Solo session while the bench was unattended. Read-only throughout.
Architecture: ColdFire/68K, big-endian, Freescale MQX RTOS -- not ARM as the
vector table first suggested. The tell is 4E 5E 4E 75 4E 56 (UNLK A6 / RTS /
LINK A6) throughout both images, plus an MQX_OK assertion.
CB vs BD: a byte diff is useless (68% of bytes differ -- separately linked
builds, everything relocated). A string-set diff is position-independent and
shows 17,128 strings shared, with almost every "unique" string being the same
message at a different source line:
CB: MONITOR[3268]: STATUS_BATTERY_LOW
BD: MONITOR[3258]: STATUS_BATTERY_LOW
Consistently 10 lines apart across five different MONITOR messages, so one
~10-line block differs in the monitor module and essentially nothing else. The
only functional string unique to either build is CITIZEN (a printer brand) in
BD. This corroborates the bench A/B from the other direction: the split is a
tiny code delta, not two protocol stacks.
The SUB dispatch is a 68K switch jump table, so byte-pattern hunting will not
isolate the write opcodes -- that needs a disassembler.
Call-home config field names recovered from the firmware's own debug dump:
Enable, DialString, Retries, SessionTimeout, WaitForConnection, WarmupTime,
PowerSave -- seven fields for the 126-byte SUB 0x2C block. SessionTimeout and
PowerSave have no Series III equivalent, and Series III's scheduled-time fields
are absent, consistent with scheduling moving into the THOR-downloaded
scheduler. AT+CSQ is present, so the firmware speaks AT to the modem directly.
All five bench events downloaded and decoded over USB: each arrived at exactly
its declared size, every channel equal length, timestamps sequential.
Two gaps recorded:
- No content-based record-type discriminator. read_idf_file() dispatches on the
.IDFH/.IDFW filename suffix, which does not exist over the wire, and the
first 64 bytes of a histogram and a waveform are byte-identical. The protocol
supplies one instead: SUB 0x0A returns 0x1E for a histogram and 0x00 for a
waveform, so the type must be carried from the chain walk.
- The 0x0C peak float runs 2-5% above max(Tran,Vert,Long) and is not the vector
sum either. Its offset was inferred from a byte marker rather than
established, so it may not be the peak at all. Marked do-not-rely-on.
Co-Authored-By: Claude Opus 5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01Ru8Lg9HkkYvX9VWWo65SmL
33 KiB
Micromate Protocol Reference — Thor / Micromate Series IV, live wire protocol
Sibling to instantel_protocol_reference.md
(Series III, "the Rosetta Stone") and
idf_protocol_reference.md (Series IV file
format). This document covers the Series IV live device protocol — what
the unit says over the wire, as opposed to what it writes into a .IDFW.
Status (2026-09-23): opening session. Everything below was established in a single bench session against one unit. Treat it as a strong start, not a settled spec — in particular, everything here comes from one unit, over USB, with no events stored.
The headline
A Micromate running the Blastware firmware answers Series III command frames.
An unmodified Series III POLL (SUB 0x5B), built by
minimateplus.framing.build_bw_frame with no changes at all, produced a
complete two-step probe/data cycle. Ten Series III read commands were then
tried and all ten answered, every one obeying the Series III response-SUB
rule.
⚠ That qualifier is load-bearing, and it was discovered after the fact. Instantel ships the Micromate in two firmware lines:
| firmware | Instantel's own description |
|---|---|
11.0CB |
"Utilize with Blastware" |
11.0BD |
"Utilize with THOR, Vision, Vision II" |
The bench unit reports 11.0CB — the Blastware build. So the clean
Series III behaviour above is very likely because the unit is in Blastware
mode, not because the Micromate natively speaks Series III. Nothing here
should be assumed to hold on a 11.0BD unit until tested.
This reframes the project. The question is no longer only "what is the Series IV protocol" but "which firmware line do we target, and does one of them let the existing Series III stack drive the whole fleet?"
Firmware — the variable nobody knew was a variable
This explains a production problem TMI has lived with: two ACH servers, two machines, and units that will not cross over. It is not a misconfiguration. Instantel ships different firmware for different host software, and the wire protocol differs with it.
Fleet audit (physical, 2026-09-22/23)
| unit | firmware | line | location |
|---|---|---|---|
| UM11719 | 11.0CB |
Blastware | Ped Bridge Loc 2 |
| UM6047 | 11.0CB |
Blastware | Brookville Loc 9 |
| UM12947 | 11.0CB |
Blastware | bench |
| UM14133 | 11.0CB |
Blastware | Pitt-Music Bldg Loc 1 |
| UM11402 | 11.0BD |
Thor | Ped Bridge Loc 1 |
| UM20147 | 11.0BD |
Thor | bench |
| UM13981 | 11.0AK |
pre-split | RKM Loc 1 |
| UM20146 | 11.0AK |
pre-split | Karns Loc 2 |
| UM12420 | 10.90GC |
pre-split | RKM Loc 2 |
4 Blastware / 2 Thor / 3 pre-split. The Blastware line is already the plurality, which makes "standardise on Blastware" less disruptive than it first appeared.
A store-derived audit (firmware is recorded in every .sfm.json sidecar as
extensions.idf_report.version) agreed with the physical audit on 7 of 9.
The two that differed — UM6047 and UM14133 — are the most recently deployed
units, reflashed after their last stored event. Useful technique: the fleet's
firmware history is reconstructable from the store without touching a unit,
but it lags reality by one deployment.
Firmware is not stable per-unit over time — five of nine have been reflashed at least once. Any fleet-wide claim needs a fresh audit.
⚠ Retraction: firmware line does NOT determine Thor compatibility
An earlier draft of this document suggested that UM12947's trouble with Thor was explained by its being on the Blastware build. That is not supported.
Ped Bridge runs UM11402 (11.0BD) and UM11719 (11.0CB) side by side, both
deployed 2026-04-20, and both call Thor successfully — UM11719 has 331
Thor-collected events in the store while on 11.0CB, through 2026-08-23.
So a Blastware-line unit does feed Thor. Whatever the CB/BD split changes, it is not "which host software can collect from it", and UM12947's specific problem remains unexplained.
What is actually established: a 11.0CB unit answers Series III command
frames. Whether a 11.0BD unit does is untested — and UM20147 (11.0BD)
is on the bench, which makes that a direct A/B away.
The version is readable over the wire from SUB 0x01 as two separate ASCII
runs — "0CB" then "11" — with no single concatenated string.
The strategic fork
- Option A — standardise the fleet on the Blastware line. Every unit,
MiniMate and Micromate alike, then speaks Series III, and the existing
minimateplus/stack drives all of it. One protocol, one call-home receiver. Dramatically cheaper if it holds up. - Option B — reverse-engineer the Thor line (
11.0BD) and support both.
Option A is the shortcut, but it is unproven and carries real unknowns, all of which are cheap to answer on the bench and expensive to discover later:
- What file format does a
11.0CBunit produce? If it emits Blastware binaries rather than.IDFW/.IDFH, the (now exact) Series III decoder applies and the IDF codec becomes a legacy path. Not a loss — we have both — but it changes what the ingest pipeline sees. - Does Thor still work with a
11.0CBunit? If not, flashing a production unit breaks data collection until the replacement path exists. - Are any Micromate-specific capabilities lost on the Blastware line?
- Is the flash reversible in the field, and what does it cost in downtime?
How this was obtained
No Thor, no modem, no Windows machine. The Micromate exposes its protocol on a USB CDC-ACM virtual serial port:
ID 2504:0300 Instantel Inc. MICROMATE COM PORT
driver: cdc_acm ATTRS{serial}=="V1.00"
→ /dev/ttyACM0
Plain CDC ACM, so no vendor driver and no proprietary USB layer — any host that can open a serial port can talk to the unit.
Baud is irrelevant over USB. Identical byte-for-byte responses at 38400 and 115200; CDC-ACM ignores the line rate. TMI provisions Micromate modem links at 115200 (Series III uses 38400) — that matters for the cellular path, not for USB.
The device never speaks first. 20 s of passive listening on an idle open port produced zero bytes. It is strictly request/response.
Physical + framing layer
Requests — Series III format, unmodified
Every frame in this session was produced by build_bw_frame(sub, offset) with
no Series IV changes, and the device accepted all of them:
[ACK 0x41] [STX 0x02] [10 10] [flags 00] [SUB] [00] [00] [offset] [params×10] [chk] [ETX 0x03]
⚠ Only the doubled BW_CMD (10 10) form has been exercised. Whether other
literal 0x10 bytes inside params require stuffing is untested — none of
the probes sent carried one.
Responses — Series III minus the DLE prefix
Series III: [DLE 0x10] [STX 0x02] … [chk] [ETX 0x03]
Micromate: [STX 0x02] … [chk] [ETX 0x03] ← no leading DLE
This single byte matters operationally: Blastware's parser locates frames by
scanning for DLE+STX, so it will never find a frame boundary in Micromate
traffic no matter what else is correct. That is a structural reason a
Micromate cannot call into a Blastware ACH server, independent of any baud
mismatch.
Response payload header
[0] CMD 0x00 same as Series III
[1] flags 0xC5 ← Series III uses 0x10. Constant across all 10 SUBs.
[2] SUB 0xFF − request_SUB
[3] PAGE_HI
[4] PAGE_LO
[5+] data
Checksum — the DLE-aware variant
chk = sum(b for b in payload if b != 0x10) & 0xFF
Confirmed on every frame captured. The POLL probe response contains no
0x10 and so cannot distinguish plain SUM8 from the DLE-aware form; the
POLL data response contains a 0x10 at payload offset 42, and only the
DLE-aware rule matches there. This is the same checksum Series III uses for
its 5A bulk-stream and write frames — not the plain SUM8 of ordinary
Series III reads.
The probe response carries the data length
Series III hardcodes DATA_LENGTHS per SUB. On the Micromate the probe
response tells you, as a uint16 BE at payload[8:10]:
⚠ Corrected 2026-09-23. An earlier draft read this as a single byte at
payload[9]. That is right only while the high byte is zero, and it is
catastrophically wrong for SUB 0x1A, whose real length is 0x082C = 2092
— read as a byte it gives 44, a 47× under-read. Always read the pair.
| SUB | command | payload[9] |
Series III constant |
|---|---|---|---|
0x15 |
serial number | 0x0A |
0x0A ✓ |
0x01 |
device info | 0x98 |
0x98 ✓ |
0x1C |
monitor status | 0x2C |
0x2C ✓ |
0x06 |
storage range | 0x24 |
0x24 ✓ |
0x2C |
call-home config | 0x7E |
0x7C ✗ differs by 2 |
0x08 |
event index | 0x5A |
— |
0x1E |
event header | 0x08 |
— |
0x1A |
compliance config | 0x082C (2092) |
— |
0x0A |
waveform header | 0x00 |
— (no event context) |
0xFE |
full config | 0x00 |
— (see note) |
Four of four known Series III lengths match exactly. Read the length from the probe rather than hardcoding it — it is free, and it already caught the call-home divergence.
Confirmed commands (read-only)
All ten below answered with a correct 0xFF − SUB response. Nothing that
writes, erases, or changes monitoring state has been sent to a unit.
| SUB | RSP | Command | Data proven |
|---|---|---|---|
0x5B |
0xA4 |
POLL / handshake | yes — ID block |
0x15 |
0xEA |
Serial number | yes — UM12947 |
0x01 |
0xFE |
Device info | yes — 152 B |
0x2C |
0xD3 |
Call-home config | yes — 126 B |
0x1C |
0xE3 |
Monitor status | yes — 44 B |
0x06 |
0xF9 |
Event storage range | yes — 36 B |
0x08 |
0xF7 |
Event index | probe only |
0x1E |
0xE1 |
Event header / first key | probe only |
0x0A |
0xF5 |
Waveform header | probe only |
0x1A |
0xE5 |
Compliance config | probe only |
Decoded so far
SUB 0x15 — serial. ASCII, null-terminated: UM12947.
SUB 0x5B / 0x01 — identification strings.
Instantel\0 and MM/ISEE/S/IO (MicroMate / ISEE standard). 0x01 also
carries eight consecutive 3f 80 00 00 float32 values (= 1.0f) — almost
certainly per-channel calibration/scale factors, by analogy with Series III's
geo_hardware_constant. Unverified.
SUB 0x1C — monitor status. Series III field offsets apply unchanged:
| field | offset | read |
|---|---|---|
| battery × 100 | payload[-10:-8] uint16 BE |
0x017D → 3.81 V |
| memory total | payload[-8:-4] uint32 BE |
15,000,000 |
| memory free | payload[-4:] uint32 BE |
15,000,000 (empty) |
| date | payload[18:22] |
day 23, month 9, year 0x07EA = 2026 |
The battery reading independently corroborates: Thor's own event reports for
these units print BatteryLevel : 3.8 volts.
SUB 0x2C — call-home config. Contains the ASCII string RADIO RING.
Worth flagging: that is the exact string seen in the RV50 ALEOS_SERIAL
debug during the BE12599 incident —
'ATQ1^MATE0^MATS0=2^M^MRADIO RING^M'. So this block holds the modem dial /
answer strings, and it is the most directly relevant command to the
call-home-receiver goal. Field layout not yet mapped; Series III's map
(raw[5] enabled, raw[6:46] dial string) is a starting hypothesis only, and
the length already differs (0x7E vs 0x7C).
SUB 0x06 — storage range. All zeros on this unit, consistent with
memory free == memory total. Series III reads first/last event keys from
the final 8 bytes; untestable until the unit holds events.
The event chain — walked end to end (2026-09-23, 5 events)
With 5 events on the bench unit (4 waveform + 1 histogram), the Series III browse walk works unmodified:
1E (all-zero params) -> first key + size
0A (key) -> partial record, histogram only
0C (key) -> 210-byte waveform record
1F (all-zero params/browse) -> next key + size
... repeat ...
1F -> all-zero key = NULL SENTINEL, chain ends
The sentinel terminated correctly after exactly 5 events.
Event keys are sequential, not addresses
055d4a81 055d4a82 055d4a83 055d4a84 055d4a85
This is a real divergence. Series III keys are flash-buffer addresses
(01110000, 011121F2, …) that advance by the event's byte length, which is
why its 5A chunk walk is address-arithmetic. Micromate keys are a plain
incrementing counter. Any port of the Series III download walk must not
assume key arithmetic means anything.
The 4 bytes after the key are the event's size
1E/1F return [key 4B][size 4B]. Series III uses that slot as an offset
to the next key; here it is a byte count:
| key | size | kind |
|---|---|---|
055d4a81 |
4,076 | histogram |
055d4a82 |
11,032 | waveform |
055d4a83 |
11,502 | waveform |
055d4a84 |
13,424 | waveform |
055d4a85 |
8,746 | waveform |
Consistent with real file sizes (corpus .IDFH ≈ 3.7–25 KB, .IDFW ≈
8.6–15.8 KB), and the histogram is unmistakably the small one. ⚠ Inferred,
not proven: the sizes sum to 48,780 while monitor status reports 57,344 bytes
used, so ~8.5 KB of overhead is unaccounted for.
SUB 0x0C — waveform record, and it carries the job metadata
Length 0xD2 = 210 bytes — identical to Series III. Contents confirmed
across all 5 events:
- the event key, echoed
- date + time (
17 09 07 ea→ 23 Sep 2026, then10 21→ 16:33 — matching the actual bench recording time) - title note
"Location" - the project string —
"Univ of Pitt-1st Yr Housing-Loc1 Ruskin" - serial
"UM12947" - channel labels
Tran/Vert/Long/Mic— the same labels Series III uses, and the same label-relative float32 layout - per-event float32 peaks: 3.5152, 1.3720, 2.3542, 3.5152, 0.4227 in/s across the five events (varied deliberately during recording)
This closes the biggest open question for the call-home-receiver goal.
The job identity strings (project / client / operator / setup) that
today arrive only via Thor's .txt sidecar — and which no amount of sample
decoding can reconstruct — are available over the wire from 0x0C. A
direct-to-SFM event need not arrive with blank metadata.
SUB 0x0A — partial record, histogram only
0x0A returned len = 0x1E (30 B) for the histogram and len = 0x00 for all
four waveforms. The histogram payload carries two timestamps and the
ASCII string "\r Vert: 0.300 in/s" — structurally the Series III
monitor-log partial record (0x2C type), which likewise holds a start/stop
pair and a "Geo: <float> in/s" trigger string.
So on the Micromate the division of labour is: 0x0A describes interval-style
records, 0x0C describes triggered events. Series III uses 0x0A's
response length (0x46 vs 0x2C) to tell real events from boundaries;
that discriminator does not apply here.
DLE stuffing in responses — confirmed present
Earlier marked untested. The 0x0C timestamp field contains 10 10, which
destuffs to a single 0x10 and yields a sensible clock reading. Responses
are DLE-stuffed, so a parser must destuff before applying field offsets.
A/B: Blastware build vs Thor build (2026-09-23)
UM12947 (11.0CB) and UM20147 (11.0BD) were each put on the bench and given
the identical read-only sweep. Both answer Series III command frames.
UM12947 11.0CB |
UM20147 11.0BD |
|
|---|---|---|
| POLL answers | ✓ | ✓ |
| All 10 read SUBs answer | ✓ | ✓ |
response_SUB = 0xFF − req |
✓ | ✓ |
| DLE-aware checksum valid | ✓ | ✓ |
| Two-step probe/data read | ✓ | ✓ |
| ID string | MM/ISEE/S/IO |
MM/ISEE/S |
| flags byte | 0xC5 |
0x03 |
0x1C length |
0x2C |
0x30 |
The firmware line does not change the wire protocol. One protocol stack can drive the whole fleet regardless of which build a unit is on. This is the single most consequential finding so far: "standardise the fleet on one firmware" becomes an optional convenience rather than a prerequisite for building a call-home receiver.
The two differences that do exist
1. The flags byte identifies the build. Response payload[1] is 0xC5
on the Blastware line and 0x03 on the Thor line, constant across all ten
SUBs on both units. That makes firmware line detectable from any response,
without reading device info. ⚠ Two units, one each — treat as a strong
hypothesis, not a proven encoding.
Note 0x03 is ETX, so on Thor-line units it arrives DLE-escaped as 10 03.
A parser that fails to destuff will mis-locate every field by one byte on
exactly half your fleet.
2. SUB 0x1C (monitor status) is 4 bytes longer on the Thor line —
0x30 vs 0x2C — with four extra trailing bytes (0f a0 00 00, purpose
unknown).
⚠ This breaks relative-to-end parsing. Series III reads battery and
memory from the end of the 0x1C block ([-10:-8], [-8:-4], [-4:]).
Those offsets are correct on 11.0CB and wrong on 11.0BD — applying them
blindly to UM20147 yields a battery reading of 577.92 V. Parse forward
from the declared length instead of backward from the end.
With the offsets shifted by 4, UM20147 reads correctly: battery 3.81 V, memory 15,000,000 total and free (no events stored).
Divergences from Series III (running list)
- No
DLEprefix on responses — bareSTX. - Response flags byte is
0xC5, not0x10. - Call-home config is 126 bytes, not 124.
- Data lengths are discoverable from the probe response at
payload[9]. 4b. Event keys are a sequential counter, not flash addresses. 4c.1E/1Freturn the event size, where Series III returns an offset. 4d.0x0Avs0x0Csplit by record type, not by the0x46/0x2Clength discriminator Series III uses. 4e. Responsepayload[1](flags) encodes the firmware line —0xC5Blastware,0x03Thor — where Series III has a constant0x10. - Modem serial rate is 115200, not 38400 (per TMI provisioning practice; not independently verified here).
The firmware images are unencrypted — and they document themselves
ref-stuff/micromate-firmware/MICROMATE(CB).BIN and MICROMATE(BD).BIN,
~2.77 MB each and within 192 bytes of one another.
- Entropy 6.08 bits/byte — neither encrypted nor compressed. Plain code and data.
- Header is a big-endian vector table, handlers at
0x4010_30xx. - ~16,700 extractable strings, including the developers' own debug
printfformat strings with function names intact.
This is a legitimate interoperability reference for hardware TMI owns, and it short-circuits work I had scoped as "only answerable from a live modem capture".
The call-home state machine, verbatim
ACH_NOT_STARTED → ACH_IDLE → ACH_INITIALIZING → ACH_CONNECTING
→ ACH_CONNECTED → ACH_TRANSFER_DATA
→ ACH_RETRY / ACH_QUITTING (also ACH_STARTED)
Supporting strings:
ACH: Entry StartCallHome()
ACH: CallHome_task ; CheckAliveTime CANCEL ; TimeBetweenRetries = %d
ACH: CallHome_task ; !ExpectedCommunicationsDetected() CANCEL ; TimeBetweenRetries = %d
ACH: CallHomeCommectionCompleteProcessing() ACH=%s EAMWC=%s
ACH: %s() three attempts and it's over
ACH: %s() Send CMD_START_MONITOR
ACH: %s() Send CMD_STOP_MONITOR
ACH: Start Ignore request, Call Home is in progress
What this tells us without a single captured packet:
- Retry limit is three — "three attempts and it's over".
ExpectedCommunicationsDetected()gates the session. If the host does not say something the unit recognises, the call is cancelled and rescheduled afterTimeBetweenRetries. A homebrew receiver must satisfy this check or units will retry forever — which is exactly the failure mode seen on BE12599.- The unit stops monitoring to call home and restarts afterwards
(
Send CMD_STOP_MONITOR/CMD_START_MONITOR). Relevant to any wedged-unit rescue: the monitoring state around a call is the device's own doing, not ours. - Calls are not re-entrant — "Call Home is in progress" is ignored.
Internal command table
CMD_CALLHOME, CMD_CALLHOME_CANCEL, CMD_CALLHOME_CONNECTION_CONFIRMED,
CMD_CALLHOME_CONNECTION_COMPLETE, CMD_CALLHOME_SET_SCHEDULE,
CMD_CALLHOME_CLEAR_SCHEDULE, CMD_STOP_CALLHOME_FILETRANSFER,
CMD_DUTYCYCLE_AUTOCALLHOME, CMD_PURGE_EVENT_FLASH.
CONNECTION_CONFIRMED as a distinct state from CONNECTION_COMPLETE implies
a handshake the host must complete before data flows — the concrete shape
of ExpectedCommunicationsDetected().
Event delivery — the mechanism, probably
All Events Uploaded
Mark/Unmark File Delete Marked Events Marked Events were Deleted
MONITOR::MESG PURGE_EVENT_FLASH BEGIN / END
A marking mechanism exists, alongside a distinct "all uploaded" terminal state. 🔶 Inferred: events are marked as transferred rather than deleted on send, and purging is a separate explicit act. If so, a receiver that fails to mark would see the same events re-offered every call — the question that gates a safe homebrew receiver. Not yet confirmed; needs either a live call-home capture or disassembly around these strings.
A full user manual is embedded
The firmware carries its own HTML help, which documents configuration we would otherwise have to infer:
- Modem mode:
Generic(through a modem) vsUSB to PC. - Modem baud: 9600 / 19200 / 38400 / 57600 / 115200 / 230400 — "must match the expected rate of the PC or modem". Confirms 115200 is a setting, not a fixed rate.
- Modem relay + warmup (0–300 s), auxiliary mode, warning/alarm hold.
- Record modes: Waveform, Waveform Manual, Histogram, Histogram-Combo; sample rates 1024 / 2048 / 4096.
- A scheduler downloaded from THOR that can start/stop monitoring, change
record mode, trigger a call home, or run a self check on a daily/weekly
schedule. Pairs with
CMD_CALLHOME_SET_SCHEDULE.
Still worth doing
Diffing the two images should isolate exactly what the CB/BD split changes —
we know the wire protocol is not it, and the flags byte (0xC5 vs 0x03)
gives a concrete anchor to search for.
SUB 0x5A — bulk download. It streams the .IDFW file verbatim.
The complete read path works with no Instantel software in the loop.
It needs no arming sequence
Series III ignores a 5A probe unless preceded by
1E → 0A → 1E(token 0xFE) → 0C → 1F(token 0xFE) → POLL × 3. The Micromate
answers a bare 5A request with nothing before it. That whole ritual is
gone.
The offset word is a LENGTH, not a position
This is the key divergence. Series III walks chunks by absolute flash
address, stepping 0x0200 per request. On the Micromate the offset word
requests how much to send:
offset_word = 0x1000 + 2 × pages pages = ceil(event_size / 512)
offset_word |
pages | destuffed file bytes |
|---|---|---|
0x1002 |
1 | 518 |
0x1004 |
2 | 1,030 |
0x1006 |
3 | 1,541 |
0x102C |
22 | 11,033 — the whole event |
event_size comes from the chain walk (the 4 bytes after the key in
1E/1F). One request returns the entire event; there is no chunk loop,
no STRT end-offset parsing, and no TERM frame. Over-requesting is safe —
0x1030 (24 pages) returned exactly the same bytes as 0x102C, so the device
caps at the real size.
Params are the Series III probe form: [0x00][key4][6 × 0x00].
The payload is the .IDFW file, byte for byte
[18-byte frame header] [ .IDFW file ] [chk] [ETX] ← raw wire
^ destuffed offset 16
The file begins 00 12 01 00 00 00 "Instantel\0" — _THOR_PREFIX +
_INSTANTEL_TAG from micromate/idf_file.py. The first 32 bytes are
identical to a production .IDFW pulled from the store.
⚠ Responses are DLE-stuffed. Destuff before locating the file, or the raw byte count overshoots (11,781 raw → 11,049 destuffed for an 11,032-byte event).
End-to-end proof
Event 055d4a82 downloaded over USB and fed straight to read_idf_file():
serial UM12947
timestamp 2026-09-23 16:33:19
samples Tran 3072 Vert 3072 Long 3072 MicL 3072
peaks Tran 0.2433 Vert 1.3706 Long 0.2672 in/s
All four channels equal length, and the timestamp matches the 0x0C record
for the same key. Independent cross-check: 0x0C reports a stored peak
of 1.3720 for this event; the decoded samples give 1.3706 — two
unrelated paths agreeing to 0.1%.
Consequence: no new codec work is needed. The bytes off the wire are the
same bytes thor-watcher forwards today, so /db/import/idf_file ingests a
directly-downloaded event unchanged. Everything the IDF decoder already does
per-sample-exact applies.
What a full read now looks like
1E → first key + size
0C(key) → project/client/operator, timestamp, peaks
5A(key, 0x1000+2×ceil(size/512)) → the whole .IDFW
1F → next key + size (until null sentinel)
Setups are FILES, not a config block
Series III has one compliance config you overwrite. Series IV keeps named setup files on an on-device filesystem, with a pointer to the current one. From the firmware:
csetup.MMB the current setup
factory.MMB Factory Default Setup File
callhome.MMB call-home config is a file too
"Current Setup File: " "Can Not Delete Active Setup File"
GetSelectedSetupFilePathName() CSelectSetupFiles CSaveSetupFile
Names are up to 20 characters and may contain spaces, hyphens, underscores. The unit's help text describes selecting, renaming and deleting them, and the event list records which setup file produced each event.
Filesystem primitives exist internally (NS_ReadFile_internal,
NS_WriteFile_internal, NS_SeekFile_internal), but no generic
file-transfer command is exposed on the wire — the only file-transfer string
is CMD_STOP_CALLHOME_FILETRANSFER. So setups are unlikely to be pushed as
raw .MMB blobs over the protocol.
SUB 0x1A reads the whole active setup — 2,092 bytes
Structurally close to Series III's ~2,126-byte compliance block, and it carries everything a setup consists of:
- the setup file name —
Univ of Pitt-1st Yr. Housing-Loc1 Ruskin.MMB - all four title note/value pairs —
Location,Client,Company,General Notes, with their strings - the sensor location string (
Loc 1) - per-channel labels and units:
Tran in./s.,Vert in./s.,Long in./s.,Mic psi (L),LMic psi (L),SMic (A)
Note LMic / SMic — linear and sound-level microphone variants that
Series III does not have.
This is the read half of setup management, and it means a setup can be round-tripped: read the active config, modify, write it back. The write half is not yet attempted.
Static analysis of the firmware (2026-09-23, solo session)
Architecture
ColdFire / 68K, big-endian, Freescale MQX RTOS — not ARM as the vector
table first suggested. The giveaway is the function epilogue/prologue
4E 5E 4E 75 4E 56 = UNLK A6 / RTS / LINK A6, littered through both
images, plus an MQX_OK assertion string.
CB vs BD: the same source, ~10 lines apart
A byte diff is useless — 68% of bytes differ because the two are separately linked builds with everything relocated. A string-set diff is position-independent and tells the real story: 17,128 strings shared, and almost every "unique" string is the same message with a different source line number:
CB: MONITOR[3268]: STATUS_BATTERY_LOW
BD: MONITOR[3258]: STATUS_BATTERY_LOW ← consistently 10 lines apart
The offset is exactly 10 across STATUS_BATTERY_LOW, STATUS_BATTERY_CRITICAL,
Battery Critical Exit Monitor, histogram interval size of 0 and
Offsets by channel — so one ~10-line block differs in the monitor module and
essentially nothing else. The only functional string unique to either build is
CITIZEN (a receipt-printer brand) in BD.
This corroborates the bench A/B from the other direction: the CB/BD split is a tiny code delta, not two protocol stacks. Whatever drives Instantel to ship two downloads, it is not a different wire protocol.
⚠ The SUB dispatch is a 68K switch jump table (CMPI.L bounds check →
MOVE.W (table,PC,Dn) → JMP (d8,PC,Xn)). Byte-pattern hunting will not
isolate the write opcodes — that needs a real disassembler.
Call-home config field names, from the firmware's own debug dump
CallHome.Enable = %s
CallHome.DialString = "%s"
CallHome.Retries = %d
CallHome.SessionTimeout = %d
CallHome.WaitForConnection = %d
CallHome.WarmupTime = %d
CallHome.PowerSave = %s
Seven fields, which is what the 126-byte SUB 0x2C block has to encode. Note
SessionTimeout and PowerSave have no Series III equivalent, and Series III's
scheduled-time fields (time1/time2 hour/min) are absent here — consistent
with Series IV moving scheduling into the THOR-downloaded scheduler instead.
Also present: AT+CSQ (signal quality), so the firmware talks AT to the modem
directly.
All five bench events, downloaded and decoded
Read-only, over USB, no Instantel software:
| key | declared size | got | decoded |
|---|---|---|---|
055d4a81 |
4,076 | 4,076 | histogram, 1 interval, 16:33:16 |
055d4a82 |
11,032 | 11,032 | waveform, 3072 × 4 ch, 16:33:19 |
055d4a83 |
11,502 | 11,502 | waveform, 3072 × 4 ch, 16:33:27 |
055d4a84 |
13,424 | 13,424 | waveform, 3072 × 4 ch, 16:33:34 |
055d4a85 |
8,746 | 8,746 | waveform, 2048 × 4 ch, 16:33:36 |
Every event arrived at exactly its declared size, every channel came out equal length, and the timestamps are sequential across the recording session.
⚠ Gap: no content-based record-type discriminator
read_idf_file() decides waveform vs histogram from the filename suffix
(.IDFH / .IDFW) — and there is no filename when downloading over the wire.
The first 64 bytes of a histogram and a waveform are byte-identical, so the
header does not obviously carry the type either.
The protocol does supply one: SUB 0x0A returns length 0x1E for a
histogram and 0x00 for a waveform (see the event-chain section). Any
direct-download implementation must carry the type from the chain walk rather
than inferring it from the payload, or find the type field inside the IDF
header.
⚠ Unresolved: the 0x0C peak float
The float32 extracted from 0x0C runs 2–5% above max(Tran, Vert, Long) from
the decoded samples:
| key | 0x0C float |
max channel |
|---|---|---|
…81 |
3.5152 | 3.4419 |
…82 |
1.3720 | 1.3706 |
…83 |
2.3542 | 2.2227 |
…84 |
3.5152 | 3.4419 |
…85 |
0.4227 | 0.4198 |
It is not peak vector sum either (computed PVS runs higher than both). The field may not be the peak at all — its offset was inferred from a byte marker, not established. Do not rely on it until it is pinned properly.
Worth noting the histogram (…81) and the loudest waveform (…84) report
identical peaks to four decimals, in both measures. That is self-consistent:
the histogram's single 1-minute interval spans the whole thumping session, so
its maximum should equal the loudest event in it.
⚠ Untested and unsafe-until-agreed
Nothing below has been sent to a unit, and nothing should be without an explicit decision:
- Writes (
0x68–0x83), call-home write (0x7E/0x7F) - Erase (
0xA3/0xA2) - Start / stop monitoring (
0x96/0x97) 0x1F(advance event pointer) — non-destructive on Series III but it does move device state, so it is parked with the rest
Also unknown:
- Whether
0x10bytes inside request params need stuffing - Everything about the call-home session — the device-initiated direction has not been observed at all. Specifically: how a unit announces itself, and how it learns an event was accepted so it stops re-sending it. That last question gates any homebrew receiver and cannot be answered over USB.
Session provenance
Unit UM12947, firmware 11.0CB (Blastware line), on the bench via
USB, zero events stored (memory free == total). Read commands only.
Every response in this document was checksum-validated.
⚠ Firmware is the single biggest caveat on this document. Every finding here
is from one unit on the Blastware build. A 11.0BD unit has not been
touched.
An empty unit is a real limitation: 0x08, 0x1E, 0x0A and 0x06 all have
event-dependent payloads that could not be exercised. Recording a couple of
events on the bench unit would unlock the entire event-walk half of the
protocol.