GET /device/monitor/status?serial=UM...&run_state=true adds
detail.run_state = {code, name}. For the ring-path bench test: the code
during and after a call-home (0x01 for ~18 s after a stop, inferred to be
call-home attempts; 0x03 = calling home) is what the test watches. Opt-in
because it is one more round trip (~0.65 s on cellular); a plain status read
is unchanged. Series 3 has no 0x49 and ignores it.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Second start through SFM, BE9441 2026-10-26 23:54:41: acked 0x69, and Brian saw
'starting, performing sensor check, recording' -- but the read-back's single
SESSION_RESET + POLL went out 2.3 s after the ack, inside the ~30 s sensor
check, and was never answered or re-sent (504 after 120 s, audit ca7b5b42).
Brian: every unit is set to sensor-check before monitoring, ~30 s, then
'Recording...'.
- MiniMateProtocol.wake(attempt_s, deadline_s): SESSION_RESET + POLL re-sent
every attempt_s until the unit answers, then the POLL data step; a named
timeout at the deadline. Reads only. MiniMateClient.wake() wraps it.
- The Series-3 change: after a START, the first read waits S3_SENSOR_CHECK_S
(30 s) before asking; every read then wakes with retries (every 5 s, up to
60 s). A stop is unchanged -- its read-back was confirmed three times tonight.
Tests: a unit deaf for 4 probes is asked a 5th time and answers; a unit that
never answers ends in the named timeout after ~6 probes; a start's first read
waits the sensor check, a stop's does not; every read is preceded by a wake.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
BE9441, 2026-10-26 23:41:48 (through the bridge): SFM's start (0x96, Blastware's
own bytes) was acked 0x69 and the unit began monitoring -- Brian saw it -- but
the read-back sent one bare 0x1C two seconds later and got nothing for 120 s,
so the change reported 'sent and acknowledged, but reading the state back
failed' (504, audit 0c1cf813). Nothing was retried.
A monitoring MiniMate ignores a bare request until SESSION_RESET + POLL
(measured 2026-04-08). The read-back had relied on the connect's wake; a stop's
read-back (18:53, the unit idle by then) never needed it. Every state read in
the change now goes through _woken_state: client.poll() (41 03, POLL, 41 03,
POLL -- the connect's own sequence), then 0x1C.
Test: after the send, every 0x1C read is preceded by a POLL. No codec, store or
DB change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Measured on BE9441 tonight, through the recording bridge:
- Monitoring flag 0x0B while recording a histogram (confirmed at the unit), 0x00
idle. Only 0x10 had been measured as monitoring (BE11529, 2026-04-08), so the
status showed 'not monitoring' and the change path refused the state as
unknown(0x0B) -- refused, not skipped, which is the design working. 0x0B joins
S3_RUN_STATES and S3_MONITORING_FLAGS as measured. The shared status's
'monitoring' now follows the change path's reading: None for an unmeasured
flag, never a guessed False.
- Free memory 0x000D0E04 arrived as 0d 0e 10 04. The parser keeps the 10 of a
10 02/03/04 escape, so every field counted from the end read one byte off:
419.84 V and 251,655,168 bytes for 6.76 V and 983,028. New
framing.unescape_reply() (also used by parse_strt_end_key) runs first.
Tests: the two recorded replies (idle 16:24, monitoring 18:47) decode to the
unit's own values; each escaped value in the last ten bytes decodes; the status
mapping for 00 / 10 / 0B / unmeasured. No codec, store or DB change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
CLAUDE.md: the KNOWN BUG entry marked fixed (887e75d) with what settled it, and
retraction notes under the 5A sections that described the misreading (partial
stuffing, offset_hi 0x10 raw, 11-byte params, metadata pages 0x1002/0x1004).
Series-III reference: a 2026-10-26 (3) row. Spec: the MiniMate check updated.
Code comments that called the page bug open, updated; the adapter's struct.error
mapping kept as a defence, its message and test no longer naming the old bug.
Verified live on BE9441 the same day: all 8 events downloaded by serial, the 4
that cross a 64 KB page included, each whole and matching the unit's own peaks.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Review of 36f4e5d (measured): with the metadata-page bug put back, 4 of 7
BE12599 non-first waveforms decoded to nothing, and the check stored them
unverified as 'a histogram' -- the worst form of the bug it was built for.
Now: no waveform decode must decode as a histogram (stored unverified) or it is
refused; a waveform's record chain must end on its END record (0x06), as the
file reader requires -- cut-short streams have none; the structure checks run
for every unit and only the peak comparison is limited to a unit confirmed in
normal range (geo_range == 0; an unread range is no longer assumed normal);
a clipped blast at full scale on both sides counts as a match. Tests pin the
one-step tolerance, both ranges, the END record and the neither case.
No DB or schema change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Series 4 downloads were checked (decoded PVS against the unit's); Series 3's
were stored unchecked (verified=None), so BE9441's two bad files went in as
successes at the bench (2026-10-25). verify_s3_download(): if the waveform
decodes, all three geophone channels must be the same length and each peak must
be the unit's own 0C peak to one 0.005 in/s step, or the download is refused
(MalformedReply, "not saved"). Unverified (stored, as before): no waveform
decode -- a histogram; 0 of ~430 fixture histograms decode as one -- or a
sensitive-range unit, whose scale is still open. The reason goes in
detail["check"].
A test puts the metadata-page bug back and the check refuses the file it
produced. Session-replay helpers move to tests/minimate_replay.py.
No codec, store or DB change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
The final fix wave's AuditIncomplete landed in _exchange's `except DeviceError`,
which keeps the line -- so on the nine ring / link-fault paths (and a refusal) a
failed audit write turned a hang-up into a held Series-4 line, blocking a THOR
call-home for up to the 5-minute idle hang-up and making "SFM hung up" false.
AuditIncomplete now carries release_line, set by every finish whose path hangs
up anyway, and _exchange honours it. Also pins the Series-3 change path's
retry=False again (the rewritten lost-ack test no longer could).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Final fix wave for phase 3 (start/stop monitoring), per the whole-phase review.
- I1 + M2: every failure after the send has started is a ChangeUnconfirmed
(audit_id, sent = "acknowledged" | "unknown", the cause's HTTP status) whose
message says the frame was sent and names the audit record -- the ring in
place of the ack, the four lost-ack follow-ups, and the three read-back
paths. Series4Session releases the line on it. Every audit finish goes
through one helper; a finish that fails is AuditIncomplete (500) and still
says what happened to the unit. The line is kept.
- I2: device_commands gains `sent` (1 acknowledged, 0 nothing sent, NULL
unknown) and `states` (before / seen / after), with an idempotent ADD COLUMN
in _migrate for a table created earlier on this branch.
- M1: a RADIO RING already buffered is caught before 0x96/0x97 goes out
(pre_send = read_unsolicited, one 10 ms transport read). That is a
CallHomeInProgress with sent=0, not ChangeUnconfirmed.
- M3: Series 3 reads the raw 0x1C flag (new MonitorStatus.state_flag); a byte
other than 0x00/0x10, or a short section, is unknown(...) and refused.
- M8: even_if_already on a unit already in the state notes that the read-back
proves nothing.
- M10: a scan pins `.protocol.execute(` to sfm/device_access/s4.py; both
scans fail closed on a file that does not parse.
- M11: docs corrected (CLAUDE.md phase-2 sentence, phase-3 bullet, REST list;
micromate/protocol.py scope and the 0x49 read_state docstring; s4.py; route
docstrings). "No command has ever been originated" is NOT retracted.
- M12: GET /db/device_commands normalises the serial.
- Minors: unused imports, S3 dry-run frames and 120 s timeout pinned, test
rename, serial-cache and short-0x49 tests, _RUN_STATES comment.
Schema: two NULLable columns on the branch-new device_commands table, added
by _migrate on open; no backfill. No codec change, TOOL_VERSION unchanged.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- Series4Session's docstring now states which errors keep the line: an
op raises a DeviceError only for a verdict on a healthy link (not
found, stale, unnameable), and _exchange keeps the line for those; any
library error means the link is suspect and hangs up. DecodeMismatch
is a library error and hangs up -- conservative. Later ops must keep
to it.
- Test: a micromate TimeoutError mid-download, on a fresh DeviceAccess
(no retry path), is NoAnswer, closes the client once and leaves no held
line. It pins existing behaviour; a mutation that keeps the line on
library errors makes it fail.
- README, event rows corrected in place: another unit's id is 422 (not
409); a stale id is 409 without downloading, both series; an event
already stored -- including by THOR's watcher -- returns its row and
changes nothing (detail.already_stored); MiniMate lists give
record_type null, on the normal timeout, and fail on a lost reply.
No code behaviour change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
minimateplus's _extract_record_type returns "Waveform" for every 0C
record layout -- it cannot tell a histogram -- so a MiniMate histogram
was listed as "waveform". The fake's "Histogram" was a value the real
client never produces, which is why no test saw it.
- _s3_ref: record_type is always None, with
detail.record_type_note = "Series 3's event record does not say
waveform or histogram; not decoded yet".
- default_s3_events(): both events "Waveform", as the real client gives.
- The shared-shape contract test allows exactly {"record_type"} to be
None for Series 3 and nothing for Series 4.
Rows SFM stores for Series 3 still carry "Waveform" from the decode (the
same as the legacy blastware_file route); parked in the spec. No codec,
store or DB change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
One lost 1F reply made the list return 1 of 3 events as a 200, and on a
download gave a false 404. The list also ran on the 120 s download
timeout, so each lost reply stalled the walk 120 s with the unit locked.
- get_events() strict mode (headers_only, or only_key): a lost 0A reply,
a lost 1F(browse) reply on the success path and on the partial-record
path all re-raise their ProtocolError; a 1F that hands back a key
already walked raises ProtocolError("event chain did not advance ...").
Legacy calls (none of the new options) behave exactly as before --
pinned by tests of the same three faults under get_events().
- Series3Adapter.list_events runs on the default timeout (10 s on TCP);
download_event keeps 120 s.
- FakeProto gains optional lost-reply injection; its default is unchanged.
The repeat-key guard sits at the loop head, which is reached only for a
non-null key -- equivalent to checking each 1F's key, in one place.
Read-only; nothing new on the wire. No codec, store or DB change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Series3Adapter.download_event compared the timestamp only after
download_event_by_key had run the whole 5A. A stale id cost a full
download of the wrong event (minutes on cellular), could surface the
64 KB-page 502 instead of 409, and a lost 0C reply was reported as "the
unit was erased".
- get_events() gains only_key_timestamp: with only_key, if the target's 0C
says another time (or none), the target takes the existing bypass --
no 1F(arm), no POLL, no 5A. A new branch only; it adds no command.
- download_event_by_key(key4, *, expected_timestamp=None) passes it on.
- The adapter converts the id's stamp to that ISO form and checks, in
order: no event -> 404; no readable 0C timestamp -> 502 MalformedReply
("could not read ... record (0C)"); a different time -> 409 StaleEvent;
no frames -> 502.
- The fake's download mirrors the real walk (no frames on a mismatch) and
records what it was asked for.
- Legacy get_events() (none of the new options) is pinned by a test of
its exact call sequence and 5A arguments, written before the change.
Read-only; nothing new on the wire. No codec, store or DB change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
THOR's watcher (and Blastware's ACH) store every event through
/db/import/idf_file and /db/import/blastware_file, with the paired report.
Storing a download of the same event afterwards overwrote that richer copy:
on Series 4 the file, .h5 and sidecar were rewritten (extensions dropped);
on Series 3 the row lost PVS, project strings and ZC frequencies and was
repointed at a second, review-less file set.
persist_download now looks up (serial, timestamp) first. A row there is
returned as it stands -- inserted false, detail.already_stored true -- with
no store call, no insert_events and no sidecar write. Never merge.
- DownloadedEvent gains `timestamp` (the unit's own event time) and
`record` (Series 4's 0x0C fields); both adapters set them.
- The Series-3 `event is None` guard now runs before backfill/save, with a
message that is true there ("nothing was stored").
- Tests build the watcher's state exactly as the import endpoints do and
prove the row dict, the sidecar bytes and the file mtimes are unchanged.
No schema change, no migration, no codec change.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
The /device guard let every non-UM serial through before its held-line
check. FastAPI drops an undeclared `serial`, so `?serial=BE18193` on a
legacy route still ran Series-3 code at an address holding a Series-4
line -- the hole the guard exists to close. Serial-aware routes go
straight to DeviceAccess, which refuses that case itself; everything
else is checked. baud is no longer parsed (it is not in the address
key, and pydantic's coercion made "38400.0" a second way past).
Also, from the same review:
- RingWatchTransport clears its frame state on every write, so a stray
STX on an idle line cannot hide the ring that answers the next request.
- Series4Session.is_open reads _client once; held_serial() calls it
without the unit lock, and a concurrent hang-up between two reads was
an AttributeError (a 500).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44