Measured at the bench 2026-11-03 (BE9441): while monitoring, a MiniMate answers
the 0x2C call-home read with its DEVICE-INFO block ('Instantel', 'MiniMate Plus',
its serial) -- as it answers 1A -- and idle again, minutes later, the call-home
block.  SFM writes back only a block it just read, so a monitoring MiniMate's
call-home cannot be changed; the change core already refused (byte 5 read as
'unknown(0x6E)', nothing sent), but said so cryptically.

- minimateplus.protocol.is_call_home_block(raw): the block starts with its own
  length 0x7C and is N (or N + 2) bytes unescaped; the device-info block fails it.
- The Series-3 change raises UnitStateUnknown (409) with 'did not return its
  call-home settings -- a MiniMate answers the call-home read with its
  device-info block while monitoring; stop monitoring first ... nothing sent',
  for a dry run too.  call_home_state returns enabled = None with that reason.
- The page's callHomeRows shows the reason in place of 'Unknown (byte 5 = ...)'.

A blind write while monitoring stays untested on purpose.