Measured on BE9441 tonight, through the recording bridge:
- Monitoring flag 0x0B while recording a histogram (confirmed at the unit), 0x00
idle. Only 0x10 had been measured as monitoring (BE11529, 2026-04-08), so the
status showed 'not monitoring' and the change path refused the state as
unknown(0x0B) -- refused, not skipped, which is the design working. 0x0B joins
S3_RUN_STATES and S3_MONITORING_FLAGS as measured. The shared status's
'monitoring' now follows the change path's reading: None for an unmeasured
flag, never a guessed False.
- Free memory 0x000D0E04 arrived as 0d 0e 10 04. The parser keeps the 10 of a
10 02/03/04 escape, so every field counted from the end read one byte off:
419.84 V and 251,655,168 bytes for 6.76 V and 983,028. New
framing.unescape_reply() (also used by parse_strt_end_key) runs first.
Tests: the two recorded replies (idle 16:24, monitoring 18:47) decode to the
unit's own values; each escaped value in the last ten bytes decodes; the status
mapping for 00 / 10 / 0B / unmeasured. No codec, store or DB change.