> Measured: **60.5 s**. Recorded rather than deleted — the prediction was sharp
> and it was wrong.
### Modem settings live in `SysParm.cfg`, not in the setup — and that is deliberate
**Settled by experiment (2026-09-25).** The unit's *call modem* setting was
changed on the keypad from `generic` to `USB to PC`. Then the **active setup was
switched from `TEST1` to `test2` on the device**, and the setting was checked
again: **unchanged.**
So it does not travel with the setup. It is device-global, in the `SysParm.cfg`
the firmware strings already hinted at, alongside `SysPref.bMonitorScheduler`.
That closes a hypothesis this document was pursuing: the ~102 bytes a `.MMB`
**file** carries beyond the 2,090-byte config block are **not** where this lives.
Those bytes remain unexplained, but they are no longer a candidate for it.
⚠ It also means the earlier observation "the call-home block (`0x2C`) was
byte-identical across 218 samples today" **does not bear on this** — the last
`0x2C` sample was at 13:30:23 and the keypad change came around 13:35. No sample
exists on the far side of it.
#### Why the split is the right design
The operator's reading, and it is convincing: **you do not want modem settings
reachable remotely.** Get them wrong over the air and you lose the connection
you would need to put them back — the unit must then be visited physically.
That explains the shape of the protocol rather than treating it as an oddity:
| | reachable over the wire? |
|---|---|
| Call Home **enable**, dial string, retries, timings (`0x2C` / `0x7E`) | **yes** — recoverable if wrong |
| **Modem type** and the rest of `SysParm.cfg` | **no** — a wrong value is unrecoverable remotely |
`SUB 0x2C` is not an incomplete view of the modem configuration. It is the
deliberately-chosen subset that is **safe to change remotely**, and the
unreachable remainder is unreachable on purpose.
#### Design principle for SFM
There is a class of setting whose misconfiguration **destroys the channel you
would use to fix it** — modem type, baud, serial parameters, the call-home
destination itself. For those:
1. **Do not expose them for remote write**, or gate them behind an explicit
confirmation that names the risk.
2. If they must be writable, use a **commit/confirm with automatic rollback** —
apply, require the unit to call back within N minutes, revert if it does not.
3. **Always allow reading them**, so an operator can diagnose a unit they cannot
reconfigure.
Instantel chose option 1. Given the failure mode, that is defensible, and worth
copying rather than improving on.
### The connection dial sets a COUNT, not a period
Three configurations, each measured over many cycles: