S3_ALARM_WRITE_MEASURED = True. At the bench on BE9441, with Brian's go,
the alarm went 1.0 -> 1.5 -> 1.0 in/s: each write acknowledged, read back
as sent, and shown in the unit's own setup menu. The write moved exactly
channel +39 of the three geophone blocks (the f32 at +38), where the
event files put it.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
From the whole-branch review of phase 5:
- The preview's revision now covers the changes asked for, so a live
change must be exactly what was previewed (any API caller, not only the
page). It covers what the preview showed -- the settings block (and a
Micromate's setup name) -- not the blocks a write only echoes, which are
re-read at send and sent as read (a MiniMate's trigger block byte 5
differs by session, #46).
- The units' limits hold on the block that would result: a range, rate or
mode change alone can no longer leave a held trigger, alarm or record
time over them.
- Every MiniMate block read is logical, so push_config_raw no longer
unescapes again (a job number of 4098, 10 02, was refused at Send); the
preview runs the write's own block check, so a block it would refuse is
shown as blocked.
- A MiniMate's monitoring refusal says a stop makes it call in (spec).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
s4_change_settings runs the shared change core with a read-back verdict
for its state (as asked / unchanged / differs). It reads the active
setup's name (unstripped), the preferences, trigger and setup blocks and
the schedule; patches the setup through micromate.setup_fields; sends
THOR's setup push (0xDA 68 73 82 83 71 72, now enableable) once; reads
the setup back. Refused before sending: a monitoring unit, a value the
unit does not allow (422), a stale preview (SettingsChanged, 409), and --
shown as 'blocked' in the preview -- the factory setup, a schedule that
loads another setup, an unreadable schedule, or a block SFM has not seen
(ChangeRefused, 409). s4_settings carries the editable fields.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
micromate/setup_fields.py: the fields an operator can change (record
mode, sample rate, interval, record time, geophone trigger, Series 3's
range and -- read-only until benched -- alarm, job number and its report
flag, the four notes by slot), the units' own limits from Instantel's
manuals, field-only byte patching (bit fields bit by bit; Series 3's
[26:28] as Blastware writes it; Series 4's CRC recomputed), the
read-back verdict compared by bytes, and the preview revision.
All 11 Blastware compliance writes reproduce from the block before them
(BE11529's first also shows Blastware re-rendering two mic levels
through its dB value; SFM echoes the unit's bytes); Series 4 keypad
fixtures reproduce pairwise.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44