docs(series4): open the Micromate live-protocol reference

First bench session against a Micromate over USB. The headline: the unit
answers Series III command frames unmodified.

An untouched Series III POLL (SUB 0x5B), built by build_bw_frame with no
changes, completed a full two-step probe/data cycle. Ten Series III read
commands were then tried and all ten answered, every one obeying the
response_SUB = 0xFF - request_SUB rule.

Confirmed this session:

- Transport is a plain USB CDC-ACM port (2504:0300, "MICROMATE COM PORT").
  No vendor driver, no Thor, no Windows box needed. Baud is ignored over USB
  (identical responses at 38400 and 115200).
- Device never speaks first -- 20 s idle listen produced nothing.
- Responses are Series III framing MINUS the leading DLE: bare
  [STX][payload][chk][ETX]. This alone means Blastware can never find a frame
  boundary in Micromate traffic, since its parser scans for DLE+STX.
- Response flags byte is 0xC5, not Series III's 0x10.
- Checksum is the DLE-aware variant (SUM8 excluding 0x10 bytes) -- the same
  one Series III uses for 5A and write frames, not the plain SUM8 of its
  ordinary reads. Disambiguated by the POLL data frame, which contains a 0x10.
- The probe response carries the data length at payload[9]. Four of four
  known Series III lengths match; call-home config differs (0x7E vs 0x7C).
- Series III monitor-status field offsets apply unchanged: battery 3.81 V
  (Thor's own reports say 3.8), memory 15,000,000 total and free, date
  23 Sep 2026.
- SUB 0x2C carries the string "RADIO RING" -- the same string seen in the
  RV50 ALEOS debug during the BE12599 incident. That block holds the modem
  dial/answer strings and is the most relevant command to the call-home goal.

Read commands only. Nothing that writes, erases, or changes monitoring state
has been sent to a unit; those are listed as unsafe-until-agreed.

Caveat recorded in the doc: one unit, over USB, with zero events stored, so
the event-walk commands (0x08, 0x1E, 0x0A, 0x06) could only be probed, not
exercised.

Co-Authored-By: Claude Opus 5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01Ru8Lg9HkkYvX9VWWo65SmL
This commit is contained in:
2026-09-23 15:59:17 -04:00
co-authored by Claude Opus 5
parent 154186a6cd
commit 095834183e
+231
View File
@@ -0,0 +1,231 @@
# Micromate Protocol Reference — Thor / Micromate Series IV, live wire protocol
Sibling to [instantel_protocol_reference.md](instantel_protocol_reference.md)
(Series III, "the Rosetta Stone") and
[idf_protocol_reference.md](idf_protocol_reference.md) (Series IV *file*
format). This document covers the Series IV **live device protocol** — what
the unit says over the wire, as opposed to what it writes into a `.IDFW`.
**Status (2026-09-23): opening session.** Everything below was established in
a single bench session against one unit. Treat it as a strong start, not a
settled spec — in particular, everything here comes from **one unit, over USB,
with no events stored**.
---
## The headline
**The Micromate answers Series III command frames.**
An unmodified Series III `POLL` (`SUB 0x5B`), built by
`minimateplus.framing.build_bw_frame` with no changes at all, produced a
complete two-step probe/data cycle from a Micromate. Ten Series III read
commands were then tried and **all ten answered**, every one obeying the
Series III response-SUB rule.
This is not a new protocol. It is close enough to Series III that large parts
of `minimateplus/protocol.py` may apply more or less directly, and the work
ahead is mapping *divergence* rather than mapping from scratch.
---
## How this was obtained
No Thor, no modem, no Windows machine. The Micromate exposes its protocol on
a **USB CDC-ACM virtual serial port**:
```
ID 2504:0300 Instantel Inc. MICROMATE COM PORT
driver: cdc_acm ATTRS{serial}=="V1.00"
→ /dev/ttyACM0
```
Plain CDC ACM, so no vendor driver and no proprietary USB layer — any host
that can open a serial port can talk to the unit.
**Baud is irrelevant over USB.** Identical byte-for-byte responses at 38400
and 115200; CDC-ACM ignores the line rate. TMI provisions Micromate *modem*
links at **115200** (Series III uses 38400) — that matters for the cellular
path, not for USB.
**The device never speaks first.** 20 s of passive listening on an idle open
port produced zero bytes. It is strictly request/response.
---
## Physical + framing layer
### Requests — Series III format, unmodified
Every frame in this session was produced by `build_bw_frame(sub, offset)` with
no Series IV changes, and the device accepted all of them:
```
[ACK 0x41] [STX 0x02] [10 10] [flags 00] [SUB] [00] [00] [offset] [params×10] [chk] [ETX 0x03]
```
⚠ Only the doubled `BW_CMD` (`10 10`) form has been exercised. Whether other
literal `0x10` bytes inside params require stuffing is **untested** — none of
the probes sent carried one.
### Responses — Series III *minus the DLE prefix*
```
Series III: [DLE 0x10] [STX 0x02] … [chk] [ETX 0x03]
Micromate: [STX 0x02] … [chk] [ETX 0x03] ← no leading DLE
```
This single byte matters operationally: Blastware's parser locates frames by
scanning for `DLE+STX`, so it will **never find a frame boundary** in Micromate
traffic no matter what else is correct. That is a structural reason a
Micromate cannot call into a Blastware ACH server, independent of any baud
mismatch.
### Response payload header
```
[0] CMD 0x00 same as Series III
[1] flags 0xC5 ← Series III uses 0x10. Constant across all 10 SUBs.
[2] SUB 0xFF − request_SUB
[3] PAGE_HI
[4] PAGE_LO
[5+] data
```
### Checksum — the DLE-aware variant
```python
chk = sum(b for b in payload if b != 0x10) & 0xFF
```
Confirmed on every frame captured. The `POLL` probe response contains no
`0x10` and so cannot distinguish plain SUM8 from the DLE-aware form; the
`POLL` **data** response contains a `0x10` at payload offset 42, and only the
DLE-aware rule matches there. This is the same checksum Series III uses for
its `5A` bulk-stream and write frames — not the plain SUM8 of ordinary
Series III reads.
### The probe response carries the data length
Series III hardcodes `DATA_LENGTHS` per SUB. On the Micromate the **probe
response tells you**, at `payload[9]`:
| SUB | command | `payload[9]` | Series III constant |
|---|---|---|---|
| `0x15` | serial number | `0x0A` | `0x0A` ✓ |
| `0x01` | device info | `0x98` | `0x98` ✓ |
| `0x1C` | monitor status | `0x2C` | `0x2C` ✓ |
| `0x06` | storage range | `0x24` | `0x24` ✓ |
| `0x2C` | call-home config | `0x7E` | `0x7C` ✗ **differs by 2** |
| `0x08` | event index | `0x5A` | — |
| `0x1E` | event header | `0x08` | — |
| `0x1A` | compliance config | `0x2C` | — |
| `0x0A` | waveform header | `0x00` | — (no event context) |
| `0xFE` | full config | `0x00` | — (see note) |
Four of four known Series III lengths match exactly. **Read the length from
the probe rather than hardcoding it** — it is free, and it already caught the
call-home divergence.
---
## Confirmed commands (read-only)
All ten below answered with a correct `0xFF − SUB` response. Nothing that
writes, erases, or changes monitoring state has been sent to a unit.
| SUB | RSP | Command | Data proven |
|---|---|---|---|
| `0x5B` | `0xA4` | POLL / handshake | yes — ID block |
| `0x15` | `0xEA` | Serial number | yes — `UM12947` |
| `0x01` | `0xFE` | Device info | yes — 152 B |
| `0x2C` | `0xD3` | Call-home config | yes — 126 B |
| `0x1C` | `0xE3` | Monitor status | yes — 44 B |
| `0x06` | `0xF9` | Event storage range | yes — 36 B |
| `0x08` | `0xF7` | Event index | probe only |
| `0x1E` | `0xE1` | Event header / first key | probe only |
| `0x0A` | `0xF5` | Waveform header | probe only |
| `0x1A` | `0xE5` | Compliance config | probe only |
### Decoded so far
**`SUB 0x15` — serial.** ASCII, null-terminated: `UM12947`.
**`SUB 0x5B` / `0x01` — identification strings.**
`Instantel\0` and `MM/ISEE/S/IO` (MicroMate / ISEE standard). `0x01` also
carries eight consecutive `3f 80 00 00` float32 values (= `1.0f`) — almost
certainly per-channel calibration/scale factors, by analogy with Series III's
`geo_hardware_constant`. **Unverified.**
**`SUB 0x1C` — monitor status. Series III field offsets apply unchanged:**
| field | offset | read |
|---|---|---|
| battery × 100 | `payload[-10:-8]` uint16 BE | `0x017D` → **3.81 V** |
| memory total | `payload[-8:-4]` uint32 BE | 15,000,000 |
| memory free | `payload[-4:]` uint32 BE | 15,000,000 (empty) |
| date | `payload[18:22]` | day 23, month 9, year `0x07EA` = 2026 |
The battery reading independently corroborates: Thor's own event reports for
these units print `BatteryLevel : 3.8 volts`.
**`SUB 0x2C` — call-home config.** Contains the ASCII string `RADIO RING`.
Worth flagging: that is the *exact* string seen in the RV50 `ALEOS_SERIAL`
debug during the BE12599 incident —
`'ATQ1^MATE0^MATS0=2^M^MRADIO RING^M'`. So this block holds the modem dial /
answer strings, and it is the most directly relevant command to the
call-home-receiver goal. Field layout **not yet mapped**; Series III's map
(`raw[5]` enabled, `raw[6:46]` dial string) is a starting hypothesis only, and
the length already differs (`0x7E` vs `0x7C`).
**`SUB 0x06` — storage range.** All zeros on this unit, consistent with
`memory free == memory total`. Series III reads first/last event keys from
the final 8 bytes; untestable until the unit holds events.
---
## Divergences from Series III (running list)
1. **No `DLE` prefix on responses** — bare `STX`.
2. **Response flags byte is `0xC5`**, not `0x10`.
3. **Call-home config is 126 bytes**, not 124.
4. **Data lengths are discoverable** from the probe response at `payload[9]`.
5. **Modem serial rate is 115200**, not 38400 (per TMI provisioning practice;
not independently verified here).
---
## ⚠ Untested and unsafe-until-agreed
Nothing below has been sent to a unit, and nothing should be without an
explicit decision:
- **Writes** (`0x68`–`0x83`), **call-home write** (`0x7E`/`0x7F`)
- **Erase** (`0xA3` / `0xA2`)
- **Start / stop monitoring** (`0x96` / `0x97`)
- `0x1F` (advance event pointer) — non-destructive on Series III but it does
move device state, so it is parked with the rest
Also unknown:
- Whether `0x10` bytes inside request params need stuffing
- Whether the bulk waveform stream (`5A` on Series III) exists here, and
whether it is the transport for `.IDFW` bodies we already decode
- Everything about the **call-home session** — the device-initiated direction
has not been observed at all. Specifically: how a unit announces itself,
and **how it learns an event was accepted so it stops re-sending it.**
That last question gates any homebrew receiver and cannot be answered over
USB.
---
## Session provenance
Unit **UM12947**, on the bench via USB, **zero events stored** (memory free ==
total). Read commands only. Every response in this document was checksum-
validated.
An empty unit is a real limitation: `0x08`, `0x1E`, `0x0A` and `0x06` all have
event-dependent payloads that could not be exercised. Recording a couple of
events on the bench unit would unlock the entire event-walk half of the
protocol.