docs(plan): 1c bench fixes, Batch 1 -- #1 settings decode, #9 read-back reconnect, #8 downloads via the file import path; #15 retracted
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
1 parent
342c69602d
commit
0dbd0b506d
1 file changed
+172
@@ -0,0 +1,172 @@
|
||||
# SFM 1c bench fixes, Batch 1 — Implementation Plan
|
||||
|
||||
> **For agentic workers:** REQUIRED SUB-SKILL: Use superpowers:subagent-driven-development (recommended) or superpowers:executing-plans to implement this plan task-by-task. Steps use checkbox (`- [ ]`) syntax for tracking.
|
||||
|
||||
**Goal:** Fix the three reliability findings from the 2026-10-29 bench that put wrong or missing data in front of a user: MiniMate settings read from escaped bytes (#1), a MiniMate start reported failed because the read-back did not survive a dropped modem line (#9), and MiniMate downloads stored without the file reader (#8).
|
||||
|
||||
**Architecture:** Three independent fixes on `feat/series4-sfm-parity` (merged up with `dev` at `342c696`), each behind its own tests on recorded BE9441 bytes. #1 changes only the four anchor-relative fields of the live `SUB 1A` decode; #9 adds one guarded reconnect to the Series-3 start read-back; #8 stores Series-3 downloads through the same Blastware-file import path production uses (`WaveformStore.save_imported_bw`), the way Series 4 already uses `save_imported_idf`.
|
||||
|
||||
**Tech Stack:** Python 3.10, pytest; `minimateplus` (client, framing, blastware_file), `sfm` (device_access, event_ingest, waveform_store).
|
||||
|
||||
**Spec:** `docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md` → *Left open by the 1c bench (2026-10-29)*, items 1, 8, 9 (and 15, retracted below).
|
||||
|
||||
## Global Constraints
|
||||
|
||||
- Branch `feat/series4-sfm-parity` only; push each commit; never `main`; no `CHANGELOG.md` edits (feature branch).
|
||||
- **Reliability before speed** (Brian, 2026-10-29): no change removes a check; a start/stop is still sent exactly once, never retried.
|
||||
- **File boundary with the S3 decode-checker session:** do not edit `minimateplus/event_file_io.py`, `histogram_codec.py`, `waveform_codec.py`, `bw_xml_report.py`, `decode_check*.py`.
|
||||
- `docs/instantel_protocol_reference.md` is CRLF: edit it with the Edit tool only.
|
||||
- Tests: `/home/serversdown/seismo-relay/.venv/bin/python -m pytest -q -p no:cacheprovider`; the baseline is exactly 16 known failures (15 `test_bw_ascii_report.py`, `test_event_file_io.py::test_save_imported_bw_without_report_works_unchanged`).
|
||||
- Store consequence (state it in the commit and docs): #8 changes how **new** live downloads are stored; existing rows and files are untouched; no backfill, no migration.
|
||||
|
||||
## Review Focus
|
||||
|
||||
1. A modem that closes the line cleanly (FIN) during the read-back shows as silence, not an `OSError`: no reconnect, the result stays "sent, not checked", nothing re-sent. (Task 2 test.)
|
||||
2. A MiniMate on a serial cable: never reconnect in the read-back — opening its port reboots the unit (measured 2026-10-25). (Task 2 test.)
|
||||
3. A settings read while the unit is monitoring returns the device-info block, not the settings block (measured 2026-10-29): mode, sample rate, interval and record time must come back `None`, not garbage. (Task 1 test.)
|
||||
4. Downloading an event twice, or one the watcher already imported, must not add a second row. (Task 3 tests.)
|
||||
5. The channel block (trigger level, range, scale) and the note strings must read exactly as before: only the four anchor fields move to unescaped bytes. (Task 1 test.)
|
||||
|
||||
## Retraction carried by this plan
|
||||
|
||||
**#15 is not a bug.** `device_commands.sent` is documented as `1 acknowledged, 0 nothing sent, NULL unknown` (`sfm/database.py:148`, `change.py:12-14`); a stop answered by a ring instead of its ack is exactly "unknown", and the row's `error` text already says the frame was sent. Task 4 retracts it in the spec.
|
||||
|
||||
---
|
||||
|
||||
### Task 1: The live settings decode reads its four anchor fields from unescaped bytes (#1)
|
||||
|
||||
**Files:**
|
||||
- Modify: `minimateplus/client.py` — `_decode_compliance_config_into` (~line 2312), its anchor block (~2385–2440)
|
||||
- Create: `tests/test_s3_compliance_anchor_fields.py`
|
||||
|
||||
**Interfaces:** Produces nothing new; `_decode_compliance_config_into(data: bytes, info: DeviceInfo) -> None` keeps its signature. `info.compliance_config` fields `recording_mode`, `sample_rate`, `histogram_interval_sec`, `record_time` change meaning only for escaped replies.
|
||||
|
||||
Measured layout once escapes are removed (`framing.unescape_reply`), from every distinct recorded reply: **mode `anchor−7` (u8), sample rate `anchor−6` (u16 BE), histogram interval `anchor−4` (u16 BE, s), record time `anchor+6` (f32 BE)**. The channel block and strings are correct on the raw bytes and must stay there (unescaped, `"Tran"+28` reads 0.0 instead of 6.2061).
|
||||
|
||||
- [ ] **Step 1: Write the failing tests** — a helper collects every `SUB 0xE5` frame from `tests/fixtures/minimate-bench-10-25-26/*_rx.bin` and `tests/fixtures/minimate-bench-10-26-26-histogram/unit_replies.bin` whose `data` contains the anchor, de-duplicated by the 18 bytes around it (`S3FrameParser().feed`).
|
||||
|
||||
```python
|
||||
@pytest.mark.parametrize("fixture,expect", [ # (mode, sample_rate, interval_s, record_time_s)
|
||||
("10-25", (1, 4096, 5, 5.0)), # Continuous
|
||||
("10-26a", (4, 2048, 2, 20.0)), # Histogram + Continuous
|
||||
("10-26b", (3, 4096, 900, 20.0)), # Blastware's 16:26 write: Histogram, 15 min
|
||||
("10-26c", (3, 4096, 2, 20.0)), # Blastware's 16:27 write: back to 2 s (tonight's unit)
|
||||
])
|
||||
def test_anchor_fields_read_from_unescaped_bytes(fixture, expect): ...
|
||||
# decode the frame's data with _decode_compliance_config_into(DeviceInfo())
|
||||
# assert (cc.recording_mode, cc.sample_rate, cc.histogram_interval_sec, round(cc.record_time, 2)) == expect
|
||||
|
||||
def test_channel_block_still_reads_from_raw_bytes():
|
||||
# data = the 10-26c frame-B data + the next E5 frame whose data contains b"Tran"
|
||||
# assert cc.trigger_level_geo == pytest.approx(0.04, abs=1e-4)
|
||||
# assert cc.geo_adc_scale == pytest.approx(6.2061, abs=1e-3) and cc.geo_range == 0
|
||||
|
||||
def test_a_reply_without_the_settings_block_reads_as_unknown():
|
||||
# the recorded SUB 0xFE (device info) reply data -- what a monitoring BE9441 returned for 1A
|
||||
# assert all four anchor fields are None
|
||||
```
|
||||
|
||||
The fixture labels are the test's own: number the de-duplicated replies in file order and map them to the four expectations above (assert there are exactly four).
|
||||
|
||||
- [ ] **Step 2: Run** `…pytest -q tests/test_s3_compliance_anchor_fields.py` → FAIL: `10-25` reads mode 0, `10-26c` reads sample rate 0 and interval 4098 (today's decoder on escaped bytes).
|
||||
- [ ] **Step 3: Implement** — in the anchor block, search `_ANCHOR` in `logical = unescape_reply(data)` (same `0:150` window) and read the four fields from `logical` at the offsets above; leave every other read on `data`. Update the layout comment to the measured logical layout and note the old "anchor−7 constant 0x10" was the escape before `04` at 1024 sps.
|
||||
- [ ] **Step 4: Run** the new tests, then the full suite → only the 16 known failures.
|
||||
- [ ] **Step 5: Commit** `fix(minimateplus): read a MiniMate's mode, sample rate, interval and record time after removing escapes (#1)`.
|
||||
|
||||
### Task 2: The Series-3 start read-back reconnects once after a dropped line (#9)
|
||||
|
||||
**Files:**
|
||||
- Modify: `sfm/device_access/s3.py` — `change_monitoring`'s `read_state` (~line 230)
|
||||
- Modify: `tests/device_access_fakes.py` — `FakeMiniMateClient`: add `read_serial()` (returns `serial_reported`, default the unit's serial; records `"read_serial"`) and `wake_errors=` (exceptions raised by successive `wake()` calls)
|
||||
- Test: `tests/test_device_access_change.py`
|
||||
|
||||
**Interfaces:** Consumes `MiniMateClient.read_serial()` (exists), `normalise_serial` (`sfm/device_access/series.py`). No new public names.
|
||||
|
||||
Behaviour: in `read_state`, an `OSError` from `_woken_state(client)` (Broken pipe / reset / aborted — measured: `[Errno 32] Broken pipe` at BE9441's first wake after the sensor check) **on a TCP address, the first time only**: log a warning "the line dropped during the read-back — reconnecting to read the state (nothing is re-sent)", `client.close()`, `client.open()`, require `normalise_serial(client.read_serial()) == serial` (else raise `WrongUnit`), then `_woken_state(client)`. A second `OSError`, any other error, or a serial-cable address propagates as today.
|
||||
|
||||
- [ ] **Step 1: Write the failing tests**
|
||||
|
||||
```python
|
||||
def test_series3_start_read_back_reconnects_after_a_dropped_line():
|
||||
fake = FakeMiniMateClient(monitor_states=[False, True], wake_errors=[BrokenPipeError(32, "Broken pipe")])
|
||||
r = access3(fake).start_monitoring(TCP3, "BE18193")
|
||||
assert (r.outcome, r.verified, r.state_after) == ("sent", "ok", "monitoring")
|
||||
assert fake.writes == ["start_monitoring"] and fake.opened == 2 and "read_serial" in fake.calls
|
||||
|
||||
def test_series3_read_back_reconnects_only_once():
|
||||
fake = FakeMiniMateClient(monitor_states=[False], wake_errors=[BrokenPipeError(32, "x")] * 2)
|
||||
with pytest.raises(ChangeUnconfirmed): access3(fake).start_monitoring(TCP3, "BE18193")
|
||||
assert fake.writes == ["start_monitoring"] and fake.opened == 2
|
||||
|
||||
def test_series3_read_back_reconnect_refuses_a_different_unit():
|
||||
fake = FakeMiniMateClient(monitor_states=[False], wake_errors=[BrokenPipeError(32, "x")], serial_reported="BE11529")
|
||||
with pytest.raises(ChangeUnconfirmed) as e: access3(fake).start_monitoring(TCP3, "BE18193")
|
||||
assert e.value.http_status == 409 and fake.writes == ["start_monitoring"]
|
||||
|
||||
def test_series3_read_back_does_not_reconnect_on_a_serial_cable():
|
||||
# address with port=, no host: opening a MiniMate's port reboots it
|
||||
... assert fake.opened == 1 and raises ChangeUnconfirmed
|
||||
|
||||
def test_series3_read_back_silence_is_not_a_dropped_line():
|
||||
fake = FakeMiniMateClient(monitor_states=[False], wake_errors=[NoAnswer("quiet")])
|
||||
... raises ChangeUnconfirmed; fake.opened == 1; writes == ["start_monitoring"]
|
||||
```
|
||||
|
||||
(`serial_reported` applies to `read_serial()` only; `connect()` keeps returning the real serial so the pre-send identity check passes. Use the module's existing `access3`, `TCP3` helpers; add a serial-cable `UnitAddress` beside `TCP3`.)
|
||||
|
||||
- [ ] **Step 2: Run** them → the reconnect tests FAIL (no reconnect today); the last two may already pass — keep them as guards.
|
||||
- [ ] **Step 3: Implement** `read_state` as described (a closed-over `reconnected` flag).
|
||||
- [ ] **Step 4: Run** `tests/test_device_access_change.py`, then the full suite → only the 16 known failures.
|
||||
- [ ] **Step 5: Commit** `fix(sfm): a MiniMate start's read-back reconnects once when the modem dropped the line (#9)`.
|
||||
|
||||
### Task 3: MiniMate downloads are stored through the Blastware-file import path (#8)
|
||||
|
||||
**Files:**
|
||||
- Modify: `minimateplus/blastware_file.py` — add `blastware_file_bytes(event: Event, a5_frames: list) -> bytes`; `write_blastware_file` writes its result (no behaviour change)
|
||||
- Modify: `sfm/device_access/models.py` — `DownloadedEvent.bw_bytes: Optional[bytes] = field(default=None, repr=False)` (Series 3: the assembled Blastware file)
|
||||
- Modify: `sfm/device_access/s3.py` — `download_event` fills `bw_bytes` from the verified event's frames
|
||||
- Modify: `sfm/waveform_store.py` — `save_imported_bw(..., source_kind: str = "bw-import")` writes it into the sidecar's `source.kind`
|
||||
- Modify: `sfm/event_ingest.py` — `persist_download`'s Series-3 branch
|
||||
- Test: `tests/test_event_ingest.py`, `tests/test_5a_blastware_walk_histogram.py`
|
||||
|
||||
**Interfaces:** Produces `blastware_file_bytes(event, a5_frames) -> bytes`; `DownloadedEvent.bw_bytes`; `save_imported_bw(bw_bytes, source_path, *, serial_hint=None, bw_report_text=None, source_kind="bw-import") -> tuple[Event, dict]`.
|
||||
|
||||
Behaviour: Series 3 mirrors Series 4 — `ev, rec = store.save_imported_bw(dl.bw_bytes, source_path=Path(dl.filename), serial_hint=dl.serial, source_kind="sfm-live")`, then the existing `db.insert_events(...)`. No `backfill_series3_event` on this path (the file's setup block supplies sample rate and record time, #44/#45); no overlay of the live `0C` peaks (data comes from the binary — they stay the download *check*). A download with no `bw_bytes` raises `IngestError("…the download carried no event file; nothing was stored")`. The "Waveform" label on a live histogram stays until the checker session's #41 (record type from `STRT[18]`) lands on `dev`.
|
||||
|
||||
- [ ] **Step 1: Write the failing tests**
|
||||
|
||||
```python
|
||||
# tests/test_5a_blastware_walk_histogram.py, beside the existing K441LTON.900 comparison:
|
||||
def test_blastware_file_bytes_is_blastwares_file():
|
||||
# the walk's frames for key 0111f210 -> blastware_file_bytes(ev, frames) == (FIX/"blastware"/"K441LTON.900").read_bytes()
|
||||
|
||||
# tests/test_event_ingest.py
|
||||
def test_a_series3_histogram_download_is_stored_from_its_file(tmp_path):
|
||||
bw = (HIST_FIX / "K441LTON.900").read_bytes() # byte-identical to SFM's download (sha f10f39a8…)
|
||||
out = persist_download(DownloadedEvent(event_id="BE9441:0111f210:20261026170036", serial="BE9441",
|
||||
series=3, filename="K441LTON.900", verified=None, bw_bytes=bw,
|
||||
timestamp=datetime.datetime(2026, 10, 26, 17, 0, 36)), store=WaveformStore(tmp_path / "w"), db=SeismoDb(tmp_path / "s.db"))
|
||||
# assert the stored file's sha256 == hashlib.sha256(bw).hexdigest()
|
||||
# assert plot_json_from_hdf5(<h5>)["channels"]["Tran"]["values"] (non-empty: the histogram decoded)
|
||||
# assert the sidecar's source.kind == "sfm-live" and the row's sample_rate == 4096
|
||||
|
||||
def test_the_small_mic_off_histogram_is_stored_with_samples(tmp_path): # K441LTOK.QZ0, refused on 10-29
|
||||
def test_a_repeat_series3_download_adds_no_row(tmp_path): # persist twice: one row; second inserted False
|
||||
def test_a_series3_download_with_no_file_stores_nothing_and_says_so(tmp_path) # replaces the no-event test
|
||||
```
|
||||
|
||||
Rewrite `test_a_series3_event_is_backfilled_then_stored_as_series3` to the new contract (bytes in, `save_imported_bw` called with `source_kind="sfm-live"`, no backfill); keep `test_downloading_an_event_the_watcher_stored_writes_nothing_series3` passing unchanged.
|
||||
|
||||
- [ ] **Step 2: Run** them → FAIL (`bw_bytes` / `blastware_file_bytes` / `source_kind` do not exist).
|
||||
- [ ] **Step 3: Implement** the five file changes above.
|
||||
- [ ] **Step 4: Run** `tests/test_event_ingest.py tests/test_5a_blastware_walk_histogram.py tests/test_device_access_events.py tests/test_server_device_access.py`, then the full suite → only the 16 known failures.
|
||||
- [ ] **Step 5: Commit** `fix(sfm): store a MiniMate download through the Blastware-file import path, as production does (#8)` — body states: new live downloads only; existing rows untouched; no backfill or migration.
|
||||
|
||||
### Task 4: Docs
|
||||
|
||||
**Files:** the spec (*Left open by the 1c bench*), `docs/instantel_protocol_reference.md` (CRLF — Edit tool), `CLAUDE.md` ("Where things stand").
|
||||
|
||||
- [ ] **Step 1:** Spec: mark #1, #8, #9 fixed (commit ids); **retract #15 in place** (by design: `sent` NULL = sent without acknowledgement); #2 fixed by the `dev` merge (`342c696`).
|
||||
- [ ] **Step 2:** Instantel reference, a dated change-log row + §7.6.1 note: the measured logical `1A` layout (mode −7, rate −6, interval −4, record time +6 after unescaping; the old "anchor−7 constant 0x10" was the escape before `04`); while monitoring, BE9441 answered the `1A` reads with its device-info block (measured 2026-10-29); the channel block is read on raw bytes, so a trigger/alarm float containing `02/03/04/10` would read shifted — open, unmeasured.
|
||||
- [ ] **Step 3:** CLAUDE.md: one line under the 1c entry — Batch 1 done (#1, #8, #9), #15 retracted.
|
||||
- [ ] **Step 4: Commit** `docs: 1c bench Batch 1 -- #1, #8, #9 fixed; #15 retracted; the 1A layout after unescaping`.
|
||||
Reference in new issue
Block a user