... # written = encode(0.3); settings_state(before, written, written) == "as asked"
def test_settings_state(): ... # "as asked" / "unchanged" / "differs"; S4 with a bad CRC -> "differs"
def test_revision_is_stable_and_content_sensitive(): ...
- Step 2: Run them. Expected: FAIL (ModuleNotFoundError:
micromate.setup_fields). - Step 3: Implement
micromate/setup_fields.pyandwith_crc. Decode via the registry (not viadecode_setup, whose notes are keyed by label). Note labels come from the unit and are display only. - Step 4: Run them. Expected: PASS. Then mutation-check
S3_BLASTWARE_FIXED(remove it: the Blastware-write test fails), delete the.pyc, and restore. - Step 5: Run the full suite and compare with the baseline. Commit:
feat(codec): one settings codec for both series, on the shared layout.
Task 3: Micromate — read the schedule file
Measured basis: reference The scheduler, and a generic file transfer and The schedule record format.
0x94 <path>opens a file for read;offset= the path's length, and the payload is the path, unpadded.0x48reads the next page, repeated until an all-zero 11-byte response.- The file is 260-byte records.
[4]is the half-hour slot.[6]is the action bitmask: 2 start-with-setup, 4 stop, 8 self-check, 16 ACH.[7]is the name length;[8:260]is the name.
Files:
- Create:
micromate/schedule.py - Modify:
micromate/protocol.py(read_file),micromate/client.py - Modify:
docs/micromate_protocol_reference.md: one line under the scheduler section saying SFM reads it (read-only) from 2026-11-03. - Test:
tests/test_micromate_schedule.py(new)
Interfaces:
-
Produces:
SCHEDULE_PATH = "\\system\\schedule\\schedule.dat".MicromateProtocol.read_file(path: str) -> bytes: the concatenated content of each0x48page after the page-0 descriptor, stopping at the all-zero response; at most 64 pages, elseProtocolError.MicromateClient.read_schedule() -> bytes.MicromateClient.get_active_setup_raw() -> str: the0x41name up to its NUL, not stripped.MicromateClient.get_prefs_content() -> bytes(0x08),get_trigger_content() -> bytes(0x2E) andget_setup_content() -> bytes(0x1A), each_content(...)of its read.@dataclass(frozen=True) class ScheduleEntry:index: int,time: str("HH:MM"),day: int,action_bits: int,actions: tuple[str, ...],setup: str.parse_schedule(body: bytes) -> list[ScheduleEntry]: 260-byte records; a record with[4:8]all zero and no name ends the list.actionsnames the bits 2/4/8/16 asstart_with_setup,stop,self_checkandcall_home; any other bit becomesunknown(0xNN).schedule_conflicts(entries, active_setup: str) -> list[ScheduleEntry]: start entries whose setup, compared case-insensitively afterstrip(), is not the active setup.
-
Step 1: Write the failing tests. Parse the bodies from
bridges/captures/9-24-26 - micromate2/raw_s3_20260924_191835_read_scheduler.bin(frames 0–4), and from the tworaw_*_20260925_*captures, withMicromateFrameParser.
def test_the_read_capture_parses(): # 2 records: start-with-setup 07:30 (40-char name), call_home 19:30 (no name)
def test_the_five_entry_schedule_parses(): # 07:30 start TEST1.mmb, 08:00 stop, 08:30 self_check, 09:00 start test2.mmb, 19:30 call_home
def test_conflicts_are_case_insensitive(): # active "test1.MMB" -> only the test2.mmb start conflicts
def test_no_schedule_records_is_no_conflict(): assert schedule_conflicts(parse_schedule(bytes(524)), "x.mmb") == []
def test_read_file_sends_the_path_unpadded_and_stops_at_zeros(): ... # a fake protocol: 0x94 offset 0x001D, then 0x48 until all-zero
def test_get_active_setup_raw_keeps_spaces(): ... # " LOC 4 .mmb " survives
- Step 2: Run them. Expected: FAIL (ModuleNotFoundError:
micromate.schedule). - Step 3: Implement the Produces block.
read_filebuilds0x94withbuild_request(0x94, len(path), bytes(10), path.encode("ascii")). Both SUBs are already inREAD_SUBS. - Step 4: Run them. Expected: PASS.
- Step 5: Run the full suite and compare with the baseline. Commit:
feat(micromate): read the schedule file (read-only).
Task 4: Change-path pieces and the Series 4 settings change
Files:
- Modify:
sfm/device_access/errors.py,sfm/device_access/change.py,micromate/opcodes.py,sfm/device_access/s4.py,tests/device_access_fakes.py - Modify:
tests/test_micromate_execute.py,tests/test_micromate_opcodes.py. Their over-ceiling example0x71becomes enableable here; use0xA8(delete one event) instead. - Test:
tests/test_device_access_s4_settings.py(new)
Interfaces:
- Consumes:
- Task 2:
editable_fields,field_values,check_changes,encode_setup,settings_state,revision. - Task 3:
read_schedule,parse_schedule,schedule_conflicts,get_active_setup_raw, and the content getters. micromate.writes.plan_setup_push(prefs, trigger, setup, name).
- Task 2:
- Produces:
class SettingsChanged(DeviceError),http_status = 409.class ChangeRefused(DeviceError),http_status = 409.settings_change() -> ChangeKind("change settings", "as asked", "applied").ENABLEABLE_WRITES = frozenset({0x96, 0x97, 0x7E, 0x7F, 0xDA, 0x68, 0x73, 0x82, 0x83, 0x71, 0x72}).s4_settings(client, serial) -> UnitSettings. It keeps every current field, plusdetail["editable"] = editable_fields(setup, 4). The revision comes only from a preview (dry run), never from this read.s4_change_settings(client, serial, *, changes: dict, revision: Optional[str], dry_run: bool, audit, address, caller, sleep, verify_timeout_s, verify_interval_s) -> ChangeResult.
s4_change_settings behaviour, through run_change (even_if_already=False):
read_state, first call:- Read the run state.
calling_homeandunknown*are returned as they are.monitoringraisesUnitStateUnknown(f"{serial} is monitoring -- stop it first (a stop makes it call in if call-home is on) -- nothing sent").
- Read the name (raw), prefs, trigger and setup.
check_changes(aSettingsValueErrorbecomesBadRequest).- When live,
revision([name.encode("ascii"), prefs, trigger, setup])must equal the request's. Otherwise raiseSettingsChanged("the unit's settings changed since the preview -- preview again; nothing sent"). image = encode_setup(setup, 4, changes).- Read the schedule; a read error is kept as a blocking reason.
- Return
settings_state(setup, image, setup, series=4).
- Read the run state.
- Later calls: read the setup and return
settings_state(before, image, now, series=4). describe_plan:plan_setup_push(prefs, trigger, image, name).describe(), plusfields={f: {label, old, new}}for the changed fields,revision,warningsandblocked.warningsalways includes: "this reads the unit's keypad edit buffer: if someone is in its menu, their unsaved edits are sent too".blocked:- the name is
factory.MMB(case-insensitive): "this unit is running its factory setup; load any other setup on its keypad, then try again"; - each schedule conflict: "this unit's schedule loads at HH:MM, which would replace this change";
- a schedule read error: "could not read this unit's schedule () -- SFM will not change a unit whose schedule it cannot check";
plan_setup_pushraisingValueError(a block size or header THOR's offsets were never seen at, or a failed checksum): "this unit's -- SFM has not seen a block like it".
- the name is
- When
blockedis non-empty,stepsis[], and a live call raisesChangeRefused("; ".join(blocked) + " -- nothing sent"). - When the schedule names the active setup,
warningsadds: "the unit's schedule loads this setup at HH:MM; the change survives it".
send:client.protocol.execute(plan), ass4_change_call_homedoes.snapshot:fields:{f: {label, old, new, now}};before/afterhex of the setup;changed_offsets;prefsandtrigger: before hex only.
is_refusal,is_ringandpre_send: as ins4_change_call_home.
Validation order: a live call without revision is refused before any I/O, in Task 6's DeviceAccess (not here).
- Step 1: Write the failing tests. Extend
FakeMicromateClientwith:setup(default:tests/fixtures/micromate-setup/um20147_baseline.bin's last 2,092 bytes);prefs,triggerandactive_name(default"test2.mmb");schedule_body(defaultbytes(524)) andschedule_error;executeapplies a setup-push plan's0x71data toself.setup(re-adding the CRC withwith_crc), so a read-back sees it.
def test_s4_dry_run_previews_old_to_new_and_sends_nothing(): # fields == {"geo_trigger_ips": {"label": ..., "old": 0.5, "new": 0.3}}; revision present; fake.executed == []
def test_s4_trigger_change_sends_the_push_and_verifies(): # subs [0xDA,0x68,0x73,0x82,0x83,0x71,0x72]; outcome sent/ok; state_after "as asked"
def test_s4_stale_revision_sends_nothing(): # SettingsChanged; fake.executed == []; audit has no rows
def test_s4_already_as_asked_is_skipped(): # outcome skipped, nothing executed
def test_s4_monitoring_is_refused(dry_run in (True, False)): # "stop it first" and "nothing sent"
def test_s4_factory_setup_is_refused_live_and_shown_in_preview():
def test_s4_schedule_conflict_blocks(): # schedule starts "other.mmb" at 07:00 -> ChangeRefused naming other.mmb and 07:00
def test_s4_schedule_on_the_active_setup_warns_only():
def test_an_unreadable_schedule_refuses():
def test_no_schedule_records_is_no_conflict(): # bytes(524) -> sent
def test_s4_a_lost_ack_records_what_the_unit_holds(): # execute_error -> outcome failed, sent True, snapshot present
def test_s4_unseen_block_size_is_blocked(): # prefs 91 B -> preview blocked; live ChangeRefused; nothing executed
def test_setup_push_subs_are_now_enableable():
- Step 2: Run them. Expected: FAIL (ImportError:
s4_change_settings). - Step 3: Implement the Produces block and update
s4_settings. In the two opcode tests, swap the over-ceiling example to0xA8. - Step 4: Run them, plus
tests/test_micromate_execute.py,tests/test_micromate_opcodes.py,tests/test_micromate_writes.py(THOR's setup-push frames still reproduce) andtests/test_device_access_s4_call_home.py. Expected: PASS. - Step 5: Run the full suite and compare with the baseline. Commit:
feat(s4): change a Micromate's settings by serial, audited, schedule-checked.
Task 5: The Series 3 settings change
Files:
- Modify:
sfm/device_access/s3.py,minimateplus/client.py(push_config_rawpre-check),tests/device_access_fakes.py - Test:
tests/test_device_access_s3_settings.py(new)
Interfaces:
-
Consumes:
- Task 1:
read_compliance_content,read_settings_blocks,is_compliance_content. - Task 2: the codec.
- Task 4:
settings_change,SettingsChanged,ChangeRefused. MiniMateClient.push_config_raw(event_index, compliance, trigger, waveform).
- Task 1:
-
Produces:
push_config_rawraisesImageRefused(fromminimateplus.protocol) when its pre-check refuses a block, before any frame.Series3Adapter.settings(address, serial) -> UnitSettings. It is built onread_compliance_contentand the codec, with the same top-level fields as today, plusdetail["editable"]. A non-settings answer raisesUnitStateUnknown(S3_SETTINGS_WHILE_MONITORING).S3_SETTINGS_WHILE_MONITORING = "did not return its settings -- a MiniMate answers this way while monitoring; stop monitoring first".Series3Adapter.change_settings(address, serial, *, changes, revision, dry_run, audit, caller, sleep, verify_timeout_s, verify_interval_s) -> ChangeResult, throughrun_changeandself._run(address, op, retry=False):read_serial:client.connect().serial.read_state, first call:- Wake, then
read_settings_blocks(). - A non-settings
complianceraisesUnitStateUnknown(f"{serial} {S3_SETTINGS_WHILE_MONITORING} -- nothing sent"). check_changes, and the live revision check over[event_index, compliance, trigger, waveform].- Compute
image = encode_setup(...)and returnsettings_state(...).
- Wake, then
- Later calls: wake,
read_compliance_content(), then the verdict. describe_plan:stepsin Blastware's order: 68 (event index, as read), 73, 71 (settings block, 2,092 B), 72, 82 (trigger, as read), 83, 69 (waveform data, as read), 74, 72. Each step hassub,whatanddata_len.fields,revisionandwarningsas Series 4. The warning: "the two bytes after the settings block go out as 00 00 (Blastware sends its own constant; effect unmeasured)".
send:client.push_config_raw(event_index, image, trigger, waveform).is_refusal:isinstance(e, ImageRefused).snapshot: as Series 4, without prefs and trigger.
-
Step 1: Write the failing tests. Extend
FakeMiniMateClientwith:settings_blocks: the default compliance is BE9441's assembled 10-26 read 2;event_index,triggerandwaveformare the0x68,0x82and0x69payloads ([16:]) ofWRITES[13:26]intests/test_minimate_write_frames.py, each minus its two trailing bytes (a write is the read plus two);push_config_rawrecords its four arguments and makes the compliance the newread_compliance_content;monitoring_settings=Truereturns a device-info-shaped block.
def test_s3_settings_read_uses_the_codec(): # editable includes geo_range; record_mode "histogram"; notes by slot with unit labels
def test_s3_dry_run_previews_and_sends_nothing():
def test_s3_trigger_change_sends_the_units_own_blocks(): # push args: event_index/trigger/waveform == the read blocks; compliance == encode_setup(...)
def test_s3_change_verifies_by_read_back(): # outcome sent/ok, state_after "as asked", snapshot fields old/new/now
def test_s3_stale_revision_sends_nothing():
def test_s3_monitoring_is_refused(dry_run in (True, False)):
def test_s3_a_refused_image_is_audited_as_nothing_sent(): # push_config_raw raises ImageRefused -> row sent 0, "refused before sending"
def test_push_config_raw_precheck_raises_image_refused(): # a 2,089-byte compliance block -> ImageRefused, no frame written
- Step 2: Run them. Expected: FAIL (AttributeError:
change_settings). - Step 3: Implement the Produces block.
- Step 4: Run them, plus
tests/test_s3_call_home.py,tests/test_minimate_write_frames.pyandtests/test_device_access_*. Expected: PASS, except the tests Task 6 deletes or rewrites. If one fails here only because it pinsapply_config, leave it for Task 6 and note that in the ledger. - Step 5: Run the full suite and compare with the baseline. Commit:
feat(s3): change a MiniMate's settings by serial on the shared codec.
Task 6: Device access, the route, and retiring the legacy write
Files:
- Modify:
sfm/device_access/access.py,sfm/server.py,minimateplus/client.py - Delete:
poc_set_project.py. Removeexp_compliance_only(and its registration) fromexperiments.py.- Ruling (spec: "the legacy write is retired"): both write a live unit through the retired encoder, which #46 found wrong. Git keeps them.
- Modify tests:
tests/test_s3_compliance_anchor_fields.py: remove its write-side tests (the encoder is gone).tests/test_minimate_write_frames.py:test_apply_config_writes_the_units_own_trigger_blockmoves to Task 5'stest_s3_trigger_change_sends_the_units_own_blocks; delete it here.tests/test_server_series3_guard.py: dropapply_configfromWRITESand its fake method.tests/test_server_device_access.py:/device/confignow 410.
- Test:
tests/test_device_access_settings.py(new),tests/test_server_settings_change.py(new)
Interfaces:
-
Consumes: Task 4's
s4_change_settingsand Task 5'sSeries3Adapter.change_settings. -
Produces:
DeviceAccess.change_settings(address, serial, *, changes: dict, revision: Optional[str], dry_run: bool = False, caller: str = "api") -> ChangeResult:normalise_serial, thenseries_for_serial;changesmust be a non-empty dict, elseBadRequest;- a live call without
revisionraisesBadRequest("preview first: a live change needs the revision from its preview -- nothing sent"); - a live call with no audit log raises
NotSupported(asset_call_home); _dispatch(..., change=True)withverify_timeout_s=_SETTINGS_VERIFY_TIMEOUT_S = 2.0andverify_interval_s=_SETTINGS_VERIFY_INTERVAL_S = 1.0.
POST /device/settings/change:- query:
serial(required),host,tcp_port,port,baud,dry_run,caller; - JSON body
{"changes": {...}, "revision": "..."}; - returns
_change_response(...); - added to
_SERIAL_AWARE_PATHS.
- query:
POST /device/configandPOST /device/config/projectreturn 410 with{"detail": "retired: change a unit's settings by serial with POST /device/settings/change (audited, previewed, read back)"}. Remove the cache-dirty hooks that only these routes called; keep the read routes.- Delete
MiniMateClient.apply_config,set_project_infoand_encode_compliance_config, and any import left orphaned.
-
Step 1: Write the failing tests:
def test_live_change_without_revision_is_refused_before_io(): # factory never called
def test_empty_changes_is_a_bad_request():
def test_change_settings_dispatches_by_series_with_change_true():
def test_route_dry_run_returns_fields_and_revision(): # 200, body.plan.fields, body.plan.revision
def test_route_verified_change_is_200_and_mismatch_409():
def test_route_unconfirmed_carries_sent_and_audit_id():
def test_route_refusal_is_409_with_reason(): # ChangeRefused / SettingsChanged / monitoring
def test_legacy_config_routes_are_gone(): # both 410, nothing reaches a client factory
def test_settings_change_is_serial_aware(): # in _SERIAL_AWARE_PATHS; a UM serial goes to the Series-4 path
- Step 2: Run them. Expected: FAIL (404 / AttributeError).
- Step 3: Implement the Produces block, the deletions, and the test updates listed under Files.
- Step 4: Run them, plus
tests/test_server_*.py. Expected: PASS. Check thatgrep -rn "apply_config\|set_project_info\|_encode_compliance_config" --include=*.py .finds only comments and history. - Step 5: Run the full suite and compare with the baseline. Commit:
feat(sfm): POST /device/settings/change; retire the legacy config write.
Task 7: The page — edit, preview, send, "the unit now holds"
Files:
- Modify:
sfm/static/device_routing.js,sfm/static/sfm_webapp.js,sfm/sfm_webapp.html - Modify:
CLAUDE.md(the "Where things stand" Series-4 parity line: phase 5 built, bench pending), and the spec's status line ("Built 2026-11-xx; bench pending"). - Test:
tests/test_webapp_device_routing.py(extendEXPORTSand cases),tests/test_webapp_static.py
Interfaces:
-
Consumes:
UnitSettings.detail.editable/.revision, and theChangeResultshapes from Tasks 4–6. -
Produces (DOM-free, in
device_routing.js):DEVICE_ROUTES.settings_change = ['POST', '/device/settings/change', true]; removeconfig_write.controlsFor(unit):configWrite: serialMode && (series === 3 || series === 4);reasons.configWrite: 'By serial only: Preview, then Send (audited).'settingsFormFields(st) -> Array<{field, label, kind, value, unit?, min?, max?, choices?, maxlength?, readonly?}>: fromst.detail.editable, in that order.settingsChanges(fields, edited) -> object: only fields whose edited value differs. Numbers compare as numbers; read-only fields are ignored.settingsPreviewRows(body) -> Array<[label, old, new]>, frombody.plan.fields.settingsBlockedText(body) -> string:body.plan.blockedjoined;''if none.settingsResultRows(body) -> Array<[label, old, new, now]>, frombody.detail.snapshot.fields.
-
Page behaviour (
sfm_webapp.jsglue; HTML idscfg-edit,cfg-preview-btn,cfg-send-btn,cfg-preview,cfg-result):- The serial-mode Config pane heading reads "On the unit — read HH:MM".
- The form is built from
settingsFormFields. - Preview POSTs a dry run and shows old → new, warnings and blocked reasons. It keeps
revisionand enables Send only when nothing is blocked and something changes. - Send asks
confirm()naming the serial and the fields, then POSTs live with the kept revision and shows "the unit now holds" rows. Errors go throughchangeErrorText. - Editing any field after a preview disables Send until the next preview.
- The by-host legacy form loses its write button. Its read stays.
- No inline script; listeners via
addEventListener.
-
Step 1: Write the failing Node-run cases in
tests/test_webapp_device_routing.py:controlsForgivesconfigWritetrue by serial for both series, and false by host;settingsChangesreturns only differing fields and compares0.30with0.3as equal;settingsPreviewRows/settingsResultRows/settingsBlockedTexton bodies shaped as Task 6 returns them;deviceUrl('settings_change', unit)→/device/settings/change?serial=…;config_writeis no longer a route.
In
tests/test_webapp_static.py, assert the page has no inline<script>body and loads both static files. -
Step 2: Run them.
…pytest tests/test_webapp_device_routing.py tests/test_webapp_static.py -q. Expected: FAIL (settingsChangesis not a function). -
Step 3: Implement the Produces block and the page behaviour.
-
Step 4: Run them. Expected: PASS. Then run the page under the jsdom harness pattern from phase 4 (
smoke/page_files_load.jsin the scratchpad): the page loads with no console error, and a stubbed dry-run body renders the preview rows. -
Step 5: Update
CLAUDE.mdand the spec status line. Run the full suite and compare with the baseline. Commit:feat(webapp): edit a unit's settings by serial: preview, send, read back.
After the build: final review, then the bench
The bench is not a task: it runs after the whole-branch review and its fixes, with Brian at the units, and each send needs his explicit go. Run SFM with ringkit/sfm_parity.py (port 8299; never 8200). Order:
- Read-only, both units:
GET /device/settingsby serial; Brian matches it against each unit's screen.- UM20147:
read_schedule(measure what a unit with no schedule answers, and whether0x08shows the scheduler switch).
- Dry runs of a trigger change on each unit. Read the preview with Brian.
- One field per unit, with Brian's go:
- the trigger (UM20147 0.5 → 0.3; BE9441 0.04 → 0.05);
- read back; Brian checks the unit's menu;
- change it back, with another go.
- A notes change on each (go).
- A several-field "backup unit" change on each (go).
- Refusals:
- a dry run while monitoring (expect the refusal);
- UM20147 with a schedule naming another setup, if Brian sets one up in THOR (expect the block).
- Alarm level (Series 3):
- one write with Brian's go, read back and checked on BE9441's menu;
- only then flip
S3_ALARM_WRITE_MEASURED, in its own commit naming the bench result.
Record everything in the spec's bench section and both protocol references (measured), then the devlog.