Brian Harrison ea4475c9ad client: use anchor-relative offsets for record_time and sample_rate
Fixed a 1-byte offset jitter that produced garbage values when the
device was set to "faster" (4096 Sa/s) mode.

Root cause: 4096 = 0x1000, so the sample_rate bytes in the raw S3
frame are `10 10 00` (DLE-escaped).  After DLE unstuffing → `10 00`
(2 bytes vs 3 for 1024/2048), making frame C 1 byte shorter and
shifting all subsequent field offsets by -1.

Fix: locate the stable 10-byte anchor `01 2c 00 00 be 80 00 00 00 00`
(max-record-limit constant + first two alarm-level floats) and read:
  sample_rate  = uint16_BE at anchor - 2
  record_time  = float32_BE at anchor + 10

Offline-validated against all five logged hex dumps (1071 and 1070
byte cfg, record times 3/5/8 s, sample rates 1024 and 4096):
  all five: correct values with anchor approach.