23 more snapshots of UM20147, keypad-driven.
MEASURED: loading a .MMB replaces the active config byte-for-byte -- three
loads, each moving exactly the fields that differ, the loaded block identical
to the file's saved state, no load marker. The setup name's spelling
(.MMB/.mmb) depends on whether it was saved or loaded.
MEASURED: job number [40:42], report flags [39] (0x02 job no., 0x08 scaled
distance; 0x01 unknown); distance [30:34] in metres and charge weight
[34:38] in kg, converted bit-exactly with ft / 3.2808 and lb / 2.21 -- the
firmware's own constants, not the textbook ones. Scaled distance is not
stored. Location confirmed at [144].
MEASURED: 0x08 is the preferences block (display units at content[31]),
not an event index; switching units moves nothing in 0x1A. 0x1A, 0x08 and
0x2E share one checksum rule (N = content[0:2]; CRC over [0:N-2] at
[N:N+2]); 0x2C has none. 0x2C session timeout is at [118:120], encoding
unknown; modem baud is in none of the six blocks.
CORRECTED IN PLACE: the 9-24 notes-label offset (0x8A -> 0x7A) and
channel-block start (0x06BC -> 0x06AC), both 16-byte slips.
RULE: never write modem baud or modem type remotely -- the remote link is
the modem port.
No codec, store or DB change; no backfill.