fix(minimateplus): the 0x1C status reads a histogram run as monitoring, and unescapes before counting from the end

Measured on BE9441 tonight, through the recording bridge:

- Monitoring flag 0x0B while recording a histogram (confirmed at the unit), 0x00
  idle.  Only 0x10 had been measured as monitoring (BE11529, 2026-04-08), so the
  status showed 'not monitoring' and the change path refused the state as
  unknown(0x0B) -- refused, not skipped, which is the design working.  0x0B joins
  S3_RUN_STATES and S3_MONITORING_FLAGS as measured.  The shared status's
  'monitoring' now follows the change path's reading: None for an unmeasured
  flag, never a guessed False.
- Free memory 0x000D0E04 arrived as 0d 0e 10 04.  The parser keeps the 10 of a
  10 02/03/04 escape, so every field counted from the end read one byte off:
  419.84 V and 251,655,168 bytes for 6.76 V and 983,028.  New
  framing.unescape_reply() (also used by parse_strt_end_key) runs first.

Tests: the two recorded replies (idle 16:24, monitoring 18:47) decode to the
unit's own values; each escaped value in the last ten bytes decodes; the status
mapping for 00 / 10 / 0B / unmeasured.  No codec, store or DB change.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-10-26 18:53:11 -04:00
1 parent 8acb9512a9
commit d990466407
5 files changed
+114 -16

No files matched your search

+15 -2
View File
@@ -34,7 +34,7 @@ import re
import struct
from typing import Optional
from .framing import S3Frame
from .framing import S3Frame, unescape_reply
from .models import (
CallHomeConfig,
ComplianceConfig,
@@ -60,6 +60,11 @@ from .transport import SerialTransport, BaseTransport
log = logging.getLogger(__name__)
# SUB 0x1C state-flag values measured while a unit was monitoring: 0x10 (BE11529,
# 2026-04-08, 98 frames) and 0x0B (BE9441 recording a histogram, 2026-10-26).
# Idle is 0x00 on both.
S3_MONITORING_FLAGS = frozenset({0x10, 0x0B})
# ── Module-level constants ────────────────────────────────────────────────────
# Trigger config payload hardcoded from 3-11-26 BW TX capture (BW frame 108).
@@ -2682,6 +2687,10 @@ def _decode_monitor_status(data: bytes) -> MonitorStatus:
mem_total 0x000EFFF2 = 983026 bytes ≈ 960 KB
mem_free 0x000E9E52 = 958034 bytes ≈ 935 KB
"""
# Unescape first (2026-10-26): BE9441's free memory 0x000D0E04 arrived as
# `0d 0e 10 04`, and with the parser's kept 10 every field counted from the
# end read one byte off -- 419.84 V and 251,655,168 bytes for 6.76 V and 983,028.
data = unescape_reply(data)
# The data section starts at offset 11 (after the S3 section header).
section = data[11:] if len(data) > 11 else data
@@ -2694,7 +2703,11 @@ def _decode_monitor_status(data: bytes) -> MonitorStatus:
# Confirmed from byte diff of all 144 0xE3 data frames in 4-8-26/2ndtry capture:
# section[1] = 0x00 in all IDLE frames, 0x10 in all MONITORING frames.
# (section[6] also changes but has non-binary values 0xea/0x07 — device-specific.)
is_monitoring = len(section) > 1 and section[1] == 0x10
#
# 2026-10-26: 0x0B on BE9441 recording a HISTOGRAM (confirmed at the unit);
# 0x00 on the same unit idle. Only measured values count as monitoring --
# sfm.device_access.s3.S3_RUN_STATES refuses any other for a change.
is_monitoring = len(section) > 1 and section[1] in S3_MONITORING_FLAGS
# The same byte, raw -- but only from a real section: for data of 11 bytes
# or fewer, `section` above falls back to the header itself.
state_flag = data[12] if len(data) > 12 else None
+21 -10
View File
@@ -600,20 +600,31 @@ def bulk_waveform_term_v2(
STRT_MARKER = b"STRT"
def unescape_reply(data: bytes) -> bytes:
"""A reply's data with each kept escape removed: `10 02/03/04` → `02/03/04`.
The unit escapes 02, 03, 04 and 10 in its replies exactly as requests do
(748/748 replies, 2026-10-26). S3FrameParser collapses `10 10` but keeps the
`10` of `10 02/03/04` (inner-frame markers to it), so its `.data` carries an
extra byte before each 02/03/04 value -- harmless for fields read before the
escape, one byte off for any read after it, or counted from the end."""
out, i = bytearray(), 0
while i < len(data):
if data[i] == DLE and i + 1 < len(data) and data[i + 1] in (STX, ETX, 0x04):
i += 1
continue
out.append(data[i])
i += 1
return bytes(out)
def parse_strt_end_key(a5_data: bytes) -> Optional[int]:
"""The event's FULL 32-bit end key from the STRT record in an A5 reply's
data, or None. The page byte is part of it: an event that ends in the next
64 KB page reads e.g. 0x011216C4 here, where `parse_strt_end_offset` gave
0x16C4. Read after removing the reply's inner DLE pairs (10 02 / 10 03 /
10 04 → 02 / 03 / 04), which the parser keeps as two bytes."""
out, i = bytearray(), 0
while i < len(a5_data):
if a5_data[i] == DLE and i + 1 < len(a5_data) and a5_data[i + 1] in (STX, ETX, 0x04):
i += 1
continue
out.append(a5_data[i])
i += 1
pos = bytes(out).find(STRT_MARKER)
0x16C4. Read after `unescape_reply`."""
out = unescape_reply(a5_data)
pos = out.find(STRT_MARKER)
if pos < 0 or pos + 10 > len(out):
return None
return int.from_bytes(out[pos + 6:pos + 10], "big")
+7 -3
View File
@@ -27,8 +27,10 @@ S3_MODES = {0x00: "Single Shot", 0x01: "Continuous", 0x03: "Histogram", 0x04: "H
S3_EVENT_TIMEOUT = 120.0 # what the legacy event-download routes use
# The 0x1C flag byte (section[1]): the only two values ever measured.
S3_RUN_STATES = {0x00: "idle", 0x10: "monitoring"}
# The 0x1C flag byte (section[1]): the only values ever measured. 0x10 is
# BE11529 monitoring (2026-04-08); 0x0B is BE9441 recording a histogram,
# confirmed at the unit (2026-10-26). Anything else is refused for a change.
S3_RUN_STATES = {0x00: "idle", 0x10: "monitoring", 0x0B: "monitoring"}
def s3_run_state(status: Any) -> str:
@@ -263,8 +265,10 @@ class Series3Adapter:
def op(client: Any) -> UnitStatus:
client.poll()
ms = client.get_monitor_status()
# The change path's reading, not a guess: an unmeasured flag is None.
monitoring = {"monitoring": True, "idle": False}.get(s3_run_state(ms))
return UnitStatus(
serial=serial, series=3, monitoring=ms.is_monitoring,
serial=serial, series=3, monitoring=monitoring,
battery_volts=round(ms.battery_v, 2) if ms.battery_v is not None else None,
memory_total_bytes=ms.memory_total, memory_free_bytes=ms.memory_free,
device_time=None, event_count=None,
+8 -1
View File
@@ -686,10 +686,17 @@ def test_series3_state_names_come_from_the_raw_flag(flag, state):
assert r.state_before == state
@pytest.mark.parametrize("flag,monitoring", [(0x00, False), (0x10, True), (0x0B, True), (0x07, None)])
def test_series3_status_shows_the_change_paths_reading(flag, monitoring):
# 0x0B: BE9441 recording a histogram (2026-10-26) once showed "not monitoring"
st = access3(FakeMiniMateClient(monitor_states=[flag])).status(TCP3, "BE18193")
assert st.monitoring is monitoring
def test_series3_monitor_status_decode_keeps_the_raw_flag():
from minimateplus.client import _decode_monitor_status
tail = bytes.fromhex("02a8000effc0000e9e52") # battery + memory: the last 10 bytes
for flag, monitoring in ((0x00, False), (0x10, True), (0x07, False)):
for flag, monitoring in ((0x00, False), (0x10, True), (0x0B, True), (0x07, False)):
st = _decode_monitor_status(bytes(11) + bytes([0x00, flag]) + bytes(34) + tail)
assert (st.state_flag, st.is_monitoring) == (flag, monitoring)
assert _decode_monitor_status(bytes(12)).state_flag is None # no section[1]
+63
View File
@@ -0,0 +1,63 @@
"""SUB 0x1C monitor status, decoded from BE9441's own replies (2026-10-26).
Recorded by the bridge (bridges/captures/10-26-26 - BE9441 blastware histogram/):
* 16:24:54, idle, before the histogram run: flag 0x00.
* 18:47:41, during a histogram run Brian confirmed at the unit: flag **0x0B**. Only
0x10 had ever been measured as "monitoring" (BE11529, 2026-04-08), so the status
read "not monitoring" and the change path refused the state as unknown.
* The same 18:47 reply's free memory, 0x000D0E04, holds a 04. The unit escapes it
(10 04) and the parser keeps both bytes, so the fields counted from the end came
out one byte off: 419.84 V and 251,655,168 bytes, against 6.76 V and 983,028.
"""
from __future__ import annotations
import os
import sys
import pytest
sys.path.insert(0, os.path.dirname(os.path.dirname(os.path.abspath(__file__))))
from minimateplus.client import _decode_monitor_status # noqa: E402
from minimateplus.framing import S3FrameParser # noqa: E402
from sfm.device_access.s3 import s3_run_state # noqa: E402
IDLE_1624 = bytes.fromhex(
"1002001010e300002c0000000000000000000000001a0a07ea0010101836000000000000010107cb00070000010107cb0013000000001002a4000efff4000e0de4fa03")
MONITORING_1847 = bytes.fromhex(
"1002001010e300002c00000000000000000000000b1a0a07ea00122f29000000000000010107cb00070000010107cb0013000000001002a4000efff4000d0e10043103")
def _status(wire: bytes):
frames = S3FrameParser().feed(wire)
assert len(frames) == 1 and frames[0].sub == 0xE3
return _decode_monitor_status(frames[0].data)
def test_idle_reply():
st = _status(IDLE_1624)
assert (st.state_flag, st.is_monitoring, s3_run_state(st)) == (0x00, False, "idle")
assert (st.battery_v, st.memory_total, st.memory_free) == (6.76, 983028, 921060)
def test_a_histogram_run_reads_monitoring():
st = _status(MONITORING_1847)
assert (st.state_flag, st.is_monitoring, s3_run_state(st)) == (0x0B, True, "monitoring")
def test_an_escaped_byte_in_the_memory_field_does_not_shift_the_fields():
st = _status(MONITORING_1847)
assert (st.battery_v, st.memory_total, st.memory_free) == (6.76, 983028, 855556)
@pytest.mark.parametrize("value", [0x02, 0x03, 0x04, 0x10])
def test_every_escaped_value_in_the_last_ten_bytes_decodes(value):
# free memory ending in each escaped value, encoded as the unit does
st0 = _status(IDLE_1624)
# the reply's last data byte (free memory's low byte, 0xE4) swapped for `value`
assert IDLE_1624[-3] == 0xE4
frames = S3FrameParser().feed(IDLE_1624[:-3] + bytes([0x10, value]) + IDLE_1624[-2:])
st = _decode_monitor_status(frames[0].data)
assert st.memory_free == (st0.memory_free & ~0xFF) | value
assert (st.battery_v, st.memory_total) == (6.76, 983028)