serversdown a7983d2958 fix: correct DOD field parsing and stop measurement-time resets
Two device-data bugs surfaced while scoping the live-feed work:

1. DOD parser misalignment. DOD's response has no leading counter and
   includes LE + LN1-LN5, but the parser reused the DRD field map
   (parts[0]=counter). That shifted everything: Lp was stored as the
   counter, Leq as Lp, LE as Leq, and LN1 as Lpeak (visible because
   "Lpeak" came out below Lmax, which is impossible). Parse DOD with its
   own map: Lp=0, Leq=1, Lmax=3, Lmin=4, Lpeak=10 (channel 1 = main).

2. measurement_start_time reset on every live-stream open/close. The DOD
   path tags state "Start"; the DRD stream path tags "Measure". The
   transition detector treated only "Start" as measuring, so opening the
   stream ("Start"->"Measure") read as a stop (cleared start time) and
   closing it ("Measure"->"Start") read as a start (reset to now). Every
   viewer reset the elapsed measurement time. Treat {"Start","Measure"}
   both as measuring.

LN1/LN2 (L1/L10) parsing + model/serialization is the next step.