A Micromate event read without Thor's report (SFM's own downloads) came out with sample rate, pre-trigger and record time 0, empty notes, and a 1970-01-01 time; every Micromate .h5, report or not, stored pre-trigger 0. - The settings were found by the bytes be 80 00 00 00 00 -- a -0.25 s pre-trigger and Fixed stop mode, not a marker -- so any other pre-trigger left them unread. The event's settings record is now found by its own signature 00 2a 04 00 00 (fields at the live 0x1A block's head offsets). - The event time is read at settings record + 0x34, not two fixed offsets that miss a header shape. - The four notes are read at 0x23c ([label 22][value 42] x 4). - IdfEvent.to_minimateplus_event sets pretrig_samples. - save_imported_idf fills the pre-trigger and notes from the file when the report lacks them. Measured against Thor's own reports on 2,027 corpus files: sample rate and notes 2,027 / 2,027; record time and pre-trigger 291 / 291 waveforms; event time = filename on 2,384 / 2,384. The issues' 20 UM12947 bench files now store 4096 S/s, 4096 pre-trigger samples (1.0 s), 3 s and their notes. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01ALX2R9LdWawLzUxwewZGA1