fix(sfm): a Micromate's call-home is refused while monitoring, and only a real call-home block is ever edited

Measured at the bench 2026-11-03 (UM20147): monitoring, a Micromate returns its
real call-home block and ACKS a call-home write, but does not apply it -- three
read-backs unchanged, and its own menu says 'Feature disabled while monitoring'.
SFM now refuses up front (UnitStateUnknown, 409: '... is monitoring -- a Micromate
ignores call-home changes while monitoring ...; stop monitoring first -- nothing
sent'), dry run included, instead of sending a write the unit ignores.

And the block must be the call-home block: micromate.writes.is_call_home_content
checks its header is its own length (00 7C, N = 124 -- 00 7C DC on every read,
UM12947 and UM20147).  plan_call_home_enabled refuses any other 126 bytes, the
change refuses before sending, and call_home_state says why (enabled = None).
At the bench this was checked by eye before the monitoring write.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
serversdownandClaude Opus 5.5 committed 2026-11-03 15:32:02 -05:00
1 parent 5326e99ac2
commit f35845d403
4 files changed
+74 -4

No files matched your search

+10
View File
@@ -101,6 +101,13 @@ def plan_stop_monitoring() -> WritePlan:
return WritePlan("stop monitoring", (WriteStep(0x97, _CONFIRM, what="stop monitoring"),))
def is_call_home_content(block: bytes) -> bool:
"""Whether 126 bytes are the 0x2C call-home content: its header is its own
length (00 7C, N = 124 -- 00 7C DC on every read on record, UM12947 and
UM20147). A bit is flipped only in a block that carries it."""
return len(block) == _CALL_HOME_LEN and int.from_bytes(block[0:2], "big") == _CALL_HOME_LEN - 2
def plan_call_home_enabled(call_home: bytes, enabled: bool) -> WritePlan:
"""Read `0x2C`, flip bit 0 of content[5], send the rest back byte for byte.
@@ -113,6 +120,9 @@ def plan_call_home_enabled(call_home: bytes, enabled: bool) -> WritePlan:
if len(call_home) != _CALL_HOME_LEN:
raise ValueError(f"call-home content must be {_CALL_HOME_LEN} B "
f"(the size THOR's offset was seen at), got {len(call_home)}")
if not is_call_home_content(call_home):
raise ValueError(f"call-home content's header is {call_home[0:2].hex()}, not its own length "
f"(00 {_CALL_HOME_LEN - 2:02x}) -- not the call-home block; nothing built")
b5 = (call_home[5] | 0x01) if enabled else (call_home[5] & 0xFE)
data = call_home[:5] + bytes([b5]) + call_home[6:]
what = "call-home " + ("enable" if enabled else "disable")
+23 -4
View File
@@ -19,12 +19,14 @@ from typing import Any, Callable, Optional, TypeVar
from micromate.client import MicromateClient
from micromate.opcodes import OpcodeRefused
from micromate.protocol import THOR_CHUNK_SIZE
from micromate.writes import plan_call_home_enabled, plan_start_monitoring, plan_stop_monitoring
from micromate.writes import (
is_call_home_content, plan_call_home_enabled, plan_start_monitoring, plan_stop_monitoring,
)
from minimateplus.transport import BaseTransport, SerialTransport, TcpTransport
from .errors import (
AuditIncomplete, CallHomeInProgress, ChangeUnconfirmed, DeviceError, EventNotFound,
MalformedReply, StaleEvent, WrongUnit, translate,
MalformedReply, StaleEvent, UnitStateUnknown, WrongUnit, translate,
)
from .change import CALL_HOME_MESSAGE, call_home_change, monitoring_change, run_change
from .events import iso_of, make_event_id, stamp_of
@@ -383,12 +385,24 @@ def _changed_offsets(before: bytes, after: bytes) -> list:
if (before[i] if i < len(before) else None) != (after[i] if i < len(after) else None)]
# Measured 2026-11-03 (UM20147): monitoring, a Micromate returns its real call-home
# block and ACKS a write, but does not apply it -- its own menu: "Feature disabled
# while monitoring". So SFM refuses before sending one it would ignore.
S4_CALL_HOME_WHILE_MONITORING = ("is monitoring -- a Micromate ignores call-home changes while "
"monitoring (its own menu: 'Feature disabled while monitoring'); "
"stop monitoring first, then change call-home")
_NOT_CALL_HOME = "returned a block that is not its call-home block (header {}, not 007c)"
def s4_call_home_state(client: Any, serial: str) -> CallHomeState:
"""Bit 0 of content[5] is the enable (UM12947 reads 0x05 / 0x04, UM20147 0x01)."""
block = client.get_call_home_content()
detail = {"byte5": f"0x{block[5]:02X}" if len(block) > 5 else None, "block": block.hex()}
if not is_call_home_content(block):
detail["reason"] = "the unit " + _NOT_CALL_HOME.format(block[0:2].hex())
return CallHomeState(serial=serial, series=4, enabled=None, dial_string=None, detail=detail)
return CallHomeState(serial=serial, series=4, enabled=bool(block[5] & 0x01),
dial_string=_dial_string(block),
detail={"byte5": f"0x{block[5]:02X}", "block": block.hex()})
dial_string=_dial_string(block), detail=detail)
def s4_change_call_home(client: Any, serial: str, *, enabled: bool, dry_run: bool,
@@ -406,7 +420,12 @@ def s4_change_call_home(client: Any, serial: str, *, enabled: bool, dry_run: boo
run = client.read_run_state().name
if run == "calling_home" or run.startswith("unknown"):
return run # the core refuses both; a write waits for a known state
if run == "monitoring" and not reads:
raise UnitStateUnknown(f"{serial} {S4_CALL_HOME_WHILE_MONITORING} -- nothing sent")
block = client.get_call_home_content()
if not is_call_home_content(block):
raise UnitStateUnknown(f"{serial} " + _NOT_CALL_HOME.format(block[0:2].hex())
+ " -- nothing sent")
reads.append(block)
return "enabled" if block[5] & 0x01 else "disabled"
+34
View File
@@ -114,3 +114,37 @@ def test_a_lost_ack_still_records_what_the_unit_holds():
assert (r.outcome, r.sent) == ("failed", True)
snap = r.detail["snapshot"]
assert snap["changed_offsets"] == [] and snap["before"] == snap["after"] # the write did not land
# Measured at the bench 2026-11-03 (UM20147): a MONITORING Micromate returns its
# real call-home block and ACKS a write, but does not apply it -- its own menu says
# "Feature disabled while monitoring". SFM refuses up front and says why, rather
# than send a write the unit will ignore.
@pytest.mark.parametrize("dry_run", [True, False])
def test_s4_call_home_is_refused_while_monitoring(dry_run):
from sfm.device_access.errors import UnitStateUnknown
fake = FakeMicromateClient(run_states=[0x02])
with pytest.raises(UnitStateUnknown) as e:
change(fake, enabled=False, dry_run=dry_run)
msg = str(e.value)
assert "monitoring" in msg and "stop" in msg and "nothing sent" in msg
assert fake.executed == []
# The block's header is its own length (00 7C, N = 124); SFM flips a bit only in a
# block that carries it, never in whatever 126 bytes came back.
def test_s4_a_block_that_is_not_the_call_home_block_sends_nothing():
from sfm.device_access.errors import UnitStateUnknown
fake = FakeMicromateClient(call_home=bytes(126))
with pytest.raises(UnitStateUnknown) as e:
change(fake, enabled=False)
assert "not its call-home block" in str(e.value)
assert fake.executed == []
def test_s4_state_of_a_block_that_is_not_the_call_home_block_says_why():
st = s4_call_home_state(FakeMicromateClient(call_home=bytes(126)), "UM20147")
assert st.enabled is None and "not its call-home block" in st.detail["reason"]
+7
View File
@@ -265,3 +265,10 @@ def test_setup_push_refuses_a_block_whose_own_length_header_is_wrong():
bad[0:2] = (int.from_bytes(bad[0:2], "big") + 1).to_bytes(2, "big")
with pytest.raises(ValueError, match="header"):
plan_setup_push(reads[0x08], reads[0x2E], bytes(bad), "x.mmb")
def test_call_home_refuses_a_block_whose_header_is_not_its_length():
# 126 bytes, but not the call-home block: its header must be its own length (00 7C).
with pytest.raises(ValueError, match="header"):
plan_call_home_enabled(bytes(126), True)