b3aff4ed66 |
feat(micromate): get_setup() -- the setup block decoded, checksum-verified, and read off a real unit
MicromateClient.get_setup() reads SUB 0x1A once and returns a MicromateSetup: record mode, sample rate, histogram interval, pre-trigger, stop mode, record time, auto-stop, active sensors, trigger sources, the three geophone trigger levels, mic type and (linear) mic trigger, job number, scaled distance, and the four notes. MEASURED FIELDS ONLY -- everything graded inferred or unknown in the reference stays in .raw. micromate/setup_block.py holds the decode as pure functions (no I/O), plus crc_ok(): one checksum rule for 0x1A, 0x08 and 0x2E. A failed checksum raises SetupChecksumError (a ProtocolError) unless verify=False; a short block raises either way. Two things the decoder gets right that a naive one would not: - the mic comes from channel block 4 (the ACTIVE mic), never blocks 5/6, which hold the stored settings of whichever mic is NOT fitted; - distance and charge weight convert back with the unit's own 3.2808 and 2.21, so 45 lb entered on the unit reads 45.0, not 44.89. TESTS: 69 new. 15 real fixtures (2,103 B each) -- 14 UM20147 reads from the 10-17/10-19 keypad passes, chosen so every field takes at least two values, plus UM12947's read from Thor's 9-24 capture (other firmware line). Expected values are what the operator read off the unit's screen, never the decoder's own output. Fixtures force-added, per the tests/fixtures/ convention. Full suite 541 passed; the 16 pre-existing failures (test_bw_ascii_report.py 15, test_event_file_io.py 1) are unchanged. VERIFIED ON HARDWARE: bridges/mm_client_check.py now prints the decoded setup; against UM20147 every line matched the unit's screen (Brian, 10-19). Also: the scratch harness uses the library checksum and checks all three blocks; mm_client_check's stale labels fixed (0x06 content[4:8] is the likely monitor-log count, not "unexplained"; 11.0BD is no longer "the first one this code has met"). No codec, store or DB change; no backfill. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44 |
||
|
|
711277131a |
feat(micromate): a config diff harness -- and the blocks turn out to be bit-stable
The Series III config map was built by changing one setting and seeing which bytes moved. Whether that method transfers to Series IV depended on an unmeasured question: how much of a config block moves ON ITS OWN. If a block carries a clock or a counter, every diff arrives polluted and there is no way to tell the real answer from the noise in a single pair. MEASURED ON UM20147 (11.0BD), FOUR READS BACK TO BACK, NOTHING TOUCHED: 0x1A compliance 2103 B 0 bytes moved 0x2C call home 137 B 0 0x2E trigger 39 B 0 0x08 event index 101 B 0 0x01 device info 163 B 0 0x1C monitor status 59 B 2 (and those two are explained below) Every config block held every byte still, so the method transfers directly and needs no noise floor subtracted. That was the open question. scratch/mm_config_diff.py drives it: `volatile` establishes the floor, `snap` saves a labelled read of all five blocks, `diff` reports changed runs with each plausible field interpretation (u8/u16be/u32be/f32be) and reports any volatile bytes separately rather than silently hiding them. CORRECTION TO content[6] OF 0x1C, long recorded only as "NOT the hour": it is a sub-second tick, ~4020 counts/second in steps of exactly 4, wrapping at 256 -- about a 64 ms period, consistent with the low 6 bits of a millisecond counter. Consequence worth knowing: a 0x1C block is never byte-identical to itself, so any "has the status block changed?" check answers yes ~15 times a second. Compare named fields, never whole blocks. I nearly recorded that byte wrong, and the trap is cheap enough to document: sampled once per second it reads as a smooth monotonic counter at ~16/s with a 16-second wrap. Twelve consecutive samples supported it. It is aliasing -- a 64 ms counter read once a second is noise that happens to look tidy. Measure a counter back to back before trusting any rate from spaced samples. Also: the settings do not have to be changed from THOR. The unit's own keypad works, which needs no Windows VM and does not contend for the modem's single TCP session. Co-Authored-By: Claude Opus 5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01Ru8Lg9HkkYvX9VWWo65SmL |
||
|
|
423608ccd1 |
perf(micromate): 16 KB per request, not 1024 -- 14x fewer round trips
The chunk-size ceiling, measured on UM20147 with every size checked byte-for-byte against a known-good download: 1024 / 2048 / 4096 / 8192 / 16384 -> served in full 32768 / 65535 -> SILENTLY CLAMPED to 16384 The clamp is the important part: a 32,768 B request returns 16,384 B of perfectly good data and no error. Nothing in the response says it was truncated; the length is the only signal. THAT DICTATES HOW THE LOOP MUST BE WRITTEN. read_event_file() now tracks its offset by BYTES RECEIVED rather than striding by chunk index. A fixed stride would either fail on a clamp or skip the bytes it never collected; tracking what actually arrived makes a clamp cost one extra request, and makes the loop self-correcting against any short response -- precisely the failure mode that has bitten the Series III side repeatedly. Consequence of that change, recorded because it reverses an earlier decision: a short chunk is NO LONGER AN ERROR. It used to raise, on the principle that a silently short event is this codebase's recurring bug. But the device returns short legitimately, and the real protection is the offset arithmetic plus the final total-length check -- which still raises on a genuinely truncated event. The test was rewritten rather than deleted, and says why. CHUNK_SIZE = 16384. For UM20147's events: 4,796 B goes 5 requests -> 1, 30,230 B goes 30 -> 2, and 72,560 B goes 71 -> 5. Over cellular at ~0.65 s per round trip that is ~46 s -> ~3.2 s on the large one. Measured on one unit over USB, so THOR_CHUNK_SIZE = 1024 stays available and the replay tests pin it -- reproducing THOR's exact traffic is one argument away, and client.download_event()/get_event() take chunk_size for a link where large responses are not surviving. Full suite: 472 passed, 16 pre-existing failures unchanged. Co-Authored-By: Claude Opus 5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01Ru8Lg9HkkYvX9VWWo65SmL |
||
|
|
d7d72cadf8 |
feat(micromate): the event chain -- walk, records, download, and a self-check
Step 4 of docs/micromate_client_spec.md. MicromateEventRef plus list_events(), iter_events(), download_event(), get_event() and decode_error(). 14 new tests; 112 micromate tests total. The load-bearing test replays THOR's captured six-event download session through iter_events() + get_event() and asserts EVERY BYTE WE EMIT MATCHES THOR'S, in order -- 74 frames -- while decoding all six events and cross-checking each waveform's peak vector sum against the float the device computed itself. Two design decisions worth recording: 1. iter_events() EXISTS BECAUSE THOR INTERLEAVES. Its captured order is 0x93 -> 1E -> 0C -> 5A*n -> 0x93 -> 1F -> 0C -> 5A*n, downloading each event before advancing the chain. list_events() walks to the end first, which is fine for browsing but leaves the device cursor parked past the event a later download addresses. 0x5A is key-addressed so it very probably does not care -- but nothing observed says either way, so the interleaved path is the one offered for downloads, and it is the one the replay test exercises. 2. get_event(verify=True) RECOMPUTES THE PEAK VECTOR SUM from the decoded samples and compares it against the device's own 0x0C float. Two independent computations over the same samples, so a disagreement means our decode is wrong. Agreement on the bench events is 0.000%. Cheap insurance in a codebase whose decode failures have historically been silent -- unhandled block tags shorten a channel and nothing raises. It is a decode-correctness check, NOT a truncation detector: a channel cut after its peak still yields the right PVS, and the docstring says so. A test corrupts a stored peak to prove the check actually fires. MicromateEventRef.uid is SERIAL:key, because the key alone is ambiguous across units, and .filename generates THOR's name (<serial>_<YYYYMMDDHHMMSS>.IDFW/H) -- returning None rather than guessing when the record type is unknown, since read_idf_file() dispatches on exactly that suffix. Recorded as a CANDIDATE, not used: 0x06 content[0:4] looks like the EVENT COUNT -- 6 on a unit holding 6 events, zeros on an empty one, and THOR reads it BEFORE the walk then downloads exactly six events without ever reading the chain sentinel. If it holds it lets a caller decide whether to walk at all, which over cellular is the useful part. Two samples on one unit, and content[4:8] reads 9 unexplained, so list_events() still walks to the sentinel: slower by one round trip and correct on evidence rather than inference. Full suite: 465 passed, 16 pre-existing failures unchanged. Co-Authored-By: Claude Opus 5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01Ru8Lg9HkkYvX9VWWo65SmL |
||
|
|
b55e4e946a |
verify(micromate): 11.0BD confirmed on hardware -- every inference held
UM20147 read over USB by mm_client_check.py. The Thor firmware line was entirely inference until now, and two of its three differences were covered only by SYNTHESISED test frames. All three held: - flags = 0x03 identifies the line -> reported firmware_line="thor". That is only reachable if `10 03` in the flags position destuffs before indexing, since 0x03 is ETX -- so the escaped-flags case is confirmed too. - the model string is shorter -> reported "MM/ISEE/S" exactly. Anchoring the search on b"MM/" rather than a fixed span is what made this work. - the 0x1C block is 4 bytes longer with the extras TRAILING, so from-the-start offsets survive -> battery 3.55 V and a clock correct to the second. That last one is the one that mattered. Series III's from-the-end offsets would have given this unit 577.92 V, which is why mm_client_check watches for an impossible voltage: it is a free self-check on exactly the inference most likely to be wrong. Also clean: serial UM20147, 5 setups (factory.MMB -> test2.mmb), active setup test2.mmb, 15,000,000 B total and free, 21 reads / 2,165 B in. "One protocol stack drives the whole fleet regardless of firmware line" -- the headline finding of 2026-09-23 -- is now demonstrated by a working client rather than by matching response SUBs. Test and docstring claims downgraded from inference to confirmed where the hardware settled them, and left as synthesised-frame notes where it did not: the BEHAVIOUR is confirmed but raw BD bytes are still not in the repo. Added --capture DIR to mm_client_check: writes a raw_bw_*/raw_s3_* pair in the layout scratch/mm_frame_parse.py already reads, so a run on an unfamiliar unit becomes a test fixture without setting up a relay. Verified by round-tripping its own output through that parser: 28 frames, 0 bad checksums. Still not covered: a download from a BD unit (UM20147 had no events stored, so the chunk walk remains CB-only), raw BD fixture bytes, a monitoring unit, a nearly-full buffer, and the inbound call-home session. Co-Authored-By: Claude Opus 5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01Ru8Lg9HkkYvX9VWWo65SmL |
||
|
|
f09b7dcaaf |
feat(micromate): client layer -- connect, state, setups; read-only
Step 3 of docs/micromate_client_spec.md: micromate/client.py, two models in micromate/models.py, 26 offline tests. Every response constant in the tests is a real captured data section from UM12947. Field offsets were measured rather than taken from the spec, which turned up one general rule and one trap: THE RESPONSE SHAPE. Every response carries an 11-byte prefix and content starts at data[11]. One rule, every command. THE TRAP: data[0] is the content length & 0xFF, with no high byte anywhere in the prefix. It is therefore correct for every response under 256 bytes -- most of them -- and then reports 44 for a 2,092-byte setup block, 30 for a 286-byte monitor-log record, and 0 for a 1,024-byte download chunk. 182 of 251 captured responses agree with a naive read; the 69 that disagree are exactly the ones >= 256 bytes. That is the third length in this protocol read too narrow, after payload[9]-vs-payload[8:10] in the probe response. The client takes content as data[11:] and lets the frame's own length bound it -- nothing needs the declared length, since the frame already knows how long it is. Also measured: - POLL content[3] is 0x50, printable as "P", immediately before "Instantel". A generic printable-run scan therefore returns "PInstantel" -- it caught a test, not a unit. Vendor comes from a fixed offset; the model is found by searching for "MM/", which is structural rather than positional and so survives the Thor line's shorter "MM/ISEE/S". - The setup walk terminates on an EMPTY NAME, not an error: 23 responses, 22 names, factory.MMB first through TEST1.mmb last. - The 0x1C clock has an unidentified byte at content[6]; the hour is at content[7]. The protocol reference's 0x1C section already had this right and names the byte -- its one-line summary in the divergences list reads as six contiguous fields and is the version not to trust. Re-verified against three captures: 19:12:25, 19:13:34 and 01:14:05 against filenames stamped 19:12:14, 19:12:14 and 01:14:03. - Battery and memory are read FORWARD from content start, never backward from the end. This block is 4 bytes longer on the Thor line; the Series III from-the-end offsets give a 11.0BD unit 577.92 V. A test appends the four trailing bytes and asserts the forward offsets survive. connect() is narrower than the spec asked. The spec said to mirror Thor's POLL -> SERIAL -> 0x49 -> POLL "because it is known-good"; measurement showed that is Thor's connection check (3 of 8 sessions) and its fourth frame repeats its first. So connect() sends the three reads that gather something, and 0x01 is not read at all -- Thor never reads it, its layout is unmapped, and firmware_line comes free from any response's flags byte. If a unit ever refuses the next command after a cold connect, put the fourth POLL back and record it. A dead clock battery yields device_time=None rather than failing the whole state read; an unreadable active setup yields active_setup=None rather than failing connect. Both are real device states. Still verified only against 11.0CB and only over USB. The BD offsets follow from the extra bytes being trailing, which is documented but not something this code has seen. Full suite unchanged at 16 pre-existing failures; 445 passed, up 26. Co-Authored-By: Claude Opus 5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01Ru8Lg9HkkYvX9VWWo65SmL |