The same 40 events as #44 also have wrong or empty pre-trigger, record time and note fields in their .h5 and sidecars:
pretrig_samples
rectime_seconds
project / client / operator / location
MiniMate BE9441 (bw-import)
66–74 (real 1024: 0.25 s at 4096)
70 (real 5 s)
empty
Micromate UM12947 (SFM download)
0 (real 4096: 1.0 s)
0 (real 3 s)
empty
Re-checked on origin/feat/series4-sfm-parity at 465bc70:
read_blastware_file gives a pre-trigger of 74;
read_idf_file gives pre_trigger_sec, record_time_sec and every project_info field as None.
How the real values are known
Record lengths: 21,504 samples = 0.25 s + 5 s at 4096 (MiniMate); 16,384 = 1 s + 3 s (Micromate).
Pre-trigger difference: on all 20 thumps, the Micromate's record starts 0.750 s before the MiniMate's (tmi-seismo's comparison, aligned by cross-correlation). That is the difference between a 1.0 s and a 0.25 s pre-trigger.
The MiniMate Plus pre-trigger is fixed at 0.25 s (716U0101 Rev 15, specifications).
The note text is in the binaries
BE9441's files contain "TestUnit-Minimate", "Test unit 1", "Terra-Mechanics Inc. - Harrison" and "Standard Recording Setup.set".
UM12947's contain "Test Unit 2", "Location", "Operator", "Company" and "thump.MMB".
The Micromate's event list (GET /device/events?serial=UM12947) did return setup: "Test Unit 2" and sensor_location: "Location" in detail. The stored .h5 and sidecar have them empty.
A pointer
The Unreleased changelog's timestamp fix (#43) reads the MiniMate's record time as a big-endian float32 in the recording-setup config block, 30 bytes before the Standard Recording Setup marker. The same value could fill rectime_seconds for a bw-import.
Evidence
The same files as #44: tmi-seismo c0b6390, bench/data/2026-10-25_phase2_garage/unit-originals/.
## Summary
The same 40 events as #44 also have wrong or empty pre-trigger, record time and note fields in their `.h5` and sidecars:
| | `pretrig_samples` | `rectime_seconds` | project / client / operator / location |
|---|---|---|---|
| MiniMate BE9441 (bw-import) | 66–74 (real 1024: 0.25 s at 4096) | 70 (real 5 s) | empty |
| Micromate UM12947 (SFM download) | 0 (real 4096: 1.0 s) | 0 (real 3 s) | empty |
Re-checked on `origin/feat/series4-sfm-parity` at `465bc70`:
- `read_blastware_file` gives a pre-trigger of 74;
- `read_idf_file` gives `pre_trigger_sec`, `record_time_sec` and every `project_info` field as None.
## How the real values are known
- **Record lengths:** 21,504 samples = 0.25 s + 5 s at 4096 (MiniMate); 16,384 = 1 s + 3 s (Micromate).
- **Pre-trigger difference:** on all 20 thumps, the Micromate's record starts 0.750 s before the MiniMate's (tmi-seismo's comparison, aligned by cross-correlation). That is the difference between a 1.0 s and a 0.25 s pre-trigger.
- **The MiniMate Plus pre-trigger is fixed at 0.25 s** (716U0101 Rev 15, specifications).
## The note text is in the binaries
- BE9441's files contain "TestUnit-Minimate", "Test unit 1", "Terra-Mechanics Inc. - Harrison" and "Standard Recording Setup.set".
- UM12947's contain "Test Unit 2", "Location", "Operator", "Company" and "thump.MMB".
- The Micromate's event list (`GET /device/events?serial=UM12947`) did return `setup: "Test Unit 2"` and `sensor_location: "Location"` in `detail`. The stored `.h5` and sidecar have them empty.
## A pointer
The Unreleased changelog's timestamp fix (#43) reads the MiniMate's record time as a big-endian float32 in the recording-setup config block, 30 bytes before the `Standard Recording Setup` marker. The same value could fill `rectime_seconds` for a bw-import.
## Evidence
The same files as #44: tmi-seismo `c0b6390`, `bench/data/2026-10-25_phase2_garage/unit-originals/`.
Series-3 half fixed on dev (merge 42fe35f, fix b4706a3). Ships in the next release. This issue stays open for the Micromate half.
What the wrong values were (measured)
Record time "70": that byte is STRT[18], the record-type marker (0x46 for a waveform, 0x0A for a histogram), not a record time.
Pre-trigger and sample count: these came from STRT bytes [16:18] and [8:10], and neither field is either value.
Blank notes: the reader stopped at the NUL that follows each label.
Where the real values are (measured)
All of these are in the recording-setup block, at offsets from the start of the setup's name (see #44 for how the name is found):
Field
Location
Verified against Blastware's _XML.XML reports (DL2, 2016–2026)
Record time
f32 BE at name − 30
6,524 / 6,524 waveforms
Pre-trigger
f32 BE at name − 36. It is the exact value Blastware prints, −0.247 s as well as −0.25 s.
6,484 / 6,484
Sample count
the decoded count
6,523 / 6,524; T438HGTZ.B30W decodes 3 samples short, a separate codec issue
Setup name + four notes
NUL-terminated label/value pairs after the name
70,042 / 70,042
The notes are read by position, not by label. Blastware lets the operator edit the labels: Seis. Loc: vs Seis Loc:, Project: with a trailing space, Location:, General:. The ASCII parser already reads them by position.
A histogram gets a sample rate only. Blastware reports neither a record time nor a pre-trigger for one.
BE9441's bench file now reads 5.0 s, 1,024 pre-trigger samples, 21,504 samples, and its notes as in the summary above.
Not done yet
Job number: 2,997 DL2 events carry one, and SFM doesn't read it yet.
The waveform-time estimate still finds the record time by the "Standard Recording Setup" name. Moving it to the new lookup would add a check for 37% of files, and that could move event times, so it needs measuring first. No DL2 waveform uses the estimate today: all 6,538 that decode as waveforms take their time from the unit's event record.
Micromate: the read_idf_file() side of this issue stays with the parity branch.
**Series-3 half fixed on `dev`** (merge `42fe35f`, fix `b4706a3`). Ships in the next release. This issue stays open for the Micromate half.
### What the wrong values were (measured)
- **Record time "70":** that byte is `STRT[18]`, the record-type marker (`0x46` for a waveform, `0x0A` for a histogram), not a record time.
- **Pre-trigger and sample count:** these came from STRT bytes `[16:18]` and `[8:10]`, and neither field is either value.
- **Blank notes:** the reader stopped at the NUL that follows each label.
### Where the real values are (measured)
All of these are in the recording-setup block, at offsets from the start of the setup's name (see #44 for how the name is found):
| Field | Location | Verified against Blastware's `_XML.XML` reports (DL2, 2016–2026) |
|---|---|---|
| Record time | f32 BE at name − 30 | **6,524 / 6,524** waveforms |
| Pre-trigger | f32 BE at name − 36. It is the exact value Blastware prints, −0.247 s as well as −0.25 s. | **6,484 / 6,484** |
| Sample count | the decoded count | **6,523 / 6,524**; `T438HGTZ.B30W` decodes 3 samples short, a separate codec issue |
| Setup name + four notes | NUL-terminated label/value pairs after the name | **70,042 / 70,042** |
- **The notes are read by position, not by label.** Blastware lets the operator edit the labels: `Seis. Loc:` vs `Seis Loc:`, `Project: ` with a trailing space, `Location:`, `General:`. The ASCII parser already reads them by position.
- **A histogram gets a sample rate only.** Blastware reports neither a record time nor a pre-trigger for one.
- **BE9441's bench file now reads 5.0 s, 1,024 pre-trigger samples, 21,504 samples,** and its notes as in the summary above.
### Not done yet
- **Job number:** 2,997 DL2 events carry one, and SFM doesn't read it yet.
- **The waveform-time estimate** still finds the record time by the "Standard Recording Setup" name. Moving it to the new lookup would add a check for 37% of files, and that could move event times, so it needs measuring first. No DL2 waveform uses the estimate today: all 6,538 that decode as waveforms take their time from the unit's event record.
- **Micromate:** the `read_idf_file()` side of this issue stays with the parity branch.
Correction to the comment above (measured 2026-10-29):T438HGTZ.B30W is not a decoder truncation. Its Tran, Vert and MicL end 3 samples before Long, and Blastware's own _XML.XML sample table does the same: its last three rows carry only the Long cell. SFM's samples match Blastware's on every row once cells are paired with the channels still running. The setup-field harness compared Tran's length (2,301) with the table's row count (2,304). So sample count is effectively 6,524 / 6,524. Corrected in the dev changelog (b023083).
**Correction to the comment above (measured 2026-10-29):** `T438HGTZ.B30W` is **not** a decoder truncation. Its Tran, Vert and MicL end 3 samples before Long, and Blastware's own `_XML.XML` sample table does the same: its last three rows carry only the Long cell. SFM's samples match Blastware's on every row once cells are paired with the channels still running. The setup-field harness compared Tran's length (2,301) with the table's row count (2,304). So sample count is effectively 6,524 / 6,524. Corrected in the dev changelog (`b023083`).
Series-3 half shipped in v0.33.0. It reached main with release PR #56 on 2026-11-01. It was deployed to prod the same day with its backfills (backfill_event_shape.py exit 0, 26,711 rows updated, 0 cleared).
This issue stays open for the Micromate half, which belongs to feat/series4-sfm-parity.
**Series-3 half shipped in v0.33.0.** It reached `main` with release PR #56 on 2026-11-01. It was deployed to prod the same day with its backfills (`backfill_event_shape.py` exit 0, 26,711 rows updated, 0 cleared).
This issue stays open for the Micromate half, which belongs to `feat/series4-sfm-parity`.
Fixed on dev (merge 0c228fa, 2026-11-20); ships in the next release, whose dev->main PR will close this.
MiniMate half: fixed in v0.33.0 (BE9441 bench files: pre-trigger 1,024 samples, record time 5 s, notes read).
Micromate half: pre-trigger, record time and the four notes are now read from the Thor file.
The 20 UM12947 bench files store 4,096 pre-trigger samples (1.0 s), 3 s, and "Test Unit 2" / "Terra-Mechanics".
Against Thor's own reports on 2,027 corpus files: record time and pre-trigger 291 / 291 waveforms, notes 2,027 / 2,027.
Also fixed: every Micromate .h5, report-paired or not, had stored pre-trigger 0. It now carries the pre-trigger in samples. The next release's scripts/backfill_thor_events.py run regenerates the stored ones.
Fixed on dev (merge 0c228fa, 2026-11-20); ships in the next release, whose dev->main PR will close this.
- **MiniMate half:** fixed in v0.33.0 (BE9441 bench files: pre-trigger 1,024 samples, record time 5 s, notes read).
- **Micromate half:** pre-trigger, record time and the four notes are now read from the Thor file.
- The 20 UM12947 bench files store 4,096 pre-trigger samples (1.0 s), 3 s, and "Test Unit 2" / "Terra-Mechanics".
- Against Thor's own reports on 2,027 corpus files: record time and pre-trigger 291 / 291 waveforms, notes 2,027 / 2,027.
- **Also fixed:** every Micromate .h5, report-paired or not, had stored pre-trigger 0. It now carries the pre-trigger in samples. The next release's scripts/backfill_thor_events.py run regenerates the stored ones.
Blocking a user prevents them from interacting with repositories, such as opening or commenting on pull requests or issues. Learn more about blocking a user.
Summary
The same 40 events as #44 also have wrong or empty pre-trigger, record time and note fields in their
.h5and sidecars:pretrig_samplesrectime_secondsRe-checked on
origin/feat/series4-sfm-parityat465bc70:read_blastware_filegives a pre-trigger of 74;read_idf_filegivespre_trigger_sec,record_time_secand everyproject_infofield as None.How the real values are known
The note text is in the binaries
GET /device/events?serial=UM12947) did returnsetup: "Test Unit 2"andsensor_location: "Location"indetail. The stored.h5and sidecar have them empty.A pointer
The Unreleased changelog's timestamp fix (#43) reads the MiniMate's record time as a big-endian float32 in the recording-setup config block, 30 bytes before the
Standard Recording Setupmarker. The same value could fillrectime_secondsfor a bw-import.Evidence
The same files as #44: tmi-seismo
c0b6390,bench/data/2026-10-25_phase2_garage/unit-originals/.Series-3 half fixed on
dev(merge42fe35f, fixb4706a3). Ships in the next release. This issue stays open for the Micromate half.What the wrong values were (measured)
STRT[18], the record-type marker (0x46for a waveform,0x0Afor a histogram), not a record time.[16:18]and[8:10], and neither field is either value.Where the real values are (measured)
All of these are in the recording-setup block, at offsets from the start of the setup's name (see #44 for how the name is found):
_XML.XMLreports (DL2, 2016–2026)T438HGTZ.B30Wdecodes 3 samples short, a separate codec issueSeis. Loc:vsSeis Loc:,Project:with a trailing space,Location:,General:. The ASCII parser already reads them by position.Not done yet
read_idf_file()side of this issue stays with the parity branch.Correction to the comment above (measured 2026-10-29):
T438HGTZ.B30Wis not a decoder truncation. Its Tran, Vert and MicL end 3 samples before Long, and Blastware's own_XML.XMLsample table does the same: its last three rows carry only the Long cell. SFM's samples match Blastware's on every row once cells are paired with the channels still running. The setup-field harness compared Tran's length (2,301) with the table's row count (2,304). So sample count is effectively 6,524 / 6,524. Corrected in the dev changelog (b023083).Series-3 half shipped in v0.33.0. It reached
mainwith release PR #56 on 2026-11-01. It was deployed to prod the same day with its backfills (backfill_event_shape.pyexit 0, 26,711 rows updated, 0 cleared).This issue stays open for the Micromate half, which belongs to
feat/series4-sfm-parity.Fixed on dev (merge
0c228fa, 2026-11-20); ships in the next release, whose dev->main PR will close this.