A mismatch becomes 'accepted' only through a written carve-out whose evidence does not rest on SFM's output. Saturation: Blastware clamps a clipped channel to the range maximum and prints OORANGE, while the file holds counts past full scale. SFM's value must sit at the ceiling (r - 0.0006 .. r * 1.01), so a Sensitive-range event read 8x high on SFM's Normal assumption stays a mismatch; a vector sum is accepted when a clipped channel under it is at the ceiling. series_pairs() is now the one place waveform rows and interval columns are paired. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
84 lines
3.2 KiB
Python
84 lines
3.2 KiB
Python
"""
|
|
minimateplus/decode_check_carveouts.py — the proven ways Blastware's report
|
|
differs from the instrument's data.
|
|
|
|
Every disagreement between SFM and Blastware is SFM's bug until shown otherwise.
|
|
A carve-out is that showing: the fields it covers, the rule it applies, and
|
|
evidence that does not rest on SFM's own output -- Blastware contradicting
|
|
itself, the instrument's own data, a third source, or Instantel documentation.
|
|
`compare` turns a covered mismatch into `accepted` only when the rule holds,
|
|
and every run counts the events each carve-out covers.
|
|
|
|
Pure: no I/O, and no import from `sfm/`.
|
|
"""
|
|
from __future__ import annotations
|
|
|
|
from dataclasses import dataclass
|
|
from typing import Callable
|
|
|
|
from .bw_xml_report import XmlReport
|
|
from .decode_check import FieldResult, SfmSide, _printed_number, series_pairs, matches_printed
|
|
|
|
|
|
@dataclass(frozen=True)
|
|
class Carveout:
|
|
id: str
|
|
fields: tuple[str, ...] # exact names, or prefixes ending in "."
|
|
evidence: str
|
|
applies: Callable[[SfmSide, XmlReport, FieldResult], bool]
|
|
|
|
|
|
_TOL = 0.0006
|
|
|
|
|
|
def _at_ceiling(value: float, r: float) -> bool:
|
|
"""SFM's value sits at the range ceiling: a decoder unit or two past nominal
|
|
full scale is the clipped count itself (32016 = 10.005 in/s). An 8x
|
|
Sensitive-range value is far above the band and does not qualify."""
|
|
return r - _TOL <= abs(value) <= r * 1.01
|
|
|
|
|
|
def _clamped(printed: str, r: float) -> bool:
|
|
if printed.strip() == "OORANGE":
|
|
return True
|
|
p = _printed_number(printed)
|
|
return p is not None and abs(p[0]) >= r - _TOL
|
|
|
|
|
|
def _saturation(sfm: SfmSide, ref: XmlReport, res: FieldResult) -> bool:
|
|
try:
|
|
r = float(ref.scalars["GeoRange"])
|
|
except (KeyError, ValueError):
|
|
return False
|
|
if res.field in ("TranPPV", "VertPPV", "LongPPV"):
|
|
v = sfm.values.get(res.field)
|
|
return v is not None and _clamped(res.bw or "", r) and _at_ceiling(float(v), r)
|
|
if res.field == "PeakVectorSum":
|
|
# A vector sum can exceed the range (11.55 on T438HGTZ.B30W); it is the
|
|
# clipped channel under it that must sit at the ceiling.
|
|
return (_clamped(res.bw or "", r) and
|
|
any(sfm.values.get(f) is not None and _at_ceiling(float(sfm.values[f]), r)
|
|
for f in ("TranPPV", "VertPPV", "LongPPV")))
|
|
got = series_pairs(sfm, ref, res.field)
|
|
if got is None or got[1]:
|
|
return False
|
|
differing = [(cell, v) for _, cell, v in got[0] if not matches_printed(v, cell)]
|
|
return bool(differing) and all(
|
|
_clamped(cell, r) and _at_ceiling(v, r)
|
|
and (cell.strip() == "OORANGE" or (v >= 0) == (_printed_number(cell)[0] >= 0))
|
|
for cell, v in differing)
|
|
|
|
|
|
SATURATION = Carveout(
|
|
id="saturation",
|
|
fields=("TranPPV", "VertPPV", "LongPPV", "PeakVectorSum",
|
|
"samples.Tran", "samples.Vert", "samples.Long",
|
|
"intervals.Tran Peak in/s", "intervals.Vert Peak in/s", "intervals.Long Peak in/s"),
|
|
evidence=("Blastware clamps a clipped channel to the range maximum and prints OORANGE; "
|
|
"the file holds counts past full scale (32016 counts = 10.005 in/s observed). "
|
|
"scratch/verify_against_ascii.py, 2026-08-28."),
|
|
applies=_saturation,
|
|
)
|
|
|
|
CARVEOUTS: tuple[Carveout, ...] = (SATURATION,)
|