From the 2026-11-04 bench:
- A MiniMate settings preview warns that a change is lost if someone is in
the unit's setup menu: a write sent with the menu open was acknowledged
and read back, then lost on Cancel or Save -- the menu writes its own
copy of the whole block back (measured both ways on BE9441).
- The two bytes after the MiniMate settings block went out as 00 00 and
the unit took them; the preview note says so instead of 'unmeasured'.
- A Micromate call-home plan no longer warns about 0x00 / 0x01: UM20147
took both, each way, on 11-03 and 11-04, matching its menu.
- read_waveform_data_raw drops the escaped-checksum's stray 0x10, as the
settings read does (it failed closed: the write's block check refused
203 bytes).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Measured at the bench 2026-11-03 (UM20147): monitoring, a Micromate returns its
real call-home block and ACKS a call-home write, but does not apply it -- three
read-backs unchanged, and its own menu says 'Feature disabled while monitoring'.
SFM now refuses up front (UnitStateUnknown, 409: '... is monitoring -- a Micromate
ignores call-home changes while monitoring ...; stop monitoring first -- nothing
sent'), dry run included, instead of sending a write the unit ignores.
And the block must be the call-home block: micromate.writes.is_call_home_content
checks its header is its own length (00 7C, N = 124 -- 00 7C DC on every read,
UM12947 and UM20147). plan_call_home_enabled refuses any other 126 bytes, the
change refuses before sending, and call_home_state says why (enabled = None).
At the bench this was checked by eye before the monitoring write.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44