serversdownandClaude Opus 5.5 a72d3a98a9 fix(minimateplus): refuse a settings write whose four fields would not read back as asked (#1 review)
_encode_compliance_config patches the raw SUB 1A bytes at offsets measured at
1024 sps (April 2026), where the escape before 04 sits between the mode and the
anchor.  b320a53 moved the decoder to the measured layout (unescaped: mode -7,
rate -6, interval -4, record time +6), which exposed the encoder: on BE9441's
recorded 10-25 Continuous / 4096 block a mode change rewrote the byte in front
of the mode, and on the 10-26 Histogram+Continuous / 2048 block a sample-rate
change landed one byte off.  The old decoder read the same wrong bytes, so such
a write looked as if it had taken.

The encoder now decodes its own result with the decoder and raises ValueError
when mode, sample rate, interval or record time would not read back as asked
(or a field not asked for would change) -- before push_config_raw sends a byte.
No write offset is moved: rewriting the encoder on unescaped bytes is #46's.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44