fix(minimateplus): the 0x1C status reads a histogram run as monitoring, and unescapes before counting from the end

Measured on BE9441 tonight, through the recording bridge:

- Monitoring flag 0x0B while recording a histogram (confirmed at the unit), 0x00
  idle.  Only 0x10 had been measured as monitoring (BE11529, 2026-04-08), so the
  status showed 'not monitoring' and the change path refused the state as
  unknown(0x0B) -- refused, not skipped, which is the design working.  0x0B joins
  S3_RUN_STATES and S3_MONITORING_FLAGS as measured.  The shared status's
  'monitoring' now follows the change path's reading: None for an unmeasured
  flag, never a guessed False.
- Free memory 0x000D0E04 arrived as 0d 0e 10 04.  The parser keeps the 10 of a
  10 02/03/04 escape, so every field counted from the end read one byte off:
  419.84 V and 251,655,168 bytes for 6.76 V and 983,028.  New
  framing.unescape_reply() (also used by parse_strt_end_key) runs first.

Tests: the two recorded replies (idle 16:24, monitoring 18:47) decode to the
unit's own values; each escaped value in the last ten bytes decodes; the status
mapping for 00 / 10 / 0B / unmeasured.  No codec, store or DB change.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-10-26 18:53:11 -04:00
1 parent 8acb9512a9
commit d990466407
5 files changed
+114 -16

No files matched your search

+7 -3
View File
@@ -27,8 +27,10 @@ S3_MODES = {0x00: "Single Shot", 0x01: "Continuous", 0x03: "Histogram", 0x04: "H
S3_EVENT_TIMEOUT = 120.0 # what the legacy event-download routes use
# The 0x1C flag byte (section[1]): the only two values ever measured.
S3_RUN_STATES = {0x00: "idle", 0x10: "monitoring"}
# The 0x1C flag byte (section[1]): the only values ever measured. 0x10 is
# BE11529 monitoring (2026-04-08); 0x0B is BE9441 recording a histogram,
# confirmed at the unit (2026-10-26). Anything else is refused for a change.
S3_RUN_STATES = {0x00: "idle", 0x10: "monitoring", 0x0B: "monitoring"}
def s3_run_state(status: Any) -> str:
@@ -263,8 +265,10 @@ class Series3Adapter:
def op(client: Any) -> UnitStatus:
client.poll()
ms = client.get_monitor_status()
# The change path's reading, not a guess: an unmeasured flag is None.
monitoring = {"monitoring": True, "idle": False}.get(s3_run_state(ms))
return UnitStatus(
serial=serial, series=3, monitoring=ms.is_monitoring,
serial=serial, series=3, monitoring=monitoring,
battery_volts=round(ms.battery_v, 2) if ms.battery_v is not None else None,
memory_total_bytes=ms.memory_total, memory_free_bytes=ms.memory_free,
device_time=None, event_count=None,